VC Training
Table Setup and Floating
9-17-2004
Agenda
• General– Air Supply Requirements
• Assembly– Assembling the System– Leveling the Table– Isolating the Table
• Installing Leveling Valves• Connecting Air Lines• Adjusting Leveling Valves
• Troubleshooting
Air Supply Requirements
• Constant air supply required during operation• Bottled nitrogen• Compressed air
• No CO2 (cause freezing)
– After initial setup air is only consumed to compensate for load changes
– Pressure requirements• Determined by dividing total payload and table weight by
isolator diaphragm area + 10 psi.
(Payload weight + Table weight)
Total isolator diaphragm area (25.9 in2/isolator)
+ 10 psi
Air Pressure Example
RS4000-510-8 (1250 lbs)Payload ( 445 lbs)Total (1695 lbs)
(445 + 1250)
(4 isolators) * (25.9 in2/isolator)
+ 10 psi 26 psi=
Approximate net weights (±10%) for 4 ft (1.2 m) wide steel tables. For 5 ft (1.5 m) wide models add 20%, for 3 ft (0.9 m) models, subtract 25%. Shipping weights can be estimated by adding 100–200 lbs (45–90 kg) for crate weight.
RS4000-510-8
Weight = 1000*1.2 = 1,250lbs
Assembly – Assembling the system
• Position isolators per sketch below (standard length tables)– Custom tables – isolator positions will be included
in a data package provided with system
16.0 (406) 6.0 (1800)
22.0 (559) 8.0 (2400)
27.0 (686) 10.0 (3000)
32.0 (813) 12.0 (3600)
37.0 (940) 14.0 (4200)
42.0 (1067) 16.0 (4800) 59.1 (1500)
47.0 (1194) 18.0 (5400) 48.0 (1200)
53.0 (1346) 20.0 (6000) 36.0 (900)
in. (mm) ft (mm) in. (mm)
DIMENSION (A) LENGTH (L) WIDTH (W)
Assembly – Assembling the system
Bolt isolators to table using clips, bolts and wrench provided
Leveling the Table
• Place a level on the table top in about the center• Adjust the support as required to level the table• When level, make sure each isolator support is
resting against the bottom of the table and is centered between the clips
Isolating the Table - Installing Leveling Valves
• 3 valves are used in all systems regardless of the number of isolators
• Leveling locations are selected in order to form the largest triangle possible
IPV-KT packaged in box IPV valve and associated mounting hardware
Isolating the Table - Installing Leveling Valves
Mounting screw holes
Control arm
Table height sensor
Over-pressure relief valve
Air supply
Metering needle valve
Isolator pressure gauge
Air line to isolator
Isolating the Table - Installing Leveling Valves
• Set the table height sensor adjustment screw to center of travel
• Attach the leveling valve to isolator using the 2 mounting screws and nut plate provided
• Rotate control arms on valves “A” to point to the outside of the table
• Rotate control arm “B” to point to the center of the 2 isolators it is controlling
Isolating the Table - Connecting Air Lines• Connect air lines to the isolator as shown below (use translucent tubing
and tee connectors.• Connect grey tubing between the air line to isolator (hose barb) and the
isolatorA razor blade works best for cutting tubing
Translucent tubing
Grey tubing
Tee connectors
Isolating the Table - Connecting Air Lines
• The adjacent diagrams detail the valve connections to the isolator
Translucent tubing
One Isolator
Two Isolators
Isolating the Table – Adjusting Leveling Valves
• Close metering needle valve on each valve. Re-open 1/8 – ¼ turn.
• Turn on Air supply and adjust to calculated pressure (DO NOT EXCEDE 100psi)
• If table does not float within 5 minutes, slowly increase pressure or adjust valve height adjustment screw to hold the control arm down further. Be sure metering needle valves are open.
• If table oscillates reduce pressure and/or close needle valve slightly
Metering needle valve
Control arm
Table height sensor
Troubleshooting