General detectors

Page 2: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

CCDs• Charge Coupled Devices invented in the 1970s• Sensitive to light from optical to X-rays• In practice, best use in optical and X-rays• CCDs make use of silicon chips• The CCD consists of (1) a p-type doped silicon

substrate, (2) the charge storage (depletion) layer, which is covered by (3) a SiO2 insulating layer; upon this is (4) an array of closely spaced electrodes, which can be set to pre-defined voltage value

Page 3: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Si array

Si arrayp-type

Si arrayn-type

Also Sb,P

Also Al,Ga

Page 4: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Electrons in a lattice do not have discrete energies. They form energy bands:

•Valence band•Conduction band

For semi-conductors, the Fermi level is just in the middle of the conduction and valence bands. At finite temperature, some electrons of the valence band can jump into the conduction band (current noise)

EG(Si)=1.1 eV (IR), EG(Ge)=0.72 eVEG(C)=5.5 eV (insulator)

Reminder of solid state physics

photon phot

on

Hole Electron(From http://hyperphysics.phy-astr.gsu.edu/hbase/hframe.html)

Page 5: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

The pn junction

Page 6: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Forward-biased pn junction

Reverse-biased pn junction

Page 7: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Transverse cut of CCD with buried channelBlecha (cours instrumentation)

4.1 eV

2.5 eV

1.8 eV

1.2 eV

Page 8: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

The electrode has positive potential to attract the generated photoelectrons in a potential well

The above MOS capacitor is 1 pixel

Page 9: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Front-illuminated CCDs

p

Inco

min

g ph

oton

s

625m

n

They have a low Quantum Efficiency due to the reflection and absorption of light in the surface electrodes. Very poor blue response. The electrode structure prevents the use of an anti-reflective coating that would otherwise boost performance.

Inco

min

g ph

oton

s

15m np

Anti-reflective coating

The QE can approach 100% . These thinned CCDs become transparent to near infra-red light and the red response is poor. Response can be boosted by the application of an anti-reflective coating on the thinned rear-side. These coatings do not work so well for front-illuminated CCDs due to the surface bumps created by the surface electrodes

Courtesy of S. Tulloch

Page 10: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Quantum Efficiency Comparison

The graph below compares the quantum of efficiency of a thick frontside illuminated CCD and a thin backside illuminated CCD.

Courtesy of S. Tulloch

Page 11: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Back-illuminated CCDs

• Thinner deadlayers higher low-E QE• Thinner active region lower high-E QE• Increased noise, charge transfer inefficiency

higher FWHM

BI CCD

FI CCD

From C. Grant, X-ray Astronomy School 2007

Page 12: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Structure of a CCD

One pixel

Channel stops to define the columns of the image

Transparenthorizontal electrodesto define the pixels vertically. Also used to transfer the charge during readout

Plan View

Cross section

The diagram shows a small section (a few pixels) of the image area of a CCD. This pattern is reapeated.

ElectrodeInsulating oxiden-type silicon

p-type silicon

Every third electrode is connected together. Bus wires running down the edge of the chip make the connection. The channel stops are formed from high concentrations of Boron in the silicon.

Courtesy of S. Tulloch

Page 13: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

RAIN (PHOTONS)

BUCKETS (PIXELS)

VERTICALCONVEYORBELTS(CCD COLUMNS)

HORIZONTALCONVEYOR BELT(SERIAL REGISTER)

MEASURING CYLINDER(OUTPUT AMPLIFIER)

CCD Analogy

Courtesy of S. Tulloch

Page 14: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Exposure finished, buckets now contain samples of rain.

Courtesy of S. Tulloch

Page 15: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Conveyor belt starts turning and transfers buckets. Rain collected on the vertical conveyoris tipped into buckets on the horizontal conveyor.

Courtesy of S. Tulloch

Page 16: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Vertical conveyor stops. Horizontal conveyor starts up and tips each bucket in turn intothe measuring cylinder .

Courtesy of S. Tulloch

Page 17: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

`

After each bucket has been measured, the measuring cylinderis emptied , ready for the next bucket load.

Courtesy of S. Tulloch

Page 18: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 19: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 20: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 21: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 22: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 23: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 24: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

A new set of empty buckets is set up on the horizontal conveyor and the process is repeated.

Courtesy of S. Tulloch

Page 25: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 26: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 27: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 28: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 29: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 30: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 31: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 32: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 33: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 34: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 35: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 36: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 37: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 38: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 39: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 40: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Courtesy of S. Tulloch

Page 41: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Eventually all the buckets have been measured, the CCD has been read out.

Courtesy of S. Tulloch

Page 42: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

pixe

l bo

unda

ry

Charge packetp-type siliconn-type silicon

SiO2 Insulating layer

Electrode Structure

pixe

l bo

unda

ry

inco

min

gph

oton

s

Charge Collection in a CCD.

Photons entering the CCD create electron-hole pairs. The electrons are then attracted towards the most positive potential in the device where they create ‘charge packets’. Each packet corresponds to one pixel

Courtesy of S. Tulloch

Page 43: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Charge Transfer in a CCD 1.

In the following few slides, the implementation of the ‘conveyor belts’ as actual electronicstructures is explained.

The charge is moved along these conveyor belts by modulating the voltages on the electrodespositioned on the surface of the CCD. In the following illustrations, electrodes colour coded redare held at a positive potential, those coloured black are held at a negative potential.

1

23

Courtesy of S. Tulloch

Page 44: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

Time-slice shown in diagram

1

2

3

Charge Transfer in a CCD 2.

Courtesy of S. Tulloch

Page 45: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 3.

Courtesy of S. Tulloch

Page 46: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 4.

Courtesy of S. Tulloch

Page 47: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 5.

Courtesy of S. Tulloch

Page 48: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 6.

Courtesy of S. Tulloch

Page 49: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 7.

Charge packet from subsequent pixel entersfrom left as first pixel exits to the right.

Courtesy of S. Tulloch

Page 50: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

1

23

+5V

0V

-5V

+5V

0V

-5V

+5V

0V

-5V

1

2

3

Charge Transfer in a CCD 8.

Courtesy of S. Tulloch

Page 51: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Photomultipliers

Gain of 107-109

From Bruker-AXS

Page 52: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Semi-conductor detectors• See introduction on semiconductors in previous slides

(conduction/valence bands)• Gamma-ray interaction produces secondary electrons; they

produce electron-hole pairs in the conduction/valence bands• Electric field separates the pairs before recombination, drifting

electrons to anode and holes to cathode. Charge is collected, which is proportional to energy deposited in detector

• Energy required to generate electron-hole pair is =(14/5)Eg+c, where 0.5≤c≤ eV

• Common detectors made of Ge (e.g., Integral SPI; Eg=0.74 eV, =2.98 eV), Si (=3.61 eV), CdTe (Integral ISGRI; Eg=1.6eV, =4.43 eV)

Page 53: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Bolometer principles

• Sensitive to energy• Not sensitive to ‘color’• Suited for almost ANY wavelength• Particularly suited for IR and FIR to sub-mm

(but cooling to mK required)

Page 54: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Bolometer priciples

Page 55: General detectors. CCDs Charge Coupled Devices invented in the 1970s Sensitive to light from optical to X-rays In practice, best use in optical and X-rays

Bolometers