204010170

Embed Size (px)

Citation preview

  • 8/14/2019 204010170

    1/48Cdigo do manual: 204010170_004 1

    O eration

    Manual

    OPERATION MANUAL

    1415 MULTIAGENT VAPORIZER

    Equipment code: 201030023

    Manual Code: 204010170_004

    Date: (AGO/2008)

  • 8/14/2019 204010170

    2/48Cdigo do manual: 204010170_004 2

    His Operation Manual has the information that is necessary for the proper usage of Multiagent Vaporizer 1415.

    Manufacturer:

    K. Takaoka Indstria e Comrcio LTDA

    Rua General Izidoro Dias Lopes, 121/141

    Bairro Vila Paulicia - So Bernardo do Campo / SP

    ZIP: 09687-100Telephone: (55 11) 4176-3500

    Fax: (55 11) 4176-3570

    Website: www.takaoka.com.br

    E-mail: [email protected]

    CGC: 61.489.381/0001-09

    I.E.: 635.547.970.118

    Suggestions, doubts or complaints:

    Call Center: (55 11) 4176 3636

    Product Registration at the Health Ministry:

    Technical Name: Vaporizer

    Trade Name: Multiagent Vaporizer 1415

    Registration Number at the Health Ministry: 10229820077

    Product Classification:

    IEC-60601-1:1988+amendment 1994 (Medical electrical equipment - Part 1: General requirements for

    basic safety and essential performance)

    IEC 60601-2-13/2003- (Medical electrical equipment - Part 2-13: Particular requirements for the safety and

    essential performance of anaesthetic systems)

    Equipment Class 1 Energized Internally Type B IPX1 Continuos Operation

    Technician in Charge:

    Eng. Mauricio Chiarioni

    CREA: Registration number 5061714921

  • 8/14/2019 204010170

    3/48Cdigo do manual: 204010170_004 3

    Product Installation Form

    EQUIPMENT: ___________________________ Code: ____________ Serial Number:_________

    DESCRIPTION CODE SERIAL NUMBER1234

    56789

    ACCESS

    ORIES

    10

    INVOICE: Original K. Takaoka [ ] YES _______________ [ ] NO ______________________Invoice Number* Representation Name

    INSTITUTION:

    Corporate Name: ____________________________________Tax Payers ID*:_______________________

    Address: _______________________________________________________________________________City: __________________________________________State:_____________ ZIP: __________________

    Person in charge of the information*:______________________________Position: ____________________

    Sector: ______________________ Phone*: ___________________ E-mail: _________________________(*) Mandatory Fields

    YOUR OPINION:

    1. Was the product delivered punctually, within the period that was agreed on? Yes No2. Did products and accessories match the order? Yes No3. Was the package damaged in any way? Yes No4. Was there any difficulty in installing the equipment? Yes No5. Is (are) the device(s) and accessories working properly? Yes No6. Were there any accessory, tube, and cable connection problems? Yes No7. Were the data, values, product description, amount, and payment conditions on theinvoice correct?

    Yes No

    8. Mention any possible inconveniences:

    Installation by: ________________________ Date ____/ ____ /_____ Sig.: __________________(Technician Name)

    Representation: _______________________ Data ____/ ____ /_____ Sig.: _________________

    Send this form to Fax (11) 4176-3570, or via registered mail to K. Takaoka.

    ATTENTION: THE WARRANTEE WILL GO INTO EFFECT AS OF THE CONFIRMATION OF THEINFORMATION CONTAINED IN THIS TERM. THIS TERM MUST BE SENT IN NO MORE THAN 30 DAYS, AS

    SET FORTH IN THE OPERATION MANUAL.If you have any doubts, or for more information, contact: SAC (55 11) 4176 3636

  • 8/14/2019 204010170

    4/48Cdigo do manual: 204010170_004 4

    TABLE OF CONTENTS

    1 WARNINGS .................... ..................... ...................... ..................... ..................... ..................... ...................... ..................... ...8

    2 QUICK INSTALLATION GUIDE............................................................................................................................................10

    3 GENERAL DESCRIPTION....................................................................................................................................................12

    4 TECHNICAL SPECIFICATION..............................................................................................................................................13

    5 OPERATING PRINCIPLE .................... ...................... ..................... ..................... ..................... ...................... ..................... .15

    6 CONTROLS AND COMPONENTS........................................................................................................................................17

    6.1 LIST OF COMPONENTS...................................................................................................................................................17

    6.2 FRONT VIEW.................................................................................................................................................................17

    6.3 REAR VIEW ..................................................................................................................................................................19

    6.4 MINI-PINOMATIC BUBBLING CHAMBER ............................................................................................................................19

    7 DISPLAY...............................................................................................................................................................................22

    7.1 SCREENS .....................................................................................................................................................................22

    7.2 MESSAGES ...................................................................................................................................................................24

    8 ASSEMBLING AND PREPARING THE VAPORIZER...........................................................................................................25

    8.1 ASSEMBLING THE ANESTHESIA MACHINE ........................................................................................................................25

    8.2 CONNECTIONS..............................................................................................................................................................27

    8.3 POWER SUPPLY............................................................................................................................................................27

    8.4 MOUNTING THE MINI-PINOMATICBUBBLING CHAMBER.................................................................................................29

    9 OPERATION....................... ..................... ..................... ..................... ...................... ..................... ..................... ................... 30

    9.1 INITIAL PROCEDURES ....................................................................................................................................................30

    9.2 FILLING THE CHAMBER ..................................................................................................................................................30

    9.3 VAPORIZATION INSTRUCTIONS .......................................................................................................................................31

    9.4 CHAMBER DRANING ......................................................................................................................................................31

    10 CLEANING AND STERILIZATION........................................................................................................................................33

    11 MAINTENANCE... ...................... ..................... ..................... ...................... ..................... ..................... ..................... ............33

    12 SYMBOLS.............................................................................................................................................................................38

    13 EMERGENCY ACTIONS.......................................................................................................................................................43

    14 WARRANTY ................... ..................... ...................... ..................... ..................... ..................... ...................... ..................... .44

  • 8/14/2019 204010170

    5/48Cdigo do manual: 204010170_004 5

    DEFINITIONS

    WarningWarns the user about the possibility of injury, death or other serious adverse reactions associated with equipment

    misuse.

    AttentionWarns the user about possible problems with the equipment associated with its misuse, such as equipment

    malfunction or failure, damage to the equipment or damage to third-party property.

    Note:Emphasizes important information

  • 8/14/2019 204010170

    6/48Cdigo do manual: 204010170_004 6

    THE COMPANY

    K. TAKAOKA is a company that for more than 48 years has dedicated its efforts to the hospital equipment industry, always in close

    cooperation with physicians. Operating mainly in the Anesthesia, Intensive Medicine, Monitoring, and Oxygen Therapy areas, K.

    TAKAOKAis proud of its leading position in the market and has a wide range of products.

    With one of its priorities being ongoing investments in the research and development of new ideas and solutions, K. TAKAOKA has

    distinguished itself through the constant introduction of technological advancements and industrial innovations in its product line, whichputs it on par with its main national and international competitors.

    Using sophisticated equipment, K. TAKAOKA designs and manufactures most of the components for its devices, and this explains the strict quality

    control they are submitted to. The company is also concerned with supplying all customers with top-quality assistance by means of its Sales and

    Technical Assistance departments.

    With representatives throughout Brazil and a firm foothold in the international market, K. TAKAOKA has deserved its customers trust over

    the years, in its high-quality standards and the great efficiency of its products and services.

    Vision:

    Being a global company.

    Mission:

    Being the national leader in manufacturing anesthesia, pulmonary ventilator, vital sign monitors, and oxygen therapy devices, contributing

    to the preservation of life, offering high technology and better service to our clients.

