Upload
solomon-cobb
View
213
Download
0
Embed Size (px)
Citation preview
Advantages of a Syringe Driven Sample Introduction Accessory for ICP SpectrometryJerry Dulude and Scott Bridger
Glass Expansion
Glass Expansion www.geicp.com
Assist components
Niagara II Plus™ Rapid Rinse accessory Sample syringe plus drive Built in switching valve Built in peristaltic pump Optional internal standard syringe plus drive
Glass Expansion www.geicp.com
ICP-MS Carryover Comparison
0.00
2.00
4.00
6.00
8.00
10.00
12.00
14.00
Blank
1
Blank
2
Blank
3
Blank
4
Blank
5
Blank
6
pp
b B
oro
n
Standard
w ith Niagara
Boron
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
pp
b M
o
Standard
w ith Niagara
Moly
0.00
0.20
0.40
0.60
0.80
1.00
1.20
Blank1 Blank2 Blank3 Blank4 Blank5 Blank6
pp
b Z
n Standard
with Niagara
Zinc
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
Blank1 Blank2 Blank3 Blank4 Blank5 Blank6
pp
b T
lStandard
with Niagara
Thallium
Glass Expansion www.geicp.com
Niagara II Plus
Selectable sample loop Optional internal standard lines Positive displacement pump Multi-stage valve Sample does not contact pump tubing
Glass Expansion www.geicp.com
Standard operation Niagara II PlusCycle time Rinse time Cycle time Rinse time
Probe to sample 5 0 5 5Uptake delay 15 0 5 5Stabilization 10 0 10 0Read 15 0 15 0Probe to rinse 5 0 0 0Rinse 30 30 0 0TOTAL 80 30 35 10
Time Saving for Niagara II Plus
Glass Expansion www.geicp.com
Additional Assist Features
Precision syringe and drive for sample Optional precision syringe and drive for
internal standard
Glass Expansion www.geicp.com
Advantages of Syringe Sample Delivery Computer control of delivery speed Broad selection of sample/IS ratio High precision for scanning systems
Glass Expansion www.geicp.com
e.g. 20x dilution with peri pump @ 0.5ml/min uptake 0.38mm ID (orange/green) 1.75mm ID (blue/green) Pump speed=2rpm
Very Noisy Signal
Glass Expansion www.geicp.com
Assist Performance on a Elan9000 ICP-MS
Uptake rate: 0.5ml/minSample: .025ml/minIS: .475ml/minDilution = 20x
* Internal standards
*
*
Glass Expansion www.geicp.com
Assist Productivity Enhancement
Without Assist With AssistCycle time Rinse time Cycle time Spray chamber
rinseProbe/uptake line rinse
Probe to sample
5 0 5 5 0
Uptake delay 15 0 5 5 0
Stabilization 10 0 4 0 4
Read 15 0 15 0 15
Probe to rinse 5 0 0 0 0
Rinse 30 30 0 0 0
TOTAL 80 30 29 10 19