Overcoming Attenuation Artifacts: How to use attenuation correction Is SPECT/CT Superior to SPECT? Raffaele Giubbini Chair and Nuclear Medicine Unit University and Spedali Civili Brescia - Italy Overcoming Attenuation Artifacts: How to use attenuation correction Is SPECT/CT Superior to SPECT? Raffaele Giubbini Chair and Nuclear Medicine Unit University and Spedali Civili Brescia - Italy

Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Embed Size (px)

Citation preview

Page 1: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Overcoming Attenuation Artifacts: How to useattenuation correction

Is SPECT/CT Superior to SPECT?

Raffaele GiubbiniChair and Nuclear Medicine UnitUniversity and Spedali Civili

Brescia - Italy

Overcoming Attenuation Artifacts: How to useattenuation correction

Is SPECT/CT Superior to SPECT?

Raffaele GiubbiniChair and Nuclear Medicine UnitUniversity and Spedali Civili

Brescia - Italy

Page 2: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Artifacts in SPECT MPI•Mechanical & electronical

–Energy window–Detector center-of-rotation and alignment –Flood field non-uniformity–Degradation in resolution with distance from the collimator

•Biological–Myocardial hypertrophy –Left bundle branch block –Cardiac position with thorax•Patient-related artifacts

–Soft-tissue attenuation •breast lateral chest-wall fat diaphragm

–Superimposed abdominal visceral activity–Motion artifactsTechnical artifacts

– Selection or cardiac axes – Selection of apex and base for polar map

Page 3: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

As a gamma camera head rotates around the body the attenuation of the projected radiation varies with the amount of tissue and the type of tissue that it had to travel through

ATTENUATION

Page 4: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

The attenuation equation is:Ax = A0e-ux

The most frequently noted effects of attenuation in myocardial SPECT are artifacts associated with breast attenuation in women and diaphragmatic attenuation in men

Attenuation artifacts are more severe using Tl201 than with Tc99m due to Thallium’s lower energies

ATTENUATION

Page 5: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Attenuation causes quantitative errors as well as distortions inthe projection profiles that are propagated into the reconstructed images. Knowledge of the attenuatingdistribution is required for attenuation correction. Solid linesdepict true profile; dotted lines depict attenuated profile.

Page 6: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Soft-tissue attenuation• Soft-tissue attenuator overlie the left ventricle • => decrease in count density • Location: depend on the position of soft-tissue

attenuator • Severity: depend on size, density, energy of

radioisotope, 360 or 180 degree acquisition • artifact: fix, reversible defect or reverse

redistribution

Page 7: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Breast attenuation• Most commonly encountered attenuation

artifact in cardiac SPECT • Extremely variable in appearance • Consider the position and configuration

of the breasts with patients in supine position

• Breast prosthesis

Page 8: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Breast attenuation• Women of average body habitus: anterior,

anteroseptal, anterolateral walls • Women with large, pendulous breasts: lateral

attenuation artifacts • Elderly women with very large, pendulous

breasts: lateral abdominal wall • Not necessarily directly proportional to breast

size, but may vary considerably according to the position, configuration, and density of the breast.

Page 9: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Diaphragmatic attenuation• Left hemidiaphragm, right ventricle

(lesser degree) => decrease count density in inferior wall of left ventricle

• Diaphragmatic elevation: obesity, pleural or pulmonary parenchymal disease, atelectasia, loss of lung volume, diaphragmatic paralysis, gastric dilation.

Page 10: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Diaphragmatic attenuation• Attenuation artifacts usually constant

(fixed) • Ascites, peritoneal dialysis: Upright

exercise => ascites fluid may shift to the pelvis => less marked attenuation than in resting images

• Barium contrast

Page 11: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

• One method of handling attenuation artifacts is simply to reduce their clinical impact through scanning techniques and knowledge of their location and severity when they occur

ATTENUATION CORRECTION

Page 12: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Conventional Correction MethodsThe most common correction methods have been a pre-reconstruction method based on work by Sorensen and a post reconstruction method developed by ChangNot used in myocardial perfusion because the thorax is too varied for a constant attenuation coefficient to be effective

