Download pptx - Insertion of MonaLisa 5

Transcript
Page 1: Insertion of       MonaLisa  5

Insertion of MonaLisa 5

Page 2: Insertion of       MonaLisa  5

plunger

blue flange

Insertion tube

IUD stem and arms

MonaLisa 5

Page 3: Insertion of       MonaLisa  5

Slide blue flange to the length of the uterus as measured by the uterine sound

Page 4: Insertion of       MonaLisa  5

Use the string to pull the IUD into the insertion tube

Page 5: Insertion of       MonaLisa  5

Insert the plunger into the tube

Page 6: Insertion of       MonaLisa  5

Insert IUD into the uterus until the blue flange is 2 cm from the cervix

Page 7: Insertion of       MonaLisa  5

Release the IUD arms by keeping the tube steady and advancing the plunger by 2 cm

Page 8: Insertion of       MonaLisa  5

Notice the ridge on the plunger at the 2 cm mark

Page 9: Insertion of       MonaLisa  5

Release IUD by withdrawing the tube, holding the plunger steady

Page 10: Insertion of       MonaLisa  5

To avoid pulling the IUD out of position, withdraw the plunger before withdrawing the tube

Page 11: Insertion of       MonaLisa  5

Cut the string about 2 cm long

Page 12: Insertion of       MonaLisa  5

For easier insertion in a curved uterus, the IUD can be curved before insertion