    Quality Policy:

    Continually improve our PRODUCTS, SERVICES and PROCESSES, involve our EMPLOYEES and VENDORS, and obtain our CLIENT

    and SHAREHOLDER satisfaction.

    K. TAKAOKA IND. E COM. LTDA.Rua General Izidoro Dias Lopes.121/141

    Bairro Vila Paulicia So Bernardo do Campo / SPZIP: 09687-100

    Telephone: (55 11) 4176 3500Fax: (55 11) 4176 3570

    E-mail: [email protected]: http://www.takaoka.com.br

  • 8/14/2019 204010170

    7/48Cdigo do manual: 204010170_004 7

    Figure 1. Vaporizer 1415

  • 8/14/2019 204010170

    8/48Cdigo do manual: 204010170_004 8

    1 WARNINGS

    Using the Equipment

    Always verify the desired bubbling flow in order for the anesthetic agent concentration in the circuit to be the desired one for the

    inhaling anesthesia.

    Always verify what type of anesthetic agent is being used by pressing the AGENT key.

    If there is any doubt regarding the type of anesthetic agent present in the bubbling chamber, replace it.

    When replacing the anesthetic agent, make sure you reposition the anesthetic agent indicator pin to the correct position. Pin

    positioning reading ensures the correct calculation of the bubbling flow table.

    Never exceed maximum chamber capacity, i.e. 100 ml.

    The chambers filling funnel must be tightly closed during anesthesia so there is no gas leakage.

    Vaporizer 1415 has two scales for bubbling flows, one for low and another one for high flows. Do not add the readings.

    During a lack of power emergency, Vaporizer 1415 can continue being used without its electronic functions. However, special care

    should be taken to prevent high anesthetic concentrations.

    Maintain the bubbling chamber empty and dry every time the vaporizer is not being used in order for the equipment to be better

    conserved.

    The Vaporizer 1415 should be kept in the upright position while it is filled with anesthetic agent.

    This operation manual should be read very carefully in order for one to use the equipment correctly and gain maximum benefit from

    all its features.

    Also, read the Operation Manuals for the other anesthetic machine parts and accessories.

    The operator must supervise the anesthesia device continually while administering gas to the patient.

    Fire and Associated Risks

    Do not use the equipment in the presence of flammable agents. Additionally, do not apply oil or grease on the equipment to avoid the

    risk of combustion.

    To reduce the risk of fire, only use tubes and circuits designed for oxygen-enriched environments. Do not use electrically conductive

    or anti-static tubes.

    In case of fire, take care of the patient's needs immediately, turn Vaporizer 1415 off, and disconnect it from the power and gas

    sources.

    Service and Tests Do not perform any internal servicing of any part of the apparatus. For regular equipment servicing or to troubleshoot any abnormality

    in its operation, call K. TAKAOKAS Authorized Technical Assistance.

    To reduce the risk of electric shocks, do not remove the cap or any part of the equipment for internal access.

    To reduce the risk of electric shock, disconnect the equipments power source before any maintenance service is performed.

    Do not perform any maintenance procedure except for those specified in the devices service manual.

    Only use parts and pieces supplied by K. TAKAOKA Ind. & Com. Ltda.

    Any attempt to change the hardware or firmware without K. TAKAOKAS formal approval in writing will automatically void all

    equipment guarantees.

    Vaporizer 1415 should be calibrated and serviced annually by an authorized K. TAKAOKA technician.

  • 8/14/2019 204010170

    9/48Cdigo do manual: 204010170_004 9

    To ensure the equipment operates safely, always run the tests and procedures prescribed in item 9.1 before using the vaporizer in

    the patients breathing circuit. If the equipment fails any of the tests, remove it from clinical use immediately. Do not use the

    equipment until the necessary repairs have been carried out.

    The K. TAKAOKA product technical features are subject to changes without prior notice due to their constant technological

    development.

    Electromagnetic Susceptibility

    The equipment may suffer interference from certain transmission machines (for example: cell phones, walkie-talkies, wirelessphones, pagers, high-frequency surgical equipment (diathermy), defibrillators, short-wave therapy devices), which could interrupt the

    equipments normal operation. Do not use these transmission devices near vaporizer 1415. Do not use Vaporizer 1415 in an

    environment where there is magnetic resonance imaging equipment.

    Power Source

    Only connect the 12-VDC cable provided with the equipment between the Pinomatic Supports 12-VDC input and the Rotameter

    Blocks 12-VDC output.

    Miscellaneous

    Do not press any key using surgical instruments or tools. Only use your fingertips to press the keys. Hard or sharp objects may

    damage them.

    Establish an appropriate cleaning and sterilization routine for the vaporizer components.

    Disposal ("waste")

    All parts and components that have contact with fluids originating from patients (for example: respiratory circuit), are potentially

    contaminated. These are named semi-critical components and they must undergo a high-level disinfections or sterilization process

    before being discarded (at the end of their service lives), or be discarded as potentially infected hospital waste.

    Discard all parts removed from the equipment in accordance with your organizations part and component disposal protocol. Follow

    local government environmental protection recommendations, particularly for electronic waste or parts.

    "FIRMWARE" Release

    Vaporizer 1415 control program has a release number that allows one to identify its level of review and it is presented when the

    equipment is started up (switched on).

    Biocompatibility

    According to the ISO 10993-1, the equipment is classified as a device that has no direct and/or indirect contact with the patients

    body; therefore it is not included in this norms scope (Clause 4.2.1).

  • 8/14/2019 204010170

    10/48Cdigo do manual: 204010170_004 10

    2 QUICK INSTALLATION GUIDE

    For quick, safe, and proper installation, follow the procedure described below (for a more detailed installation, check item 8in this manual):

    Item Procedure Figure

    01Check package integrity. If there is any damage, contact your

    representative.None

    02Open the package and check its contents. If there is any

    discrepancy, contact your representative.None

    03Remove the equipment and its accessories; check the integrity of allcomponents. If there is any apparent damage in the equipment or in

    any of its accessories, contact your representative.

    None

    04Use the 12-VDC cable connected to the rotameter block output to

    supply the pinomatic support.

    05Install Vaporizer 1415 on the pinomatic support. Check whether

    the vaporizer is firmly installed onthe support.

    06

    Open the bubbling chamber removing the funnels screw-on lidand fill the bubbling chamber with the desired anesthetic agent.

    The level must always be between 10 and 100 mlduring theanesthesia procedure.

    07Firmly close the funnel with its lid on.

    08Set-up the chamber placing the agents indicator pin to its correct

    position.

  • 8/14/2019 204010170

    11/48Cdigo do manual: 204010170_004 11

    09Install the bubbling chamber on the Mini-Pinomatic support (right-

    side of the vaporizer). Check whether the installation is firm inorder for there to be

    no leakage.

    10 Switch the ventilator on. During start up, the display shows thefirmware release that is installed.

    11The display will show a bubbling flow concentration

    correspondence table, which must be used to adjust the desiredanesthetic agent concentration.

    12Using the DISPLAY key, one can browse through other

    correspondence tables.

    13Using the AGENT key, one may check the anesthetic agent

    being used.

    14Continue reading this manual to achieve full understanding of the

    Vaporizers operation.None

  • 8/14/2019 204010170

    12/48Cdigo do manual: 204010170_004 12

    3 GENERAL DESCRIPTION

    Vaporizer 1415 was designed to integrate the K. TAKAOKAline of Anesthesia Devices. It is a rotameter-controlled universal vaporizer

    designed to use the HALOTHANE, ENFLURANE, ISOFLURANE or SEVOFLURANE agents. Vaporizer 1415 is microprocessed and has

    important features that render it an easy-to-use, versatile device.

