1 BROOKHAVEN SCIENCE ASSOCIATES

National Synchrotron Light Source II

Hard Coherent X-Ray Beamline

Lonny BermanEFAC Meeting, October 4th, 2007

with contributions by Alex Babkevich, Nigel Boulding, Scott Coburn, Simon Mochrie, Ian Robinson, Alec Sandy, Lin Yang

Page 2: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

2 BROOKHAVEN SCIENCE ASSOCIATES

Scientific Mission• Application of hard (7–20 keV) coherent x-rays to the study of nanoscale

dynamics and structure of complex materials

• Equilibrium dynamics and fluctuations about the evolution to equilibrium in colloids, polymers, membranes, concentrated proteins, glasses, …

• 3-D imaging of microscopic non-crystalline objects such as catalytically active materials, cavities within steel, defect structures in magnetic multilayers, cellular imaging

Mark A. Pfeifer, Garth J. Williams, Ivan A. Vartanyants, Ross Harder, and Ian K. Robinson, “Three-dimensional mapping of a deformation field inside a nanocrystal,” Nature 442, 63 (2006)

O. G. Shpyrko, E. D. Isaacs, J. M. Logan, Yejun Feng, G. Aeppli, R. Jaramillo, H. C. Kim, T. F. Rosenbaum, P. Zschack, M. Sprung, S. Narayanan, and A. R. Sandy; "Direct measurement of antiferromagnetic domain fluctuations," Nature 447, 68 (2007)

Page 3: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

3 BROOKHAVEN SCIENCE ASSOCIATES

Beamline Requirements and Specifications

• Coherent flux to supply photons to two different endstations that operate simultaneously, one for x-ray photon correlation spectroscopy (XPCS), one for coherent x-ray diffraction (CXD)

• Energy range 7-20 keV using either tunable silicon double crystal monochromator or “pink beam” reflected from mirror

• At 8 keV, coherent flux using a Si mono is 6x1012 ph/sec, and tranverse coherence lengths at 25 m from the source are 819 µm vertically and 65 µm horizontally (Ch. 3 of NSLS-II CDR)

• Stability• Parasitically fluctuating signals contaminate XPCS time autocorrelation analysis• Parasitically fluctuating signals significantly complicate phase retrieval• Fixed secondary horizontal source intended to improve stability for these experiments

• Long small angle XPCS station• “Smart” area detectors likely to have larger pixels than today’s

• Long beamline for CXD• High demagnification focusing for CXD and large Q XPCS• Reduced flux density for radiation sensitive samples

Page 4: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

4 BROOKHAVEN SCIENCE ASSOCIATES

Overall Beamline Layout

XPCS endstation 50 m from source

CXD endstation 250 m from source

FOE

Page 5: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

5 BROOKHAVEN SCIENCE ASSOCIATES

Conceptual Design - Beam Splitting• 2 secondary sources or pinholes in the FOE

• Multiplexed operation creates twice as much beamtime• Small horizontal-bounce mirrors used to extract coherent fraction of

the beam (in the horizontal)• 2× mirror deflections and long beamline creates sufficient horizontal

clearance

CHX FOE Beam Splitting Concept - Plan View

Hi-pass filter

Slit

Mirrors = secondary source or pinhole

½ Slit

WB Stop

ShutterWhite beam (WB)

Power-reducing aperture

Pink beam

12 mrad

Page 6: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

6 BROOKHAVEN SCIENCE ASSOCIATES

XPCS Beamline Layout

endstation enclosure

monochromator enclosure

Page 7: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

7 BROOKHAVEN SCIENCE ASSOCIATES

CXD Beamline Layout

monochromator enclosure

endstation enclosure

Page 8: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

8 BROOKHAVEN SCIENCE ASSOCIATES

Insertion Device

• U19 undulator preferred (unless U14 is available)• High beta straight section• Brighter than APS by a factor of 7 at 10 keV under

these conditions (U19, high beta); coherent flux is directly proportional to brightness

• Useful upgrade is to add a second undulator in-line

undulator source >~ 212 µmsecondary source = 67 µm coherence-defining aperture at sample = 30 µm

30 m 20 m

Filling of secondary aperture for XPCS beamline can only be accomplished if undulator source width >~ 212 µm, leading to the preference for a high beta straight section.

