Augenabteilung am St. Franziskus Hospital Münster Methods Ongoing prospective single-center study comparing 14 eyes of 7 patients treated with EPI-LASIK using the XP EPI Separator (Bausch & Lomb, Rochester, New York) to a control group of 14 eyes of 7 patients treated with LASEK. Each epithelial flap was clinically evaluated by its dimensions and completeness of epithelial separation. After laser ablation, epithelial flaps of one eye per patient were repositioned and the other one discarded. All eyes received a soft contact lens.
Citation preview
Augenabteilung am St. Franziskus Hospital Münster
The Role of the Epithelial Flap
in EPI-LASIK and LASEK
Suphi TaneriTravel Grant from B&L
ASCRS Symposium, April 2-8, 2009, San FranciscoASCRS Symposium, April 2-8, 2009, San Francisco
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
PurposePurpose• To evaluate the safety and predictability of
EPI-LASIK using an EPI Separator head coupled to the Zyoptix XP microkeratome
• To elucidate the influence of the epithelial flap in EPI-LASIK and LASEK on visual recovery and pain perception.
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
MethodsMethods• Ongoing prospective single-center study
comparing 14 eyes of 7 patients treated with EPI-LASIK using the XP EPI Separator (Bausch & Lomb, Rochester, New York) to a control group of 14 eyes of 7 patients treated with LASEK.
• Each epithelial flap was clinically evaluated by its dimensions and completeness of epithelial separation.
• After laser ablation, epithelial flaps of one eye per patient were repositioned and the other one discarded. All eyes received a soft contact lens.
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
MethodsMethodsBaseline:
EPI-LASIK
Age (Years)
Corneal Thickness (µm)
Sphere (D)
Cylinder (D)
SE (D) BCVA
average Keratometry(D)
Mean 29 550 -3.81 -0.67 -4.14 0.93 43.9SD 5 46 1.50 0.48 1.58 0.10 0.8Min 20 497 -1.50 -1.50 -6.63 0.7 42.6Max 40 611 -6.25 0.0 -1.63 1.0 45.9
LASEKAge (Years)
Corneal Thicknes (µm)
Sphere (D)
Cylinder (D)
SE (D) BCVA
average Keratometry (D)
Mean 34 530 -3.61 -0.43 -3.82 0.9 44.3
SD 4 25 1.42 0.44 1.57 0.13 1.7
Min 20 494 -1.75 -1.50 -5.88 0.7 41.6
Max 50 588 -5.75 -0.25 -2.0 1.0 47.0
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
ResultsResultsHE stain of a discarded epithelial flap
magnification 200x
Microscopical examination showed complete epithelial separation
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
ResultsResultsPAS stain of a discarded epithelial flapmagnification 200x
Microscopical examination showed complete epithelial separation
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
Results: Visual recoveryResults: Visual recovery
Mean uncorrected visual acuity was slightly better in LASEK in the abscene of an epithelial flap.In Epi-LASIK mean uncorrected visual acuity was better if the flap had been repositioned (not significant).
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
Results: Development of pain perceptionResults: Development of pain perception
Pain scores decreased continually from day 1 to day 4 in all groups and showed no significant difference between the 4 groups.
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
Results: Closing of the epitheliumResults: Closing of the epithelium
closed
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
Results: Haze and Epithelial Results: Haze and Epithelial OpacificationOpacification
• Epithelial opacification was slightly higher with repositioned flaps (Fantes scale).
LASEK Mean RangeFlap repositioned 0.43 ± 0.53 0-1Flap discarded 0.33 ± 0.52 0-1Epi-LASIKFlap repositioned 0.75 ± 0.27 0.5-1.0Flap discarded 0.33 ± 0.41 0-1.0
Augenabteilung am St. Franziskus Hospital MünsterAugenabteilung am St. Franziskus Hospital Münster
ConclusionConclusionInitial results indicate a safe and standardized fashion of performing a surface ablation using the Zyoptix XP Epi Separator.
The role of the epithelial flap in surface ablation seems negligible in the first 3 months.
Differences in late on-set haze formation may exist.