Page 13: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Sealed source transmission scan-based correction methods

• Most common sealed source is gadolinium-135• Used as a sealed line source with a collimator

on top and passed along the patient either after or at the same as the gamma camera is reading

• Important to realize that defects in transmission scan can have a detrimental effect to the attenuation corrected images

Page 14: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

A collimated planar beam of photons traverses the patient forming a complete projection profile as the source moves across the full field of view parallel to the system axis. Within the detector, an electronic mask defines a narrow spatial window” opposite the source and moving in unison. external radionuclide source approach whereby a series of 14

Page 15: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

external radionuclide source approach whereby a series of 14 collimated and shuttered line sources are positioned in an “array”opposite each detector such that the full detector field of view is irradiated simultaneously.

Page 16: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

2 point sources collimated to be incident at oblique angles on the opposing detectors. 133Ba point sources (t1/2 10.5 years, 356 keV gamma emissions) translate along the system axis simultaneously or sequentially with the emission acquisition. The transmission source is positioned such that the beam is incident on the detectors at an angle to the hole axes. Transmission photons penetrate thecollimator septa for detection.

Page 17: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Sealed SourceMany of the SPECT QC procedures transfer over to the transmission imaging systemsYou must check the transmission tomogram for artifacts (motion, gating, missing frames)Other problems arise from old line sources decayed beyond being usable and body truncation

Page 18: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential
Page 19: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

G.A.

Page 20: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Attenuation MapA CT image is a mesure of attenuation profiles in different angular projections

The reconstructed image is a two dimensional map of linear attenuation correction µ(x,y) of the X radiation

Page 21: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

µ(x,y) is depending on the energy of the X ray and it is expressed asHounsfield Units

Attenuation Map

Page 22: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Conversione #CT in funzione delle diverse energie del fascio

COEFFICIENTE LINEARE DI ATTENUAZIONE

0,00

0,05

0,10

0,15

0,20

0,25

0,30

0,35

0,40

-1500 -1000 -500 0 500 1000 1500#CT

µ (c

m-1

)

kV 140kV 100keV 140

Mix aria-acqua Mix osso-acqua

Page 23: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

0

5

10

15

20

0,0 0,1 0,2 0,3 0,4 0,5 0,6Energia (MeV)

Cortical Bone vs Soft TissueCortical Bone vs Inflated LungSoft Tissue vs Inflated Lung

99mTc 511 keV

TC

Rapporto tra i coefficienti di attenuazione lineare in funzione dell’energia del fotone per diversi tessuti

Page 24: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Beam hardening

L’aumento della energia media del fascio di raggi X è funzione:della densità dell’assorbitoredello spessore dell’assorbitore

Correzione mediantecalibrazione

Correzione mediantecalibrazione e correzione

iterativa

Page 25: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Normalization for matrix size

CT scan512×512

SPECT Image 64×64

Attenuation Image 64×64

Page 26: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential
Page 27: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

0 90 180

Page 28: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Motion artifacts: Upward creep• A patient’s rate and depth of respiration increase

markedly during dynamic exercise, resulting in more marked, increased lung volume, and a lower position of the diaphragm. During the acquisition, as the depth of respiration decreases, the height of the diaphragm will progressively rise, with a gradual upward shift of the heart

• Due to abdominal relaxation and modification of diaphragmatic excursion a upward-shift of the heart is commonly observed during acquisition even after rest injection

Page 29: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

uncorr. Stasis MDC MOCO

Page 30: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

A new source of errors:

• Misregistration of Emission and Transmission Scans

Page 31: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Stress RestStress

Page 32: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Stress Rest

Page 33: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential
Page 34: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Single Photon Emission Computerized Tomography (SPECT/CT)

Page 35: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

A multicenter evaluation of commercial attenuation compensation techniques

in cardiac SPECT using phantom models

O'Connor MK et al. J Nucl Cardiol 2002;9:361-76

Page 36: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

A multicenter evaluation of commercial attenuation compensation techniques

in cardiac SPECT using phantom models O'Connor MK et al. J Nucl Cardiol 2002;9:361-76

Page 37: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Clinical validation of SPECT attenuation correction using x-ray CT

Masood Y et al. J Nucl Cardiol 2005;12:676-86

Page 38: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

AC

AC

NC

NC

AC

AC

NC

NCPrevious Inferior MI

Page 39: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

S.A.

