6
Polystyrene analysis using MALDI-TOF MS Three polystyrene samples covering a mass range from 2 kDa to 200 kDa were analyzed using an autoflex maX MALDI-TOF MS. Abstract For all samples, several charac- teristic values were determined, including the average molecular weights M n and M w , the mono- mer mass, the sum-mass of the present end-groups and, to a certain extent, even the chemical compositions. Introduction The production of polymer materials is currently under severe economic pressure. Therefore, quality control of raw materials or finished goods is of increas- ing importance in order to avoid product failure or even health risks. A standard procedure for the detection of the average molecular mass of polymer samples is size exclusion chromatography. This technique, however, cannot reveal the chemical structures behind the detected polymer masses. Other methods like FTIR or NMR spectroscopy can be used to generate chemical information. Unlike the aforementioned tech- nologies, Matrix Assisted Laser Desorption/ Ionization-Time Of Flight Mass Spectrometry (MALDI-TOF MS) offers the unique possibility to detect the average molecular weights and concurrently delivers detailed information on the chemical composition of the sample. A detailed workflow for the analysis of synthetic polymers by MALDI-TOF MS is also given in ISO 10927:2018-10 [1]. Author: Volker Sauerland; Bruker Daltonik GmbH, Bremen, Germany. Keywords: MALDI, autoflex maX, Polymer, Polystyrene

Polystyrene analysis using MALDI-TOF MS

  • Upload
    others

  • View
    32

  • Download
    0

Embed Size (px)

Citation preview

Polystyrene analysis using MALDI-TOF MSThree polystyrene samples covering a mass range from 2 kDa to 200 kDa were analyzed using an autoflex maX MALDI-TOF MS.

Abstract

For all samples, several charac-teristic values were determined, including the average molecular weights Mn and Mw, the mono-mer mass, the sum-mass of the present end-groups and, to a certain extent, even the chemical compositions.

Introduction

The production of polymer materials is currently under severe

economic pressure. Therefore, quality control of raw materials or finished goods is of increas-ing importance in order to avoid product failure or even health risks. A standard procedure for the detection of the average molecular mass of polymer samples is size exclusion chromatography. This technique, however, cannot reveal the chemical structures behind the detected polymer masses. Other methods like FTIR or NMR spectroscopy can be used to generate chemical information.

Unlike the aforementioned tech-nologies, Matrix Assisted Laser Desorption/Ionization-Time Of Flight  Mass  Spectrometry (MALDI-TOF MS) offers the unique possibility to detect the average molecular weights and concurrently delivers detailed information on the chemical composition of the sample.

A detailed workflow for the analysis of synthetic polymers by MALDI-TOF MS is also given in ISO 10927:2018-10 [1].

Author: Volker Sauerland; Bruker Daltonik GmbH, Bremen, Germany.

Keywords: MALDI, autoflex maX, Polymer, Polystyrene

Experimental

The polystyrene samples with Mp = 1920 Da (PS1) and Mp = 210 kDa (PS3) were purchased from Polymer Standards Services GmbH (Mainz, Germany). A polystyrene sample with molecular weight Mp = 19 kDa (PS2) was purchased from Polymer Laboratories LTD (Church Stretton, UK).

All samples were dissolved in THF at a concentration of 10 g/L. DCTB, prepared at 20 g/L in THF, was used as the matrix. Silver trifluoroacetate, was prepared as the doping salt at a concentration of 0.1 M in THF. The polymer, matrix, and salt solutions were mixed in a ratio 3:20:1 for PS1 and PS2 and 3:100:1 for PS3. A droplet of the mixture (0.5 μL) was hand-spotted onto three spots of a gound steel target plate.

An autoflex® maX (Bruker Daltonics) MS system, operating in reflector mode for sample PS1 and in linear

mode for samples PS2 and PS3, was used for data acquisition. All spectra were measured in the positive ion mode and calibration was performed on polystyrenes prepared on neigh-boring sample positions.

Data acquisition and calibration were carried out with COMPASS for flex 1.4 (Bruker Daltonics), followed by polymer signal analysis using PolyTools 1.0 SR 1 (Bruker Daltonics).

Results

The oligomer distribution of the PS1 sample was found to be in the mass range of 1000 to 3300 Da as shown in Figure 1. The average monoisotopic mass difference between the indi-vidual mass signals is 104.0624 Da, which is in good agreement with the theoretical monoisotopic mass value of 104.0626 Da for C8H8. The isotopic distribution of a single oligomer is shown in Figure 1 (inset). The presence of silver in the molecular

ion may be inferred from the strong intensity of the +2 isotope. The analysis of the spectrum using PolyTools is displayed in Figure 2. The program automatically identifies the monomer unit as styrene and performs the extrapolation of the ion series towards lower masses until the residue mass is below the monomer mass. As the residual mass is the combination of the masses of the two end-groups and the cation, PolyTools determines the optimal combination of predefined editable entries. Polystyrene was defined within the PolyTools template and the end groups were found to be hydrogen and butyl, both in agreement with the supplier’s infor-mation. As silver was added during the sample preparation (and the isotopic pattern also suggests the presence of silver), the spectrum interpretation by PolyTools using the silver cation can be confirmed. In addition to the end-group analysis, PolyTools also calculates automatically

Inte

nsit

y [a

.u.]

m/z

0

0.5

1.0

x104

1.5

1750 27501500 2500 32501250 2250 30001000 2000

2142

.178

Inte

nsit

y [a

.u.]

m/z

0

0.5

1.0

x104

1.5

2.0

214621442142 21502140 2148

Figure 1: MALDI-TOF mass spectrum of polystyrene in the 2 kDa mass range obtained in reflector mode. Clearly one series of ion signals can be observed corresponding to one end-group combination present in the sample. The insert shows the isotopic distribution of the 19mer attached by Ag+

1309

.6

1101

997.

