Chapter 29. Temperature Monitoring

  • Upload
    adham08

  • View
    238

  • Download
    0

Embed Size (px)

Citation preview

  • 8/11/2019 Chapter 29. Temperature Monitoring

    1/40

    Chapter 29

    Temperature MonitoringIndications for Temperature MonitoringNumerous s tudies have shown that s ignif icant temperature changes routinely occur

    in anesthetized patients ( 1 ,2 ,3 ) . The monitoring guidelines of the American Society

    of Anesthesiologists s t ate that Every patient receiving anesthesia shall have

    temperature monitored when cl inical ly signif icant changes in body temperature are

    intended, anticipated or suspected. The Standards for Nurse Anesthesia Practice

    state that one must monitor body temperature continuously on all pediatric patients

    receiving general anesthesia and, when indicated, on all other patients ( 3A ).

    Temperature monitoring should be performed whenever large volumes of cold blood

    and/or intravenous f luids are administered, when the patient is del iberately cooled

    and/or warmed, for pediatr ic surgery of substantial durat ion, and in hypothermic or

    pyrexial patients or those with a suspected or known temperature regulatory

    problem such as malignant hyperthermia. Major surgical procedures, especial ly

    those involving body cavit ies, should be considered a strong indicat ion for

    temperature monitoring.

    Standards A U.S . s ta nd ard fo r el ec tro nic the rmo mete rs , not inc ludin g inf ra re d the rmo me ters ,

    was published in 2000 ( 4 ) . Celsius display thermometers must be graduated in

    intervals of not greater than 0.1C, while thermometers using a Fahrenheit display

    must be graduated in intervals of not greater than 0.2F. The maximum error for an

    individual reading is shown in Table 29.1 .

    There is also a U.S. s tandard for infrared thermometers ( 5 ) . This specif icat ion does

    not prescribe a method for determining cl inical accuracy.

    P.859

    TABLE 29.1 Maximum Error for Electronic Thermometers

  • 8/11/2019 Chapter 29. Temperature Monitoring

    2/40

    Temperature Maximum Error

    Celsius Scale

    Less than 35.8C 0.3F

    35.8C to less than 37.0C 0.2F

    37.0C to 39.0C 0.1F

    39.0C to 41.0C 0.2F

    Greater than 41.0C 0.3F

    Fahrenheit Scale

    Less than 96.4C 0.5F

    96.4C to 98.0C 0.3F

    98.0C to 102.0C 0.2F

    102.0C to 106.0C 0.3F

    Greater than 106.0C 0.5F

    (From ASTM 15111200, 2000).

    Technologies

  • 8/11/2019 Chapter 29. Temperature Monitoring

    3/40

    A varie ty of tec hn ologi es are av ailab le to me asure tem pe ra tu re. No ne is su i tab l e fo r

    al l s i tuat ions. Many devices simply display the temperature. These are less than

    optimal, because a high or low temperature may go unnoticed for some t ime. Mostmodern devices have alarms that can be set i f the temperature exceeds high or low

    limits . Temperature-monitoring capabil i ty is available on most physiologic monitors

    used in the operating room and perioperative areas. Frequently, they have the

    abil i ty to measure temperature at two different s i tes ( Fig. 29.1 ) . These devices

    usually have the abil i ty to trend the temperature and to transfer temperature

    information to an electronic record.

    There are a number of s tand-alone devices, both mechanical and chemical . Those

    stand-alone devices that are battery-powered should have a means to indicate

    when battery power is low. Trend i ndicators are available on s ome of these

    instruments.

    T h e r m i s t o r

    A therm istor is comp osed of a me tal ( i .e . , ma ng an es e, nickel , cob alt , i ro n , or zinc )

    oxide sintered into a wire or fused into a rod or bead ( 6 ,7 ) . There must be a source

    of current and a means to measure that current. Resistance of the metal oxide

    increases as the temperature decreases and vice versa so the resistance can be

    converted to a temperature. Advantages of thermistors include small s ize, rapid

    response t ime, continuous readings, and sensi t ivi ty to s mall changes in

    temperature. They are fair ly inexpensive. Probes c an be interchangeable and

    disposable.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    4/40

    View Figure

    Figure 29.1 Monitor with capability for monitoringtemperature at two sites and alarms. (Courtesy of FisherPaykel Healthcare.)

    T h e r m o c o u p l e

    A therm oc oup le cons is ts of an elec tri ca l c i rc uit th a t ha s two diss imilar me tal s

    welded together at their ends ( 6 ,8 ) . One of the two metal junctions remains at a

    constant temperature. The other is exposed to the area being measured, producing

    a voltage difference that is measured and converted to a temperature reading.

    Adv an ta ge s of th erm ocoup les include accu ra c y, small s iz e, rap id res po ns e t ime ,

    continuous readings, stability, and probe interchangeability. The materials are

    inexpensive, so the probes can be made disposable.

    P l at i n u m W i r e

    The electr ical resistance of plat inum wire varies almost l inearly with temperature.

    By employing an extremely small d iameter wire, rapid thermal equil ibrat ion is

    possible. Resistance is measured in a manner similar to a thermistor. These

    thermometers are accurate and giv e continuous readings. Probes can be made

    interchangeable.

    If a probe with a thermocouple, thermistor, or plat inum wire is to be used inside the

    body, i t must have

    P.860

    an outer sheath. The end of the probe designed to be inserted into the patient is

    sealed and the electr ical connection made at the o ther end. The connection needs

    to be kept d ry. If i t becomes wet, erroneous readings can result ( 9 ,10 ,11 ).

  • 8/11/2019 Chapter 29. Temperature Monitoring

    5/40

    View Figure

    Figure 29.2 The flexible adhesive-backed strip of thisliquid crystal temperature monitor has a black background.To use, the covering over the adhesive is removed, and themonitor is placed on the skin.

    L i q u i d C r y s t a l

    Certain organic compounds in thermal transformation from a solid to a l iquid s tate

    pass through an intermediate phase that exhibits anisotropic (optical ly act ive)

    propert ies ( 12 , 13 ). The term liquid crystal is used to describe this s tate. When l ight

    shines on such a material , crystals scat ter some of the l ight , producing ir idescent

    colors. The l iquid crystals are encapsulated so that the colors form let ters and

    numbers.

    A l iqu id crys tal tem pe ra ture mo ni to r is shown in Figure 29.2 . I t consists of a f lexible

    adhesive backing with plast ic-encased l iquid crystals on a black background that

    prevents reflect ion of the transmitted l ight and enhances the color resolution. The

    covering over the adhesive is removed, and the disc or s tr ip is placed on the skin.

    The l iquid crystal thermometer is available in two forms: one displays the skin

    temperature direct ly; the other has a buil t- in correct ion factor (offset) so that the

    temperature displayed est imates core temperature ( 7 , 13 ).Liquid crystal thermometers are safe, convenient , noninvasive, easy to apply and

    read, disposable, nonirr i tat ing, and inexpensive. They give fast , continuous

    readings and involve no electronic circuitry. They can be applied before induction

    of anesthesia and are easi ly transferred to the recovery area with the patient .

    Disadvantages of l iquid crystal thermometers include the need for subject ive

    observer interpretat ion and the inabil i ty to interface with a recording system. They

    are less accurate than other devices. Extreme ambient temperature, humidity, and

    air movement can cause inaccuracy ( 13 ). Other disadvantages include diff icult ies

  • 8/11/2019 Chapter 29. Temperature Monitoring

    6/40

    with adhesion secondary to skin secret ions, al lergic reactions to adhesive backing,

    inaccuracy, and imprecision ( 14 , 15 ). They are capable of measuring temperature

    only on the skin. If lef t in the sun for an extended period, an error indicat inghyperthermia can be produced. If the device is frozen, al l the numbers can be

    visible at once ( 13 ). Infrared heating lamps may cause erroneously elevated

    readings ( 12 ). There is a report of a l iquid crystal thermometer that gave a falsely

    high temperature ( 16 ).

    I n f r a r e d

    An infra red th erm om ete r is an el ec tron ic in str umen t th at mea sure s a port ion of the

    infrared radiat ion from surfaces within i ts f ield of v iew ( 17 , 18 ,19 ). Electromagnetic

    radiation is emitted by an object in proportion to its temperature. Because theprobe is placed in the outer part of the ear canal and usually has a relat ively wide

    view angle, i t detects infrared emissions from parts of both the ear canal wall and

    tympanic membrane and computes an av erage or highest temperature. The

    displayed value may be the actual temperature (unadjusted, calibration mode) or

    with an offset to est imate temperature at another si te based on selected study

    samples (si te equivalent mode) ( 5 ,7 ,17 ,20 ).