Page 40: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

uncorrected corrected

Page 41: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

109 Consecutive pts with inferior MIATN ATN

uncorrecteuncorrectedd

ATN ATN correctedcorrected

SSSSSS 14.0214.02±±7.97.9 9.49.4±±7.17.1

SRSSRS 9.59.5±±77 5.65.6±±6.16.1

SDSSDS 4.54.5±±3.23.2 3.83.8±±2.82.8

✔✔

P<0.001

P<0.001

P<n.s.

Page 42: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential
Page 43: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential
Page 44: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Kaplan-Meier survival curves (end point cardiac death) based on the locationof reversible perfusion abnormalities.

Kaplan-Meier survival curves (end point reinfarction) based on the locationof reversible perfusion abnormalities.

Elhendy A. et al. Am J Cardiol 2004;94:289–293)

Page 45: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

• Attenuation correction in cardiac SPECT: The boy who cried wolf?– G Germano, et al (J Nucl Cardiol 2007;14:25-35)

CONTROVERSIES IN NUCLEAR CARDIOLOGY

Although no one would deny that accurate AC could improve SPECT MPI, the question as to whether current implementations are sufficiently robust remains unanswered, especially with regard to the average user.

Page 46: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Clinical value of attenuation correction instress-only Tc-99m sestamibi SPECT imaging

Gary V.

Heller G et al. J Nucl Cardiol 2004;11:273-81.

Page 47: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

CONTROVERSIES IN NUCLEAR CARDIOLOGY

• SPECT attenuation correction: An essential tool to realize nuclear cardiology’s manifest destiny– E.V. Garcia (J Nucl Cardiol 2007;14:16-24.)

Sensitivity % Specificity % Normalcy %Years Pazienti NC AC NC AC NC AC

totale -media

1996 -2006 1327 86 86 57 77 75 92

AC markedly improves diagnostic accuracy over no AC regardless of :� clinical site,� radionuclide used, � camera manufacturer, � whether quantitative versus qualitative techniques are used, � whether an obese versus a non obese population is studied� whether exercise or pharmacologic stress is used

Page 48: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

What is needed?• Adequate technology• Application specialists aware of technical and physiopatological issues

• A quick automatic quality control on MPI SPECT and CT realignement

• A possibility of manual realignement for SPECT and CT

Page 49: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Shielding (1/16-inch lead) is usually required at the walls along the sides and back of the gantry

Page 50: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

EXPERIMENTAL EVALUATION

LV wall Volume 110 ccExternal souceVolume 150 ccSource-myocardial wall distance 30 - 50 mmUptake ratio LV wall ext.souce

1 01 11 21 4

Pixel size 4.5 mmrange 180°°°°step 3°°°°collimator cast HRBPF Butt. 0.35 - 5Max count density(wall) 95 counts

Page 51: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

1 : 1 1 : 21 : 0 Ratio 1 : 2 Ratio 1 : 4

Page 52: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Results:

Ratio 1 : 1

Ratio 1 : 2

Ratio 1 : 4

Page 53: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Minimize superimposed abdominal visceral activity

• Exercise Tl-201 studies: maximal exercise • Dipyridamle Tl-201 and Tc-99m sestamibistudies: add dynamic, submaximal exercise (walking, biking) – Exercise is performed immediately after the infusion

of dipyridamole, and radiotracer is injected during exercise, at least 1 minute before its cessation.

• Tc-99m sestamibi: wait 60 min, fat meal after injection, 2 glasses of water before imaging

Page 54: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential

Conclusion:• SPECT myocardial perfusion imaging

• Poor specificity caused by image artifacts.

• Recognize the sources of artifacts to avoid misinterpretations

• => Improve the specificity

Page 55: Giubbini - Attenuation correctionnucleus.iaea.org/HHW/NuclearMedicine/CardiovascularandPulmonary/I… · CONTROVERSIES IN NUCLEAR CARDIOLOGY • SPECT attenuation correction: An essential