Upload
tranlien
View
216
Download
1
Embed Size (px)
Citation preview
This lesson was derived
from pages 15 through 21
in the textbook:
Lesson 04:
Resolution and Attenuation
This lesson contains 21 slides
plus 18 multiple-choice
questions.
Resolution and
Attenuation
GOOD
GOOD POOR
Axial and lateral resolution, which are categories of spatial resolution are
parameters of an ultrasound imaging system that characterizes its ability to
detect closely spaced interfaces and to display the echoes from those
interfaces as distinct and separate objects. The better the resolution, the
greater the clarity, or sharpness, of an ultrasound image.
Interfaces not closely spaced
Closely spaced
RESOLUTION
Closely spaced
GOOD
GOOD POOR
Of the many parameters affecting the axial and lateral resolution of an
ultrasound system, the most important is the operating frequency of the
ultrasound transducer. Transducers that operate at higher frequencies provide
improved resolution. The trade-off when using higher frequencies is reduced
penetration.
Interfaces not closely spaced
Closely spaced
RESOLUTION
Closely spaced
AXIAL RESOLUTION
SCANNED STRUCTURE DISPLAYED IMAGE
Axial resolution (also called longitudinal, range or depth resolution) is the
minimum required reflector separation along the direction of propagation
required to produce separate reflections.
AXIAL RESOLUTION
The axial resolution has a value equal to one-half the spatial pulse length.
Good axial resolution is achieved with short spatial pulse lengths. Short spatial
pulse lengths are the result of higher frequency, highly damped transducers.
SPATIAL PULSE LENGTH AXIAL RESOLUTION
4 mm 2 mm
3 mm 1.5 mm
2 mm 1 mm
LATERAL RESOLUTION
Lateral resolution (also called azimuthal resolution) is the minimum reflector
separation perpendicular to the direction of propagation required to produce
separate reflections.
SCANNED STRUCTURE DISPLAYED IMAGE
LATERAL RESOLUTION
Good lateral resolution is achieved with narrow acoustic beams. A narrow
acoustic beam is the result of a long near field (Fresnel zone) and a small angle
of divergence in the far field (Fraunhofer zone). The lateral resolution has a
value equal to the beam-width.
Most transducers in use today incorporate arrays rather than single-elements.
However, understanding the basic concepts of single-element transducers
should provide a basis for a better understanding of transducer arrays.
BEAM-WIDTH
LATERAL RESOLUTION
The beam-width, or beam diameter, from a non-focused, circular piezoelectric
element narrows as it propagates through the near field to a dimension equal to
half the diameter, or aperture size, of the piezoelectric element. The beam-
width then widens (diverges) as it propagates through the far field. At a
distance from the piezoelectric element equal to twice the length of the near
field, the beam-width has a dimension equal to the element’s aperture size.
The lateral resolution at given depth has a value equal to the beam-width at
that depth.
BEAM-WIDTH
LATERAL RESOLUTION
BEAM-WIDTH
BEAM-WIDTH LATERAL RESOLUTION
4 mm 4 mm
3 mm 3 mm
2 mm 2 mm
LATERAL RESOLUTION
The angle of beam divergence decreases in the far field and the length of the
near field increases if a higher frequency transducer is used. Additionally, the
angle of beam divergence decreases in the far field and the length of the near
field increases if a transducer with a larger aperture size.
Focusing can be used to further improve the lateral resolution. Focusing
creates a beam pattern that converges within the near field to produce a very
small cross section. The reduced cross section results in an intensity increase,
which is proportional to the square of the reduction. However, beyond the focal
zone, beam divergence is greater than from a non-focused transducer.
LATERAL RESOLUTION
Focal zones from single-element transducers are fixed during the
manufacturing process and cannot be changed. Various “fixed-focus” methods
are used including curved piezoelectric elements and external acoustic lenses.
Fixed focusing is often called “mechanical” focusing.