    The instrument has an otoscopelike probe ( Fig. 29.3 ) . Disposable probe covers are

    used for hygiene and to prevent cerumen buildup. The probe should be placed

    carefully but f irmly as far as possible into the ear canal , aiming toward the

    tympanic membrane. The probe window should be clean and s hiny, and a n ew

    probe cover should be used each t ime.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    7/40

    View Figure

    Figure 29.3 The infrared thermometer's probe is insertedinto the external ear canal.

    P.861

    Patients tolerate infrared thermometry well ( 18 ). Because i t does not contact any

    surface, there should be no trauma associated with its use. Measurements are

    rapid. This technology is often used in the postanesthesia care unit and other

    cri t ical care areas.

    There are a number of problems with infrared thermometry. Poor penetration,

    improper aiming, and obstructions such as c urvatures of the ear canal can result in

    signif icantly lower temperatures. Measurements are intermit tent . This device is

    general ly not useful for monitoring in the operating room.

    M e r c u r y - g l a s s

    These thermometers are not useful for intraoperat ive anesthesia, as they have a

    long equil ibrat ion period (usually 3 to 5 minutes) and cannot be read without beingremoved from the monitoring site. In the future, they will be banned to reduce

    mercury contamination in the environment.

    If a new thermometer is used on each patient , there is a waste-disposal problem. If

    the thermometer is reused, i t must be cleaned and steri l ized. If a thermometer is

    broken, the mercury vaporizes and can present a health hazard.

    Thermal Compartments

  • 8/11/2019 Chapter 29. Temperature Monitoring

    8/40

    Al th oug h arbi tr ary, i t is some wh at us eful to div ide th e bod y into two th erm a l

    compartments: the core that includes the deep, vi tal internal organs and a shell of

    peripheral t issue that serves as insulat ion for the core.

    C o r e

    Core temperature is uniform and high compared with the rest of the body. I t

    normally varies between 35.7C and 37.8C ( 21 ,22 ,23 ). When signif icant changes

    in body heat are expected, core temperatures should be monitored.

    Sites differ in how well they reflect core temperature. The difference may depend

    on the rate of temperature change. A si te that reflects core temperature accurately

    when temperature change is s low may fai l to reflect rapid changes.

    P e r i p h e r yNormally, thermoregulatory vasoconstriction maintains a temperature gradient

    between the core and periphery (shell) of 2C to 4C. Regional temperature

    variat ions exist i n the periphery. The correlat ion between temperatures measured

    at different body si tes depends on several factors, including the stabil i ty of the

    body temperature and whether or not there have been recent cold or warm

    challenges. Skin and axil lary temperatures are usually considered shell

    temperatures.

    Monitoring SitesBody temperature can be monitored at a number of sites. The temperature can vary

    considerably in different parts of the body at any t ime. Factors influencing

    temperature at any given si te include the t issue 's heat production, the temperature

    and rate of blood flow through the area, the amount of insulation from the

    environment, and external influences on the site ( 24 ).

    The best site for temperature monitoring depends on the purpose of the

    measurement, durat ion of the surgical procedure, the surgical s i te , the anesthesia

    technique, and av ailable equipment. Considerat ions should include accuracy,

    speed, convenience, access, safety, pat ient acceptabil i ty, and cost-effect iveness.

    Abs ol ute ac curac y is us uall y no t nec essa ry , but mea surem ents mu st be c lo se

    enough to detect temperature changes that may influence treatment decisions. It

    may be helpful to monitor two si tes. The difference between core and a second si te

    can provide indirect information on blood f low (slow change = poor blood f low) and

    is helpful in guarding against an overshoot during warming or cooling.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    9/40

    Studies comparing temperature measurement among different sites are often

    confusing. Many try to relate peripheral s i tes to c ore temperature and use different

    si tes as the standard for core temperature.

    P u l m o n a r y A r t e r y

    Pulmonary artery temperature can be measured in patients who have a Swan-Ganz

    catheter with a thermistor in place. It is thought by many to be the best method of

    measuring core body temperature ( 25 ,26 ).

    Pulmonary artery temperature generally correlates well with intrathecal and jugular

    bulb temperatures, even with rapid cooling and rewarming ( 27 ,28 ). Poor correlat ion

    with brain temperature was found during profound hypothermia ( 29 ).

    Pulmonary artery readings are not rel iable during thoracotomy or cardiopulmonarybypass when there is no f low through the heart and lungs and may be direct ly

    affected by the cardioplegia used during cooling.

    E s o p h a g u s

    Esophageal temperature measurement can be accomplished by using a simple

    probe, an esophageal stethoscope with thermistor ( Fig. 29.4 ), or a gastric tube with

    the temperature sensor some distance from the end of the tube ( Fig. 29.5 ).

    P.862

    View Figure

    Figure 29.4 Esophageal stethoscope and temperature probe.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    10/40

    Temperatures in the esophagus may vary up to 4C, depending on the temperature

    probe location within the esophagus ( 7 ). The esophageal temperature should be

    measured with the sensor located in the lower third or fourth of the esophagus ( 30 ). At thi s de pth, th e es opha gus l ies be tween th e heart an d the des cen ding ao rt a.

    Placing the sensor in this posi t ion wil l minimize (but not c ompletely el iminate) the

    effect of respired gases ( 31 ). When the sensor is part of an esophageal

    stethoscope, the ideal depth of placement is 12 to 16 cm dis tal to the point of

    maximum heart sounds ( 32 ). If the probe is placed higher in the esophagus, the

    reading will be lower ( 30 ,32 ,33 ,34 ). If the probe is placed in the stomach, it may

    record temperatures higher than core, reflecting liver metabolism. In addition, the

    response time to temperature changes is slow with the probe in the stomach.

    The probe is most accu rately placed by using an electrocardiographic lead buil t into

    the probe ( 35 ,36 ). The posi t ive lead is at tached to the probe, and the negative lead

    is attached to the right shoulder. A biphasic P wave indicates that the probe tip is

    at the midatr ial level .

    In adults , the ideal posi t ion is approximately 38 to 42 cm below the central inc isors

    (32 ) or at least 24 cm below the larynx ( 30 ). For nasal insert ion, the fol lowing

    formulas can be used ( 37 ):

    L (cm) = 0.228 (standing height) - 0.194

    orL (cm) = 0 .479 (si t t ing height) - 4.44

    where L is the length from the opening to the nares.

    View Figure

    Figure 29.5 Gastric tube with temperature probe. This alsofunctions as an esophageal stethoscope.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    11/40

    P.863

    In children, the ideal dis tance in centimeters below the corniculate cart i lages is

    approximated by the fol lowing formula ( 38 ):

    10 + (2 age in years)/3 cm

    A temp era ture prob e can be ins ert ed into the es oph ag us thro ugh the drain tub e of a

    ProSeal laryngeal mask. The optimal location for the temperature probe is 15 to 20

    cm distal to the drain tube ( 39 ).

    Esophageal temperature is considered core temperature by many investigators.

    Temperatures measured in this location have shown good agreement with

    pulmonary artery temperature ( 28 , 40 , 41 ,42 ,43 ,44 , 45 ). During rapid warming or

    cooling, esophageal temperature shows less lag time than that measured at most

    other si tes ( 46 ), although some studies found that bladder temperature showed a

    closer approximation to pulmonary artery temperature than the esophageal

    temperature ( 47 ,48 ). Brain temperature may be adequately reflected by esophageal

    temperature during mild, but not profound, hypothermia ( 29 ).

    Patients having act ive or passive airway humidificat ion have sl ightly h igher

    esophageal temperatures ( 31 ,34 ).

    Contraindicat ions to use of an esophageal probe include procedures on the face,oral cavity, nose, airway, or esophagus and patients who have esophageal

    disorders. I t is poorly tolerated by awake patients . Correct placement may be

    diff icult , and probes may become displaced. Esophageal temperatures are

    unrel iable during thoracic surgery. Continuous gastr ic suctioning wil l cause a

    decrease in esophageal temperature ( 49 ). When an esophageal probe is used with

    the patient in the si t t ing or prone posi t ion, oral secret ions can track down to the

    connection between the probe and monitor cable. This ca n lead to incorrect

    readings ( 9 ) .

    N a s o p h a r y n x

    The temperature of the nasopharynx is measured with a sensor that is in contact

    with the posterior nasopharyngeal wall posterior to the soft palate. This location

    should place i t c lose to the hypothalamus.

