Development of an Intraocular Pressure Measurement System Alex Phan, Phuong Truong, Alexander Kief, Milien Dhorne, Andrew Camp, Robert N. Weinreb, and Frank E. Talke Normal Vision Intermediate Symptom Advanced Symptom To date, elevated intraocular pressure (IOP) continues to be the primary risk factor for glaucoma due to its association with optic nerve damage and irreversible blindness. Current standard care, which involves routine IOP measurements during office visits, can only provide snapshots of the patient's IOP profile. In order to improve glaucoma care and determine the effectiveness of therapeutic treatments, there is a pressing need for frequent and reliable IOP data to better understand the relationship between elevated eye pressure and optic nerve damage. The objective of this project is to develop an optical pressure sensor that allows monitoring of the IOP on a frequent or semi-continuous basis. Motivation Lens Beam Splitter Monochromatic Light Source Camera Capsular bag IOP Sensor Optical Reader Iris Cornea Optic nerve Retina SiN diaphragm on Si substrate Epoxy SU-8 Spacer Glass Substrate 10 μm 0.4 mm Intraocular Pressure Measurement System The measurement system consists of a hand-held reader designed to capture interferometric patterns from the implanted optical sensor. As shown in the figure above, the sensor is integrated onto an intraocular lens and implanted during standard cataract surgery. By directing the monochromatic light emitted from the reader towards the sensor, one can obtain the interference pattern and, consequently, the intraocular pressure. Optical Sensing Concept The sensor is designed based on the principle of interferometry. When monochromatic light is directed towards the sensor, reflected light waves from the bottom surface of the SiN diaphragm interfere with the reflected waves from the top surface of the SiN-coated glass substrate creating bright and dark fringes. The optical read-out of the sensor is captured using a camera and post-processed using image processing algorithm to correlate fringes pattern with pressure. Experimental Setup and Image Processing Sensor Fabrication 100μm a b c d Ex-vivo Study Hand-held Reader 200μ m 30 mmHg 40 mmHg 50 mmHg 200μ m 20 mmHg 40 mmHg 60 mmHg 200μm Benchtop Handheld 500 μm 5 mm a 5 mm b In-vivo Study To demonstrate the feasibility of the system for point-of-care applications, the benchtop microscope set-up was replaced with a portable hand-held reader. The hand-held reader with a CMOS camera attachment can be directed towards the sensor implanted in the rabbit eye. The same fringe pattern can be observed in both the benchtop and the hand- held device, indicating the successful use of a portable readout system. The sensor was integrated onto two types of intraocular lenses and surgically implanted into a rabbit eye using standard cataract surgery procedures. The results show that the intraocular pressure measurement system has a sensitivity of 22 nm/mmHg and an accuracy as high as ±0.5 mmHg post implantation. Conclusion In-vivo study is currently in progress. Sensors were integrated onto IOLs with reduced foot print to minimized surgical risk. The goals of the study are: (1) evaluate the biostability and biocompatibility of the sensor and (2) investigate the feasibility of the hand-held reader in a clinical setting by integrating it with a standard slit lamp . The response of the sensors was characterized using an artificial anterior chamber (Katena) to simulate the pressure condition in the eye. Interference fringe patterns were obtained as a function of the applied pressure and captured by the CMOS camera for analysis. A custom-made image processing algorithm was developed in MATLAB software to analyze the fringe data. (a) Symmetric filtered fringe pattern (b) positions of bright and dark fringes (c) 2-D deflection profile across the diaphragm (d) 3-D surface reconstruction of the diaphragm deflection profile. A balance salt solution (BSS) column was connected to a pressure regulator and used to precisely control the hydrostatic pressure (±0.1 mmHg) inside the artificial anterior chamber. Spin coat 10 μm SU-8 Pattern SU-8 to define cavity PECVD 80 nm SiN Seal with medical grade epoxy Align and bond with SiN diaphragm Glass substrate The sensor is fabricated by bonding a glass substrate with an SU-8 spacer to a silicon nitride diaphragm as shown in the process flow illustration. Once bonded using medical- grade epoxy to form a hermetic and watertight seal, the sensor was miniaturized to 1.5 x 1.5 mm using saw dicing to remove the excess sensor footprint. 500 μm SEM image of cross section of fabricated sensor Photographs of sensor using (a) white light and (b) monochromatic light Overall, our results show a very promising approach to help monitor eye pressure in glaucoma patients. As a patient point-of-care technology, the proposed approach will enable patients to obtain accurate and more frequent measurements of their eye pressure at the convenience of their home. Frequent measurement data will equip ophthalmologists with the information necessary to make timely therapeutic interventions and improve treatment plans to preserve vision of glaucoma patients. Acknowledgements : We would like thank Ben Suen, Ray Descoteaux, and the UCSD Nano3 facility for their support and advice in fabrication and experimental set-up. We would also like to thank Dr. Gerrit Melles for helpful discussions. Finally, we would like to acknowledge the Shiley Eye Center at UCSD for their great support and use of their facilities. This research was supported in part by the Glaucoma Research Foundation and the National Research Foundation. 3mm

Development of an Intraocular Pressure Measurement System...Development of an Intraocular Pressure Measurement System Alex Phan, Phuong Truong, Alexander Kief, Milien Dhorne, Andrew

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Development of an Intraocular Pressure Measurement System...Development of an Intraocular Pressure Measurement System Alex Phan, Phuong Truong, Alexander Kief, Milien Dhorne, Andrew

