1 Physics & Instrumentation in Positron Emission Tomography Paul Vaska, Ph.D. Center for Translational Neuroscience Brookhaven National Laboratory July 21, 2006

The Physics of Positron Emission Tomography (PET)

Embed Size (px)

Citation preview

Page 1: The Physics of Positron Emission Tomography (PET)

1

Physics & Instrumentation inPositron Emission TomographyPhysics & Instrumentation in

Positron Emission Tomography

Paul Vaska, Ph.D.

Center for Translational NeuroscienceBrookhaven National Laboratory

July 21, 2006

P matrix = 0.6mm, = 0.1

Page 2: The Physics of Positron Emission Tomography (PET)

2

Non-invasive Medical Imaging Techniques

Non-invasive Medical Imaging Techniques

Anatomical• X-ray• CAT• MRI• Ultrasound

Functional• “nuclear medicine” - SPECT, PET• Optical fluorescence, …

CAT

X-Ray

MRI

Page 3: The Physics of Positron Emission Tomography (PET)

3

Positron Emission TomographyPositron Emission Tomography

Recent mainstream acceptance- relatively expensive

- cyclotron for tracer production

- detectors must stop high-energy gamma-rays

- low resolution (>2 mm), limited counting statistics

- BUT unique functional capabilities

Applications- Diagnosis of disease

- cancer (WB), cardiac, …

- Research- brain function

- animal studies

Page 4: The Physics of Positron Emission Tomography (PET)

4

Technical Challenges in PET ImagingTechnical Challenges in PET Imaging

Radiochemistry - better tracers Imaging Physics - better images by

• Detector design– Spatial resolution– Sensitivity

• Image processing– Corrections for physical effects– Image reconstruction algorithms

Data Analysis & Biological Modeling - better interpretation of images

Page 5: The Physics of Positron Emission Tomography (PET)

5

PET Imaging OverviewPET Imaging Overview

- Synthesize radiotracer- Inject radiotracer- Measure gamma-ray

emissions from isotope (~20-60 min)

- Reconstruct images of radiotracer distribution (nCi/cc)

Page 6: The Physics of Positron Emission Tomography (PET)

6

+

+ +

NucleusNucleus

NeutronsNeutrons

ProtonsProtons

ElectronsElectrons

Positron (Positron (++) Decay) Decay

18F-FDG

Page 7: The Physics of Positron Emission Tomography (PET)

7

++ Decay Decay

+

+

+

+

+

+

+

+

+

Neutron-deficient isotopes can decay by emitting Neutron-deficient isotopes can decay by emitting positronspositrons

+

anti-neutrinoanti-neutrino

positronpositron Net effect: one Net effect: one proton proton replaced by replaced by

• neutronneutron• anti-neutrinoanti-neutrino• positronpositron

Page 8: The Physics of Positron Emission Tomography (PET)

8

Positron annihilationPositron annihilation

Annihilation gives• 2x 511 keV gamma rays• 180 degrees apart• Line of response

Positron range & gamma noncollinearity

Scanner is just a photon counter!• Counts gamma-ray pairs vs.

single gammas• Time window ~ 1 ns

511 keV

511 keV

e+

e-

Page 9: The Physics of Positron Emission Tomography (PET)

9

Raw Data & Image ReconstructionRaw Data & Image Reconstruction

0 projection

0

180

90

90 p

roje

ctio

n

image reconstruction

“sinogram”

Page 10: The Physics of Positron Emission Tomography (PET)

10

Important Detector PropertiesImportant Detector Properties

- Spatial resolution- Directly controls spatial resolution

in reconstructed image- Currently ~ 1 - 5 mm- Depth-of-interaction?

- Reduces “parallax”

Page 11: The Physics of Positron Emission Tomography (PET)

11

Important Detector PropertiesImportant Detector Properties

- Detection efficiency (aka sensitivity, stopping power)- Reduces noise from counting statistics- Currently > ~ 30% (singles)

55M Events1M Events

Page 12: The Physics of Positron Emission Tomography (PET)

12

Important Detector PropertiesImportant Detector PropertiesImportant Detector PropertiesImportant Detector Properties

Random (accidental) coincidence

- Time resolution- Affects acceptance of random

coincidences- Currently ~ 1 - 10 ns- Time-of-flight (TOF)?

- c = ~ 1 ft/ns- Need << 1 ns resolution

Page 13: The Physics of Positron Emission Tomography (PET)

13

Important Detector PropertiesImportant Detector PropertiesImportant Detector PropertiesImportant Detector Properties

Scatter and Attenuation

511 keV

- Energy resolution- Scattered gammas change

direction AND lose energy- Affects acceptance of scattered

coincidences- Currently ~ 20%

- Deadtime- Handle MHz count rates!

