Upload
others
View
15
Download
0
Embed Size (px)
Citation preview
Drawings
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.Subject to change without notice. Printed in Japan.
P -
ADMAG TI SeriesAXW Magnetic Flowmeter[Size: 500 to 1000 mm (20 to 40 in.)]Flange JIS 10K (Process Connection Code CJ1)
SD 01E25D11-05EN1st Edition: Apr. 2017 (KP)7th Edition: Dec. 2019 (KP)
SD 01E25D11-05EN
Unit: mm (approx. in.)
*1:This length becomes 21 mm (0.83 in.) shorter when indicator code N is selected.*2:The thickness of optional grounding rings (4mm per one) and customer supplied gaskets are not included in “L” here.
F01.ai
L*2
Ground Terminal
(M4)Eye Bolt
(H2
Hi
)
(H3)
48 (1.89)150 (5.91)
42 (1
.65)
42 (1
.65)
ø 86 (3.38)
H1 H
r
68 (2.68)
56 (2.20)*1
232 (9.13)*1
25 (0.98)48 (1.89)
83 (3.27)
ø128
(5.0
4)
111 (4.37)θ°
(ød)øC
øD
N-øh
IntegralFlowmeter
IntegralFlowmeter
RemoteSensor
RemoteSensor
Ground Terminal(M4)
t
Direction of Cable Entry
Standard(0°)
+90°rotation
+180°rotation
-90°rotation
Optional Code RA
Optional Code RB
Optional Code RC
Integral Flowmeter
Cable Entry Display Cable Entry Display
Remote Sensor
Front side Cable Entry Back side Cable Entry
* The direction of cable entry changes as shown left depending on the designation of the optional code RA, RB, or RC.
Unless otherwise specifi ed, difference in the dimensions are specifi ed as : General tolerance = ± (Criteria of tolerance class IT18 in JIS B0401-1) / 2
[Style: S2]
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.
P -
2
SD 01E25D11-05EN
Model code: AXW - - CJ1N -
Unit: mm (approx. in.)Process Connection Code CJ1Flange Type JIS B2220 10KSize Code 500 600 700 800 900 10LSize 500 (20) 600 (24) 700 (28) 800 (32) 900 (36) 1000 (40)Lining Code H, F, U, DLay Length L 600 (23.62) 720 (28.35) 840 (33.07) 960 (37.80) 1080 (42.52) 1200 (47.24)Flange Outer Diameter øD 675 (26.57) 795 (31.30) 905 (35.63) 1020 (40.16) 1120 (44.09) 1235 (48.62)Flange Thickness (*1)(incl. lining flare)
Lining F 35 (1.38) 37 (1.46) 39 (1.54) 41 (1.61) 43 (1.69) 45 (1.77)Lining H, D 33.5 (1.32) 35.5 (1.40) 37.5 (1.48) 39.5 (1.56) 41.5 (1.63) 43.5 (1.71)Lining U 36 (1.42) 38 (1.50) 40 (1.57) 42 (1.65) 44 (1.73) 46 (1.81)
Bolt Circle Diameter øC 620 (24.41) 730 (28.74) 840 (33.07) 950 (37.40) 1050 (41.34) 1160 (45.67)Inner Diameter ød Lining H, F, D 477 (18.78) 578 (22.76) 675 (26.57) 777 (30.59) 878 (34.57) 976 (38.43)
Lining U 469 (18.46) 570 (22.44) 667 (26.26) 769 (30.28) 870 (34.25) 968 (38.11)Bolt Hole Pitch Half Angle θ 9° 7.5° 7.5° 6.4° 6.4° 6.4°Bolt Hole Diameter øh 27 (1.06) 33 (1.30) 33 (1.30) 33 (1.30) 33 (1.30) 39 (1.54)Number of Bolt Holes N 20 24 24 28 28 28Height H1 735 (28.94) 855 (33.66) 965 (37.99) 1091 (42.95) 1191 (46.89) 1325 (52.17)Height H2 401 (15.80) 452 (17.80) 501 (19.73) 552 (21.74) 604 (23.77) 655 (25.78)Height Hr 792 (31.19) 903 (35.56) 1007 (39.65) 1116 (43.93) 1217 (47.92) 1326(52.19)Inner Diameter of Eye Bolt 35 (1.38) 35 (1.38) 35 (1.38) 40 (1.57) 40 (1.57) 50 (1.97)Remote Sensor, Approx. Weight, Unit: kg (lb) (*2) 156 (344) 223 (492) 310 (683) 408 (899) 522 (1151) 689 (1519)
