2
 Sample vessels and reagents Avoid the use of colored tubes and gloves, which may contain metal traces We recommend to work with reusable vessels and beakers made of PTFE PVA: Polyvinyl alcohol acts as a lm former and leads to at, homogenous sample lms Sample preparation on carriers Pipette sample in the centre of the carrier Slightly agitate suspensions when pipetting After drying the sample, spot size should not exceed 100 µm in height and ø 10 mm Acids need to be evaporated completely Sample drying By vacuum in a desiccator (not suitable for blood and tissues) By heat in an oven at 60 °C On a Ceran™ heat plate; avoid contamination by covering the discs with a petri dish Siliconization of quartz carriers For aqueous samples the quartz carriers need to be siliconized Pipette a droplet of silicon solution (e.g. Serva™) and dry. The hydrophobic surface avoids spreading of the sample. Fast dilution procedure Dilute a sample aliquot with 0.2 Vol.-% PVA or ultrapure water (whole blood 1:1, serum 1:5) Add 10 µl IS solution (Ga, end conc. 0.2 mg/L) and homogenize Blood samples: pipette 10 µl on an untreated sample carrier and dry by heat (60 °C) Salt water: pipette 10 µl on a siliconized sample carrier, dry by vacuum and measure Cl, Br. Afterwards, the Cl matrix can be remo- ved with 5 µl HNO 3 and subsequent heating. Samples with volatile elements For samples containing volatile elements the use of xation agents is recommended Hg: add a 0.01 M thiourea solution in a ratio of 1:1 and dry by air in a clean bench The basic procedure Fill micro reaction tube with 1 ml sample Add 5 µl internal standard (IS) solution (Ga, 100 mg/L, end concentration 500 µg/L) and homogenize Pipette 10 µl on a siliconized sample carrier and dry by heat (60 °C) or vacuum Avoid contamination Use a laminar ow box as working place with air ltration for a contamination-free prepara- tion and storage of samples and equipment Store sample discs in suitable containers. During preparation open these containers and other tubes and vials just as long as necessary Fast dilution procedure Dilute 500 µl of the sample (homogenate) with 0.2 Vol.-% PVA water (1:1) Add 10 µl IS solution (Ga, 50 mg/L, end conc. 0.5 mg/L) and homogenize Pipette 10 µl on a siliconized sample carrier and dry by heat or air Digestion procedure Add 100 µl conc. nitric acid to 400 µl sample and heat to 70 °C for 1 hour After cooling add 10 µl IS solution (Ga or Y, 25 mg/L, end conc. 0.5 mg/L) and homogenize Pipette 10 µl on a siliconized sample carrier and dry by heat or air Semi-quantitative analysis of thin layers Cut out a 5 x 5 mm section of the thin layer with a ceramic knife Fix the section on the center of the carrier with some ultrapure vacuum grease Standardless quantication mode delivers relative element ratios Quantitative extraction of thin layers Put lter in a 7 ml THF vessel (e.g. Savilex), add 2 ml HNO3 (p.a.) and 20 µl Se IS solution (1g/L) Place in an ultraconic bath for 15 min Rinse lter with 1 ml HNO 3, homogenize and transfer 10 µl to sample carrier Quantitative procedure without digestion Grind sample in a mortar or ball mill Weigh about 50 mg in a tube and add 2.5 ml 1 Vol.-% aqueous Triton X-100 solution Add 10 µl IS solution (Se, 1 g/L, end conc. 200 mg/kg) and homogenize in an ultrasonic bath for 15 min Pipette 10 µl on a siliconized sample carrier and dry by vacuum Procedure for determination of element ratios Transfer some particles (recommended size < 20 µm) to the center of the carrier, x with vacuum grease, if necessary Apply for direct measurement without further treatment Sample Types and Preparation for TXRF Quantitative procedure without digestion Grind dried sample in a mortar or ball mill Weigh about 50 mg in a tube and add 2.5 ml 1 Vol.-% aqueous Triton X-100 solution Add 10 µl IS solution (Ga, 1 g/L, end conc. 200 mg/kg) and homogenize in an ultrasonic bath for 15 min Pipette 10 µl on a siliconized sample carrier and dry by vacuum Procedure for micro samples Weigh the sample in a micro cup with an ultra-balance and note the exact weight Add 500 µl Triton X-100 and 10 µl internal standard (Ga, 100 mg/L, end conc. 20 mg/kg) Homogenize, load carrier with 10 µl suspension Sample Preparation Water Potable, river, rain, sea and waste water Pure chemicals Acids, bases, solvents Body uids Blood, serum, urine Salted uids, beverages Sea, waste water, wash solutions, juices Sample Type and Examples Tissues