    TOTAL GAS FLOWis a mixture of oxygen, nitrous oxide or compressed air, regulated at the anesthesia devices rotameter block. This

    flow enters in the Multiagent Vaporizer, where it is divided in to two parts:

    1 - BUBBLING FLOW small parcel of the total flow that is controlled by the flow control button on Vaporizer 1415 and is measured by

    the respective rotameter. The flow crosses the bubbling chamber, becoming saturated with anesthetic agent steam.

    2 - DILUENT FLOW another part of the total flow that goes directly to Vaporizer 1415s output and then mixes with the bubbling flow

    that is saturated with anesthetic.

    The continual flow of fresh gases that leaves Vaporizer 1415 is, therefore, constituted by the total gas flow with a certain concentration of

    the volatile anesthetic agent steam.

    Main Features

    High luminosity liquid crystal display to view the concentration versus bubbling flow/anesthetic agent flow ratio table set-up in the

    chamber (indicated by the pin).

    The required bubbling flow calculation function dispenses the use of the calculation rule to determine the desired concentration,

    making it even easier to use the vaporizer.

    The Dilumaticsafety system blocks vaporization and the supply of anesthetic agents unless the total flow is open at the rotameter.

    The exclusive Pinomaticsystem, that allows for easy interchange between Multiagent Vaporizers and K. TAKAOKACalibrated

    Vaporizers, through a quick connection system.

    Two rotameters for the bubbling flow with differentiated scales for low (5 to 350 ml) and high (400 to 2200 ml) flows; This

    characteristic makes Vaporizer 1415 adequate for any anesthetic agent, allowing the operator to view the bubbling inside it.

    Amber glass bubbling chamber to protect the anesthetic agent from degradation due to sun light; however, allowing the operator to

    view the bubbling inside it. Capacity for 100 ml of anesthetic agent.

    Automatic recognition of the anesthetic agent indicator pins position on the chamber.

    Bubbling flow control button protected against accidental activation with a 360oprotection rim.

    Internal sensors for automatic total flow measurement.

    Internal one-way valves that ensure constant concentrations even when using controlled ventilation, avoiding an increase in

    concentration due to the pumping effect.

    Housed in a steel plate case coated with highly resistant electrostatic powder painting. This box performs as an extra heat exchange

    surface, increasing the anesthetics thermal stability.

    During a lack of power emergency, Vaporizer 1415 can continue being used without its electronic functions.

    12-VDC power supply. The 12-VDC cable comes with the equipment for connection.

  • 8/14/2019 204010170

    13/48Cdigo do manual: 204010170_004 13

    4 TECHNICAL SPECIFICATION

    Classification

    IEC-60601-1:1988+amendment 1994 (Medical electrical equipment - Part 1: General requirements for

    basic safety and essential performance)

    IEC 60601-2-13/2003- (Medical electrical equipment - Part 2-13: Particular requirements for the safety and

    essential performance of anaesthetic systems) Equipment Class 1 Energized Internally

    Type B IPX1 Continuos Operation

    Anesthetic Agents

    HALOTHANE

    ENFLURANE

    ISOFLURANE

    SEVOFLURANE

    The anesthetic agent must be indicated in the Mini-Pinomatic bubbling chamber through the indicator pin. The pins position is read

    electronically to determine the proper CONCENTRATION x BUBBLING FLOW table calculation.

    MINI-PINOMATIC Bubbling Chamber

    Anesthetic agent tank . ........................................................................................................................100 ml

    Scale....................................................................................................................................................100 ml

    Minimum grading..................................................................................................................................2.5 ml

    Precision...............................................................................................................................................1.0 ml

    Minimum work level...............................................................................................................................10 ml

    Maximum work level............................................................................................................................100 ml

    Bubbling Rotameter

    Low flow rotameter...............................................................................................scale from 5 to 350 ml/min

    High flow rotameter...............................................................................................scale from 0.4 to 2.2 l/min

    Calibrated for 760 mmHg and 20oC (not compensating the pressure).

    Reading is taken in the center of the sphere.

    Safety Device

    Dilumaticprotection system against the absence of total flow.

    Flow control with a 360oprotection rim against accidental activation.

    Graphic Display

    Type of display .....................................liquid crystal with back lighting.

    Two 24-character 5 x 7-point lines.

    Bubbling Flow Calculation

    The table shows the concentration versus necessary bubbling flow.

    Concentration ............................... .................. ...................... ..................... ......... .... 0.5% to 9.0%, resolution of 0.5%.

    Bubbling flow ..................... ...................... ..................... ............................5 to 100 ml/min, resolution of 5 ml/min.

    ... 100 ml/min to 1.0 l/min, resolution of 50 ml/min.

    1.0 l/min to 2.0 l/min, resolution of 0.1 l/min.

    Total Flow Reading

    Total flow ........................................................0.5 to 15 l/min, resolution of 0.1 l/min. Precision of 0.2 l/min.

  • 8/14/2019 204010170

    14/48Cdigo do manual: 204010170_004 14

    Serial Outlet for Microcomputers

    Type.......................................................................................................................................RS 232C serial port, 9 pins, male, isolated

    Data transmission rate ................... ..................... ..................... ..................... ...... ................................................19.200 bps

    Data format ..................... ...................... ..................... ..................... .................... .........................................................1N81

    Electrical Features

    Supply...................................................................................................................................................12Vdc

    Vaporizer power consumption ........................................................................................ 6 VA (maximum)

    Notes:The hospital's power system should be properly grounded and in accordance with the ABNT NBR 13534 regulation - Instalaes eltricas

    em estabelecimentos assistenciais de sade - Requisitos de segurana. Failure to comply with this recommendation can result in injury to

    the patient or operator or equipment damage, and it voids the Vaporizer's warranty.

    Dimensions and Weight

    Vaporizer

    Height 320 mm

    Width 226 mm

    Depth 190 mm

    Weight 6.7 Kgf

  • 8/14/2019 204010170

    15/48

  • 8/14/2019 204010170

    16/48Cdigo do manual: 204010170_004 16

    Figure 5.1 Vaporizer 1415 Operation Principle

    D F

    B F

    T F

    B F V F

    T F

    TF = Total FlowBF = Bubbling FlowDF = Dilute FlowVF = Vapor Flow

    V F

  • 8/14/2019 204010170

    17/48Cdigo do manual: 204010170_004 17

    6 CONTROLS AND COMPONENTS

    6.1 List of Components

    The components listed below are supplied with Vaporizer 1415:

    Item Code Description

    1 204010170 Operation Manual

    2 202010813 Mini-Pinomatic torn off Bubbling Chamber

    Notes:When you receive the device, if any of the components above is missing or damaged, contact the K. TAKAOKAS authorized

    representative immediately, as there are different warranty periods for the various components. See Chapter 14.

    To purchase optional or replacement parts, contact the authorized K.TAKAOKAdistributor.

    6.2 Front View

    The items below refer to Ventilator 1415s front view (Figure 6-1).

    1 - AGENT Key

    When depressed, this key shows the anesthetic agent indicated by the Mini-Pinomatic bubbling chamber indicator pin.

    WarningThe electronic recognition system reads the agent indicator pin position and DOES NOT IDENTIFY THE ANESTHETIC

    AGENT.

    2 - Liquid Crystal Display Manual

    This display shows the data calculated in form of a CONCENTRATION x BUBBLING FLOW table according to the following identified

    parameters:

    Position of the anesthetic agent indicator pin;Total flow originating from the rotameter.3 - High Flow Rotameter

    Rotameter that measures the bubbling flow as adjusted by the control (6). This rotameter should be used to read flows between 0.4 and

    2.2 l/min. Reading is taken in the middle of the sphere.

