Water Injections in GC -

Does Water Cause Bleed?

Page 2: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Eberhardt KuhnApplications ChemistApril 4, 2001

Practical Advice and Useful Tips for the Analysis ofSemivolatile Organics by GC and GC/MS

11:00 a.m. ESTTelephone Number: 904-779-4779Chair Person: Lisa Lloyd

Page 3: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Practical Advice and Useful Tips for the Analysis ofSemivolatile Organics by GC and GC/MS

11:00 a.m. ESTTelephone Number: 904-779-4779Chair Person: Lisa LloydStarts in Five Minutes

Eberhardt KuhnApplications ChemistApril 4, 2001

Page 4: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Practical Advice and Useful Tips for the Analysis ofSemivolatile Organics by GC and GC/MS

11:00 a.m. ESTTelephone Number: 904-779-4779Chair Person: Lisa LloydStarts in One Minute

Eberhardt KuhnApplications ChemistApril 4, 2001

Page 5: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Why Inject Water?

•Convenient

• Aqueous samples (waste water,drinking water, etc.)

• Biological samples

•Necessary

• Purge & Trap

Page 6: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Potential Problems with Water

•Injector Issues

• Large expansion volume

• Backflash

Page 7: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Potential Problems with Water

•Detector Issues

• Extinguish FID flame

• Decrease sensitivity of ECD

Page 8: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Potential Problems with Water

•Solvent - stationary phase mismatch

• poor wettability of many stationary phasesby water

• “puddles”

Page 9: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Potential Problems with Water

•Damage to stationary phase

• Change in Retention Times

• Change in Selectivity

• Increase in Bleed

Page 10: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

What is Normal Column Bleed?

Normal background signalgenerated by the elution ofnormal degradation products ofthe column stationary phase

Page 11: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

● thermodynamic equilibrium process occurs tosome degree in all columns

● polysiloxane backbone releases low molecularweight, cyclic fragments

● occurs at low level in low temperature, O2-free,clean system

● increased by increased temperature, oxygenexposure or chemical contamination

�����������

Page 12: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Bleed: Why Does It Happen?“Back Biting” Mechanism of Product Formation

+

Repeat

Si Si Si Si Si Si SiO O O O O O OH

CH3 CH3 CH3 CH3 CH3 CH3 CH3

CH3 CH3 CH3 CH3 CH3 CH3 CH3

OHSiSi O Si O Si OCH3 CH3 CH3 CH3

CH3 CH3 CH3 CH3Si

O O

O

CH3H3C

H3C

H3C

SiSiCH3

CH3

OO O O O

OSi

HO

H3C

CH3 CH3 CH3 CH3

Si SiSi Si Si SiCH3

CH3

CH3

CH3CH3 CH3

CH3

CH3CH3

Cyclic products arethermodynamically

more stable!

Page 13: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Dial 1-904-779-4779 for e-Seminar Audio

Low Bleed Stationary Phases DB-35ms vs. DB-35

* 5ng Decachlorobiphenyl

DB-35ms

8.00 10.00 12.00 14.00 16.00 18.00 20.00

320°C

DB-35

8.00 10.00 12.00 14.00 16.00 18.00 20.00

300°C

CLP Pesticides Analysis

S/N = 10

S/N = 3

Page 14: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Dial 1-904-779-4779 for e-Seminar Audio

Low Bleed Stationary Phases Spectral Purity

DB-35msPrimary Ion Primary Ion

DB-35

Page 15: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

06000

7000

8000

9000

1.0e4

1.1e4

1.2e4

1.3e4

5 10 15 20 25Time (min.)

24 pA / 260°C

DB-624 30M x .53mm I.D., 3.0µm

12 pA / 320°C

DB-1 30m x .32mm I.D., .25µm

Column Bleed is Influenced by:

Phase type

Temperature

Column dimensions

Page 16: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

FILM THICKNESSBleed

More stationary phase = More degradation products

Page 17: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

What is a Bleed Problem?

An abnormal elevated baseline at high temperature

IT IS NOT

A high baseline at low temperature

Wandering or drifting baseline at any temperature

Discrete peaks

Page 18: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Goals of StudyWho cares?

