Local sensitivity analysis of cardiovascular system parameters R. Gul 1 & S. Bernhard 1,2 1 Fachbereich Mathematik, Freie Universit¨ at Berlin, Germany 2 Department of Electrical Engineering and Information Technology, Pforzheim University, Germany Abstract Cardiovascular disease is one of the major problems in today’s medicine and the number of patients increase worldwide. To treat these types of diseases, prior knowledge about function and dysfunction of the cardiovascular system is essential for identifying the disease in an early stage. Mathematical modeling is a powerful tool for prediction and investigation of the cardiovascular system. It has been shown, that the Windkessel model, drawing an analogy between electrical circuits and fluid flow, is an eective method to model the human cardiovascular system. The aims of this work are the derivation of a computational cardiovascular model for the arm arteries, and to analyze the behavior of the vascular network structure by parameter sensitivity analysis. Sensitivity analysis is essential for parameter estimation and simplification of cardiovascular models. In optimal experiment design (OED) sensitivity analysis is used to construct experiments and corresponding models that allow the interpretation of cardiovascular measurements in an eective manner. In this paper we have applied sensitivity analysis to a linear elastic model of the arm arteries to find sensitive parameters and their confidence intervals that guide us to the estimation of cardiovascular network parameters. To calculate the percentage eect on the measurable state variables pressure and flow, with respect to percentage change in cardiovascular input parameters, we use norms. This method allows us to quantify and verify results obtained by sensitivity analysis. The sensitivities with respect to flow resistance, arterial compliance and flow inertia, reveal that the flow resistance and diameter of the vessels are the most Modelling in Medicine and Biology X 155 www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press doi:10.2495/BIO130141

Local sensitivity analysis of cardiovascular system parameters

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Local sensitivity analysis of cardiovascular system parameters

Local sensitivity analysis of cardiovascularsystem parameters

R. Gul1 & S. Bernhard1,21Fachbereich Mathematik, Freie Universitat Berlin, Germany2Department of Electrical Engineering and Information Technology,Pforzheim University, Germany

Abstract

Cardiovascular disease is one of the major problems in today’s medicine and thenumber of patients increase worldwide. To treat these types of diseases, priorknowledge about function and dysfunction of the cardiovascular system is essentialfor identifying the disease in an early stage.

Mathematical modeling is a powerful tool for prediction and investigation ofthe cardiovascular system. It has been shown, that the Windkessel model, drawingan analogy between electrical circuits and fluid flow, is an effective method tomodel the human cardiovascular system. The aims of this work are the derivationof a computational cardiovascular model for the arm arteries, and to analyze thebehavior of the vascular network structure by parameter sensitivity analysis.

Sensitivity analysis is essential for parameter estimation and simplification ofcardiovascular models. In optimal experiment design (OED) sensitivity analysisis used to construct experiments and corresponding models that allow theinterpretation of cardiovascular measurements in an effective manner. In thispaper we have applied sensitivity analysis to a linear elastic model of the armarteries to find sensitive parameters and their confidence intervals that guideus to the estimation of cardiovascular network parameters. To calculate thepercentage effect on the measurable state variables pressure and flow, with respectto percentage change in cardiovascular input parameters, we use norms. Thismethod allows us to quantify and verify results obtained by sensitivity analysis.

The sensitivities with respect to flow resistance, arterial compliance and flowinertia, reveal that the flow resistance and diameter of the vessels are the most

Modelling in Medicine and Biology X 155

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

doi:10.2495/BIO130141

Page 2: Local sensitivity analysis of cardiovascular system parameters

sensitive parameters. Those parameters play a key role in diagnoses of severestenosis and aneurysms. In contrast, wall thickness and elastic modulus are foundto be less sensitive.Keywords: computational cardiovascular model, cardiovascular parameters,sensitivity analysis, Windkessel model.

1 Introduction

With growing interest in the prediction and diagnosis of cardiovascular diseases,different mathematical models have been developed and applied. Windkesselmodels (electrical analogy to fluid flow) have shown to be an effective approachin modeling the human cardiovascular system [1–4]. Westerhof et al. [3] studiedthe design, construction and evaluation of an electrical model. Quarteroni et al. [1]introduced a multiscale approach, where local and systemic models are coupledat a mathematical and numerical level. He also introduced the Windkessel modelsfor different inlet and outlet conditions.