    Al th oug h some s tudie s sho w a goo d c orr elat ion of nasop haryn gea l te mp era tu re wi th

    core temperature ( 29 , 46 , 50 ,51 ,52 ), other s tudies have found the correlat ion less

    sat isfactory ( 53 ,54 ,55 ).

  • 8/11/2019 Chapter 29. Temperature Monitoring

    12/40

  • 8/11/2019 Chapter 29. Temperature Monitoring

    13/40

    Electronic rectal thermometers may have two measurement modes: dwell or

    monitor. In the dwell mode, the temperature is displayed continuously. The dwell

    mode requires a minimum of 2 to 3 minutes to reach a stable temperature. Thepredictive mode estimates the temperature on the basis of the curve of temperature

    rise. This mode requires only 30 seconds. A comparison of the measurements of

    the two modes yielded similar values during steady-state condit ions and mild

    temperature f lux.

    Disposable probe covers can be used to avoid c ross contamination. Probes should

    be inserted to at least 8 cm in adults and 3 cm in children ( 65 ). The depth of

    insert ion can be marked on the probe. I t should be checked for placement after the

    patient is moved. The probe should be securely taped to the patient 's buttocks, and

    the lead wire should be secured to a garment or sheet to avoid probe displacement.

    Rectal temperature is influenced by heat-producing flora, the temperature of the

    blood returning from the legs, and insulat ion by feces. Rectal temperature is

    usually somewhat higher than that measured at more c entral s i tes during s teady-

    state condit ions ( 51 , 66 ,67 ,68 ,69 , 70 , 71 ). The rectum is not a vascular area, and the

    lag time may be prolonged with shifting temperature ( 26 ,52 , 55 , 72 ,73 ).

    The rectum is usually accessible and monitoring relat ively noninvasive. Rectal

    temperature is not influenced by ambient temperature. I t is general ly disl iked by

    patients as uncomfortable, by hospital personnel as cumbersome, and by b oth asaesthetical ly object ionable. Probes are p rone to extrusion during recovery from

    anesthesia. Other disadvantages are the relat ive inaccessibi l i ty during surgery and

    the r isk of bacterial contamination. Contraindicat ions include gynecologic and

    urologic procedures. Bowel perforat ion is a r isk ( 74 ). A pararectal abscess and

    pneumoperitoneum have been reported with its use ( 75 ,76 ).

    Ty m p an i c M e m b r an e

    Two different technologies ut i l ize the ear. One uses a probe that contacts the

    tympanic membrane, and the other measures infrared radiation.

    Contact

    The anatomical posi t ion of the tympanic membrane is deep within the skull and

    separated from the internal carotid artery by only the narrow air-f i l led c left of the

    middle ear and a thin shell of bone, making i t an at tract ive si te for temperature

    measurements. The tympanic membrane and the hypothalamus share a common

    blood supply, so measuring temperature at this s i te may reflect thermal information

    at the primary si te of thermoregulat ion ( 17 ).

  • 8/11/2019 Chapter 29. Temperature Monitoring

    14/40

    Temperature can be measured by inserting a thermistor or thermocouple probe into

    the external auditory canal un ti l i t contacts the tympanic membrane. Because of the

    danger of perforating the membrane, clinicians tend not to place the probe farenough into the canal . If i t does not touch the membrane, the readings wil l not be

    accurate. In the awake patient , i t should be inserted unti l the patient feels the

    thermocouple touch the tympanic membrane; appropriate placement is confirmed

    when the patient easi ly detects a gentle rubbing of the at tached wire ( 77 ). The

    lower anterior quarter of the membrane should be used ( 78 ). After the probe is

    inserted, the aural canal should be occluded with cotton wool and cov ered

    externally to prevent air movement from cooling the probe ( 79 ,80 ).

    Contact tympanic membrane probes are shown in Figure 29.7 . The sensor is

    enclosed in soft foam. I t usually has a widened segment, piece of foam, or a

    feather or barb to hold i t in place after insert ion. After insert ion, the reading should

    stabil ize quickly. If i t does not , the probe should be s lowly advanced unti l the

    reading stabil izes.

    Numerous studies have shown a good correlation between tympanic membrane

    temperature and temperature measured in the esophagus, pulmonary artery, or

    urinary bladder ( 31 , 71 ,78 ,79 , 80 , 81 ,82 ,83 ,84 ,85 ). Temperature may be less

    accurate during rapid changes ( 29 , 40 ,79 ).

    Head posit ion can cause as ymmetric c hanges in the tympanic temperatures ( 86 ).Upon assuming a lateral position, the temperature on the lower side increases

    while that on the upper side decreases.

    Adv an ta ge s of ty mp an ic mem bran e te mp erat ure mo ni tori ng inc lud e c le anline ss an d

    convenience. I t is tolerated by c onscious patients , making i t useful for

    postoperative monitoring. The si te is readily accessible during most surgical

    procedures.

    Complications have been reported. Oozing from the ear, trauma to the external

    auditory canal with subsequent external ot i t is , and perforat ion of the membrane

    have been reported ( 81 ,87 ,88 ). These complicat ions occurred before the advent of

    f lexible cotton-t ipped probes that are less traumatic ( 7 ). Recommended methods to

    avoid trauma include otoscopic inspection of the canal and drum before insertion,

    stopping insert ion as soon as resistance is fel t , and placement in awake patients to

    assess discomfort . Care should be taken that the

    P.865

    probe is not pushed into the canal when the head is moved.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    15/40

    View Figure

    Figure 29.7 Tympanic membrane temperature probes.

    Contraindicat ions include any ear abnormali ty that would prevent correct

    placement, a skull f racture that passes through the osseous meatus, and

    perforation of the tympanic membrane.

    View Figure

    Figure 29.8 Disposable probe for measuring skintemperature.

    InfraredInfrared ear thermometry (infrared emission detection thermometry, infrared

    tympanic membrane thermometry, infrared tympanic thermometry) is performed by

    insert ing an otoscopelike probe into the external ear canal ( 19 ). The t ip is usually

    covered by a disposable cover. The device detects the amount of infrared heat

  • 8/11/2019 Chapter 29. Temperature Monitoring

    16/40

    emitted. The wide angle is too large to measure only the tympanic membrane, so it

    also reads the temperature of the ear canal .

    Since these thermometers are known to give temperatures lower than coretemperatures, some manufacturers provide offsets to estimate the temperature at

    other si tes ( 89 ). These offsets vary for different brands ( 17 ,26 ,90 , 91 ).

    A tug on th e ea r to s tra ig hte n th e cana l wh ile ta k ing a reading wa s foun d by som e

    investigators to improve correlation with core temperatures ( 92 ,93 , 94 , 95 ,96 ). Other

    studies did not f ind that an ear tug is helpful ( 20 ,69 ). The device should be inserted

    with a gentle back and forth motion, and f irm but gentle pressure should be applied

    to seal the canal from ambient air ( 94 ).

    Many studies have compared temperatures obtained by infrared ear thermometers

    to temperatures obtained f rom other si tes

    (19 ,25 ,41 , 42 , 44 ,52 ,58 , 66 , 69 ,70 ,83 ,90 ,91 , 93 ,97 ,98 , 99 , 100 ,101 ,102 ,103 ,104 ,105 ,106

    , 107 , 108 ,109 ,110 ,111 ,112 ,113 ,11 4 ,11 5 ,116 , 117 ,118 ,119 ,120 , 121 ,122 ,12 3 ). The

    correlat ions vary from excellent to poor. There is considerable va riat ion among

    different instruments and with different users of the same instrument

    (20 ,109 , 114 ,120 ). A difference of 1.6C between the two ears has been reported

    (114 ). Readings are affected by the ambient temperature ( 12 4 ,125 ). If the face is

    cooled, there will be a decrease in the reading ( 126 ). Acute ot i t is media wil l not

    influence the reading unless there is suppurat ion, in whichP.866

    case the reading wil l be sl ightly increased ( 98 ,10 2 ,117 ,12 7 ).

    Adv an ta ge s of in frare d te mp era tu re mo ni tori ng in c lud e spe ed an d red uc ed po ten tia l

    for cross contamination. I t is noninvasive, requires no clothing removal , and is well

    tolerated by conscious patients . The si te is usually readily accessible. The ease

    with which the measurements can be obtained has made these the thermometers of

    choice in many pre- and postoperat ive units . Because the readings are intermit tent

    and subject to such variability, this technology is not recommended for

    intraoperative use.