LATERAL RESOLUTION
Although the focus of a single-element transducer cannot be changed,
transducer arrays, produce sound beams with dynamic (multiple) transmit focal
zones at all imaging depths along the two-dimensional scanning plane. The
use of a dynamic aperture minimizes variations in beam width.
LATERAL RESOLUTION
HIGH-FREQUENCY TRANSDUCERS
BETTER RESOLUTION
GREATER ATTENUATION
POORER PENETRATION
LOW-FREQUENCY TRANSDUCERS
POORER RESOLUTION
LESS ATTENUATION
BETTER PENETRATION
Although improved resolution is obtained when high frequency transducers are
used, there is greater attenuation, which is the reduction of sound energy as it
passes through most materials. Many factors contribute to attenuation,
including reflection and beam divergence. However, the major cause of
attenuation is absorption, which is the result of heat conversion due to friction
created by vibrating tissue.
RESOLUTION vs. PENETRATION
(in tissue)
a = - 0.5 dB per cm per MHz
ATTENUATION COEFFICIENT
Attenuation in human soft tissue averages 0.5 dB per centimeter for each
megahertz. A loss totaling 3 dB is equivalent to a loss of one-half of the sound
energy. A 3 dB loss is present at the half intensity depth (H.I.D).
Attenuation is very high in bone with high and low frequencies. Conversely,
attenuation is low in fluids (e.g., blood) and very low in water, even when high
frequencies are used. Regardless of the medium, penetration is inversely
proportional to the transducer’s frequency.
In human soft tissue (assuming attenuation of -0.5 dB per cm per MHz), the
H.I.D. (in cm) may be determined by dividing 6 by the transducer frequency
(in MHz).
HALF INTENSITY DEPTH
(in tissue)
H.I.D. = 6 divided by frequency
TRANSDUCER FREQUENCY ATTENUATION PENETRATION HALF INTENSITY DEPTH
Increase Increase Decrease Decrease
Decrease Decrease Increase Increase
ATTENUATION COEFFICIENTS IN TISSUE(based on - 0.5 dB per cm per MHz)
Frequency -dB per cm Half-Intensity-Depth
2 MHz 1 3 cm
2.25 MHz 1.125 2.67 cm
2.5 MHz 1.25 2.4 cm
3 MHz 1.5 2 cm
3.5 MHz 1.75 1.71 cm
4 MHz 2 1.5 cm
5 MHz 2.5 1.2 cm
7 MHz 3.5 0.86 cm
7.5 MHz 3.75 0.8 cm
10 MHz 5 0.6 cm
15 MHz 7.5 0.4 cm
2 MHz 2.25 MHz 2.5 MHz
TRANSDUCER FREQUENCIES
5 MHz 7 MHz 7.5 MHz
10 MHz 12 MHz 15 MHz
3 MHz 3.5 MHz 4 MHz
ADULT LIVER
2.5 MHz 4 MHzAlthough frequencies in the range of 3 MHz to 5 MHz are normally used for
general purpose imaging (abdominal, obstetrical, gynecologic, and cardiac), a
frequency of 2.25 MHz or 2.5 MHz may be needed for adequate sound
penetration of a larger than normal patient.
Although frequencies in the range of 3 MHz to 5 MHz are normally used for
general purpose imaging (abdominal, obstetrical, gynecologic, and cardiac), a
frequency of 2.25 MHz or 2.5 MHz may be needed for adequate sound
penetration of a larger than normal patient.
3.5 MHz 5 MHz
ADULT LIVER AND RIGHT KIDNEY
7.5 MHz
THYROIDBREAST VASCULAR OPHTHALMIC
7 MHz 12 MHz10 MHz
SMALL PARTS
A frequency of 5 MHz or higher may be used if it is not necessary to display
echoes from deeper structures. Frequencies above 5 MHz are typically used
for thyroid, breast, peripheral vascular, ophthalmic, endocavity,
musculoskeletal, intraoperative, and some general purpose pediatric studies.
