4
7090 SERIES Nitrogen & Sulfur GC detectors

7090 GC-NSD brochure - ken-pd.com · Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig) Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig) Outputs 1 mV, 1 V, 10 V Combustion

  • Upload
    lambao

  • View
    215

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 7090 GC-NSD brochure - ken-pd.com · Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig) Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig) Outputs 1 mV, 1 V, 10 V Combustion

7090 SERIESNitrogen & Sulfur GC detectors

Page 2: 7090 GC-NSD brochure - ken-pd.com · Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig) Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig) Outputs 1 mV, 1 V, 10 V Combustion

P r o d u c t f e a t u r e sPatented Pyro-chemiluminescent™nitrogen and sulfur technology

� one injection — two results

� add to FID for one injection — three results

� highly selective for nitrogen and sulfur

� equimolar response for all nitrogen and sulfur bearing compounds

Maximum uptime

� long-term operation on one calibration

� 14" dual zone furnace for optimal oxidation and reduction

� complete sample combustion enhances short- and long-term

stability, equimolarity, and selectivity

� long ceramic tube life: 6 –12 months

� oil-less vacuum pump operates maintenance-free for several years

The leadership continues…Building upon years of innovation, analytical leadership, and

customer satisfaction, Antek sets the standard for simultaneousnitrogen and sulfur GC detection.

Utilizing accuracy and precision of our Pyro-chemilumines-cent™ technology, the 7090NS Nitrogen/Sulfur GC Detectordelivers equimolar response for all nitrogen- and sulfur-containing compounds. Chemiluminescence is the clean, fast,interference-free method for determining all bound nitrogenand sulfur in a variety of matrices.

R-S + R-N + O 2 —� SO 2 + NO + other oxidesR-S + R-N + H 2 —� H 2 S + NO + other reduced speciesH 2 S + O 3 —� SO 2 * —� SO 2 + h υ (200–400 nm)NO + O 3 —� NO 2 * —� NO 2 + h υ ' (600–900 nm)

T h e A n t e k A d v a n t a g e

� Experiencedwe are the experts at nitrogen and sulfur analysis technologywith over 25 years experience

� Recognizedmethods of choice by regulatory agencies and method writinginstitutions

� Responsiveour experienced sta� of highly trained chemists will help yousolve your analytical problems or �t the 7090 Series to yourapplication

� Cost-e�ectiveAntek’s GC detectors o�er a low-cost alternative for selec-tive multi-element detection

N i t r o g e n , N i t r o s a m i n e s , S u l f u rAntek Instruments, Inc. o�ers the most advanced detectors

for GC, SFC, and SimDis. With the Model 7090’s versatility,users can buy one detector now—nitrogen or sulfur—and addthe other later.

The Model 7090N Chemiluminescent Nitrogen Detector isspeci�c for nitrogen compounds and easily converts to nitro-samine-selective mode without any hardware change. TheModel 7090S Sulfur Chemiluminescent Detector is speci�c forsulfur compounds. Our Model 7090NS o�ers unprecedentedsimultaneous detection of nitrogen and sulfur compounds.

Conventional element enhanced detectors are less selectiveand therefore still have problems with interfering compounds.The 7090 Series eliminates those problems by detecting nitro-gen and/or sulfur compounds only—without halogen,phosphorous, or hyrocarbon interferences that have complicat-ed this work in the past. This unique detector also gives a linearresponse, and has sensitivity that is stable over time, providingmore reliable results.

7090 SERIES

Page 3: 7090 GC-NSD brochure - ken-pd.com · Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig) Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig) Outputs 1 mV, 1 V, 10 V Combustion

Month 6

S a m p l e d a t a

Above : The Model 7090 GC-NSD is ideal for a variety of applicationsincluding agriculture, biotechnology, biochemical, food, solid phasesynthesis, organic synthesis, petrochemical, pharmaceutical, and polymer/plastics. At left : ANTEK’s 7090 Series retro�ts with most GCs on themarket. Shown here with Hewlett Packard’s Model 6890.