Within the Windkessel model the hemodynamic state variables (pressure (p)and flow (q)), are interrelated to the model parameters like elastic modulus (E),vessel length (l) and diameter (d), wall thickness (h), the density of blood (ρ) andthe network structure. Provided that the model parameters estimated correspondto the cardiovascular measurements, the Windkessel model is a good way to studyvascular parameters, which are difficult to measure directly.

The basis for robust parameter estimation is on the one hand an optimalexperimental measurement setup and on the other hand the development of modelsthat describe the hemodynamic state variables in a set of relevant parameters thatcan be estimated with high accuracy. The design consists of several logical steps,dealing with questions like:• How does the optimal measurement setup to identify structural vascular

parameters look like?• What are the parameters, variables and experimental measurement locations

within a clinical setting?• Which vascular system parameters are most influential on the hemodynamic

state variables pressure and flow?• Which vascular system parameters are insignificant and may be fixed or

eliminated?Sensitivity analysis is a powerful approach to find sensitive and therefore importantcardiovascular system parameters. The important parameters are further usedto design the measurement setup and to modify the computational model forfurther analysis. Sato et al. [5] studied the effects of compliance, volume andresistance on cardiac output using sensitivity analysis. Yu et al. [6] used parametersensitivity to construct a simple cardiovascular model. Leguy et al. [7] appliedglobal sensitivity analysis on the arm arteries and showed that the elastic modulusis most sensitive parameter, while arterial length is a less sensitive parameter.In optimal experimental design the information matrix (like, fisher information

156 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 3: Local sensitivity analysis of cardiovascular system parameters

matrix (FIM)) is used for the set of parameters. This information matrix dependson parameter sensitivity analysis.

The methods developed in this paper, are seen as the first step towardscardiovascular system identification from cardiovascular measurements. Withinthis work we derive a computational cardiovascular model for the arm arteriesby using the Windkessel approach. In a first instance we apply local sensitivityanalysis to study the effects of cardiovascular parameters on the hemodynamicstate variables. Finally we will apply the concept of norms to quantify and compareour results.

2 Derivation of the model equations

Under the assumption that the arterial tree is decomposed into short arterialsegments of length l with a constant circular cross-section and linear elastic wallbehaviour, the following one dimensional flow equations can be derived fromthe linearized Navier–Stokes equation, the equation of continuity and the shell-equation for thin walled, linear elastic tubes [1, 8]

−∂p∂x= Rq+L∂q

∂t, (1)

−∂q∂x=

pZ +C

∂p∂t. (2)

Within these equations the state variables are the inflow q and the relative pressurep. The viscosity and inertial forces of the blood flow are described by the viscousflow resistance R and blood inertia L per unit length respectively. The elasticproperties of the wall are modeled by a compliance C per unit length, whilethe outflow is modeled by a leakage 1

Z per unit length [8]. Integration of thetwo partial differential eqns. (1) and (2) along flow axis leads to a system ofequations (3) commonly used to describe electrical circuits (see Figure 1). In thistype of model each segment of the arterial system is described by a set of twoequations that are known as three element Windkessel equations. Here (pin,qout)are the boundary conditions for non-terminal nodes. To model a mean venouspressure with a value of 15mmHg for terminal nodes the equation system is setupby including an additional terminal resistance Z and using boundary conditions(pin, pout). The matrix form of the Windkessel eqns. with boundary conditions is

dXdt= AX+B (3)

For non-terminal segments (Figure 1:left) X = (qin, pout)T

A =

⎛⎜⎜⎜⎜⎜⎝

−RL

−1L

1C 0

⎞⎟⎟⎟⎟⎟⎠ , B(pin,qout) =

⎛⎜⎜⎜⎜⎜⎝

pinL

− qoutC

⎞⎟⎟⎟⎟⎟⎠

Modelling in Medicine and Biology X 157

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 4: Local sensitivity analysis of cardiovascular system parameters