    S k i n

    Skin temperature can be measured by using a l iquid crystal device or f lat d isc or t ip

    of a lead containing a thermocouple or thermistor ( 6 ) ( Fig. 29.8 ). Some skin probes

    have a special backing to at tach to the body and material to insulate the sensor

  • 8/11/2019 Chapter 29. Temperature Monitoring

    17/40

    from ambient condit ions. An opaque dressing and/or tape over the sensor may

    decrease the effect of environmental factors.

    Skin temperature is most commonly measured at the forehead, because this sitehas a fai r ly good blood f low, and there is not much underlying fat . The back, chest ,

    anterior abdominal wall, fingers, toes, and the antecubital space inside the elbow

    have also been used.

    The correlat ion between skin temperature and core temperature is controversial .

    Skin surface temperature is typically 2C to 4C less than core temperature. Some

    manufacturers build in an offset to compensate for this difference ( 128 ). In some

    studies, skin temperature has been found to correlate well with core temperature

    (129 , 13 0 ). However, most invest igat ions have found that skin temperature does not

    accurately reflect core temperature, and there may not be a relat ionship between a

    change in skin and co re temperature ( 29 ,31 , 61 ,78 , 10 8 , 13 1 ,132 , 13 3 ).

    Monitoring skin temperature carries few risks. The si te is e asi ly accessible. The

    main disadvantage is that skin temperature is a poor est imate of core body

    temperature and may give e rroneous information ( 16 ). Skin temperature readings

    are affected by ambient temperature, skin surface warming devices, intraoperat ive

    changes in cardiac output , and regional vasoconstr ict ion ( 134 ,135 , 13 6 , 13 7 ). The

    usefulness of skin temperature monitoring as a screening device for malignant

    hyperthermia is l imited, as cutaneous vasoconstr ict ion may occur with thissyndrome ( 13 8 ).

    Skin temperature may be used to evaluate the quali ty of a regional block. A rise in

    skin temperature is an indicat ion that the block is successful . Another use is in

    microsurgery. An increase in skin temperature may indicate that blood flow to that

    area has increased.

    A x i l l a

    To measure axillary temperature, a mercury-glass thermometer or a probe with a

    thermocouple or thermistor is posi t ioned over the axil lary artery and the arm

    adducted ( 31 ). Significant differences have been found between mercury and

    electronic axil lary temperatures ( 68 , 111 ). Temperature should not be measured on

    the same side where a blood pressure cuff is on the upper arm. Equil ibration may

    take as long as 10 to 15 minutes ( 72 ,13 8 ).

    Al th oug h a few s tu die s fou nd sat is fa c to ry corr elati on wi th mo re centra l s i te s

    (133 , 13 9 ,140 , 141 , 14 2 ), most s tudies show poor correlat ion

    (20 ,25 ,29 , 45 , 51 ,58 ,68 , 69 , 98 ,11 5 ,116 , 12 9 , 14 3 ,144 ,14 5 ,14 6 ,147 , 14 8 , 14 9 ).

  • 8/11/2019 Chapter 29. Temperature Monitoring

    18/40

    Axil la ry temp eratu re me asureme nts are eas y to tak e and are no t obj ec tiona b le to

    the patient or nursing staff . The si te is usually accessible. However, i t is not

    considered accurate or rel iable in adults . I t i s used most frequently on infants andchildren because the patient 's small s ize and greater surface vasculature make a

    relatively uniform temperature ( 17 , 31 ). Readings are influenced by contact with the

    probe, skin perfusion, exposure to the environment, skin warming devices, and

    proximity of the probe to the axil lary artery ( 20 , 26 ,56 , 14 8 ).

    M o u t h

    Sublingual temperature is measured by placing a p robe in one of the pockets on

    either s ide of the frenulum of the tongue ( 15 1 ). The patient 's mouth should be

    closed and enough t ime al lowed for the reading to be accurate. This area containssmall muscular arteries that respond to masticat ion or hot or co ld l iquids by

    expanding or contract ing.

    Correlat ion with temperatures measured at more c entral s i tes varies s omewhat but

    in general is fair ly sat isfactory ( 20 ,58 , 77 ,113 , 11 5 ,15 1 ,152 , 15 3 ).

    Sublingual temperatures are well tolerated by patients . Readings are not affected

    by the presence or absence of teeth, in tubation, administrat ion of oxygen, a

    nasogastr ic tube on continuous suction, or the temperature of inspired gases

    (145 , 15 0 ,154 ). Sublingual temperature may be inaccurate if the patient is a mouth

    breather or has tachypnea ( 15 5 ). Warm and cold ambient temperatures have a

    small effect on oral temperatures ( 12 4 ,125 ).

    Tr a c h e a

    Temperatures in the trachea can be measured by using a tracheal tube with the

    temperature sensor in the c uff . Studies differ in how well temperatures measured at

    this s i te correlate with those at other s i tes ( 156 ,157 ,158 ,15 9 ).

    P.867

    In g u i n a l A r e a

    The inguinal area has been used to measure temperature ( 45 ,72 ,160 ). The leg is

    f irst abducted and the femoral pulse determined. The sensor is placed just lateral

    to the femoral artery, and the leg is adducted to c reate a seal . A s ignif icant

    drawback to this s i te is the t ime required to reach equil ibrium.

    H y p o p h a r y n x

  • 8/11/2019 Chapter 29. Temperature Monitoring

    19/40

    A therm oc oup le can be a tt ac hed to the cuf f of a supra glott ic dev ice to meas ure

    temperature in the hypopharynx ( 161 ,162 ). Temperatures measured in this location

    have correlated well with those at other s i tes .

    Hazards of Thermometry

    D am a g e t o t h e M o n i t o r i n g S i t e

    Tympanic membrane and rectal perforation and trauma to the nose, external

    auditory canal, rectum, and esophagus have been reported

    (74 ,75 ,76 , 81 , 87 ,88 ,16 3 ,164 ). Probes can have sharp edges that could p resent a

    hazard to the patient ( 16 5 ).

    B u r n sBurns can occur at the measurement si te i f the probe acts as a ground for the

    electrosurgical apparatus ( 166 , 16 7 , 16 8 ). No insulat ion can c ompletely block radio

    frequency currents , and if there is no other sat isfactory return path, the current can

    burn through insulat ion ( 16 9 ). Using a battery-operated device is no guarantee of

    electr ical safety, because the chassis may be g rounded through a metal support .

    Temperature probes should be examined before use to detect damage to the

    insulat ion. Esophageal burns may be avoided by insert ing the probe via a small

    tracheal tube ( 170 ). The probe may be pulled back into the tube during periods of

    maximal electr ical act ivi ty.

    In c o r r e c t In f o r m a t io n

    A faulty prob e can cau se an inco rr ec t tempe ra tu re to be displ ay ed

    (171 , 17 2 ,173 , 174 ). Secret ions or f luids in the connection between the probe and

    reading instrument can result in falsely elevated readings ( 9 , 10 ). If the battery in a

    battery-powered device is depleted, the unit may s top functioning or give incorrect

    information ( 6 ). Using jacks that fit the receptacle but have internal electronics

    incompatible with the monitor can give false readings.P r o b e C o n t a m i n a t io n

    Reusable temperature probes may be a s ource of bacterial or v iral pathogens even

    if protect ive covers are used ( 6 ,175 ).

    F a u l t y Pr o b e s

    Af ter a p ro be is re mov ed f ro m i ts pa ckagin g, i t shoul d be ins pe cted . A part o f th e

    sheath on a p robe could break off and be aspirated.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    20/40

  • 8/11/2019 Chapter 29. Temperature Monitoring

    21/40

    11. Wiegert PE. An unusual cause of misleading temperature readings.

    Anes th es iolog y 1990 ;72 :2 08 .

    [Full text Link] [CrossRef]

    [Medline Link]

    12. Bjoraker DG. Liquid crystal temperature indicators. Anesthesiol Rev

    1 990 ;1 7:5 05 6.

    [Medline Link]

    13. MacKenzie R, Asbury AJ. Clinical evaluation of l iquid c rystal skin

    thermometers . Br J Anaes th 1994;72:246249.

    [CrossRef]

    [Medline Link]

    14. Lake CL. Monitoring of temperature and heart and lung sounds. In: Lake CL,

    Hines RL, Bli t t CD. Clinical monitoring. Practical applicat ions for anesthesia and

    cr i t ica l care . Phi lade lphia : WB Saunders Company, 2001:357370.

    15. Markin DA, Henry DA, Baxter SS, et al . Comparison between two types of body

    surface temperature devices: eff iciency, accuracy and cost . J Post Anesth Nurs

    1 990 ;5 :3 337 .

