A WIRELESS- IMPLANTABLE MICRO SYSTEM FOR CONTINUES BLOOD GLUCOSE MONITORING

ANOOP PRoll No:07Reg. No:CTAHEEC013

Page 2: A Wireless- Implantable Micro System

ABSTRACT

• Remotely powered implantable micro system • Consists of a micro fabricated glucose

biosensor flip-chip bonded to a transponder chip

• Transponder chip measures the output signal of the glucose biosensor and transmits to the external reader

• Chip can measure the sensor output current ranging from 1 nA to 1 A

Page 3: A Wireless- Implantable Micro System

LITERATURE SURVEY

1. NON-CONTINOUS BLOOD GLUCOSE MONITORING • Consist of a blood glucose meters• small drop of blood is needed to place on

test strip

Page 4: A Wireless- Implantable Micro System

LITERATURE SURVEY

2. Continues wired blood glucose monitoring• Used conventional electrochemical glucose

sensors• Large sizes• Wired

Page 5: A Wireless- Implantable Micro System

DESIGN

Page 6: A Wireless- Implantable Micro System

DESIGN

• Microfabricating the glucose sensor• Inegrating all the interfacing circuits into a

integral chip(TRANSPONDER CHIP)• Attaching the microfabricated glucose sensor

to the transponder chip using flip chip bonding

Page 7: A Wireless- Implantable Micro System

SYSTEM OVERVIEW AND FUNCTIONING

Page 8: A Wireless- Implantable Micro System

SYSTEM OVERVIEW AND FUNCTIONING

• Monitoring system consisting of an implantable microsystem and an external reader.

• Microsystem measures the blood glucose concentration and transmits the measured data to the external reader

Page 9: A Wireless- Implantable Micro System

GLUCOSE SENSOR

• Amperometric electrochemical biosensor• Consists of three electrodes: 1)a working electrode(WE) 2)a reference electrode(RE) 3)a counter electrode(CE)

Page 10: A Wireless- Implantable Micro System

ELECTRODES

1. WORKING ELECTRODE• Act as a surface on which the electrochemical

reaction takes place between the glucose and an enzymatic layer

• It catalyzes the oxidation of glucose to gluconic acid

Page 11: A Wireless- Implantable Micro System

ELECTRODES

2. REFERENCE ELECTRODE• The reference electrode is used to measure the

solution (blood or interstitial fluid) potential

Page 12: A Wireless- Implantable Micro System

ELECTRODES

3. COUNTER ELECTRODE• An inert conductor (gold) that supplies the current

required for the electrochemical reaction

Page 13: A Wireless- Implantable Micro System

TRANSPONDER CHIP

1. RF FRONT-END BLOCK

Page 14: A Wireless- Implantable Micro System

TRANSPONDER CHIP

2. DATA-ACQUISITION

Page 15: A Wireless- Implantable Micro System

EXTERNAL READER

Page 16: A Wireless- Implantable Micro System

ADVANTAGES

• Small size of the implant• No need of oxygen• Use of passive telemetry link

Page 17: A Wireless- Implantable Micro System

CONCLUSION

• Miniaturization procedure for a wireless-implantable microsystem for continuous blood glucose monitoring

• Low cost procedure• Microsystem implanted just underneath the

skin