For terminal segments (Figure 1:right) X = (qin,qout)T

A =

⎛⎜⎜⎜⎜⎜⎝

−RL

−1L

1C − 1

ZC

⎞⎟⎟⎟⎟⎟⎠ , B(pin, pout) =

⎛⎜⎜⎜⎜⎜⎝

pinL

− poutZC

⎞⎟⎟⎟⎟⎟⎠

The electrical parameters for i-th segment in the arterial tree are related to thephysiological parameters of the fluid and vessel wall by:

Ri =8νlπr4 , Li =

ρlπr2 , Ci =

2πr2lEh. (4)

Figure 1: Linear elastic model for fluid flow in nonterminal vessel segments (left)and for terminal vessel segments (right).

2.1 Vascular model of the arm arteries

2.1.1 Network structure and model equationsEach segment of arm arteries in a network structure as given in Figure 2 isrepresented by an electrical circuit as shown in Figure 1.

Applying Kirchhoff’s current and voltage laws to the arterial structure given inFigure 2, with number of terminals Nt = 3 and number of segments Ns = 15, weobtain a system of coupled ordinary differential equations for pressure and flow.Flow equations:

q1 =pin− p1−R1q1

L1, q2 =

p1− p2−R2q2

L2, q3 =

p2− p3−R3q3

L3

q4 =p3− p4−R4q4

L4, q5 =

p4 − p5−R5q5

L5, q6 =

p5− p6−R6q6

L6

q7 =p6− p7−R7q7

L7, q8 =

p7 − p8−R8q8

L8, q9 =

p8− p9−R9q9

L9˙

q10 =p9− p10−R10q10

L10, q11 =

p6− p11−R11q11

L11, q12 =

p11− p12−R12q12

L12

q13 =p11− p13−R13q13

L13, q14 =

p13− p14−R14q14

L14, q15 =

p14− p15−R15q15

L15

158 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 5: Local sensitivity analysis of cardiovascular system parameters

Pressure equations:

p1 =q1−q2

C1, p2 =

q2−q3

C2, p3 =

q3−q4

C3, p4 =

q4−q5

C4

p5 =q5−q6

C5, p6 =

q6−q11−q7

C6, p7 =

q7−q8

C7, p8 =

q8−q9

C8

p9 =q9−q10

C9, p10 =

q10− (p10− pout)/Z1

C10, p11 =

q11−q12−q13

C11

p12 =q12− (p12− pout)/Z2

C12, p13 =

q13−q14

C13, p14 =

q14−q15

C14,

p15 =q15− (p15− pout)/Z3

C15

Figure 2: Simplified anatomy of the arm arteries (left) and model geometry withNs = 15 and Nt = 3 (right).

2.2 State-space representation

The state-space representation is a useful approach to describe the dynamics inarterial networks efficiently [9]. In state-space form, we have a system of twoequations: an equation for determining state xt of the system (state equation), andanother equation to describe the output yt of the system (observation equation).The matrix form can be written as

xt = Axt−1+But, (5)

yt = Cxt +Dut. (6)

Here xt is the state vector of the system, ut the input vector and yt the observationvector. The dynamics of the system is described by the state dynamics matrixA ∈ M(n× n). The input matrix B ∈ M(n× i) specifies the time dependency of thein- and outflow boundary values and the observation matrix C ∈ M(m×n) definesthe observation locations within the state-space system, i.e. the nodal location in

Modelling in Medicine and Biology X 159

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 6: Local sensitivity analysis of cardiovascular system parameters

the network. Here m denotes the number of observations. Finally the input toobservation matrix D ∈ M(m × i) adds the influence of the input vectors to theobservation vectors. Besides the computational advantage the state-space formallows the integration of experimental measurements (observations) into the modelbuilding process. This step is essential for the adjacent model parameter estimationfrom experimental measurements, that we have planned in a future study.