    [Medline Link]

    16. Marsh ML, Sessler DI. Failure of intraoperative l iquid-crystal temperaturemoni tor ing . Anes th Analg 1996;82:11021104.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    17. Anonymous. Infrared thermometers. An earful of innovation. Technol Anesth

    1 992 ;12 :1 7 .

    18. Weiss ME, Pue AF, Smith J III . Laboratory and hospital test ing of new infrared

    tympanic thermometers . J Cl in Eng 1991;16:137144.

    [Medline Link]

    19. Edge G, Morgan M. The Genius infrared tympanic thermometer. An evaluation

    for c l in ica l use . Anaes thes ia 1993;48:604607.

    [CrossRef]

    [Medline Link]

    20. Erickson RS, Meyer LT. Accuracy of infrared ear thermometry and other

    tempera ture methods in adul ts . Am J Cr i t Care 1994;3 :4054.

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    22/40

    21. Sessler DI. A proposal for new temperature monitoring and thermal

    management guidel ines . Anes thes io logy 1998;8912981300.

    22. Mitchell AM, Kennedy RR. Preoperative core temperatures in elective surgicalpatients show an unexpected skewed distr ibution. Can J Anesth 2001;48:850853.

    23. Terndrup TE. An appraisal of temperature assessment by infrared emission

    detect ion tympanic thermometry. Ann Emerg Med 1992;21:14831492.

    [CrossRef]

    [Medline Link]

    24. Blainey CG. Site select ion in taking body temperature. Am J Nurs

    1 97 4; 74 :1 85 9 18 60 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    25. Schmitz T, Bair N, Falk M, et al. A comparison of five methods of temperature

    measurement in febri le intensive care patients . Am J Cri t Care 1995;4:286292.

    [Medline Link]

    26. Schuman AJ. The accuracy of infrared auditory canal thermometry in infants

    and chi ldren . Cl in Pedia t r 1993;32:347354.

    [Medline Link]

    27. Crowder CM, Tempelhoff R, Theard A, et al. Jugular bulb temperature:comparison with brain surface and core temperatures in neurosurgical pat ients

    dur ing mi ld hypothermia . J Neurosurg 1996;85:98103.

    [Medline Link]

    28. Kumar M, Murray M, Werner E, et al. Monitoring intrathecal temperature: does

    core temperature reflect intrathecal temperatures during aortic surgery? J

    Cardio thorac Vasc Anes th 1994;8 :3539.

    29. Stone JG, Young WL, Smith CR, et al . Do s tandard monitoring si tes reflect t rue

    brain temperature when profound hypothermia is rapidly induced and reversed?

    Anes th es iolog y 1995 ;82 :3 44 35 1 .

    30. Whitby JD, Dunkin LJ. Temperature differences in the oesophagus. Br J

    Anaes th 19 69;41 :61 5 61 8.

    [CrossRef]

    [Medline Link]

    P.868

  • 8/11/2019 Chapter 29. Temperature Monitoring

    23/40

    31. Bissonnette B, Sessler KI, LaFlamme P. Intraoperative temperature monitoring

    si tes in infants and children and the effect of inspired gas warming on esophageal

    tempera ture . Anes th Analg 1989;69:192196.[Full text Link]

    [CrossRef]

    [Medline Link]

    32. Kaufman RD. Relationship between esophageal temperature gradient and heart

    and lung s ounds heard by esophageal s tethoscope. Anesth Analg 1987;66:1046

    1048.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    33. Whitby JD, Dunkin LJ. Temperature differences in the oesophagus. Br J

    Anaes th 19 68;40 :99 1 99 5.

    [CrossRef]

    [Medline Link]

    34. Siegel MN, Gravenstein N. Passive warming of airway gases (art if icial nose)

    improves accuracy of esophageal temperature monitoring. J Clin Monit 1990;6:89

    92.

    [Medline Link] 35. Brengelmann GL, Johnson JM, Hong PA. Electrocardiographic verif icat ion of

    esophageal temperature probe posi t ion. J Appl Physiol 1979;47:638642.

    [Medline Link]

    36. Freund PR, Brengelmann GL. Placement of esophageal stethoscope by acoustic

    cri teria does not consistently yield an optimal location for the monitoring of core

    tempera ture . J Cl in Moni t 1990;6 :266270.

    [Medline Link]

    37. Mekjavic IB, Rempel ME. Determination of esophageal probe insertion length

    based on s tanding and s i t t ing he ight . J Appl Phys io l 1990;69:376379.

    [Medline Link]

    38. Whitby JD, Dunkin LJ. Oesophageal temperature differences in children. B r J

    Anaes th 19 70;42 :10 13 1 015 .

    [CrossRef]

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    24/40

    39. Mitchell S, Brimacombe J, Keller C. Feasibi l i ty, accuracy and optimal location

    for oesophageal core temperature measurements using the ProSeal laryngeal mask

    ai rway dra in tube . Anaes th In tens Care 2003;31:282285.[Medline Link]

    40. Shiraki K, Konda N, Sagawa S. Esophageal and tympanic temperature

    responses to core blood temperature changes during hyperthermia. J Appl Physiol

    1 986 ;6 1:9 81 02 .

    [Medline Link]

    41. Stavem K, Saxholm H, Smith-Erichsen N. Accuracy of infrared ear thermometry

    in adul t pa t ien ts . In tens ive Care Med 1997;23:100105.

    [CrossRef]

    [Medline Link]

    42. Harasawa K, Kemmotsu O, Mayumi T, et al. Comparison of tympanic,

    esophageal and blood temperatures during mi ld hypothermic cardiopulmonary

    bypass: a s tudy using an i nfrared emission detect ion tympanic thermometer. J Clin

    M on i t 1 99 7; 13 : 19 2 4.

    [CrossRef]

    [Medline Link]

    43. Horrow JC, Rosenberg H. Does urinary catheter temperature reflect core

    tempera ture dur ing card iac surgery? Anes thes io logy 1988;69:986989.44. Robinson J , Charl ton J , Seal R , et al . Oesophageal , rectal , axil lary, tympanic

    and pulmonary artery temperatures during cardiac surgery. Can J Anaesth

    1 99 8; 45 :3 17 3 23 .

    [Medline Link]

    45. Lefrant J-Y, Muller L, Emmanuel Coussaye J, et al. Temperature measurement

    in intensive c are patients: comparison of urinary bladder, oesophageal , rectal ,

    axil lary and inguinal methods v ersus pulmonary artery core temperature. Intensive

    Care Med 1999 ;29 :414418 .

    [Medline Link]

    46. Bone ME, Feneck RO. Bladder temperature as an est imate of body temperature

    dur ing card iopulmonary bypass . Anaes thes ia 1988;43:181185.

    [Medline Link]

    47. Russell SH, Freeman JW. Comparison of bladder, oesophageal and pulmonary

    artery temperatures in major abdominal surgery. Anaesthesia 1996;51:338340.

    [Full text Link]

    [CrossRef]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    25/40

    [Medline Link]

    48. Lil ly JK, Boland JP, Zekan S. Urinary bladder temperature monitoring. A new

    index of body core tempera ture . Cri t Care Med 1980;742744.49. Nelson EJ, Grissom TE. Continuous gastr ic suctioning decreases measured

    esophageal temperature during general anesthesia. J Clin Monit 1996;12:429

    432.

    [CrossRef]

    [Medline Link]

    50. Moorthy SS, Winn BA, Jal lard MS, et al . Monitoring urinary bladder

    tempera ture . Hear t Lung 1985;14:9093.

    [Medline Link]

    51. Cork RC, Vaughan RW, Humphrey LS. Precision and accuracy of intraoperative

    tempera ture moni tor ing . Anes th Analg 1983;62:211214.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    52. Sandstrom K, Nilsson K, Andreasson S, et al . Jugular bulb temperature

    compared with non-invasive temperatures and cerebral arteriovenous oxygen

    saturat ion differences during open heart surgery. Paed Anaesth 1999;9:123128.

    53. Whitby JD, Dunkin LJ. Cerebral oesophageal and nasopharyngeal temperatures.Br J Anaes th 1971 ;43 :673676 .

    [CrossRef]

    [Medline Link]

    54. Grocott H, Newman M, Croughwell N, et al. Continuous jugular venous versus

    nasopharyngeal temperature monitoring during hypothermic cardiopulmonary

    bypass for card iac surgery. J Clin Anes th 1997;9 :312316.