• Frequent Inquiries

• Water can cause problems

• Quantitative data

• Establish “guidelines”

Page 19: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

STATIONARY PHASES

•Dimethylpolysiloxane (DB-1)

•Polyethylene glycol (DB-Wax)

•Cyanopropylphenyl (DB-225)

•Cyclodextrin (CycloSil B)

•Divinylbenzene/ethylene glycoldimethacrylate (HP-PLOT U)

Page 20: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Experimental Conditions

• Instrument: Agilent 6890 with Autoinjector

• Injector: 250o C, 1:5 split

• Injection volume: 1 µL

• Detector: FID, 300o C

• Carrier gas: H2 at 40 cm/sec

Page 21: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Experimental Conditions

• Oven: 130o C, 60o C, or 200o C* isothermal

• Column Dimensions: 30 m x .53 mm I.D. x 1.0 µm#

• 1,000 water injections at each temperature on each column

• Bleed profile after 250, 500, and 1,000 water injections

• Test mix after 250, 500, and 1,000 water injections

• **for Cyclosil B and DB-Wax only #CycloSil B: 30 m x .32 mm x .25 µm

Page 22: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Test MixesCompound List for Chromatograms DB-1

Peak Compound

1 2-chlorophenol

2 undecane

3 2,4-dimethylaniline

4 1-undecanol

5 tetradecane

6 acenaphthalene

7 pentadecane

Page 23: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Test MixesCompound List for Chromatograms DB-Wax

Peak Compound

12345678910

2-octanonetetradecane1-octanolmethyl deconatemethyl undecanoatenaphthalene1-decanolmethyl dodecanoate2,6-dimethylaniline2,6-dimethylphenol

Page 24: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Test MixesCompound List for Chromatograms DB-225

Peak Compound

1 tetradecane

2 2-chlorophenol

3 hexadecane

4 naphthalene

5 2,4-dimethylaniline

6 1-undecanol

7 octadecane

8 ethyl dodecanoate

Page 25: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Test MixesCompound List for Chromatograms CycloSil B

Peak Compound

123456789

dodecane(R)-linalool(S)-linaloolnaphthalene1-nonanolgamma-heptalactonegamma-heptalactonemethyl decanoatetetradecane

Page 26: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Test MixesCompound List for Chromatograms HP- PLOT U

Peak Compound

1 methanol

2 ethanol

3 pentane

4 ethyl ether

5 acetone

6 hexane

7 ethyl acetate

Page 27: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Break Number 1

• For Questions and Answers

• Press *1 on Your Phone to

• Ask a Question

Page 28: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-1

Parameter Before Inj. After Inj.

Ret. Factor

Ret. Index 1

Ret. Index 2

Theor. plates

Bleed (pA)

14.6 14.5

1349.88 1350.02

1427.77 1428.16

1448 1474

12.8 11.2

Page 29: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-Wax

Parameter Before Inj. After Inj.

Ret. Factor

Ret. Index 1

Ret. Index 2

Theor. plates

Bleed (pA) Bleed 2 Bleed 3 (final)

12.6 12.6

1149.54 1149.73

1163.44 1163.71

1277 1261

50.542.244.8

97.841.032.1

Page 30: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

DB-Wax before and after 2,000 water injections

Page 31: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-225

Parameter Before Inj. After Inj.

Ret. Factor

Ret. Index 1

Ret. Index 2

Theor. plates

Bleed (pA)

11.5 11.4

1622.30 1621.26

1711.51 1711.03

1101 1110

15.0 114.0

Page 32: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results CycloSil B

Parameter Before Inj. After Inj.

Ret. Factor

Ret. Index

Resolution

Theor. plates

Bleed (pA)

7.8 7.6

1306.3 1306.0

1.9 1.3

2631 2025

28.4 15.1

Page 33: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results HP-PLOT U

Parameter Before Inj. After Inj.

Ret. Factor

Ret. Index

Bleed (pA)

5.2 5.3

538.0 540.0

950 982

74.2 35.6

Theor. plates

Page 34: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-1Asymmetry

Compound Before Inj. After Inj.

chlorophenol

dimethylaniline

undecanol

0.48 0.47

0.47 0.46

0.51 0.46

Page 35: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Asymmetry (Skew)

Skew numbers are based on a modifiedGaussian-peak model*. It relies more onthe zeroth and first moment calculations,which are less influenced by baselinenoise.