Homogenates of hair, kidney, liver, lung Multi-phase mixtures Liquid nutrients, food homogenates Plant and wooden material Algae, leaves, braid, moss, needles, roots Food and nutrients Fish, meat, vegetables Oils, pigments Fuels, crude oil, fat, grease, creams, inks Thin layers Contaminations, lms, foils, precipitates Soils, minerals Sediments, sewage sludge, ores, silicates Particles, hi-tech materials Fines, nanoparticles, aerosols, ue-ash Golden Rules of TXRF Why wait for Sample Prep! www.s2picofox.com QuantitativeTXRFanalysisof particles: particle size must be below 70 µm, for best accuracy a size of < 20 µm is recommended The representativity of small amounts of inhomogenous samples might be limited In case of small sample amounts the internal standard (IS) is used for dilution Example: 10 µl of whole blood is diluted with 10 µl of IS solution (0,4 mg/L). After homoge- nization 10 µl of the mixture is pipetted on a carrier Remarks Triton X-100 is a common detergent, which is used to adjust the viscosity of aqueous solutions For matrix-rich samples the calibration for light elements (Z < 20) should be optimized Alternatively a fast disruption procedure can be applied to tissues: – Treat tissue in buffer in an ultrasonic bath – Separate particles by centrifugation – Add IS to supernatant and transfer to disc For the analysis of oils and greases a pipette might not be suitable: Weigh about 500 mg dissolved or liquid sample  – Add 20 µl organic IS (1 g/L) and homogenize Wipe a thin sample lm on the disc (cotton bud)  – Measure without drying Detection Limits Detection limits for NIST fresh water reference standard Detection limits for urine, whole blood and serum of non-digested reference standards Detection limits for plant standards prepared as suspension Detection limits for sh standards prepared as suspension or after digestion Detection limits for glass samples prepared as suspension Detection limits for industrial grease samples dissolved in organic solvents Detection limits down to 0.1 µg/L Total element content of solution and particulate matter will be determined Detection limits below 20 µg/L for metals Fast sample preparation by a simple dilution step, no digestion required Fast sample preparation by a simple dilution step possible; improved detection limits after treatment with nitric acid High linear range from 50 ppb to 5 % Preparation of a suspension after grinding avoids time-consuming digestion Easy analysis of wet and lubricious samples TXRF Benets Due to the high sensiti vity of TXRF, the “organic lm“ method delivers instant results Non-destructive measurements of lms and foils possible Extremely fast preparation for determinin g relative element ratios of particles Fast element screening prior to an ICP-MS measuremen t (Bruker820-MS)     ©    2    0    1    1    B   r   u    k   e   r    N   a   n   o    G   m    b    H  .    O   r    d   e   r    N   o  .    D    O    C   -    P    8    1   -    E    X    S    0    0    4  .    P    h   o    t   o   s   :       B    l   o   o    d   s   a   m   p    l   e       (    C    C    )    J  .    H   e   u   s   e   r  ,       W    h   e   a    t    b   e   r   r    i   e   s       (    C    C    )    Z   a   n    d    l   a   n    d  ,       P   e   a   c    h       (    C    C    )    L   u   c    i   e   n    M   o   n    f    i    l   s  ,       C    i    t   r   u   s    f   r   u    i    t       (    C    C    )    H  .    Z   e    l    l  ,       C    h    i   m   n   e   y       (    C    C    )    S    t   a    h    l    k   o   c    h   e   r  ,       D    i   e   s   e    l   r   a    i   n    b   o   w       (    C    C    )    J   o    h   n  ,       P    i   g   m   e   n    t   s       (    C    C    )    D   a   n    B   r   a    d   y  ,       D   e   s    i   c   c   a    t    i   o   n   c   r   a   c    k   s       (    C    C    )    H   a   n   n   e   s    G   r   o    b   e  . The maximum sample volume applied on a carrier is about 20 µl, the minimum volume depends on the precision of the pipette Multiple pipetting/drying cycles for sample enrichment are possible Bruker AXS GmbH Karlsruhe · Germany Phone +49 (7 21) 5 95-28 88 Fax +49 (7 21) 5 95-45 87 [email protected] Bruker AXS Inc. Madison, WI · USA Phone +1 (608) 276-3000 Fax +1 (608) 276-3006 [email protected] Bruker Nano GmbH Berlin · Germany Phone +49 (30) 670990-0 Fax +49 (30) 670990-30 [email protected] www.bruker.com