    [CrossRef]

    [Medline Link]

    55. Sakuragi T, Mukai M, Dan K. Deep body temperature during the warming phase

    of card iopulmonary bypass . Br J Anaes th 1993;71:583585.

    [CrossRef]

    [Medline Link]

    56. Bissonnette B. Temperature monitoring in p ediatr ic anesthesia. In: Pulleri ts J ,

    Holzman R, eds. Anesthesia Equipment for Infants and Children. Boston: Lit t le ,

    Brown & Company, 1992 :6376 .

  • 8/11/2019 Chapter 29. Temperature Monitoring

    26/40

    57. Singleton RJ, Ludbrook GL, Webb RK, et al . Physical injuries and

    environmental safety in anaesthesia: an analysis of 2000 incident reports . Anaesth

    In tens Ca re 1993;21 :659663.[Medline Link]

    58. Erickson RS, Ki rklin SK. Comparison of ear-based, bladder, oral and axil lary

    methods for core temperature measurement. Cri t Care Med 1993;21:15281534.

    [CrossRef]

    [Medline Link]

    59. Earp JK, Finlayson DC. Relationship between urinary bladder and pulmonary

    artery temperatures: a prel iminary study. Heart Lung 1991;20:265270.

    [Medline Link]

    60. Grealy B, Peter JV, Moran JL, et al . Is urinary temperature measurement a

    rel iable method for determining body temperature in cri t ical ly i l l pat ients? Anaesth

    In tens Ca re 2001 ;29 :189 .

    61. Earp JK, Finlayson DC. Urinary bladder/pulmonary artery temperature ratio of

    less than 1 and shivering in cardiac surgical pat ients . Am J Cri t Care 1992;1:43

    52.

    [Medline Link]

    62. Cheng W-P, Marino M, Romagnoli A, et al. Temperature measurement during

    CPB. Anes thes io logy 2001;95:A384.63. Imaizumi H, Tsunoda K, Ichimiya N, et al. Urinary bladder temperature

    moni tor ing in extens ive ly burned pa t ien ts . J Cl in Moni t 1993;9 :99100.

    [CrossRef]

    [Medline Link]

    64. Glosten B, Sessler DI, Faure E , et al . Bladder vs tympanic core temperature

    measurement dur ing cesarean sec t ion. Anes thes iology 1990;73:A960.

    65. Holtzclaw BJ. Monitoring body temperature. AACN Clin Issues Cri t Care Nurs

    1 993 ;4 :4 455 .

    [Full text Link]

    [Medline Link]

    66. Shinozaki T, Deane R, Perkins FW. Infrared tympanic thermometer. Evaluation

    of a new c l in ica l thermometer. Cr i t Care Med 1988;16:148150.

    [CrossRef]

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    27/40

    67. Rabinowitz RP, Cookson ST, Wasserman SS, et al . Effects of anatomic si te ,

    oral s t imulation and body posit ion on est imates of body temperature. Arch Intern

    M ed 1 99 6; 15 6: 77 7 78 0.[CrossRef]

    [Medline Link]

    68. Heidenreich T, Giuffre M. Postoperative temperature measurement. Nurs Res

    1 99 0; 39 :1 53 1 55 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    69. Erickson RS, Woo TM. Accuracy of infrared ear thermometry and tradit ional

    temperature methods in young children. Heart Lung 1994;23:181195.

    [Medline Link]

    70. Milewski A, Ferguson KL, Terndrup TE. Comparison of pulmonary artery, rectal,

    and tympanic membrane temperatures in adult intensive care unit pat ients . Clin

    Ped iat r 1991 ;30[Supp l 4 ] :1316 .

    71. Klein DG, Mithell C, Petr inec A, et al . A comparison of pulmonary artery, rectal ,

    and tympanic membrane temperature measurement in the ICU. Heart Lung

    1 99 3; 22 :4 35 4 41 .

    [Medline Link] 72. Kunnel MT, O'Brien C, Munro BH, et al . Comparisons of rectal , femoral ,

    axil lary, and skin-to-mattress temperatures in stable neonates. Nurs Res

    1 98 8; 37 :1 62 1 64 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    73. Kern FH, Greeley WJ. Con: monitoring of nasopharyngeal and rectal

    temperatures is not an adequate guide of brain cooling before deep hypothermic

    c i rcula tory ar res t . J Cardio thor Vasc Anes th 1994;8 :363365.

    74. Merenstein GB. Rectal perforat ion by thermometer. Lancet 1970;1:1007.

    [CrossRef]

    [Medline Link]

    75. Di Paola I , Macneil P . Rectal thermometer complicat ion. Anaesth Intens Care

    1 985 ;13 :4 41 .

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    28/40

    76. Greenbaum EI, Carson M, Kincannon WN, et al. Rectal thermometer-induced

    pneumoper i toneum in the newborn . Pedia t r ics 1989;44:539542.

    [Medline Link] 77. Kober A, Scheck T, Flesdi B, et al . Effect iveness of resist ive heating

    compared with passive warming in treating hypothermia associated with minor

    t rauma: a randomized t r ia l . Mayo Cl in Proc 2001;76:369375.

    [Medline Link]

    78. Brinnel H, Cabanac M. Tympanic temperature is a core temperature in humans.

    J The rm Bio l 1989 ;14 :4753 .

    [CrossRef]

    79. Walpoth BH, Galdikas J , Leupi F, et al . Assessment of hypothermia with a new

    tympanic thermometer. J Cl in Moni t 1994;10:9196.

    [Medline Link]

    80. Zweifler RM, Voorhees ME, Mahmood A, et al . Rectal temperature reflects

    tympanic temperature during mild induced hypothermia in nonintubated subjects . J

    Neurosurg Anes thes io l 2004;16:232235.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    81. Webb GE. Comparison of esophageal and tympanic temperature monitoringdur ing card iopulmonary bypass . Anes th Analg 1973;52:729733.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    82. Ferrara-Love R. A comparison of tympanic and pulmonary artery measures of

    core tempera tures . J Pos t Anes th Nurs 1991;6:161164.

    [Medline Link]

    83. Sato KT, Kane NL, Soos G, et al . Reexamination of tympanic membrane

    tempera ture as a core tempera ture . J Appl Phys io l 1996;80:12331239.

    [Medline Link]

    84. Nishikori T, Atarashi K, Negishi C, et al . Bladder temperature recording is a

    reliable method for core temperature measurement during laparoscopic

    cholecys tec tomy. Anes thes io logy 2001;95:A548.

    85. Pujol A, Fusciardi J , Ingrand P, et al . Afterdrop after hypothermic

    cardiopulmonary bypass: the value of tympanic membrane temperature monitoring.

    J Cardio thorac Vasc Anes th 1996;10:336341.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    29/40

    [CrossRef]

    [Medline Link]

    86. Ogawa T, Sugenoya J , Ohnishi N, e t al . Effects of b ody and head posi t ions onbi la tera l d i fference in tympanic tempera ture . Eur J Appl Phys io l 1993;67:354359.

    [CrossRef]

    [Medline Link]

    87. Tabor MW, Blaho DM, Schriver WR. Tympanic membrane perforation.

    Complication of tympanic thermometry during general anesthesia. Oral Surg

    1 98 1; 51 :5 81 5 83 .

    88. Wallace CT, Marks WE, Adkins W Y, et al . Perforat ion of the tympanic

    membrane. A complication of tympanic thermometry during anesthesia.

    Anes th es iolog y 1974 ;41 :2 90 29 1 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    89. Schuman AJ. Tympanic thermometry: temperatures without tears. Contemp

    P ed ia tr 1 9 91 ;8 :5 4 73 .

    [Medline Link]

    90. Jakobsson J , Nilsson A, Carlson L. Core temperature measured in the auricular

    canal: comparison between four different tympanic thermometers. Acta Anaes th es io l Sc and 199 2;36 :8 19 824.

    [Medline Link]

    91. Matsukawa T, Ozaki M, Hanagata K, et al. A comparison of four infrared

    tympanic thermometers with tympanic membrane temperatures measured by

    thermocouples. Can J Anaes th 1996;43:12241228.

    [Medline Link]

    92. Pransky SM. The impact of technique and conditions of the tympanic membrane

    upon inf rared tympanic thermometry. Cl in Pedia t r 1993;30:5052.

    [Medline Link]

    93. Terndrup T, Rajk J. Impact of operator technique and device on infrared

    emission detect ion tympanic thermometry. J Emerg Med 1992;10:683688.