In the current study parameter values were taken from the literature [4]. Thestate vector xt contains the flow and pressure functions at all network locations,whereas the output vector yt contains the flow and pressure at selected nodesi. For a m = 4 dimensional observation vector, the output vector is e.g. y(t) =(q5(t), p5(t),q6(t), p6(t))T where y ∈ Rm at nodes 5 and 6. The state-space systemfor the arm artery given in Figure 2 is then defined by

Ai j =

⎧⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎨⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎩

−R i+12

L i+12

i = 1,3,5, . . . ,29, j = i

1C i

2

i = 2,4,6, . . . ,30, j = i−1

−1L i+1

2

i = 1,3,5, . . . ,29, j = i+1

1L i+1

2

i = 3,5,7, . . . ,29, j = i−1 and i � 21,25

also for i = 21, j = 12 and i = 25, j = 22−1C i

2

i = 2,4,6, . . . ,30, j = i+1 and i � 20,24

also for i = 12, j = 21 and i = 22, j = 25−1

ZkC i2

k = 1,2,3, i = 20,24,30, j = i

0 otherwise

Bi j =

⎧⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎨⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎪⎩

1Li

i = j = 11

Zj−1C i2

i = 20, j = 2

1Zj−1C i

2

i = 24, j = 3

1Zj−1C i

2

i = 30, j = 4

0 otherwise

Ci j =

⎧⎪⎪⎨⎪⎪⎩

1 i = 1,2,3,4, j = i+80 otherwise

and Di j = 0.

3 Methods of local sensitivity analysis

To understand the interdependence of the state variables and the parameters ofthe cardiovascular model, it is not enough to compute a solution, but also toquantify the sensitivity of the model parameters. Sensitivity analysis is a usefultool to quantify the variation in state variables at different nodes in the vascularnetwork caused by a change in model parameters. Due to the interdependence of

160 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 7: Local sensitivity analysis of cardiovascular system parameters

the electrical analog parameters RCL, we also discuss the linear independent basisparameters ELdh. Further, we are interested in how the node location within thenetwork influences the results. Therefore we discuss three different scenarios. Thesensitivity results obtained are compared to the 2-norm of the distance vector ofthe state variables of two time series.

3.1 Sensitivity with respect to E, l, d and h

The cardiovascular model consists of a system of ODEs with a parameter set θ andan initial condition yi(0), given by

dydt= fi(yi, θ, t) i = 1,2, . . . ,n. (7)

In local sensitivity analysis, parameters are varied segmentwise by some portionaround a fixed value and the effects of individual perturbations on the observationsare studied [10]. Using differential calculus the sensitivity coefficients are

S i =∂yi

∂θ= limΔθ→0

yi(θ+Δθ)− yi(θ)Δθ

, (8)

where yi is the ith model output and θ is the model input parameter. There existsa variety of methods to compute the sensitivity coefficients in eqn. (8), within thiswork we use the method of external differentiation:

S i =∂yi

∂θ� yi(θ+Δθ)− yi(θ)

Δθ(9)

Applied to our network structure, this equation produces a set of two sensitivitytime series S i(t) (one for pressure and one for flow) per parameter and per networknode (see Figure 3).

3.2 Sensitivity with respect to R, C and L

To find the sensitivity of the electrical analog parameters R, C and L oncardiovascular pressure and flow, we solve eqn. (7) numerically using theCVODES solver, which is a part of SUNDIALS software suit [11, 12]. Thecomputational method is also based on external differentiation.

3.3 Network structure and sensitivity

To study the influence of the vascular network structure, we use a physiologicalnetwork structure with identical (non-physiological) Windkessel elements, i.e. theparameters Ri,Ci and Li are identical for each node. This allows us to analyzethe influence of the network structure onto sensitivity values at different nodelocations. Therefore we compute the nodal sensitivity time series S i for all nodalparameters as described in previous section. For each node, n, we obtain two

Modelling in Medicine and Biology X 161

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 8: Local sensitivity analysis of cardiovascular system parameters

sensitivity time series (one for the pressure and one for the flow) per parametervariation. The total number of possible parameter variations is 3n2, so we endup with a set of 6n2 time series. To reduce the complexity, we average the timedependency by computing the mean of the absolute value of every time series.The 6n2 real values are used for further analysis in section 4. They are displayedinto 6 matrices that characterize the sensitivity of pressure and flow in the networkstructure based on changes in the electrical parameters RCL. Each cell in Figure 5represents the mean absolute value of time series (S i). Due to the fact that we usethe pressure as an input and output boundary condition, the change in pressurewith any parameter at all inlet and terminal nodes will be zero.