Integral Flowmeter, Height H3 438 (17.24) 489 (19.26) 538 (21.20) 590 (23.21) 641 (25.24) 692 (27.25)Integral Flowmeter, Maximum Height Hi 839 (33.01) 950 (37.39) 1054 (41.49) 1163 (45.77) 1264 (49.77) 1373 (54.04)Integral Flowmeter, Approx. Weight, Unit: kg (lb) 159 (351) 226 (498) 313 (690) 411 (906) 525 (1157) 692 (1526)
*1: The tolerance of the flange thickness “t” is as follows. • Size 500 to 1000 mm (20 to 40 in.): +3/-2 mm (+0.12/-0.08 in.) *2: When submersible type or optional code DHC is selected, waterproof glands with union joints and cables are attached. When the cable length
is 30-meters, add 9.5 kg (20.9 lb) to the weight in the table.
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.
P -
3
SD 01E25D11-05EN
Terminal Configuration and WiringRemote Sensor:<To be wired to Remote Transmitter>
F02.ai
Flow Signal Output
Excitation Current Input
Description
Protective Grounding(Outside of the terminal box)
Terminal SymbolABC
EX1EX2
Note: When submersible use or optional code DHC is selected, waterproof glands with union joints and cables are attached.
Integral Flowmeter:<To be wired to Power Supply and I/Os>
F03.ai
M4 Screw Type
Clamp Type
Terminal Symbol
N/-L/+
I/O4 -I/O4 +I/O3 -I/O3 +I/O2 -I/O2 +I/O1 -I/O1 +
Selected Input/Output
Functional Grounding
Shorting Screw(Need to be fixed for normal operation)
Protective Grounding(Inside and outside of the terminal box)
Description
Power Supply
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.
P -
4
SD 01E25D11-05EN
Flange JIS 10K (Process Connection Code CJ1) [Style: S1] Unit: mm (approx. in.)
*1:This length becomes 21 mm (0.83 in.) shorter when indicator code N is selected.*2:The thickness of optional grounding rings (4mm per one) and customer supplied gaskets are not included in “L” here.
F04.ai
L*2
Ground Terminal
(M4)Eye Bolt(H
2
Hi
)
(H3)
48 (1.89)150 (5.91)
42 (1
.65)
42 (1
.65)
ø 86 (3.38)
H1 H
r
68 (2.68)
56 (2.20)*1
232 (9.13)*1
25 (0.98)48 (1.89)
83 (3.27)
ø128
(5.0
4)
111 (4.37)θ°
(ød)øC
øD
N-øh
IntegralFlowmeter
IntegralFlowmeter
RemoteSensor
RemoteSensor
Ground Terminal(M4)
t
Direction of Cable Entry
Standard(0°)
+90°rotation
+180°rotation
-90°rotation
Optional Code RA
Optional Code RB
Optional Code RC
Integral Flowmeter
Cable Entry Display Cable Entry Display
Remote Sensor
Front Side Cable Entry Back Side Cable Entry
* The direction of cable entry changes as shown left depending on the designation of the optional code RA, RB, or RC.
Unless otherwise specified, difference in the dimensions are specified as : General tolerance = ± (Criteria of tolerance class IT18 in JIS B0401-1) / 2
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.