    [CrossRef]

    [Medline Link]

    94. Tourangeau A, MacLeod F, Breakwell M. Tap in on ear thermometry. Can Nurs

    1 993 ;1 2:2 42 8.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    30/40

    95. Alexander D, Terndrup T. Temperature measurement. Pediatrics 1992;90:649

    650.

    [Medline Link] 96 . Freed GL, Fra ley JK. In reply. Pedia t rics 1992;90:650.

    97. Latman NS, Hans P, Nicholson L, et al . Evaluation of cl inical thermometers for

    accuracy and re l iab i l i ty. Biomed Ins t rum Technol 2001;35:259265.

    [Medline Link]

    98. Muma BK, Treloar DJ, Wurmlinger K, et al . Comparison of rectal , axil lary, and

    tympanic membrane temperatures in infants and young children. Ann Emerg Med

    1 991 ;2 0:4 14 4.

    [CrossRef]

    [Medline Link]

    99. Yetman RJ, Coody DK, West MS, et al. Comparison of temperature

    measurements by an aural infrared thermometer with measurements by traditional

    rec ta l and axi l lary techniques . J Pedia t r 1993;122:769773.

    [CrossRef]

    [Medline Link]

    100. Petersen-Smith A, Barber N, Coody DK, et al. Comparison of aural infrared

    with tradit ional rectal temepratures in chi ldren from bir th to age three years. J

    Ped iat r 1994 ;125 :8385 .[Medline Link]

    101. Brennan DF, Falk JL, Rothrock SG, et al . Reliabil i ty of infrared tympanic

    thermometry in the detect ion of rectal fever in children. Ann Emerg Med

    1 995 ;2 5:2 13 0.

    [CrossRef]

    [Medline Link]

    102. Terndrup TE, Wong A. Influence of ot i t is media on the correlat ion between

    rec ta l and audi tory canal tempera tures . Am J Dis Chi ld 1991;145:7578.

    [Medline Link]

    103. Green MM, Danzl DF, Praszkier H. Infrared tympanic thermography in the

    emergency depar tment . J Emerg Med 1989;7 :437440.

    [CrossRef]

    [Medline Link]

    104. Terndup TE, Melewski A. The performance of two tympanic thermometers in a

    pedia t r ic emergency depar tment . Cl in Ped 1991;30(suppl ):1823.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    31/40

    105. Bernardo L, Clemence B, Henker R, et al . A c omparison of aural and rectal

    temperature measurements in moderately and severely injured children. J Emerg

    N ur s 1 99 6; 22 :4 03 4 08 .[Full text Link]

    [CrossRef]

    [Medline Link]

    106. Chamberlain JM, Grander J , Rubinoff JL, et al . Comparison of a t ympanic

    thermometer to rectal and oral thermometers in a pediatric emergency department.

    C l in Ped 1991;30 [Supp l 4 ] :2429 .

    107. Stewart JV, Webster D. Re-evaluation of the tympanic thermometer in the

    emergency department . Ann Emerg Med 1992;21:158160.

    [CrossRef]

    [Medline Link]

    108. Patel N, Smith CE, Pinchak AC, et al . Comparison of esophageal , tympanic,

    and forehead skin temperatures in adult patients . J Clin Anes 1996;8:462468.

    109. Romano MJ, Fortenberry JD, Autreu E, et al. Infrared tympanic thermometry in

    the pedia t r ic in tensive care uni t . Cri t Care Med 1993;21:11811185.

    [CrossRef]

    [Medline Link]

    P.869

    110. Rotello LC, Crawford L, Terndrup TE. Comparison of infrared ear thermometer

    derived and equilibrated rectal temperatures in estimating pulmonary artery

    tempera tures . Cr i t Care Med 1996;24:15011506.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    111. Summers S. Axillary, tympanic and esophageal temperature measurement:

    descript ive comparisons in postanesthesia pat ients . J Postanesth Nurs 1991;6:420

    425.

    [Medline Link]

    112. Kenney RD, Fortenberry JD, Surrat t SS, et al . Evaluation of an infrared

    tympanic membrane thermometer in pediatric patients. Pediatrics 1990;85:854

    858.

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    32/40

    113. Giuliano KK, Scott SS, Ell iot S, et al . Temperature measurement in cri t ical l y i l l

    oral ly intubated adults: a c omparison of pulmonary artery c ore, tympanic, and oral

    methods . Cri t Care Med 1999;27(10) :21882193.[Full text Link]

    [CrossRef]

    [Medline Link]

    114. Cupitt JM, Badsha Z. Temperature measurementswhich method is best?

    Anaes th es ia 2002;57 :61 9.

    115. Darm RM, Hecker RB, Rubal BJ. A comparison of noninvasive body

    temperature monitoring devices in the PACU. J Post Anesth Nurs 1994;9:144

    149.

    [Medline Link]

    116. Davis K. The accuracy of tympanic temperature measurement in children.

    Ped iat r Nur s 1993;19 :267272.

    [Medline Link]

    117. Kelly B , Alexander D. Effect of ot i t is media on infrared tympanic thermometry.

    C l in Ped 1991;30 [Supp l 4 ] :4648 .

    118. Lattavo K, Britt J, Dobal M. Agreement between measures of pulmonary artery

    and tympanic temperatures . Res Nurs Health 1995;18:365370.

    [CrossRef] [Medline Link]

    119. Weiss ME. Tympanic infrared thermometry for full-term and pre-term neonates.

    Cl in Pedia t r 1991;30[Suppl 4] :4245.

    120. Imamura M, Matsukawa T, Ozaki M, et al. The accuracy and precision of four

    infrared aural canal thermometers during cardiac surgery. Acta Anaesthesiol Scand

    1 99 8; 42 :1 22 2 12 26 .

    [Medline Link]

    121. Craig JV, Lancaster GA, Taylor S, et al. Infrared ear thermometry compared

    with rectal thermometry in children: a systematic review. Lancet 2002;360:603

    609.

    [CrossRef]

    [Medline Link]

    122. Childs C, Harrison R, Hodkinson C. Tympanic membrane temperature as a

    measure of core tempera ture . Arch Dis Chi ld 1999;80:262266.

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    33/40

    123. Bock M, Hohlfeld U, von Engeln K, et al. The accuracy of a new infrared ear

    thermometer in patients undergoing cardiac surgery. Can J Anesth 2005;52:1083

    1087.124. Doyle F, Zehner WJ, Terndrup TE. The effect of ambient temperature extremes

    on tympanic and oral tempera tures . Am J Emerg Med 1992;10:285289.

    [CrossRef]

    [Medline Link]

    125. Zehner WJ, Terndrup TE. The impact of moderate ambient temperature

    variance on the relat ionship between oral rectal and tympanic membrane

    tempera tures . Cl in Ped 1991;30[Suppl 4] :6164.

    126. Thomas KA, Savage MV, Brengelmann GL. Effect of facial cooling on tympanic

    tempera ture . Am J Cr i t Care 1997;6 :4651.

    [Medline Link]

    127. Kenney RD, Fortenberry JD, Surrat t SS, et al . Evaluation of an infrared

    tympanic membrane thermometer in pediatric patients. Pediatrics 1990;85:854

    858.

    [Medline Link]

    128. Sessler DI. Temperature monitoring and management during neuraxial

    anes thes ia . Anes th Analg 1999;88:243245.

    [Full text Link] [CrossRef]

    [Medline Link]

    129. Allen GC, Horrow JC, Rosenberg H. Does forehead l iquid crystal temperature

    accura te ly ref lec t core tempera ture? Can J Anaes th 1990;37:659662.

    130. Brull SJ, Cunningham AJ, Connelly NR, et al . L iquid crystal skin thermometry:

    an accura te re f lec tion of core tempera ture? Can J Anaes th 1993;40:375381.

    131. Lacoumenta S, Hall GM. Liquid c rystal thermometry during anaesthesia.

    Anaes th es ia 1984;39 :54 5 6.

    [CrossRef]

    [Medline Link]

    132. Vaughan MS, Cork RC, Vaughan RW. Inaccuracy of liquid crystal thermometry

    to identify core temperature trends in postoperat ive adults . Anesth Analg

    1 98 2; 61 :2 84 2 87 .

    [CrossRef]

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    34/40

  • 8/11/2019 Chapter 29. Temperature Monitoring

    35/40

    142. Mayfield SR, Bhatia J, Nakamura KT, et al. Temperature measurement in term

    and pre term neonates . J Pediat r 1984;104:271275.

    [CrossRef] [Medline Link]

    143. Anagnostakis D, Matsaniotis N, Grafakos S, e t al . Rectal-axil lary temperature

    difference in febri le and afebri le infants and chi ldren. Clin Pediatr 1993;32:268

    272.