Figure 3: Time series of flow resistance R sensitivity coefficient S i for pressure andflow at node 7.

3.4 Sensitivity analysis by using norms

To obtain a suitable measure for sensitivity we calculated the mean Euclideandistances of observations made in a model with different parameter sets θ1, θ2.

‖θ1, θ2‖:= meant∈T‖ yi(θ2, t)− yi(θ1, t) ‖2

‖yi(θ1, t)‖2 i = 1,2,3, . . . ,2×Ns,

4 Results and discussion

Our local sensitivity analysis was carried out as described in section 3. The resultsare structured into the following sections:

4.1 Sensitivities with respect to physiological parameters

The sensitivity for pressure and flow are obtained by a variation of thecardiovascular parameters E, l,d and h of arm arteries by ±10%. From Figure 4it is directly evident, that diameter and length are most sensitive parameters, whilethe elastic modulus and wall thickness are comparatively less sensitive parameters.

162 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 9: Local sensitivity analysis of cardiovascular system parameters

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.81

1.5

2

2.5

3

3.5

4

4.5

5x 10

−6 E sensitivity on flow at node 7

time [s]

flow

[ml/s

]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.80.8

0.9

1

1.1

1.2

1.3

1.4

1.5x 10

4 E sensitivity on pressure at node 7

time [s]

pres

sure

[Pa]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.81

1.5

2

2.5

3

3.5

4x 10

−6 l sensitivity on flow at node 7

time [s]

flow

[ml/s

]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.80.8

0.9

1

1.1

1.2

1.3

1.4

1.5x 10

4 l sensitivity on pressure at node 7

time [s]

pres

sure

[Pa]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.81

1.5

2

2.5

3

3.5

4

4.5

5x 10

−6 d sensitivity on flow at node 7

time [s]

flow

[ml/s

]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.80.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6x 10

4 d sensitivity on pressure at node 7

time [s]

pres

sure

[Pa]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.81

1.5

2

2.5

3

3.5

4

4.5

5x 10

−6 h sensitivity on flow at node 7

time [s]

flow

[ml/s

]

−10% +10%Base values

0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.80.8

0.9

1

1.1

1.2

1.3

1.4

1.5x 10

4 h sensitivity on pressure at node 7

time [s]

pres

sure

[Pa]

−10% +10%Base values

Figure 4: ±10% change in E, l, d and h in arm artery at node 7.

4.2 Sensitivities with respect to fluid dynamical parameters

The sensitivities w.r.t. R,C and L were calculated by forward sensitivity analysis(FSA) using the SUNDIALS software. The sensitivity pattern obtained byvariation of the viscous flow resistance R in any segment of brachial artery (seeFigure 5 (top)) indicates a strong local (within brachial artery) influence on flowand has significant global influence on all following nodes of the brachial, ulnarand radial arteries. In contrast, changing R in the parallel association of the ulnar

Modelling in Medicine and Biology X 163

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 10: Local sensitivity analysis of cardiovascular system parameters

and radial arteries have negligible local and global effects, because in parallelarteries the total flow resistance is given by the fraction 1

Rtotal= 1

Rulnar+ 1

Rradial, i.e.

due to the increment in total diameter the flow resistance reduces. Physically achange of R in one branch redirects the flow into the other branch while the overallflow is maintained. The sensitivity of flow resistance in parallel branches is thussmaller than in series connections.

In contrast to the flow resistance, the sensitivity of arterial compliance C inbrachial part has small influence (locally and globally) on pressure and flow (seeFigure 5 (middle)). This can be explained by the fact, that in series segmentsthe total compliance is given by the sum of all segmental compliances in seriesCtotal =

∑6i=1 Ci. The total compliance is larger than the individual compliances,

thus a change of C in any node has a small effect on pressure and flow in the wholearm artery. In contrast a variation of the arterial compliance C in the ulnar andradial arteries have a large (local) effect on pressure and large (global) effect onflow especially in the brachial artery.