Sample Type Preparation Txrf

Embed Size (px)

DESCRIPTION

preparação de amostra TXRF

Citation preview

  • Sample vessels and reagents

    Avoid the use of colored tubes and gloves, which may contain metal traces

    We recommend to work with reusable vessels and beakers made of PTFE

    PVA: Polyvinyl alcohol acts as a film former and leads to flat, homogenous sample films

    Sample preparation on carriers

    Pipette sample in the centre of the carrier Slightly agitate suspensions when pipetting After drying the sample, spot size should not exceed 100 m in height and 10 mm

    Acids need to be evaporated completely

    Sample drying

    By vacuum in a desiccator (not suitable for blood and tissues)

    By heat in an oven at 60 C On a Ceran heat plate; avoid contamination by covering the discs with a petri dish

    Siliconization of quartz carriers

    For aqueous samples the quartz carriers need to be siliconized

    Pipette a droplet of silicon solution (e.g. Serva) and dry. The hydrophobic surface avoids spreading of the sample.

    Fast dilution procedure

    Dilute a sample aliquot with 0.2 Vol.-% PVA or ultrapure water (whole blood 1:1, serum 1:5)

    Add 10 l IS solution (Ga, end conc. 0.2 mg/L) and homogenize

    Blood samples: pipette 10 l on an untreated sample carrier and dry by heat (60 C)

    Salt water: pipette 10 l on a siliconized sample carrier, dry by vacuum and measure Cl, Br. Afterwards, the Cl matrix can be remo-ved with 5 l HNO3 and subsequent heating.

    Samples with volatile elements

    For samples containing volatile elements the use of fixation agents is recommended

    Hg: add a 0.01 M thiourea solution in a ratio of 1:1 and dry by air in a clean bench

    The basic procedure

    Fill micro reaction tube with 1 ml sample Add 5 l internal standard (IS) solution (Ga, 100 mg/L, end concentration 500 g/L) and homogenize

    Pipette 10 l on a siliconized sample carrier and dry by heat (60 C) or vacuum

    Avoid contamination

    Use a laminar flow box as working place with air filtration for a contamination-free prepara-tion and storage of samples and equipment

    Store sample discs in suitable containers. During preparation open these containers and other tubes and vials just as long as necessary

    Fast dilution procedure

    Dilute 500 l of the sample (homogenate) with 0.2 Vol.-% PVA water (1:1)

    Add 10 l IS solution (Ga, 50 mg/L, end conc. 0.5 mg/L) and homogenize

    Pipette 10 l on a siliconized sample carrier and dry by heat or air

    Digestion procedure

    Add 100 l conc. nitric acid to 400 l sample and heat to 70 C for 1 hour

    After cooling add 10 l IS solution (Ga or Y, 25 mg/L, end conc. 0.5 mg/L) and homogenize

    Pipette 10 l on a siliconized sample carrier and dry by heat or air

    Semi-quantitative analysis of thin layers

    Cut out a 5 x 5 mm section of the thin layer with a ceramic knife

    Fix the section on the center of the carrier with some ultrapure vacuum grease

    Standardless quantification mode delivers relative element ratios

    Quantitative extraction of thin layers

    Put filter in a 7 ml THF vessel (e.g. Savilex), add 2 ml HNO3 (p.a.) and 20 l Se IS solution (1 g/L)

    Place in an ultraconic bath for 15 min Rinse filter with 1 ml HNO 3, homogenize and transfer 10 l to sample carrier

    Quantitative procedure without digestion

    Grind sample in a mortar or ball mill Weigh about 50 mg in a tube and add 2.5 ml 1 Vol.-% aqueous Triton X-100 solution

    Add 10 l IS solution (Se, 1 g/L, end conc. 200 mg/kg) and homogenize in an ultrasonic bath for 15 min

    Pipette 10 l on a siliconized sample carrier and dry by vacuum

    Procedure for determination of element ratios

    Transfer some particles (recommended size < 20 m) to the center of the carrier, fix with vacuum grease, if necessary

    Apply for direct measurement without further treatment

    Sample Types and Preparation for TXRF

    Quantitative procedure without digestion

    Grind dried sample in a mortar or ball mill Weigh about 50 mg in a tube and add 2.5 ml 1 Vol.-% aqueous Triton X-100 solution

    Add 10 l IS solution (Ga, 1 g/L, end conc. 200 mg/kg) and homogenize in an ultrasonic bath for 15 min

    Pipette 10 l on a siliconized sample carrier and dry by vacuum

    Procedure for micro samples

    Weigh the sample in a micro cup with an ultra-balance and note the exact weight

    Add 500 l Triton X-100 and 10 l internal standard (Ga, 100 mg/L, end conc. 20 mg/kg)