High frequency ultrasound (above 20 MHz) is used for dermatology,
stomatology, and intravascular ultrasound.
Answers to the following
EIGHTEEN practice
questions were derived
from material in the
textbook:
Question 1
The minimum reflector separation required to produce
separate echoes is
the spatial resolution of the ultrasound system
a function of TGC
the dynamic range
the attenuation coefficient
the reflection coefficient
Page 15
Question 2
Axial resolution is affected by all of the following EXCEPT
frequency
damping
spatial pulse length
wavelength
focusing
Page 16
Question 3
Far field beam divergence can be reduced on a
single-element transducer by using
electronic focusing
a transducer with a smaller element diameter
a higher frequency transducer
a lower frequency transducer or a smaller
element diameter
adjustable focusing
Pages 17 and 18
Question 4
The area between the face of an unfocused single-element
transducer and the point where the beam starts to diverge
is the
fraunhofer zone
refraction zone
focal plane
far zone
near field
Page 17
Question 5
Assuming a fixed frequency, when the diameter of an
unfocused transducer is increased, the
far field divergence increases
penetration decreases
length of the near field increases
length of the near field decreases
attenuation increases
Page 18
Question 6
The fresnel zone is the
focal zone of a focused transducer
far zone
region where the near field changes to the far zone
distance from the face of a non-focused transducer
to the beginning of the far field
the fraunhofer zone
Page 17
Question 7
Assuming no losses due to attenuation, which reflector
provides the strongest echo?
A
B
C
D
E
Page 18
Question 8
A focused, curved, single piezoelectric element
can be used for CW Doppler
can be dynamically focused
produces a beam pattern that is determined during
manufacturing
can be electronically focused
can be used in a convex array transducer
Page 18
Question 9
Which of the following does NOT affect lateral resolution?
focusing
beam width
element diameter
frequency
Spatial pulse length
Pages 16 and 17
Question 10
Higher frequency transducers provide
improved lateral resolution
smaller Doppler shifts
improved axial resolution and reduced attenuation
increased penetration
poor axial resolution
Pages 15 and 19
Question 11
Which of the following does NOT contribute to attenuation?
beam divergence
scattering
absorption
reflection
constructive interference
Pages 4 and 19
Question 12
As the frequency of sound increases,
the amount of scatter is increased
the attenuation decreases
the amount of scatter decreases
the penetration increases
there will be a decrease in the number of specular reflectors
Page 19
Question 13
Which of the following transducers provides the maximum
penetration?
10.0 MHz
7.5 MHz
2.25 MHz
3.5 MHz
5.0 MHz
Pages 19 and 20
Question 14
Which transducer configuration is the best choice for
imaging superficial structures?
high frequency, far focus
high frequency, near focus
low frequency, far focus
low frequency, near focus
low frequency, mid focus
Page 21
Question 15
The average attenuation of ultrasound energy in the patient
is approximately
2.0 dB per cm per MHz
10.0 dB per cm per MHz
20.0 dB per cm per MHz
5.0 dB per cm per MHz
0.5 dB per cm per MHz
Page 19
Question 16
If sound from a 3 MHz transducer has 3 dB of attenuation
after traveling through 2 cm of tissue, what is the amount of
attenuation of sound from a 5 MHz transducer after
traveling through 1 cm of the same tissue?
5 dB
1 dB
2 dB
2.5 dB
3 dB
Page 19
Question 17
The half-value-layer or the half-intensity-depth
is the depth where the intensity is 50% of the originally
transmitted intensity
increases as the frequency of the transducer increases
is the thickness of the matching layer in an array
is the range of frequencies contained in an ultrasound
pulse
is the attenuation coefficient in tissue
Page 19
Question 18
Lateral resolution is determined mainly by
reflector size
beam diameter
pulse duration
bandwidth
spatial pulse length
Page 17
END OF LESSON 04