Figure 5: Effects of viscous flow resistance R (top), vessel compliance C (middle)and blood inertia L (bottom) on pressure and flow in the arm arteries.

164 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 11: Local sensitivity analysis of cardiovascular system parameters

From eqn. (4) it is obvious that viscous resistance and blood inertia are inverselyrelated to r4 and r2 respectively. Which means in large arteries blood inertia playsan important role, while in small arteries viscous resistance is more important. Avariation of blood inertia in the first node of the brachial artery has large influenceon pressure and flow of all following nodes of the arm arteries (see Figure 5(bottom)). However, we observe only minor local (within brachial artery) effectson flow when we change L in each segment of the brachial artery. Further, dueto the fact that the total inductance 1

Ltotal= 1

Lulnar+ 1

Lradialreduces at the furcation,

the flow and pressure in the ulnar and radial arteries are less sensitive with respectto L.

4.3 Sensitivities with respect to norms

Finally we compare the results obtained by sensitivity analysis with those obtainedby using norms (see Table 1 and 2). We found that the diameter and length of vesselare most influential parameters and that the norm computed for the wall thicknessand elastic modulus has identical values.

5 Conclusion

The sensitivity analysis is the first step towards the estimation of model parametersfrom experiments. It identifies the parameters in our model that are importantto describe the dynamic behaviour of the system, i.e. it defined the parametersthat should be estimated correctly and on the other hand those parameters thatare less important. Sensitivity analysis thus allows the design of problem specificexperiments, i.e. just gather the information that is required to generate a predictivemodel that describes the actual health status of a patient. It therefore improves themodel quality and thus the ability for diagnosis and prediction in cardiovascularphysiology. This finally benefits the medical doctor in decision making.

In this work we applied different methods of sensitivity analysis to a lumpedparameter Windkessel model of the arm arteries. The results indicate a strongdependence of the pressure and flow state variables onto a variation in vesseldiameter and length. According to the elastic properties and the thickness of thearterial wall, a much lower sensitivity was found.

Further, we have used the concept of norms to compare the variation in statevariables according to parameter changes. We found a good agreement to theresults obtained by sensitivity analysis.

The methods applied, give satisfactory results if the cardiovascular parametersare independent, in the real scenarios however, they are often interdependent likee.g. the observation of a high correlation between the extension of the elastic wallsand the tangential tension caused by transmural pressure. To study these type ofeffects in a more general way, we plan to apply global sensitivity analysis to aclosed loop cardiovascular system model, which deals with variations in manyparameters at a time.

Modelling in Medicine and Biology X 165

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 12: Local sensitivity analysis of cardiovascular system parameters

Tabl

e1:±1

0%ch

ange

inE

,l,d

and

hat

node

7of

the

arm

arte

ry(s

eeFi

gure

2)an

dth

eirc

orre

spon

ding

perc

enta

gech

ange

inflo

wat

each

node

.

N0

N1

N2

N3

N4

N5

N6

N7

N8

N9

N10

N11

N12

N13

N14

N15

q 0q 1

q 2q 3

q 4q 5

q 6q 7

q 8q 9

q 10

q 11

q 12

q 13

q 14

q 15

E0.

450.

440.

420.

360.

340.

270.

240.

360.

390.

190.

180.

180.

110.

240.

140.

12l

0.54

0.53

0.47

0.43

0.39

0.32

0.31

0.95

0.86

0.48

0.46

0.25

0.13

0.34

0.19

0.15

d1.

961.

761.

751.

551.

431.

161.

061.

341.

190.

800.

780.

870.

450.

100.

630.

54h

0.45

0.44

0.42

0.36

0.34

0.27

0.24

0.36

0.39

0.19

0.18

0.18

0.11

0.24

0.14

0.12

Tabl

e2:±1

0%ch

ange

inE

,l,d

and

hat

node

7of

the

arm

arte

ry(s

eeFi

gure

2)an

dth

eirc

orre

spon

ding

perc

enta

gech

ange

inpr

essu

reat

each

node

.