    Homogenize, load carrier with 10 l suspension

    Sample Preparation

    WaterPotable, river, rain, sea and waste water

    Pure chemicalsAcids, bases, solvents

    Body fluidsBlood, serum, urine

    Salted fluids, beveragesSea, waste water, wash solutions, juices

    Sample Type and Examples

    TissuesHomogenates of hair, kidney, liver, lung

    Multi-phase mixturesLiquid nutrients, food homogenates

    Plant and wooden materialAlgae, leaves, braid, moss, needles, roots

    Food and nutrientsFish, meat, vegetables

    Oils, pigmentsFuels, crude oil, fat, grease, creams, inks

    Thin layersContaminations, films, foils, precipitates

    Soils, mineralsSediments, sewage sludge, ores, silicates

    Particles, hi-tech materialsFines, nanoparticles, aerosols, flue-ash

    Golden Rules of TXRF

    Why wait for Sample Prep!www.s2picofox.com

    Quantitative TXRF analysis of particles: particle size must be below 70 m, for best accuracy a size of < 20 m is recommended

    The representativity of small amounts of inhomogenous samples might be limited

    In case of small sample amounts the internal standard (IS) is used for dilution

    Example: 10 l of whole blood is diluted with 10 l of IS solution (0,4 mg/L). After homoge-nization 10 l of the mixture is pipetted on a carrier

    Remarks Triton X-100 is a common detergent, which is used to adjust the viscosity of aqueous solutions

    For matrix-rich samples the calibration for light elements (Z < 20) should be optimized

    Alternatively a fast disruption procedure can be applied to tissues: Treat tissue in buffer in an ultrasonic bath Separate particles by centrifugation Add IS to supernatant and transfer to disc

    For the analysis of oils and greases a pipette might not be suitable: Weigh about 500 mg dissolved or liquid sample Add 20 l organic IS (1 g/L) and homogenize Wipe a thin sample film on the disc (cotton bud) Measure without drying

    Detection Limits

    Detection limits for NIST fresh water reference standard

    Detection limits for urine, whole blood and serum of non-digested reference standards

    Detection limits for plant standards prepared as suspension

    Detection limits for fish standards prepared as suspension or after digestion

    Detection limits for glass samples prepared as suspension

    Detection limits for industrial grease samples dissolved in organic solvents

    Detection limits down to 0.1 g/L Total element content of solution and particulate matter will be determined

    Detection limits below 20 g/L for metals Fast sample preparation by a simple dilution step, no digestion required

    Fast sample preparation by a simple dilution step possible; improved detection limits after treatment with nitric acid

    High linear range from 50 ppb to 5 %

    Preparation of a suspension after grinding avoids time-consuming digestion

    Easy analysis of wet and lubricious samples

    TXRF Benefits Due to the high sensitivity of TXRF, the organic film method delivers instant results

    Non-destructive measurements of films and foils possible

    Extremely fast preparation for determining relative element ratios of particles

    Fast element screening prior to an ICP-MS measurement (Bruker 820-MS)

    2

    011

    Bru

    ker

    Nan

    o G

    mbH

    . Ord

    er N

    o. D

    OC

    -P81

    -EX

    S0

    04.

    Pho

    tos:

    B

    lood

    sam

    ple

    (CC

    ) J.H

    euse

    r, W

    heat

    berr

    ies

    (CC

    ) Zan

    dlan

    d,

    Pea

    ch

    (CC

    ) Luc

    ien

    Mon

    fils,

    C

    itrus

    fru

    it (C

    C) H

    . Zel

    l, C

    him

    ney

    (CC

    ) Sta

    hlko

    cher

    , D

    iese

    lrain

    bow

    (C

    C) J

    ohn,

    P

    igm

    ents

    (C

    C) D

    an B

    rady

    , D

    esic

    catio

    n cr

    acks

    (C

    C) H

    anne

    s G

    robe

    .

    The maximum sample volume applied on a carrier is about 20 l, the minimum volume depends on the precision of the pipette

    Multiple pipetting/drying cycles for sample enrichment are possible

    Bruker AXS GmbH

    Karlsruhe GermanyPhone +49 (7 21) 5 95-28 88Fax +49 (7 21) 5 95-45 [email protected]

    Bruker AXS Inc.

    Madison, WI USAPhone +1 (608) 276-3000Fax +1 (608) [email protected]

    Bruker Nano GmbH

    Berlin GermanyPhone +49 (30) 670990-0Fax +49 (30) [email protected]

    www.bruker.com