    4 - Low Flow Rotameter

    Rotameter that measures the bubbling flow as adjusted by the control (6). This rotameter should be used to read flows between 5 and 350

    ml/min. Reading is taken in the middle of the sphere.

    WarningDo not add the readings from both rotameters (3) and (4), since both indicate the same flow. Use the rotameter suited for

    the flow range that is being measured low or high.

    5 - Bubbling Chamber Support for the MINI-PINOMATIC

    System

    Bubbling Chamber support with a flow input orifice and a flow output orifice, to be coupled to the bubbling chambers upper base, by this

    constituting Vaporizer 1415s MINI-PINOMATICquick coupling system.

    6 - Mini-Pinomatic Bubbling Chamber

    The bubbling chamber has a quick coupling system to allow for anesthetic agent exchanges during the patients anesthesia procedure (for

    example: induction with SEVOFLURANE and maintenance with ISOFLURANE). See more details in item 6.4.

    7 - Bubbling Flux Control

    Button that controls bubbling flow in Vaporizer 1415, which corresponds to the parcel of the total flow that crosses the Mini-Pinomatic

    bubbling chamber (5). The value of this flow will be read in the respective flowmeters (3) and (4). I t opens counterclockwise.

  • 8/14/2019 204010170

    18/48Cdigo do manual: 204010170_004 18

    WarningThe bubbling flow determines the actual concentration Vaporizer 1415 provides directly. Therefore, it is important that

    the flow this button adjusts read at rotameter (3) and (4) equals the bubbling flow required for the desired

    concentration, as shown by the liquid crystal display (2).

    8 - On/Off Key

    Key that turns Vaporizer 1415s electric part on or off.

    AttentionThis key only influences the vaporizers electronic functions (liquid crystal display and control electronics). The anesthetic

    gas supply system functions independently of the electronic part, and in an emergency it can work even without electronic

    supply.

    9 - DISPLAY Key

    When this key is depressed, it presents the next concentration versus bubbling flow sequence.

    The process is cyclical. After the last screen, the cycle goes back to showing the 0.5% to 3% table.

    Figure 6-1: Front View of Vaporizer 1415

  • 8/14/2019 204010170

    19/48Cdigo do manual: 204010170_004 19

    6.3 Rear View

    The following items refer to Vaporizer 1415s rear view (Figure 6-2).

    1 - Quick Coupling for the PINOMATIC System

    System with a flow input orifice, a flow output orifice, a guiding pin, two electrical contacts and a grounding contact to be coupled to

    the anesthesia devices PINOMATICquick coupling system.

    2 - RS-232-C Connector

    Serial communication connector between the Vaporizer and the PC, used to update the equipment control and diagnosis programdynamically. The serial cable and the PC are not supplied with the Vaporizer. They are optional accessories. The serial cables

    specifications are detailed in Chapter 4.

    3 - Identification Tag

    Tag showing Vaporizer 1415s model and serial number for identification purposes.

    Figure 6-2: Rear View of Vaporizer 1415

    6.4 Mini-Pinomatic Bubbling Chamber

    The following items refer to the Vaporizer 1415s Mini-Pinomatic Bubbling Chamber (Figure 6-3).

    1 - Bubbling Chamber

    The bubbling chamber has a quick coupling system to allow for anesthetic agent exchanges during the patients anesthesia procedure (for

    example: induction with SEVOFLURANE and maintenance with ISOFLURANE).

    Transparent amber glass chamber used to receive the liquid anesthetic agent and where vaporizing occurs. It has volume grading (in ml),

    for the liquid level in it to be read.

  • 8/14/2019 204010170

    20/48Cdigo do manual: 204010170_004 20

    AttentionThe agent level inside the chamber must be between 10 and 100 ml during the anesthesia procedure.

    2 - Quick Coupling for the Mini-Pinomatic

    System

    System with a flow input orifice and a flow output orifice to be coupled to Vaporizer 1415s lateral support.

    3 - Anesthetic Agent Indicator Pin

    A special pin recognized electronically by Vaporizer 1415, which, according to the position of the insertion in the upper part of the

    Mini-Pinomatic bubbling chamber indicates which anesthetic agent is in it.

    WarningThe electronic recognition system reads the agent indicator pin position and DOES NOT IDENTIFY THE ANESTHETIC

    AGENT PRESENT IN THE BUBBLING CHAMBER.

    THE PIN IS SPECIAL. If it is lost, request replacement from your TAKAOKA authorized representative immediately.

    Firmly screw-on the agent indicator pin (clockwise).

    4 - Anesthetic agent identification tag

    This is a tag that allows one to insert and screw-on the anesthetic agent indicator pin correctly. The tag follows anesthetic agent color

    identification.

    HALOTHANE

    ISOFLURANE

    SEVOFLURANE

    ENFLURANE

    5 - Funnel Lid

    Screw-on lid to close the bubbling chamber. The lid is placed onto its position by turning it clockwise.

    WarningThe lid must be firmly closed during the anesthesia procedure to prevent any gas leakage. Make sure the funnels

    plastic washer seal is in good condition to prevent gas leakage (Figure 9-1).

    6 - Bubbling Chamber Filling Funnel

    Funnel used to drain and fill the bubbling chamber with anesthetic agents.

    7 - Product Identification Tag

    Tag showing bubbling chamber model information.

  • 8/14/2019 204010170

    21/48Cdigo do manual: 204010170_004 21

    Figure 6-3: Mini-Pinomatic Bubbling Chamber

    1

    2

    3

    4

    5

    6

    7

  • 8/14/2019 204010170

    22/48Cdigo do manual: 204010170_004 22

    7 DISPLAY

    Vaporizer 1415s front panel has a display that shows important alphanumeric information to regulate Vaporizer 1415 and for better

    anesthesia control. The microprocessor performs the required bubbling flow calculation, dispensing the use of a calculation rule to

    determine the desired concentration.

    The most important information the computerized display shows is the anesthetic agent concentrationversus flow bubblingrelation

    table. These required bubbling flow values appear below the concentration values shown on the display.

    The microprocessor calculates the display information theoretically, based on values measured by internal sensors and on data supplied

    by the operator, according to the following table:

    Parameter used for the calculations Value measured by the internal sensor Value adopted by the operator

    Desired concentration X

    Type of agent X

    Total Flow X

    7.1 Screens

    7.1.1 Start-up

    As soon as Vaporizer 1415 is turned on, its computerized display shows a system start-up screen (Figure 7-1). This screen has

    information on the anesthetic agent being used and on the software release number. The latter is important for maintenance purposes.

    After a short period of t ime, the display will progress automatically to the first concentration versus bubbling flow table screen.

    Figure 7-1: Start-up Screen

    7.1.2 Concentration x bubbling flow relation tables

    This is the most important information displayed, and these are the required bubbling flows shown under the values of the different

    concentrations. These values guide the operator on the bubbling flow that must be opened manually on Vaporizer 1415 for the respective

    concentrations to be reached. The bubbling flow is expressed in ml/minute.

    Vaporizer 1415 allows anesthetic agent concentrations ranging from 0.5 to 9.0% to be adjusted in the mixture. These concentrations are

    divided into three screens:

    0.5 to 3.0% (Figure 7-2) 3.5 to 6.0% (Figure 7-3) 6.5 to 9.0 % (Figure 7-4)

    To go through the three possible screens, press the DISPLAY key repeatedly. The process is cyclical. After the last screen, the cycle goes

    back to showing the 0.5% to 3.0% table.