4

2246

.240

2142

.178

2038

.114

1934

.051

2350

.303

2454

.369

1829

.989

2558

.431

1725

.925

2662

.494

1621

.863

2766

.557

1517

.800

2870

.623

1413

.739

2974

.686

75

3078

.747

1205

.611

3182

.819

.547

3286

.860

88

the average molecular weights Mn (2137.67 Da) and Mw (2212.31 Da), the dispersity ĐM=1.035, the degree of polymerization (DP, 20.54), and the percentage of the total ion signal covered by the individual ion series.

The spectrum shown in Figure 3 was obtained from sample PS2 using linear mode of the autoflex maX. The average distance of 104.19 Da (average masses) between the signals in the spectrum is in good

agreement with the theoretical average mass value of 104.15 Da for C8H8. The interpretation by PolyTools (Figure 4) confirms the expected combination of end-groups. Due to the lower mass accuracy of linear mode collection, the spectrum interpretation shows a residual mass of 0.23 Da, which is typical for a polymer in this mass range.

Even at much higher masses, e.g. between 100 - 300 kDa, still the average molecular weights can be calculated by PolyTools (Figure 5). As shown in Figure 6, at this high mass range the individual oligomers cannot be resolved. It is noteworthy that higher charged species appear in the spectrum such as one polystyrene molecule carrying two silver ions or a cluster of two polystyrene molecules sharing three silver ions (as the doubly charged dimer appears at the same mass as the singly charged monomer).

Figure 2: PolyTools interpretation of the spectrum of polystyrene in the 2 kDa mass range. Monomer detection, average mass calculations and end group analysis are done automatically

Figure 3: Spectrum of PS2 in the mass range 14 to 22 kDa acquired using the linear mode of the autoflex maX. In this mass range the isotopic resolution is lost, but the oligomers can be separated as shown in the insert

1755

9.1

1693

4.5

1724

6.8

1787

2.0

1662

1.7

1818

3.9

1620

5.2

1849

6.2

1870

4.6

1589

2.6

1568

4.2

1901

7.1

1537

1.9

1943

3.9

1505

8.7

1974

7.1

1464

1.8

2005

9.0

2037

0.6

Inte

nsit

y [a

.u.]

m/z

0

1000

2000

3000

14,000 15,000 16,000 17,000 18,000 19,000 20,000 21,000 22,000

Inte

nsit

y [a

.u.]

m/z

1000

2000

3000

4000

16,900 16,950 17,05017,000

0

1693

4.5

1703

8.7

Styrene

Figure 4: PolyTools interpretation of the spectrum of PS2. Using a slightly higher mass tolerance the monomer mass and the end-groups/cation are detected correctly

Figure 5: Using an artificially created peaklist, PolyTools is able to calculate average mass values for the spectrum of PS3

Figure 6: Linear mode MALDI-TOF spectrum of the PS3 sample in the mass range of 100 to 280 kDa. Oligomeric species cannot be separated anymore

[M+2Ag]2+

[2M+3Ag]3+

[M+Ag]+

Inte

nsit

y [a

.u.]

m/z

0

500

1000

1500

100,000 120,000 140,000 160,000 180,000 200,000 220,000 240,000 260,000 280,000

Conclusion

The combination of a high-resolution MALDI-TOF MS system and PolyTools provides detailed information on

• Average molecular weight determination

• Automatic detection of monomer

• End-group characterization

together with

• Single end-group analysis using MS/MS [2]

• Copolymer analysis [3]

• Detection of impurities

Bru

ker

Dal

toni

cs is

con

tinua

lly im

prov

ing

its p

rodu

cts

and

rese

rves

the

rig

ht

to c

hang

e sp

ecifi

catio

ns w

ithou

t no

tice.

© B

ruke

r D

alto

nics

04

-202

0, M

T-12

6, 1

8773

95

Bruker Daltonik GmbH

Bremen · GermanyPhone +49 (0)421-2205-0

Bruker Scientific LLC

Billerica, MA · USA Phone +1 (978) 663-3660

For Research Use Only. Not for Use in Clinical Diagnostic Procedures.

[email protected] – www.bruker.com

Learn More

You are looking for further Information? Check out the link or scan the QR code for more details.

www.bruker.com/autoflex

References

[1] Plastics - Determination of the molecular mass and molecular mass distribution of polymer species by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF-MS) (ISO 10927:2018); German version EN ISO 10927:2018

[2] Knop et al. (2010). Systematic MALDI-TOF CID Investigation on Different Substituted mPEG 2000, Macromol. Chem. Phys. 211, 677–684

[3] Weidner et al. (2007). A novel software tool for copolymer characterization by coupling of liquid chromatography with matrix-assisted laser desorption/ionization time-of-flight mass spectrometry, Rapid Commun. Mass Spectrom. 21, 2750–2758