P -
5
SD 01E25D11-05EN
Model code: AXW - - CJ1N -
Unit: mm (approx. in.)Process Connection Code CJ1Flange Type JIS B2220 10KSize Code 500 600 700 800 900 10LSize 500 (20) 600 (24) 700 (28) 800 (32) 900 (36) 1000 (40)Lining Code H, F, U, DLay Length L 600 (23.62) 720 (28.35) 840 (33.07) 960 (37.80) 1080 (42.52) 1200 (47.24)Flange Outer Diameter øD 675 (26.57) 795 (31.30) 905 (35.63) 1020 (40.16) 1120 (44.09) 1235 (48.62)Flange Thickness (*1)(incl. lining flare)
Lining F 35 (1.38) 37 (1.46) 39 (1.54) 41 (1.61) 43 (1.69) 45 (1.77)Lining H, D 33.5 (1.32) 35.5 (1.40) 37.5 (1.48) 39.5 (1.56) 41.5 (1.63) 43.5 (1.71)Lining U 36 (1.42) 38 (1.50) 40 (1.57) 42 (1.65) 44 (1.73) 46 (1.81)
Bolt Circle Diameter øC 620 (24.41) 730 (28.74) 840 (33.07) 950 (37.40) 1050 (41.34) 1160 (45.67)Inner Diameter ød Lining H, F, D 477 (18.78) 578 (22.76) 675 (26.57) 777 (30.59) 878 (34.57) 976 (38.43)
Lining U 469 (18.46) 570 (22.44) 667 (26.26) 769 (30.28) 870 (34.25) 968 (38.11)Bolt Hole Pitch Half Angle θ 9° 7.5° 7.5° 6.4° 6.4° 6.4°Bolt Hole Diameter øh 27 (1.06) 33 (1.30) 33 (1.30) 33 (1.30) 33 (1.30) 39 (1.54)Number of Bolt Holes N 20 24 24 28 28 28Height H1 735 (28.94) 855 (33.66) 965 (37.99) 1091 (42.95) 1191 (46.89) 1325 (52.17)Remote Sensor, Height H2 408 (16.06) 459 (18.07) 509 (20.04) 560 (22.05) 611 (24.06) 662 (26.06)Remote Sensor, Maximum Height Hr 800 (31.50) 911 (35.87) 1016 (40.00) 1124 (44.25) 1225 (48.23) 1334 (52.2)Inner Diameter of Eye Bolt 35 (1.38) 35 (1.38) 35 (1.38) 40 (1.57) 40 (1.57) 50 (1.97)Remote Sensor, Approx. Weight, Unit: kg (lb) (*2) 156 (344) 223 (492) 310 (683) 408 (899) 522 (1151) 689 (1519)
Integral Flowmeter, Height H3 438 (17.24) 489 (19.26) 538 (21.20) 590 (23.21) 641 (25.24) 692 (27.25)Integral Flowmeter, Maximum Height Hi 839 (33.01) 950 (37.39) 1054 (41.49) 1163 (45.77) 1264 (49.77) 1373 (54.04)Integral Flowmeter, Approx. Weight, Unit: kg (lb) 159 (351) 226 (498) 313 (690) 411 (906) 525 (1157) 692 (1526)
*1: The tolerance of the flange thickness “t” is as follows. • Size 500 to 1000 mm (20 to 40 in.): +3/-2 mm (+0.12/-0.08 in.) *2: When submersible type or optional code DHC is selected, waterproof glands with union joints and cables are attached. When the cable length
is 30-meters, add 9.5 kg (20.9 lb) to the weight in the table.
All Rights Reserved. Copyright © 2013, Yokogawa Electric Corporation.
P -
6
SD 01E25D11-05EN
Terminal Configuration and WiringRemote Sensor:<To be wired to Remote Transmitter>
F05.ai
Flow Signal Output
Excitation Current Input
Description
Protective Grounding(Outside of the terminal box)
Terminal SymbolABC
EX1EX2
Note: When submersible use or optional code DHC is selected, waterproof glands with union joints and cables are attached.
Integral Flowmeter:<To be wired to Power Supply and I/Os>
F06.ai
M4 Screw Type
Clamp Type
Terminal Symbol
N/-L/+
I/O4 -I/O4 +I/O3 -I/O3 +I/O2 -I/O2 +I/O1 -I/O1 +
Pulse/Status Output 1
Current Output 1
Selected Input/Output
Functional Grounding
Shorting Screw(Need to be fixed for normal operation)
Protective Grounding(Inside and outside of the terminal box)
Description
Power Supply