  • 8/14/2019 204010170

    23/48Cdigo do manual: 204010170_004 23

    Figure 7-2: First Screen: CONCENTRATION from 0.5% to 3.0%

    Figure 7-3: Second Screen: CONCENTRATION from 3.5% to 6.0%

    Figure 7-4: Third Screen CONCENTRATION of 6.5 to 9.0%

    7.1.3 Agent

    Another important piece of information the display shows is the type of anesthetic agent that is in the bubbling chamber (Figure 7-5). The

    system recognizes this automatically according to the ANESTHETIC AGENT INDICATOR PINposition.To access the type of anesthetic

    agent, press the AGENT key.

    Figure 7-5: ANESTHETIC AGENT presentation screen

  • 8/14/2019 204010170

    24/48Cdigo do manual: 204010170_004 24

    7.2 Messages

    The display may show two messages:

    7.2.1 NO DILUENT FLOW

    This means there is no TOTAL FLOWregulated by the Rotameter; therefore, the Dilumatic safety system will stop the vaporization. This

    message will be shown at the bubbling flow site (Figure 7-6).

    Figure 7-6: NO DILUENT FLOW message

    7.2.2 VAPORIZER WITHOUT CHAMBER!!!

    This means that the Mini-Pinomatic Bubbling Chamber or ANESTHETIC AGENT INDICATOR PIN are not positioned precisely on theirrespective housing areas, therefore, the microprocessed internal safety system will block vaporization. This message will be shown on the

    central portion of the display (Figure 7-7).

    Figure 7-7: Message: VAPORIZER WITHOUT CHAMBER!!!

  • 8/14/2019 204010170

    25/48Cdigo do manual: 204010170_004 25

    8 ASSEMBLING AND PREPARING THE VAPORIZER

    When you receive Vaporizer 1415, follow steps 1 to 3 in item 2, Quick Installation Guide carefully.

    More information will be provided on mounting and preparing Vaporizer 1415 on the Anesthesia Device Cabinet using the PINOMATIC

    support below.

    8.1 Assembling the Anesthesia Machine

    The exclusive PINOMATIC system allows easy interchanges between the Multiagent Vaporizers and the different K. TAKAOKACalibrated Vaporizers through its quick coupling connection (Figure 8-1). The PINOMATICSupport components are listed below for a

    better understanding of their operation.

    Figure 8-1: PINOMATICSupport

    1 Guide Pin Orifice

    Orifice for Vaporizer 1415s connection block guide pin.

    2 Power Contacts

    Two power contact points with Vaporizer 1415s connection block. The electrical current passes from the Pinomatic Support to the

    Vaporizer through these contacts.

    3 Vaporizer Inlet Connection

    Connection nozzle for total flow entry in Vaporizer 1415. This nozzle has a floating system and sealing rings for perfect coupling.

    AttentionMake sure all Pinomatic

    support connection nozzles are clean and its rubber disks are in perfect condition.

    4 Vaporizer Outlet Connection

    Connection nozzle for total flow output in Vaporizer 1415. This nozzle has a floating system and sealing rings for perfect coupling.

    5 Fuse

    Fuse to protect Rotameter 1415s electrical part. The fuse compartment cover is threaded and can be easily opened for

    replacement purposes.

    6 12-Volt Source Input

    12-VDC inlet to supply Vaporizer 1415s electronic components. This inlet must be connected to the Rotameter Block 12-VDC outlet

    through the appropriate cable that is provided with the equipment.

    2

    7

  • 8/14/2019 204010170

    26/48Cdigo do manual: 204010170_004 26

    7 Ground Contact

    A ground contact with Vaporizer 1415s connection block. This contact grounds Vaporizer 1415.

    Mounting the PINOMATICSupport on the anesthesia devices rule must be done as follows:

    1. Attach the PINOMATIC support to the anesthesia machines ruler, as described in Figure 8-2. Use a -inch Phillips or ring

    screwdriver (approximately 13 mm) to fasten the four screws.

    2.

    Figure 8-2: PINOMATICsupport fixture

    3. Couple the desired vaporizer on the Anesthesia Devices PINOMATICsupport simply by gently placing it on its position, lowering it

    from top to bottom (Figure 8-3). If no vaporizer is coupled, connect the rigid joining tube (that comes with Vaporizer 1415),

    interconnecting the Supports two upper nozzles (Figure 8-4).

    Figure 8-3: Mounting Vaporizer 1415 on the PINOMATICsupport

    PINOMATIC

    Support

    AttachingRuler

    Screw

    Housing

    PINOMATIC Support

    AttachingRuler

    Vaporizer1415

    PINOMATIC quickconnection system

  • 8/14/2019 204010170

    27/48Cdigo do manual: 204010170_004 27

    Figure 8-4: Rigid tube

    8.2 Connections

    Mounting the PINOMATICSupport connections on the Anesthesia Devices Cabinet must be done as follows:

    1 - Firmly interconnect the Rotameters output to the PINOMATICsupports lower INPUT connection (Figure 8-5), making sure to

    avoid possible gas leakage or flow obstruction.

    2 - Firmly interconnect the PINOMATIC supports OUTPUT to the respective input connection on the Anesthesia Device base or on

    the Rotameter, depending on the equipment model (Figure 8-5), making sure to avoid possible gas leakage or flow obstructions.

    Figure 8-5: PINOMATICsupport connections

    WarningThe two lower PINOMATICsupport connections must be interconnected to the Rotameter or to the Anesthesia Device

    by means of two distinct tubes (a braided and a transparent one) preventing accidental inversions. These tubes are

    supplied with the equipment.

    8.3 Power Supply

    1 - Initially, check if the on/off key (7) located on Vaporizer 1415s side panel is on the OFFposition.

    2 - Connect an end of the 12-VDC cable (Figure 8-6) to the Rotameter Block 12-VDC output.

    Figure 8-6: 12-VDC Cable.

    3 - Connect the other end of the 12-VDC cable to the 12-VDC input (Figure 8-7) on the right side of the PINOMATICSupport.

    Gases enteringPINOMATIC

    Support

    Gases thatleave

    Vaporizer 1415

  • 8/14/2019 204010170

    28/48Cdigo do manual: 204010170_004 28

    Figure 8-7: PINOMATICSupport 12-VCD input connection.

    PINOMATIC

    Support 12-VCD inputconnection

  • 8/14/2019 204010170

    29/48Cdigo do manual: 204010170_004 29

    WarningFailure to comply with these electrical recommendations may result in injury to the patient, operator or equipment and

    will void Rotameter 1415s warranty.

    8.4 Mounting the MINI-PINOMATIC Bubbling Chamber

    1 - Mount the bubbling chamber on Vaporizer 1415s MINI-PINOMATICsupport simply by gently placing it on its position from top to

    bottom (Figure 8-8).

    Figure 8-8: Mounting Vaporizer 1415s Bubbling Chamber.

  • 8/14/2019 204010170

    30/48Cdigo do manual: 204010170_004 30

    9 OPERATION

    Before using the device, follow the procedure below to make sure Vaporizer 1415 is operating properly.

    AttentionBefore starting any procedure, make sure that the control button for the bubbling flow is fully closed (button fully

    turned counter-clockwise).

    Notes:The figures between parentheses refer to Vaporizer 1415s front view (fv) and to the bubbling chambers (bc) numbering.

    9.1 Initial Procedures

    1 - Check that the total gas flow of the anesthesia devices Rotameter is initially closed.

    2 - Make sure the PINOMATICsupport and its interconnections and tubes are properly and firmly attached. Always be alert to the

    possibility of gas leakage or flow obstruction.

    3 - Feed Multiagent Vaporizer 1415 with the correct electric current, as provided for in the instructions.

    4 - Turn Vaporizer 1415s electric components on using the on/off key (7-fv), and check if the computerized display turns on (2-fv).

    5 - Check whether the bubbling chamber (5-fv) is totally dry or, if the chamber already has an anesthetic agent in it, check what type

    of agent it is and if it does not show signs of aging.