511 keV 400 keV

Page 14: The Physics of Positron Emission Tomography (PET)

14

ScintillationCrystal

PMTPre-Amplifier+ Electronics

Gamma photon converts to optical photons (proportional to gamma energy, typ. 1000’s)

photons are collected at the end of the crystal

light is converted to an electrical signal & amplified

Front-end electronics condition the signal for further processing

Prototypical PET DetectorPrototypical PET Detector

Gamma Ray

Optical reflector

Page 15: The Physics of Positron Emission Tomography (PET)

15

New DevelopmentsNew Developments

• Detectors

• Multimodality imaging

• Specialized applications

Page 16: The Physics of Positron Emission Tomography (PET)

16

Scintillator NaI(Tl) BGO GSO LSO LuAP LPS LaBr (ns) 230 300 60 40 18 30 35

(cm-1) 0.35 0.95 0.70 0.86 0.95 0.70 0.47E/E (%) 6.6 10.2 8.5 10.0 ~15 ~10 2.9

Rel. light output (%) 100 15 25 70 30 73 150175

25

New Developments: DetectorsNew Developments: Detectors

• Scintillators• No perfect choice - tradeoffs

• Also practical qualities• Rugged?• Hygroscopic?• Cost?

Page 17: The Physics of Positron Emission Tomography (PET)

17

New Developments: DetectorsNew Developments: Detectors

• Photosensors• Photomultiplier tubes• Avalanche photodiodes

• Arrays, position-sensitive• Compact but noisier

• Silicon photomultipliers• Very new• Best of both?

APD array

PMT

SiPM

Page 18: The Physics of Positron Emission Tomography (PET)

18

New Developments: DetectorsNew Developments: Detectors

• Solid-state detectors• Direct conversion,

no photodetector • Great dE/E &

spatial resolution• Poorer timing &

stopping power• CZT Z2

Z1

Sa2

Sa1

Sc

Page 19: The Physics of Positron Emission Tomography (PET)

19

New Developments: DetectorsNew Developments: Detectors

• Pb converters & ionization

HIDAC Pb-walled straws (50 cm long)

Page 20: The Physics of Positron Emission Tomography (PET)

20

New Developments: DetectorsNew Developments: Detectors

• 3D gamma-ray event positioning• Depth of interaction• Reduces parallax problem

LSOslab

APDslab

s

vs.

LSO slab

crystal holder

APD

decoupling capacitor

HV filter capacitor

Current-limiting resistor

signal output connector

SHV connector

unused APD slot

Page 21: The Physics of Positron Emission Tomography (PET)

21

New Developments: DetectorsNew Developments: Detectors

• Time of flight using LaBr3n

o T

OF

300 p

s T

OF

1 Mcts 5 Mcts 10 Mcts

Page 22: The Physics of Positron Emission Tomography (PET)

22

New DevelopmentsNew Developments

• Multimodality imaging• PET/CT• PET/MRI

• Specialized applications• Brain, breast, prostate• Small animal - microPET• Arterial input function

• Humans - wrist scanner• Animals - microprobe

• Awake rat brain - RatCAP

Page 23: The Physics of Positron Emission Tomography (PET)

23

RatCAP: Rat Conscious Animal PET

Eliminate anesthesia in preclinical neuroscience using PET in order to:

• Remove confounding effects of anesthetic on neurochemistry

• Enable stimulation in animal PET

• Enable correlations of behavior and neuro-PET

Page 24: The Physics of Positron Emission Tomography (PET)

24

Architecture

Detector blocks x12• LSO 2.2 x 2.2 x 5 mm in 4 x 8 array• 1:1 coupling to APD• ASIC - single all digital output

Timestamp & Signal Processing Module• Programmable real-time logic (FPGA)• 1 ns bins (debugging, now 10 ns)

Data acquisition• PCI card in standard PC• Up to 70 MB/s = ~10 Mcps singles• Offline software for coincidences, corrections, recon, …

TSPM TDC

PCI card

ASIC

opticaldifferential

RatCAP

Page 25: The Physics of Positron Emission Tomography (PET)

25

Architecture

RatCAP

TSPM

LSO APD

ASICs

all interconnections

38 mm FOV

72 mm OD optical links to PCI

high voltage

194 g

data, clock, power

18 m

m a

xial F

OV

Page 26: The Physics of Positron Emission Tomography (PET)

26

Performance

Spatial resolution (FWHM @ CFOV)• FBP: 2.1 mm• MLEM: <1.5 mm

Energy resolution: 23% FWHM

Time resolution: 14 ns FWHM• window = 30 ns

Sensitivity (point @ CFOV): 0.7%

Peak Noise Equivalent Count rate: 14 kcps @ 5 Ci/cc

1st prototype: LLD = 150 keV average, variable

Page 27: The Physics of Positron Emission Tomography (PET)

27

Imaging Conditions

Anesthetized 250-350 g rats

Limited DAQ livetime >> long scans for statistics

Artifacts

Page 28: The Physics of Positron Emission Tomography (PET)

28

F-18 Fluoride Bone Scan

1.3 mCi fluoride

RatCAP microPET R4

Page 29: The Physics of Positron Emission Tomography (PET)

29

C-11 Raclopride

1.8 mCi raclopride

In the RatCAP

Page 30: The Physics of Positron Emission Tomography (PET)

30

C-11 Methamphetamine

Time-activity curve for striatum

Page 31: The Physics of Positron Emission Tomography (PET)

31

Thanks!Thanks!