Study of water equivalence of the PRESAGE dosimeter for 3D proton beam dosimetry 1 Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006, Australia 2 Department of Radiation Oncology, Royal Prince Alfred Hospital, Sydney, NSW 2050, Australia Tina Gorjiara 1 , Zdenka Kuncic 1 , Robin Hill 1, 2 Samuel Blake 1 and Clive Baldock 1

Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Embed Size (px)

Citation preview

Page 1: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Study of water equivalence of the PRESAGE dosimeter for 3D

proton beam dosimetry

1 Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006,Australia

2 Department of Radiation Oncology, Royal Prince Alfred Hospital, Sydney, NSW 2050,Australia

Tina Gorjiara1, Zdenka Kuncic1 , Robin Hill1, 2 Samuel Blake1 and Clive Baldock1

Page 2: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

History of Proton Therapy

› Dr Robert Wilson (1914-2000)- A Harvard University physicist who played a

central role on the development of the atomicbomb

- Published a paper in 1946 that first proposed themedical use of protons for cancer therapy

› In 1954- The University of Berkeley began using proton

technology after the construction of a cyclotron totreat cancer patient

Page 3: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Proton therapy

3

http://procure.com/media/factsheets/sciencehttp://www.medical.siemens.com/webapp/wcs/stores/servlet/

Page 4: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Dosimetry in proton therapy› According to IAEA recommendations, the dosimetry of proton beams can

be performed with an ionization chamber dosimeter .

› However, because of the steep dose gradients, dose measurements aredifficult close to the Bragg peak.

› In addition, the ion chambers only provide point dose information andwould require a large number of them to provide two or three dimensionaldosimetric information.

› Therefore, 3D dosimetry techniques such as polymer gel dosimetry haspotential to be used as dosimeters to improve spatial resolution in protontherapy dosimetry.

Page 5: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Gel Dosimetry5

a) b) c)

d) e) f)

Page 6: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

What is PRESAGE?

› The PRESAGE dosimeter is composed of radiochromiccomponents (leuco dyes) and halogen-containing freeradical initiators.

› It undergoes a color change when exposed to ionizingradiation.

Page 7: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Advantages of PRESAGE as a dosimeter

› Being able to provide three-dimensional dose information.

› Insensitivity to oxygen contamination.

› Being machinable and molded to variety of shapes andsizes.

› Absorbing rather than scattering light which facilitates highaccuracy readout by optical CT.

Page 8: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Different formulations of PRESAGE

8

Material

Effective

atomic

number

Density

(g/cm³)wH wC wN wO wS wCl wBr

PRESAGE®

Formulation A7.69 1.050 0.089 0.62 0.050 0.21 0.0038 0.031 0.0023

PRESAGE®

Formulation B7.74 1.050 0.091 0.62 0.049 0.21 0.019 0.0021 0.0047

PRESAGE®

Formulation C8.65 1.101 0.089 0.61 0.045 0.22 - 0.033 0.0084

PRESAGETM 9.70 1.112 0.090 0.61 0.045 0.22 - 0.023 0.018

Water 7.42 1.000 0.11 - - 0.89 - -

Page 9: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

9

Page 10: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Aim

› Since Monte Carlo simulations can provide accuratepredictions of dose (GOLD STANDARD) , in this work, wehave calculated the relative dose response of the fourPRESAGE dosimeters to proton beam and compared it withdose response of water.

10

Page 11: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Method

› The GEANT4 Hadrontherapy example was modified accordingly tocalculate depth doses in the PRESAGE® Formulation A, PRESAGE®Formulation B, PRESAGE® Formulation C, PRESAGETM and water.

› Mean energy of proton beam was set to 62 MeV.

11

Page 12: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Method

12

40 cm

40 cm

Page 13: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Method

› To compare the depth doses in each dosiemeter, the water equivalentthickness (WET) was calculated according to IAEA technical report 398 foreach dosimeter using the following formula:

› tm is thickness of dosimeters and Rw and Rm are the proton range inwater and PRESAGE dosimeter, respectively.

13

m

wm R

RtWET ≈

Page 14: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

14

Page 15: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

15

Page 16: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

16

Page 17: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

17

Page 18: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

18

Page 19: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

19

Page 20: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

20

Page 21: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

21

Page 22: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

22

Page 23: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

23

Page 24: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

24

0

1000

2000

3000

4000

5000

6000

7000

8000

0 0.5 1 1.5 2 2.5 3 3.5

Dep

osite

d en

ergy

(MeV

)

Water equivalent thickness (cm)

Secendary protons

PRESAGE® Formulation AWater

Page 25: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

25

0.5

1

1.5

2

2.5

3

3.5

0 0.5 1 1.5 2 2.5 3 3.5

Dep

osite

d en

ergy

(MeV

)

Water equivalent thickness (cm)

Secendary Neutrons

PRESAGE® Formulation AWater

Page 26: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Results

26

0

5000

10000

15000

20000

0 0.5 1 1.5 2 2.5 3 3.5

Dep

osite

d en

ergy

(MeV

)

Water equivalent thickness (cm)

Secondary electrons

PRESAGE® Formulation A

Water

Page 27: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

Conclusion

› PRESAGE® dosimeters with new formulas, specially formulation A, are

more water equivalent than the currently available PRESAGE dosimeter.

› Thus more suitable for accurate 3D dosimetry of proton beams.

› Further experimental studies are planned to investigate possible saturation

in PRESAGE for high LET radiation fields.

27

Page 28: Investigation of radiological properties of PRESAGE ...web/@eng/@phys/...Dosimetry in proton therapy › According to IAEA recommendations, the dosimetry of proton beams can be performed

28

Thanks!