    WarningIf you are in doubt about the type of anesthetic agent that is in the bubbling chamber, empty (discard the agent), dry,

    and fill the chamber with the proper agent.

    6 - Fill the bubbling chamber (5-fv) according to instructions in item 9.2.

    7 - Identify the anesthetic agent present in the bubbling chamber correctly (5-fv) through the anesthetic agent identification pin (3-bc)

    on the upper part of the chamber.

    8 - Check whether the anesthetic agent level in the bubbling chamber (5-fv) is appropriate for the foreseen anesthesia time (the level

    must always be between 10 and 100 ml).9 - Check whether the funnel lid is tightly screwed on the bubbling chamber (5-bc).

    10 - Check whether the Mini-Pinomatic chamber (5-fv) mounting is correct and firm on its respective support (9-fv) on the right side of

    Vaporizer 1415 Plus.

    9.2 Filling the Chamber

    For initial bubbling chamber filling or refilling with the liquid anesthetic agent, follow this procedure:

    1 - Close the bubbling flow control (6-fv).

    2 - Remove the Mini-Pinomatic bubbling chamber (5-fv) from its support (9-fv) on the right side of Vaporizer 1415.

    3 - Remove the funnel threaded cover (5-bc).

    4 - Fill the bubbling chamber (5-fv) or complete its level with the proper anesthesia agent. The level must be between 10 and 100 ml.5 - Check whether the plastic sealing washer located on the lid of the funnel (5-bc) is in perfect condition (Figure 9-1) to avoid gas

    leakage during the anesthesia procedure.

    6 - Firmly close the funnel with its lid (5-bc).

    7 - Check whether the Mini-Pinomatic chamber (5-fv) mounting is correct and firm on its respective support (9-fv) on the right side of

    Vaporizer 1415.

  • 8/14/2019 204010170

    31/48Cdigo do manual: 204010170_004 31

    WarningMake sure that the bubbling chamber is filled with the correct agent. Identify the Mini-Pinomatic chambers content

    through the indicator pin located on its upper part.

    The funnel lid must be tightly closed so there is no gas leakage.

    Never exceed the maximum chamber capacity, i.e., 100 ml.

    Figure 9-1: Location of the filling funnel lid-sealing washer.

    9.3 Vaporization Instructions

    After completing the initial procedures and when both patient and equipment are ready to start the anesthesia procedure, follow the steps

    described below:

    1 - Turn Vaporizer 1415 on using its ON/OFF key (7-fv), located on the right side of the vaporizer.

    2 - Depress the AGENT key (1-fv) and make sure the agent that is shown is the same as the one selected in the Mini-Pinomatic

    bubbling chamber through its indicator pin.

    3 - Through the DISPLAY key (8-fv), select the desired CONCENTRATION versus BUBBLING FLOW table.

    WarningWhen there is a divergence between the indicator pin and the agent shown on the display, dont use the

    CONCENTRATION versus BUBBLING FLOW table as the values that are shown will not match the agent in the bubbling

    chamber.

    4 - In the anesthesia devices Rotameter, regulate the total gas flow to be used.

    5 - Set the bubbling flow button (6-fv) according to the desired concentration.

    WarningChanges made to the total fresh gas flow or to the anesthetic agents temperature inside the bubbling chamber will cause

    changes to the concentration value that is being supplied. In this case, conveniently readjust the bubbling flow according to

    the value shown in the CONCENTRATION versus BUBBLING FLOW table.

    6 - The anesthetic agent level in the bubbling chamber (5-fv) can be filled during anesthesia, according to the CHAMBER FILLING

    procedure described above.

    7 - To stop the vaporization, completely close the bubbling flow control (6-fv).

    8 - Completely drain and dry the bubbling chamber (5-fv) for better Mini-Pinomatic chamber conservation (Item 9.4).

    9.4 Chamber Draning

    To drain the anesthetic liquid in the bubbling chamber totally or partially, follow the procedure below:

    AttentionTo collect the anesthetic agent, only use a flask destined specifically for the agent that is being discarded. The flask

    must also be marked with the name of this agent.

    1 - Close the bubbling flow control button (6-fv).

    2 - Remove the Mini-Pinomatic bubbling chamber (5-fv) from its support (9-fv) on the right side of the Vaporizer.

    3 - Remove the funnel lid (5-fv).

    Filling

    Funnel

  • 8/14/2019 204010170

    32/48Cdigo do manual: 204010170_004 32

    4 - Slowly pour the chambers content into the flask of origin.

    5 - Maintaining its lid off (5-fv), couple the bubbling chamber (5-fv) to the support (9-fv).

    6 - Keeping the bubbling flux control button (6-fv) open, allow it to drain for a few minutes at a high total flow rate through Vaporizer

    1415, so it will dry completely. Use a flow value between 5 and 10 l/min open in the Anesthesia Device Rotameter Block.

    7 - Once the drying procedure has been concluded, close the total gas flow and the bubbling flow control button (6-fv).

    8 - Firmly place the funnel with its lid on it (5-fv).

    9 - Mount the Mini-Pinomatic bubbling chamber (5-fv) on its respective support (9-fv) on the right side of Vaporizer 1415.

  • 8/14/2019 204010170

    33/48Cdigo do manual: 204010170_004 33

    10 CLEANING AND STERILIZATION

    Switch Ventilator 1415 off before cleaning it. Do not immerse the Vaporizer in any kind of liquid.

    1. To clean the external parts of Ventilator 1415, use a clean, soft cloth (or sponge) moistened with isopropyl alcohol, soap and water,

    or a suitable germicidal solution. Be careful to avoid the accumulation of any cleaning product residue in the equipment connections.

    After the cleaning, rinse (not required for alcohol) and dry it with a clean, dry and soft cloth.

    2. To clean the Vaporizers screen, use a soft and clean, lint-free clothe. Do not use paper towels or rough cloths to avoid scratching thescreen surface.

    3. The Vaporizers bubbling chamber must be kept totally empty and dry while the Anesthesia Device is not being used to keep the

    bubbling gasket orifices from getting clogged (item 9.4).

    Do not use abrasive or corrosive agents when cleaning.

    RECOMMENDATIONS FOR PROCESSING K. TAKAOKA MECHANICAL VENTILATION AND ANESTHESIA

    EQUIPMENT COMPONENTS

    The hospital accessories used in gas anesthesia and mechanical ventilation are classified as semi-critical, due to the potential risk they

    pose for transmitting infection. Semi-critical articles are all those that come into contact with mucous membranes and are capable ofpreventing invasion of the sub-epithelial tissue and which require high-level disinfection or sterilization to ensure quality in multiple-use

    situations.

    Choosing the processing method - disinfection or sterilization - depends on the nature of the materials. The TABLEshows the methods

    recommended to process K. TAKAOKAanesthesia and mechanical ventilation equipment components, considering their composition and

    technical specifications. The recommended methods are: cleaning, chemical disinfection, and gas or chemical sterilization. The processing

    should be performed according to the step sequence shown in the following FLOWCHART.

    CLEANING A process that removes soil and organic matter from any surface or object. Cleaning is done by mechanical friction,

    immersion, cleaning machines and ultra-sound machines. It is the most important step in decontamination; all items must be washed

    before passing through any disinfection or sterilization process. No object should be sterilized if there is any organic matter (oil, fat,

    blood...) on it. Cleaning should always be done with soap and water. When the immersion method is used, preferentially use an enzyme

    detergent. Enzyme detergents have specific action on organic matter: they degrade and dissolve it in a matter of minutes. The objects

    should remain immersed for 5 minutes.

    DISINFECTION A thermal or chemical process that eliminates all microorganisms, except for those with spores. There are three

    disinfection categories: high, medium and low level.

