Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

Embed Size (px)

Citation preview

Page 1: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

Haematology

Page 2: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

ParameterPARAMETER LASERCYTE QBC VETAUTOREAD IMPEDANZ SYSTEM

Hct Yes Yes Yes

Hgb Yes Yes Yes

RBC Yes Yes

Retics Yes Up to 4%

MCV Yes Yes

RDW Yes

MCHC Yes Yes Yes

MCH Yes Yes

WBC Yes Yes Yes

LYM Yes Yes ( Small )

MONO Yes Yes ( Middle )

NEU Yes Yes Yes ( Large )

EOS Yes Dogs and Cows

BASO Yes

PLT Yes Yes Yes

PCT Yes

PDW Yes

MPV Yes Yes

FIBRINOGEN Yes

NRBC’s Yes

Page 3: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

3 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Eosinophils

Most frequent causes

Parasitic and allergic processes

Heartworm, Strongyle, Lung Flukes

Feline asthma, Flea bite dermatitis

Least frequent cause

Neoplasia, Fungal Infecion

Viral Infection ( FeLV )

Eosinopenia

When concurrent with lymphopenia is usually due to stress

Eosinophilia

Page 4: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

4 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

What can cause changes in blood values?

Infections Bacteria Viral Funghi

Inflammation Intoxication Stress Anaemia Tumors Bleeding disorders Autoimmun diseases

Page 5: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

5 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Why run Haematology inhouse?

• Blood samples consist of living cells and should therefore be analysed as quickly as possible, to avoid:

Changes in cells because of storage and transport

To minimise artefacts which can develop if the sample has contact to EDTA for a longer period of time

So treatment decisions are based on todays situation and not yesterdays

Page 6: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

6 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Fresh Sample versus 48hr Sample

Page 7: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

7 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

CBC Red cells (Erythrocytes)

White cells (Leucocytes) Granulocytes

- Neutrophiles - Eosinophiles

- Basophiles Agranulocytes

- Lymphocytes - Monocytes

Platelets (Trombocytes)

Page 8: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

8 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

White differentials

• Typical “2 part” Diff Agranulocytes Granulocytes

• Typical “3 part” Diff Lymphocytes Monocytes Granulocytes

• Typical “5 part” Diff Lymphocytes Monocytes Neutrophile Grans Eosinophile Grans Basophile Grans

Page 9: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

9 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

?What differentials do you get with..

QBC Autoread

LaserCyte

Impedance System

Page 10: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

10 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Why can’t an impedance system give you a 5 part diff SIZE MATTERS !SIZE MATTERS !

Page 11: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

11 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

LaserCyte….

Cells focused into a stream that passes through a laser light that interrogates them.

Light scatter is sensed by four photodiodes. Time of flight measurement also taken. Initial RBC count and platelet count taken. The RBC’s are then lysed using a mild agent. 5 part differential then performed qualiBeads are run with every sample. These act as an internal standard for quality assurance.

Page 12: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

12 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

LaserCyte Technologie...

Page 13: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

13 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

QBC Autoread…

The QBC uses centrifugation to separate cells by density.

The ‘float’ in the QBC tube reduces the volume of lumen in the area of the buffy coat.

This causes the cells to be spread out over a greater distance.

Cells take up the ‘acridine orange’ present in the tube (DNA and RNA/Lipoproteins).

Cell types are differentiated by the location and intensity of dye.

Page 14: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

14 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

QBC VetAutoread

Page 15: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

15 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Coulter® Principle

Impedance counters rely on the Coulter principle. When a particle passes through a narrow orifice, it causes a change in resistance. The magnitude of the change corresponds to cell size and the number of changes corresponds to cell size.

Page 16: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

16 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Interfering Substances3 Part Impedance Technology

Giant / Clumped Platelets Results in erroneously low platelet counts. Erroneously high

WBC count. Can also be misclassified as RBC’s. nRBC’s and unlysed RBC’s

Unable to classify nRBC’s. nRBC’s result in erroneously high WBC count

Excessive eosinophils, metamyelocytes, blasts and plasma cells. Can cause erroneous monocyte count

Leukaemia and chemotherapy Harsh lysing agent can damage fragile cells

Page 17: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

17 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Interfering Substances..

High WBC count WBC’s counted during RBC count. This can effect the red cell

parameters. (Hct, RDW,MCV and RBC) Agglutinated red cells

Can cause erroneously low RBC and can trap platelets. Microcytic cells and fragments

Can effect platelet count

Page 18: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

18 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

Interfering Substances….

How does the operator know if you have abnormal cells present that may affect the results ?

How do they know if the results are due to the clinical condition or erroneous reading caused by abnormal cells?

Would you trust a system that could not flag ‘suspect’ results ?

Page 19: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

19 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

ParameterPARAMETER LASERCYTE QBC VETAUTOREAD IMPEDANZ SYSTEM

Hct Yes Yes Yes

Hgb Yes Yes Yes

RBC Yes Yes

Retics Yes Up to 4%

MCV Yes Yes

RDW Yes

MCHC Yes Yes Yes

MCH Yes Yes

WBC Yes Yes Yes

LYM Yes Yes ( Small )

MONO Yes Yes ( Middle )

NEU Yes Yes Yes ( Large )

EOS Yes Dogs and Cows

BASO Yes

PLT Yes Yes Yes

PCT Yes

PDW Yes

MPV Yes Yes

FIBRINOGEN Yes

NRBC’s Yes

Page 20: Haematology. 2 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential Parameter PARAMETERLASERCYTEQBC VETAUTOREADIMPEDANZ SYSTEM HctYes HgbYes RBCYes ReticsYesUp

20 © 2004 IDEXX Laboratories, Inc., IDEXX Confidential

ABC Analyser – The Bad !

No retic information 3 Part Diff Impedance technology

Interfering substances 160 cycles include calibration / flushes / start up runs etc No data storage Few or no flags on abnormal samples Needs recalibrating every 6 months Different smart card per species