Page 9: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

9 BROOKHAVEN SCIENCE ASSOCIATES

Front End Layout

• Front end must include motorized adjustable apertures• Will typically be closed down to an opening no greater

than 100 µrad x 100 µrad• Thereafter, separate beams of up to 20 µrad x 20 µrad

each need to be extracted for the two beamlines (separation to be done in the FOE)

Page 10: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

10 BROOKHAVEN SCIENCE ASSOCIATES

White Beam Components

• Water-cooled silicon mirror and monochromator preferred for each beamline, to avoid vibrations associated with a cryo-cooling system (power density is ~100 W/mm2 but power in 20 µrad x 20 µrad is ~30 W)

• Finite-element analysis and Shadow ray tracing were carried out (and are still underway) for various cases, involving use of one and two undulators, operating at low K and at high K

• The results so far indicate that water-cooled optics will probably be acceptable, although the slope errors will still be noticeable and may have to be compensated using downstream focusing optics

• Pink beam capability is preserved into both endstations

Page 11: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

11 BROOKHAVEN SCIENCE ASSOCIATES

Specialized Beam Conditioning Optics

• Focusing optics will sometimes be used in the two endstations, just before the experimental apparatus

• For the XPCS endstation, a vertical focusing mirror (or kinoform lens) is desired to reduce the vertical coherence length in order to match the horizontal coherence length

• For the CXD endstation, a two-dimensional focusing system (could be mirror-based, zone-plate-based, refractive-lens-based) is sometimes desired to focus the beam onto a nanometers scale sample

Page 12: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

12 BROOKHAVEN SCIENCE ASSOCIATES

XPCS Endstation

mirror or refractive lens

apertures

sample chamber

10 m flight path

heavily pixellated area detector with on-board correlator electronics

Page 13: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

13 BROOKHAVEN SCIENCE ASSOCIATES

CXD Endstation

chamber containing zone plate or refractive lens, apertures, and sample positioner on goniometer possible area detector

articulation scheme

Page 14: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

14 BROOKHAVEN SCIENCE ASSOCIATES

Requirements Imposed on Conventional Facilities

• Exquisite stability, particularly relative stability between the reference point at extraction and the reference point at distant endstation must be achieved. Floor vibration threshold should be same as for the main experimental hall, 25 nm rms for the range of 4-50 Hz.

• Importance of relative vibrations between the two buildings is under study through detailed finite element analysis of wave propagation and interaction with the two structures.

• Field data measurements from monitoring stations at the desired locations are being gathered to establish spatial variability of vibration environment at the site as well as its correlation characteristics.

Page 15: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

15 BROOKHAVEN SCIENCE ASSOCIATES

Outstanding Issues

• Several key design decisions still to be made (e.g. choice of focusing optics)

• Studies of thermal distortions of optics are still underway• Distant endstation and building design need to be firmed up• Area detector for x-ray photon correlation spectroscopy does

not exist and needs a lot of R&D (separately from project)• Instrument for coherent x-ray diffraction does not exist and

needs a lot of R&D, particularly in the area of sample registration (to keep the sample always in the same spot in the beam as it is oriented on the diffractometer); XRadia is interested in developing such a system and undertaking the necessary R&D, and is already doing so for Argonne

Page 16: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

16 BROOKHAVEN SCIENCE ASSOCIATES

Cost Estimate (Burdened and Escalated)

Note: white beam mirrors and slits are in white beam components, downstream monochromators and focusing optics are in beam conditioning optics

Enclosures  1,583,317.00$ Beam Transport  1,565,088.00$ Utilities  316,364.00$ Beam Conditioning Optics  1,989,884.00$ Personnel Safety System  278,181.00$ Equipment Protection System  86,038.00$ White Beam Components  1,278,815.00$ End Station 1  3,509,501.00$ End Station 2  3,328,590.00$ Beamline Controls  386,610.00$ Beamline Control Station 35,686.00$ Satellite Building 1,392,908.00$ Beamline Management 1,529,692.00$

Grand Total 17,280,674.00$

Page 17: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

17 BROOKHAVEN SCIENCE ASSOCIATES

Alternative Concepts

• We are fixed on the design pathway described thus far, based on the CD-1 conceptual design, at least through the CD-2 review

• In the course of elaborating the advanced conceptual design, alternative concepts came to mind

• We believe that it is important that conventional SAXS should be accommodated in the XPCS endstation, especially because of the large demand for SAXS and the absence of a project beamline to accommodate it

• And we also believe that it might be feasible to accommodate the needs of the CXD (out)endstation together with the needs of the hard x-ray nanoprobe (out)endstation through the same single beamline and (out)endstation, providing a route to significant cost savings, albeit requiring extensive re-design and cost estimate revisions of both project beamlines

Page 18: 1 BROOKHAVEN SCIENCE ASSOCIATES National Synchrotron Light Source II Hard Coherent X-Ray Beamline Lonny Berman EFAC Meeting, October 4 th, 2007 with contributions

18 BROOKHAVEN SCIENCE ASSOCIATES

Summary

• Two beamlines, each extracting a coherent portion of the beam radiated by an undulator

• Deflecting mirrors are used to separate the beams and deliver them to two independent beamlines that can operate simultaneously, one for XPCS and one for CXD

• The CXD beamline needs to be long, and a separate building needs to be constructed to accommodate the distant endstation

• Lots of R&D challenges that have to be addressed!