    HIGH LEVEL DISINFECTION A process that eliminates all microorganisms except a large number of spores (bacteria, almost all fungi

    spores, TB bacilli, viruses) after an exposure time of 10 to 30 minutes. Ex.: Glutaraldehyde immersion.

    INTERMEDIATE LEVEL DISINFECTION A process that inactivates vegetative bacteria, fungi, and almost all viruses, except spores.

    Ex.: Mechanical friction with 70% Alcohol.

    LOW LEVEL DISINFECTION A process that inactivates the majority of bacteria, some fungi, some viruses, however does not affect

    more resistant bacteria like TB bacilli or spores. Used only for surfaces. Ex.: Water and detergent cleaning.

    THERMAL DISINFECTION A thermal process, which uses thermal disinfectant liquids against all plant forms, destroys some of the

    spores when used at a temperature of between 60 and 90OC. This process is done in a thermal-disinfection unit, the machine that works

    with two types of cycles for sensitive and resistant materials, using the appropriate detergent, one for sensitive materials and one for

    resistant materials.

    STERILIZATION A process that completely eliminates or destroys all microorganisms (spores, bacteria, fungi and protozoa); it is

    performed by physical processes (steam) or chemical processes (liquid Glutaraldehyde, gaseous - ethylene oxide gas and plasma -

    hydrogen peroxide). The spore is the most difficult microorganism to inactivate. Ex.: Autoclave, Hydrogen Peroxide, ethylene oxide,Glutaraldehyde (exposure of the material for 10 hours).

    Note:

  • 8/14/2019 204010170

    34/48Cdigo do manual: 204010170_004 34

    Hydrogen peroxide (oxygenated water) is a sterilization process that takes place at a maximum temperature of 45C. Materials

    that cannot be autoclaved can be sterilized with peroxide, except for those materials derived from cellulose.

    FLOWCHART

    IF PROCESSINGCONCLUDED

    IFDISINFECTION

    Flowchart of the sequential steps for processing of theAnesthesia and Mechanical Ventilation Equipment components

    DAMP CLOTH

    STORE

    STERILIZE

    PACK

    PHYSICAL MEDIUM(STEAM)

    STORE

    PACK INSTERILIZED FLASK

    DRY IN ANASEPTIC WAY

    RINSE WITHSTERILIZED H2O

    IMMERSERFULLY

    LQUID(GLUTARALDEHYDE)

    STORE

    STERILIZE

    PACK

    GASEOUS(ETHYLENE OXIDE)

    CHEMICALMEDIUM

    IFSTERILIZATION

    STORE

    STORE

    PACK

    DRY IN ANASEPTIC WAY

    IMMERSE FOR THE TIME AND ATTHE TEMPERATURE REQUIRED

    LIQUID PHYSICAL MEDIUMTHERMO DISINFECTION

    STORE

    PACK

    DRY IN ANASEPTIC WAY

    RINSE WITHSTERILIZED H2O

    FILLINTABULATIONS

    IMMERSE THE ARTICLE

    LIQUID CHEMICAL MEDIUMGLUTARALDEHYDE

    ACCORDING TO ARTICLE DESTINATION

    DRY

    RINSE

    DETERGENT SOLUTION

    CLEAN

    CONSIDER ALL CONTAMINATED

    WITH PRESENCE OF ORGANIC MATTER OR SOIL

    STORE

    STERILIZE

    PACK

    PLASMA(HYDROGEN PEROXIDE)

    STORE

    PACK

    MECHANICALFRICTION

    LIQUID CHEMICAL MEDIUM70% ALCOHOL

    FILLINTABULATION

  • 8/14/2019 204010170

    35/48Cdigo do manual: 204010170_004 35

    TABLE

    Methods recommended to process K. TAKAOKA Anesthesia and Mechanical Ventilation equipment components.

    Component Cleaning Disinfection Sterilization

    Sphygmomanometer bracket Detergent Solution NOT RECOMMENDED NOT RECOMMENDED

    Anti-pollution bag Detergent SolutionGlutaraldehyde

    Thermal disinfectionSensitive

    Ethylene OxideGlutaraldehyde or

    Hydrogen Peroxide

    Rotameter block Damp cloth 70% AlcoholSynthetic phenol

    NOT RECOMMENDED

    Sphygmomanometer cable Detergent Solution Synthetic phenol NOT RECOMMENDED

    EKG Cable / Oximeter Detergent Solution Synthetic phenol NOT RECOMMENDED

    Humidifier Chamber Detergent SolutionGlutaraldehyde

    Thermal disinfectionresistant

    Steam

    Bell Detergent SolutionGlutaraldehyde

    Thermal disinfectionSensitive

    Ethylene OxideGlutaraldehyde or

    Hydrogen Peroxide

    Canister Detergent SolutionGlutaraldehyde

    Thermal disinfectionSensitive

    Ethylene OxideGlutaraldehyde or

    Hydrogen Peroxide

    Oximeter Galvanic Cell Detergent Solution NOT RECOMMENDED NOT RECOMMENDED

    Bell Rod Set Detergent Solution Glutaraldehyde70% Alcohol

    Ethylene Oxide orGlutaraldehyde

    Air Vent Elbow Detergent Solution Glutaraldehyde70% Alcohol

    Ethylene Oxide orGlutaraldehyde

    Drains Detergent Solution

    Glutaraldehyde

    Thermal disinfectionresistant

    SteamHydrogen Peroxide

    Flow meter Damp cloth 70% AlcoholSynthetic phenol

    NOT RECOMMENDED

    Diaphragms Detergent SolutionGlutaraldehyde

    Thermal disinfectionresistant

    Steam or GlutaraldehydeHydrogen Peroxide

    Y-Shaped Connector (nozzle) Detergent SolutionGlutaraldehyde

    Thermal disinfectionsensitive

    Ethylene OxideGlutaraldehyde or

    Hydrogen Peroxide

    T-Shaped Capnograph Connector Detergent SolutionGlutaraldehyde

    Thermal disinfectionsensitive

    Ethylene OxideGlutaraldehyde or

    Hydrogen Peroxide

    Mask Detergent Solution Glutaraldehyde

    70% Alcohol

    Ethylene Oxide orGlutaraldehyde

    Monitor Module Damp cloth Synthetic phenol70% Alcohol

    NOT RECOMMENDED

    Aspirating Flask Detergent Solution Glutaraldehyde70% Alcohol Steam or Glutaraldehyde

    Ventilator Module 70% Alcohol Synthetic phenol70% Alcohol

    NOT RECOMMENDED

    Cabinet (external parts) 70% Alcohol Synthetic phenol70% Alcohol

    NOT RECOMMENDED

    Mask Strap Detergent Solution GlutaraldehydeEthylene Oxide

    Glutaraldehyde or

    Hydrogen Peroxide

  • 8/14/2019 204010170

    36/48Cdigo do manual: 204010170_004 36

    Component Cleaning Disinfection Sterilization

    Flow Sensor Detergent SolutionGlutaraldehyde

    Thermal disinfectionresistant

    Steam

    Auxiliary Thermometer Detergent Solution Synthetic phenol70% Alcohol

    Ethylene OxideHydrogen peroxide

    Injected Liquid Thermometer Detergent Solution 70% Alcohol Ethylene OxideHydrogen peroxide

    Esophageal Temperature Sensor Detergent Solution Glutaraldehyde Ethylene OxideHydrogen peroxide

    Side Stream (capnograph) Detergent Solution Glutaraldehyde Ethylene OxideHydrogen peroxide

    Canister Support Detergent Solution Glutaraldehyde70% Alcohol

    Ethylene Oxide orGlutaraldehyde

    Corrugated Tubes ** Detergent SolutionGlutaraldehyde

    Thermal disinfectionsensitive

    Ethylene Oxide orGlutaraldehyde

    Vacuum meter Detergent Solution Glutaraldehyde70% Alcohol

    NOT RECOMMENDED

    Inhalation and Exhalation Valves Detergent Solution GlutaraldehydeThermal disinfectionsensitive

    Ethylene OxideGlutaraldehyde orHydrogen Peroxide

    Calibrated Vaporizer Damp cloth Synthetic phenol70% Alcohol

    NOT RECOMMENDED

    Multi-agent Vaporizer Damp cloth Synthetic phenol70% Alcohol

    NOT RECOMMENDED

  • 8/14/2019 204010170

    37/48Cdigo do manual: 204010170_004 37

    11 MAINTENANCE

    Correct preventive Vaporizer 1415 maintenance will guarantee it can be used accurately and safely for an extended period of time.