∗ Cha

nge

inpr

essu

rew

ithre

spec

tto

any

para

met

eris

zero

,bec

ause

we

use

pres

sure

asa

boun

dary

cond

ition

.

N0

N1

N2

N3

N4

N5

N6

N7

N8

N9

N10

N11

N12

N13

N14

N15

p 0p 1

p 2p 3

p 4p 5

p 6p 7

p 8p 9

p 10

p 11

p 12

p 13

p 14

p 15

E0∗

0.02

0.03

0.05

0.06

0.08

0.08

0.13

0.13

0.14

0∗0.

090∗

0.09

0.10

0∗

l0

0.03

0.05

0.06

0.08

0.09

0.10

0.32

0.32

0.37

00.

110

0.11

0.12

0d

00.

130.

150.

210.

270.

330.

360.

580.

580.

630

0.38

00.

400.

440

h0

0.02

0.03

0.05

0.06

0.08

0.08

0.13

0.13

0.14

00.

090

0.09

0.10

0

166 Modelling in Medicine and Biology X

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press

Page 13: Local sensitivity analysis of cardiovascular system parameters

Acknowledgements

The author would like to thank the HEC/DAAD for their financial support also toDr. S. Bernhard and Rudolf Huttary for fruitful discussions and encouragement.

References

[1] Quarteroni, A., Ragni, S. and Veneziani, A., Coupling between lumped anddistributed models for blood flow problems. Computing and Visualization inScience. Volume 4, pages 111-124 (2001)

[2] Milisic, V. and Quarteroni, A., Analysis of lumped parameter models for bloodflow simulations and their relation with 1D models. Mathematical Modelingand Numerical Analysis (1999)

[3] Westerhof, N., Bosman, F., De Vries, C.J. and Noordergraaf, A., Analogstudies of the human systemic arterial tree. Journal of Biomechanics, Volume2, pages 121-143 (1969)

[4] Noordergraaf, A., Verdouw, P.D. and Boom, H.B.K., The use of an analogcomputer in a circulation model. Progress in Cardiovascular Diseases. Volume5, number 5 (1963)

[5] Sato, T., Yamashiro, S.M., Vega, D. and Grodins, F.S., Parameter sensitivityanalysis of a network model of systemic circulatory mechanics. Annals ofBiomedical Engineering. Volume 2, pages 289–306 (1974)

[6] Yu, Y.C., Boston, J.R., Simaan, M.A. and Antaki, J.F., Sensitivity analysisof cardiovascular models for minimally invasive estimation of systemicvascular parameters. Proceedings of American Control Conference, SanDiago, California, (1999)

[7] Leguy, C.A.D., Bosboom, E.M.H., Belloum, A.S.Z., Hoeks, A.P.G. and vande Vosse, F.N., Global sensitivity analysis of a wave propagation model forarm arteries. Medical Engineering and Physics. Volume 33, pages 1008–1016(2011)

[8] Jager, Gerad S., Westerhof, N., Noordergraf, A., Oscillatory flow impedancein electrical analog of arterial system: representation of sleeve effect and non-newtonian properties of blood. Circulation research. Volume XVI, pages 121–133 (1965)

[9] Bernhard, S., Al Zoukra, K. and Schutte, C., Statistical parameter estimationand signal classification in cardiovascular diagnosis. Environmental Healthand Biomedicine. Volume 15, pages 458–469 (2011)

[10] Zi, Z., Sensitivity analysis approaches applied to systems biology models.IET system biology. Volume 5, issue 6, pages 336–346 (2011).

[11] Hindmarsh, A.C., Brown, P.N., Grant, K.E. et al. SUNDIALS: suite ofnonlinear and differential/algebraic equation solvers, ACM Transactions onMathematical Software. Volume 31, Number 3, pages 363–396 (2005)

[12] Serban, R., Hindmarsh, A.C., CVODES: the sensitivity-enabled ODE solverin SUNDIALS. Proceedings of IDETC/CIE (2005)

Modelling in Medicine and Biology X 167

www.witpress.com, ISSN 1743-3525 (on-line) WIT Transactions on Biomedicine and Health, Vol 17, © 2013 WIT Press