    [Medline Link]

    144. Kresch MJ. Axil lary temperature as a screening test for fever in children. J

    Ped iat r 1984 ;104 :596599.

    [CrossRef]

    [Medline Link]

    145. Konopad E, Kerr JR, Noseworthy T, et al . A comparison of oral , axil lary, rectal

    and tympanic-membrane temperatures of intensive care patients with and without

    ora l endot racheal tube . J Adv Nurs 1994;20:7784.

    [CrossRef]

    [Medline Link]

    146. Morley CJ, Hewson PH, Thornton AJ, et al. Axillary and rectal temperature in

    ambula tory and hospi ta l ized chi ldren . Arch Dis Chi ld 1992;67:122125.

    [Medline Link] 147. Ogren JM. The inaccuracy of axil lary temperatures measured with an

    elec tronic thermometer. Am J Dis Chi ld 1990;144:109111.

    [Medline Link]

    148. Ward CF. New temperature monitoring guidelines: an observation and caveat.

    Anes th es iolog y 1999 ;91 :3 25 32 6 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    149. Weisse ME, Reagan MS, Boule L, et al . Axil lary vs rectal temperature in

    ambula tory and hospi ta l ized chi ldren . Pedia t r Infec t Dis J 1991;10:541542.

    [CrossRef]

    [Medline Link]

    150. Franceschi VT. Accuracy and f easibi l i ty of measuring oral temperature in

    cr i t ica l ly i l l adul ts . Focus Cr i t Care 1991;18:221228.

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    36/40

    151. Erickson RS. Comparison of tympanic and o ral temperatures in surgical

    pa ti ent s . Nurs Res 1991 ;40 :9093.

    [Full text Link] [CrossRef]

    [Medline Link]

    152. Talo H, Macknin ML Medendorp SV. Tympanic membrane temperatures

    compared to rectal and oral temperatures. Clin Ped 1991;30(suppl):3033.

    153. Hooker EA, Houston H. Screening for fever in an adult emergency department:

    ora l vs tympanic thermometry. South Med J 1996;89:230234.

    [Full text Link]

    [Medline Link]

    154. Fall is WM. Oral measurement of temperature in oral ly intubated cri t ical care

    pat ien ts : s ta te -of- the-sc ience review. Am J Cr i t Care 2000;9 :334343.

    [Medline Link]

    155. Tandberg D, Sklar D. Effect of tachypnea on the estimation of body

    tempera ture by an ora l thermometer. New Engl J Med 1983;303:945946.

    [Medline Link]

    156. Hayes JK, Collette DJ, Peters JL, et al. Monitoring body-core temperature

    from the trachea: comparison between pulmonary artery, tympanic, esophageal and

    rec ta l tempera tures . J Cl in Moni t 1996;12:261269.[CrossRef]

    [Medline Link]

    157. Matsukawa T, Sessler DI, Ozaki M, et al. Comparison of distal oesophageal

    temperature with deep and tracheal temperatures. Can J Anaesth 1997;44:433

    438.

    [Medline Link]

    158. Matsukawa T, Ozaki M, Sessler D, et al. Accuracy and precision of deep

    sternal and tracheal temperatures at high- and low-fresh-gas f lows. Br J Anaesth

    1 99 8; 81 :1 71 1 75 .

    [Medline Link]

    159. Yamakage M, Kawana S, Wantanabe H, et al. The utility of tracheal

    tempera ture moni toring . Anes th Analg 1993;76:795799.

    [Full text Link]

    [CrossRef]

    [Medline Link]

  • 8/11/2019 Chapter 29. Temperature Monitoring

    37/40

    160. Bliss-Holtz J . Comparison of rectal , axil lary, and inguinal temperatures in ful l-

    te rm newborn infants . Nurs Res 1989;38:8587.

    [Full text Link] [CrossRef]

    [Medline Link]

    161. Matsukawa T, Goto T, Ozaki M, et al. Core temperature monitoring with new

    venti la tory devices . Anes th Analg 2003;96:16881691.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    162. Wadhwa A, Sessler DI, Sengupta P, et al. Core temperature measurements

    through a new airway device, perilaryngeal airway (CobraPLA). J Clin Anesth

    2 00 5; 17 :3 58 3 62 .

    [CrossRef]

    [Medline Link]

    163. Dickey WT, Ahlgren EW, Stephen CR. Body temperature monitoring via the

    tympanic membrane . Surgery 1970;67:981984.

    [Medline Link]

    164. Wolfson JS. Rectal perforation in infants by thermometer. AJDC

    1 99 6; 111: 19 7 20 0.[Medline Link]

    165. Will iams DJ, Kelleher AA. Defective nasopharyngeal temperature probes.

    Anaes th es ia 2002;57 :12 23 .

    [Full text Link]

    [CrossRef]

    [Medline Link]

    166. Parker EO. Electrosurgical burn at the si te of an esophageal temperature

    probe . Anes thes io logy 1984;61:9395.

    [Medline Link]

    167. Schneider AJL, Apple HP, Braun RT. Electrosurgical burns at skin temperature

    probes . Anes thes io logy 1977;47:7274.

    [Full text Link]

    [CrossRef]

    [Medline Link]

    168. Wald AS, Mazzia VDB, Spencer FC. Accidental burns. JAMA 1971;217:916

    921.

  • 8/11/2019 Chapter 29. Temperature Monitoring

    38/40

  • 8/11/2019 Chapter 29. Temperature Monitoring

    39/40

    For the fol lowing quest ions, answer

    i f A, B, and C are correct i f A and C a re correct i f B and D a re correct i f D is correct i f A, B, C, and D are correct .

    1. Advantages of thermistors include

    A. Sma ll s ize

    B. Interchangeable and disposable probes

    C. Continuous readings

    D. Rapid response t ime

    View Answer 2. With the platinum wire thermometer

    A. Cu rr ent f lo ws in prop ort ion to th e tem pe ra ture d if fe re nce

    B. Rapid thermal equil ibrat ion is possible because of the small diameter of the wire

    C. Two wires of different metals are welded together at their ends

    D. Resistance of the wire varies with temperature

    View Answer 3. Advantages of the liquid crystal thermometers include

    A. Fa s t cont in uous re ad ing sB. Infrared lamps do not interfere with their readings

    C. They can be applied prior to induction

    D. Accuracy even at the extremes of ambient temperature

    View Answer 4. Advantages of infrared temperature monitors include

    A. Th ey do no t con tac t th e ea rd rum di re c tl y

    B. Tolerated well by patients

    C. Rapid measurement

    D. Continuous measurement

    View Answer 5. The accuracy of axillary temperature determinations is

    influenced by

    A. Prox imi ty to the ax ill ary artery

    B. Perfusion of the sk in

    C. Skin contact with the probe

    D. Patient age

    View Answer 6. Factors influencing the temperature reading in the naso-

    pharynx include

  • 8/11/2019 Chapter 29. Temperature Monitoring

    40/40

    A. Te mp eratu re of th e inspi red ga ses

    B. Leak in the tracheal tube cuff

    C. Gastr ic suctionD. Temperature in the hypothalamus

    View Answer 7. For correct use of the esophageal temperature monitor,

    A. Th e sen sor shoul d be lo cated in the lo we r third to lowe r fou rth of the esop hag us

    B. Stomach placement of the probe will provide temperatures lower than core

    C. The probe should be placed 12 to 16 cm dis tal to the point of maximum heart

    sounds

    D. Temperature of the respired gases is not a factor in the lower third of the

    esophagus

    View Answer 8. Factors that can cause esophageal temperatures to be

    unreliable include

    A. Airwa y hum idificat ion

    B. Thoracic surgery

    C. Continuous gastr ic suction

    D. Patient in the si t t ing posi t ion

    View Answer 9. When using the tympanic membrane to measure

    temperature,

    A. Th e pro be do es no t hav e to ac tu al ly contac t the me mb ra neB. Perforat ion of the membrane can occur

    C. Readings wil l be stable if the probe is in c lose proximity to the membrane

    D. The probe should have a means to hold i t in place

    View Answer 10. Factors that will cause incorrect readings from an

    infrared ear thermometer include

    A. Ce ru men

    B. Ambient air temperature

    C. Oti t is media with suppuration

    D. Oti t is media without suppuration

    View Answer 11. Factors affecting the temperature measured in the

    rectum include

    A. Cysto scopy

    B. Peri toneal lavage

    C. Rectal contents

    D. Colon pathology

    View Answer