    1 - Carry out a periodical visual inspection of Vaporizer 1415. Should there be any apparent damage, do not use the device.

    2 - At least once a month inspect the condition of the plastic washer placed in Vaporizer 1415s bubbling chamber funnel lid. If this

    washer is damaged, take steps to have it replaced to prevent gas leakages during anesthesia.

    3 - If it is not possible to switch Vaporizer 1415s electric components on normally, initially check:

    If there is power in the electric output that feeds the Rotameter Block;

    12-VDC cable integrity;

    4 - Periodically, check if the flowmeter spheres are moving freely along the entire length of their respective conical tubes and if they fall

    to zero when the flow control is closed. If this does not happen, call K. TAKAOKASauthorized Technical Assistance.

    AttentionOnly use genuine K. TAKAOKA replacement parts. Using parts that are not genuine may place the patients safety at

    risk.

    Do not perform any internal maintenance on the Vaporizer and do not open its case. For periodic service or to perform

    any maintenance on the Vaporizer, call K. TAKAOKAS authorized Technical Assistance.

  • 8/14/2019 204010170

    38/48Cdigo do manual: 204010170_004 38

    12 SYMBOLS

  • 8/14/2019 204010170

    39/48

  • 8/14/2019 204010170

    40/48Cdigo do manual: 204010170_004 40

  • 8/14/2019 204010170

    41/48Cdigo do manual: 204010170_004 41

  • 8/14/2019 204010170

    42/48Cdigo do manual: 204010170_004 42

  • 8/14/2019 204010170

    43/48Cdigo do manual: 204010170_004 43

    13 EMERGENCY ACTIONS

    In the event of an adverse event, K. TAKAOKA suggests the following procedure:

    Contact the manufacturer (Technical Assistance sector Manager) about the devices condition, and dont perform any test or

    investigation in the absence of an authorized K. TAKAOKA technician.

    Register the manufacturer, model, and serial number of all devices involved in the adverse event. Record this information in the

    patients records and/or in an uncommon event form. If the device is discardable or has discardable accessories, also record theclassification number and all of the discardable accessory numbers. It is important to keep the device and any accessory that was

    involved in the adverse event.

    Do not clean or submit the device to a chemical or physical process or fix it. These actions may affect its performance and safe use.

    Record the names of all health professionals who were present during the incident.

    Identify the device, indicating it was involved in an adverse event; record the event date, and the name of the person who tagged the

    device. Indicate on the tag that the device must not be used, cleaned, fixed or destroyed without the approval of an authority, such as

    a risk manager. If the adverse event involves more than one device, all devices involved must be tagged and stored.

    Preserve the packages of all discardable accessories involved in the event and store them with the device. The discardable

    accessory packages typically include not only the devices catalogued number, but also the lot number. Furthermore, a few

    specifications included on the package may be useful for the expert investigation.

    Before unplugging the device or removing its batteries, make sure its memory will not be lost. Many devices have computerized

    memories that may be lost if the batteries are removed or if the device is unplugged. Experts may use this memory to determine

    specifically when the conditions of the related device occurred. To determine which devices have computerized memories and how

    they must be controlled after an event, read the instruction manual or contact your clinical engineer.

    Put the device and its accessories in a safe place to prevent subsequent damage. This will keep the device from being put back into

    service; protected rooms and devices may have to be used regardless of a previous incident.

  • 8/14/2019 204010170

    44/48Cdigo do manual: 204010170_004 44

    14 WARRANTY

    K TAKAOKA IND.E COM. LTDA. guarantees the equipment it produces against manufacturing defects for a period of one year from the

    date of acquisition by the first owner. Other items provided with the equipment are listed below.

    The following is list showing the technical assistance companies authorized by K TAKAOKA IND. E COM. LTDA., both in Brazil and

    abroad, which, besides the manufacturer, have exclusive maintenance rights. Modification, violation, adjustment or maintenance by third

    parties is not permitted.

    The equipment manufactured or overhauled by K TAKAOKA IND. E COM. LTDA.has a breakable warranty seal. If this seal is broken,

    the warranty will become automatically void.

    The improper use of the equipment and/or the non-compliance with the instructions contained in this manual, the use of a voltage different

    from that specified and of parts and/or accessories which have not been approved by K TAKAOKA IND. E COM. LTDA.will void the

    warranty.

    Damage caused by accidents or Acts of God are not covered by the warranty; the same applies to batteries, fuses, filters, etc.

    Following are items provided with the equipment, as well as some optional accessories and their respective periods of warranty against

    "manufacturing defects".

    CODE ITEM PERIOD OF WARRANTY

    201030023 Vaporizer 1415 1 year

    204010170 Operation Manual No guarantee

    202010813 Separate Mini-Pinomatic Bubbling Chamber (optional) 1 year

    Multiagent Vaporizer 1415s useful life is estimated to average 5 years. This may vary according to use and to adequate preventive

    maintenance.

    Person Technically in Charge: Mauricio ChiarioniCREA: Registration number 5061714921

  • 8/14/2019 204010170

    45/48Cdigo do manual: 204010170_004 45

    K. TAKAOKA DISTRIBUTORS IN BRAZIL

  • 8/14/2019 204010170

    46/48Cdigo do manual: 204010170_004 46

    TECHNICAL ASSISTANCEK. TAKAOKA

    K. TAKAOKA Indstria e Comrcio Ltda. informs that only its Technical

    Service Centers are authorized to service the equipment it supplies.

    Service rendered by third parties pose serious risks once the origin of the

    parts employed is unknown and workmanship does not meet the stringent

    standards set by K. TAKAOKA

    We are not in a position to ensure the correct operation of the equipment we

    manufacture if it has been repaired by non-authorized third parties.

    Any requests regarding Technical Assistance Service and preventive

    maintenance, by contract or else, shall be addressed directly to K.

    TAKAOKA or to one of its exclusive and duly authorized representatives.

  • 8/14/2019 204010170

    47/48Cdigo do manual: 204010170_004 47

    AUTHORIZED K. TAKAOKA TECHNICAL ASSISTANCE

  • 8/14/2019 204010170

    48/48

    Information for Technical AssistanceThis card must be filled-out and returned together with the device.

    Name:

    Hospital:

    Address:

    Phone: District:

    ZIP: City: State-

    Information for Technical AssistanceThis card must be filled-out and returned together with the device.

    Name:

    Hospital:

    Address:

    Phone: District:

    ZIP: City: State-

    Defect description:

    Defect description:

    Rua General Izidoro Dias Lopes, 121/141 CEP 09687-100 So Bernardo do Campo- SP

    Tel.: (5511) 4176-3500 Fax.: (5511) 4176-3570

    Rua General Izidoro Dias Lopes, 121/141 CEP 09687-100 So Bernardo do Campo - SP

    Tel.: (5511) 4176-3500 Fax.: (5511) 4176-3570