*W.W.Yau, Anal. Chem., vol. 49, No. 3 (1977), pp 395-398

Page 36: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Asymmetry (Skew)

The numbers represent the deviationfrom an ideal Gaussian, i.e., perfectlysymmetrical, peak.

The smaller the number, the better.

Noticeable tailing starts at ~0.8.Obvious tailing start at ~1.2

Page 37: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Skew Factors DB-1

0.3

0.35

0.4

0.45

0.5

0.55

1 3 5 7 9

chlorophenol

dimethylaniline

undecanol

Page 38: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-WaxAsymmetry

Compound Before Inj. After Inj.

decanol

dimethylaniline

dimethylphenol

0.38 0.47

0.30 0.32

0.25 0.27

Page 39: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Results DB-225Asymmetry

Compound Before Inj. After Inj.

chlorophenol

dimethylaniline

undecanol

0.46 0.47

0.42 0.43

0.63 0.58

Page 40: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Break Number 2

• For Questions and Answers

• Press *1 on Your Phone to

• Ask a Question

Page 41: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

For water injections on all columns withbonded stationary phases, no change in:

polarityselectivityretentionefficiencyactivitybleed

is observed.

Page 42: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

For non-bonded phases, like CycloSilB, water injections can wash out partof the non-bonded stationary phase -loss of resolution, retention, and,possibly, efficiency.

Page 43: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

Amount of wash-out is dependent ontemperature: at 130o C and 200o C itis minimal, at 60o C it is noticeable.

Solubility of phase material is greaterin liquid water.

Page 44: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

Amount of wash-out is gradual!

Most likely dependent on solubility of phasein water (cyclodextrin = high).

No change in selectivity.

No increase in bleed!

Page 45: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

��������������� ����������������������

2

2+2+

&+��2+

222

2

2+

2+2+

2+

&+��2+ &+

��2+

2+2+

2+ 2+ 2+ 2+2+

2+

��*OXFRVH�PROHFXOHV�MRLQHG�WKURXJK�����JO\FRVLGLF�OLQNDJHV3ULPDU\�K\GUR[\O�JURXSV�SDUWLDOO\�EORFN�WKH�EDVH&DYLW\�VL]H�LV�DERXW�������c�LQ�GLDPHWHU+\GUR[\�JURXSV�DUH�SHUDON\ODWHG

Page 46: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

For bonded PLOT columns, no negativeeffects of injecting water were observed.

N.B. Some PLOT columns, e.g., Aluminaand Molesieve, may not be suitable forwater injections. Observe manufacturersrecommendations.

Page 47: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Summary

For all phases, the time to bleed down, or re-condition, a column after injecting water isdependent on the run temperature:

low-temperature injections take a lot longer than high-temperature injections.

Recommend periodic bake-out if injectingwater at <80o C.

Page 48: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Conclusions

Non-polar columns (bonded and cross-linked) : safe to inject and rinse with water

Polar columns (bonded and cross-linked) :water injections are safewater-rinsing not recommended

Non-bonded columns: water injections canwash out stationary phase - use with caution

Page 49: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Conclusions

Problems associated withwater injections are oftencaused by injector relatedphenomena, e.g., backflash.

Page 50: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

Solvent Expansion Volumes

Solvent Vapor volume (µL)

of 1µL liquid

Water 1010

Methanol 450

Carbon disulfide 300

Methylene chloride 285

Acetone 245

n-Hexane 140

Page 51: Water Injections in GC - Does Water Cause Bleed? · 11:00 a.m. EST Telephone Number: 904-779-4779 Chair Person: Lisa Lloyd. ... CLP Pesticides Analysis S/N = 10 S/N = 3. Dial 1-904-779-4779

J&W Scientific Technical Support

It’s Free!

1-800-552-0413

111-800-227-9770 (toll-free in US)

(916) 985-7888 (phone)

(916) 985-1101 (FAX)

web site:

www.jandw.com