Development of an Intraocular Pressure

Measurement SystemAlex Phan, Phuong Truong, Alexander Kief, Milien Dhorne, Andrew Camp, Robert N. Weinreb, and Frank E. Talke

Normal Vision Intermediate Symptom Advanced Symptom

To date, elevated intraocular pressure (IOP) continues to be the primary risk factor forglaucoma due to its association with optic nerve damage and irreversible blindness.Current standard care, which involves routine IOP measurements during office visits, canonly provide snapshots of the patient's IOP profile. In order to improve glaucoma care anddetermine the effectiveness of therapeutic treatments, there is a pressing need forfrequent and reliable IOP data to better understand the relationship between elevated eyepressure and optic nerve damage. The objective of this project is to develop an opticalpressure sensor that allows monitoring of the IOP on a frequent or semi-continuous basis.

Motivation

Lens

Beam Splitter

Monochromatic

Light Source

Camera

Capsular bag

IOP

Sensor

Optical Reader

Iris

Cornea

Optic nerve

Retina

SiN diaphragm on

Si substrate

Epoxy

SU-8 Spacer

Glass

Substrate

10 µm0.4 mm

Intraocular Pressure Measurement System

The measurement system consists of a hand-held reader designed to captureinterferometric patterns from the implanted optical sensor. As shown in the figure above,the sensor is integrated onto an intraocular lens and implanted during standard cataractsurgery. By directing the monochromatic light emitted from the reader towards the sensor,one can obtain the interference pattern and, consequently, the intraocular pressure.

Optical Sensing Concept

The sensor is designed based on the principle ofinterferometry. When monochromatic light is directedtowards the sensor, reflected light waves from thebottom surface of the SiN diaphragm interfere with thereflected waves from the top surface of the SiN-coatedglass substrate creating bright and dark fringes. Theoptical read-out of the sensor is captured using acamera and post-processed using image processingalgorithm to correlate fringes pattern with pressure.

Experimental Setup and Image Processing

Sensor Fabrication

100µm

a b

c d

Ex-vivo Study

Hand-held Reader

200µ

m

30

mmHg

40

mmHg

50

mmHg

200µ

m

20

mmHg

40

mmHg

60

mmHg

200µm

Benchtop

Handheld

500 µm

5 mm

a

5 mm

b

In-vivo Study

To demonstrate the feasibility of the system forpoint-of-care applications, the benchtop microscopeset-up was replaced with a portable hand-heldreader. The hand-held reader with a CMOS cameraattachment can be directed towards the sensorimplanted in the rabbit eye. The same fringe patterncan be observed in both the benchtop and the hand-held device, indicating the successful use of aportable readout system.

The sensor was integrated onto two types ofintraocular lenses and surgically implanted intoa rabbit eye using standard cataract surgeryprocedures. The results show that theintraocular pressure measurement system hasa sensitivity of 22 nm/mmHg and an accuracyas high as ±0.5 mmHg post implantation.

Conclusion

In-vivo study is currently in progress. Sensors wereintegrated onto IOLs with reduced foot print tominimized surgical risk. The goals of the study are:(1) evaluate the biostability and biocompatibility ofthe sensor and (2) investigate the feasibility of thehand-held reader in a clinical setting by integratingit with a standard slit lamp .

The response of the sensors was characterized using an artificial anterior chamber(Katena) to simulate the pressure condition in the eye. Interference fringe patterns wereobtained as a function of the applied pressure and captured by the CMOS camera foranalysis. A custom-made image processing algorithm was developed in MATLABsoftware to analyze the fringe data.

(a) Symmetric filtered fringe pattern (b) positions of bright anddark fringes (c) 2-D deflection profile across the diaphragm (d) 3-Dsurface reconstruction of the diaphragm deflection profile.

A balance salt solution (BSS) column was connected to a pressureregulator and used to precisely control the hydrostatic pressure(±0.1 mmHg) inside the artificial anterior chamber.

Spin coat 10 µm SU-8

Pattern SU-8 to define cavity

PECVD 80 nm SiN

Seal with medical grade epoxy

Align and bond with SiN diaphragm

Glass substrate

The sensor is fabricated by bonding a glass substrate with an SU-8 spacer to a siliconnitride diaphragm as shown in the process flow illustration. Once bonded using medical-grade epoxy to form a hermetic and watertight seal, the sensor was miniaturized to 1.5 x1.5 mm using saw dicing to remove the excess sensor footprint.

500 µm

SEM image of cross section of fabricated sensor

Photographs of sensor using (a) white light and (b) monochromatic light

Overall, our results show a very promising approach to help monitor eye pressure inglaucoma patients. As a patient point-of-care technology, the proposed approach willenable patients to obtain accurate and more frequent measurements of their eyepressure at the convenience of their home. Frequent measurement data will equipophthalmologists with the information necessary to make timely therapeuticinterventions and improve treatment plans to preserve vision of glaucoma patients.

Acknowledgements: We would like thank Ben Suen, Ray Descoteaux, and the UCSD Nano3 facility for their support and advice in fabrication and experimentalset-up. We would also like to thank Dr. Gerrit Melles for helpful discussions. Finally, we would like to acknowledge the Shiley Eye Center at UCSD for their greatsupport and use of their facilities. This research was supported in part by the Glaucoma Research Foundation and the National Research Foundation.