Versatile

� nitrogen, sulfur or simultaneous nitrogen/sulfur— buy one

detector, add a second later

� compatible with most commercially available GCs

� electronic mass �ow control for all gases

� modular designed detectors and electronics

Satisfaction guaranteed

� application solutions by experienced, knowledgable technical sta�

� one year limited warranty on parts and labor

� in-depth training courses o�ered at corporate headquarters

Heavy diesel Regular unleaded gasoline

Det

ecto

r Re

spo

nse

32 minutes 40 minutes

FID

Nitrogen

Sulfur

FID

Nitrogen

Sulfur

Det

ecto

r Re

spo

nse

Stability45 injections at 9.24 ppmS

RepeatabilityMonth 1

CS2: 7%

MeSH: 7 %

EtSH: 7 %DMS: 9 %COS: 9 %nBUSH: 9 %thiophene: 7 %iBUSH: 11 %H 2S: 22%thiophenol: 9%

% deviation

CS2: 10%

H 2S: 9.7 %

MeSH: 8.4 %DMS: 11.2 %EtSH: 10 %COS: 10.9 %thiophene: 9.9 %iBUSH: 8.8 %

nBUSH: 9.0%

thiophenol: 10.1%

Page 4: 7090 GC-NSD brochure - ken-pd.com · Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig) Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig) Outputs 1 mV, 1 V, 10 V Combustion

USA : Call (800) 444-TEST, (281) 940-1803 or Fax (281) 580-07198824 Fallbrook Drive, Houston, Texas 77064E-mail: [email protected], Web site: www.paclp.com

I-7090-99 ©Copyright 2002 Antek Instruments L.P. Printed in USA

P r o d u c t s p e c i � c a t i o n sDetection Method

Pyro-chemiluminescence™

PerformanceSensitivity : <3 pgN/sec or <1 pgS/sec (individually optimized);<30 pgN/sec and <10 pgS/sec (simultaneous detection)Selectivity: >10 7 to hydrocarbon and nitrogen/sulfur by weightLinearity : >10 4

Equimolarity : <±10%

Gases“Chromatographic” or “Zero Grade”Dry oxygen : 2–20 cc/min 99.75%, 1.5 bar (20 psig)Dry hydrogen : 50–400 cc/min 99.995%, 1.5 bar (20 psig)

Outputs1 mV, 1 V, 10 V

Combustion Zone Temperature950ºC–1100ºC in 1ºC increments

Power115 VAC 50/60 Hz 12 Amp or230 VAC 50/60 Hz 6 Amp

PhysicalDimensions: 36 cm wide (14") x 59 cm deep (23") x61 cm high (24")Weight : 28 kg (62 lb)

Patents and ApprovalsU.S. Patents : #4,018,562, #4,778,764, #4,843,016, #4,352,779,#4,678,756; U.K. Patent : #1578281; other patents are pending;ASTM : D5504, D5623; GPA : 2199-99

O r d e r i n g i n f o r m a t i o nModel 7090N Nitrogen Detector designed for sensitive, selective, equimolar quantitation of nitrogen compounds. Accessoriesinclude ceramic tubes, operation/service manual, �ttings and gas line tubing required for system operation. Requires vacuum pu mpand mounting kit.

Model 7090S Sulfur Detector designed for sensitive, selective, equimolar quantitation of sulfur compounds. Features large 14"furnace, facilitating complete combustion. Accessories include ceramic tubes, operation/service manual, �ttings and gas line t ubingrequired for system operation. Complies with GPA 2199-99, ASTM D 5504 and D 5623. Requires vacuum pump and mounting kit.

Model 7090NS Nitrogen/Sulfur Detector designed for simultaneous quantitation of nitrogen and sulfur compounds. Featureslarge 14" furnace, which facilitates complete combustion and enhances sensitivity and equimolarity. Accessories include ceramictubes, operation/service manual, �ttings and gas line tubing required for system operation. Complies with ASTM D 5504 andD 5623. Requires vacuum pump and mounting kit.

Nitrogen Detector Kit for addition to Model 7090S Sulfur Detector. P/N 92154

Sulfur Detector Kit for addition to Model 7090N Nitrogen Detector. P/N 92153

Vacuum Pump low maintenance, high capacity vacuum pump. Oil-less for safe oxygen operation. Includes vacuum kit. P/N 71304

Mounting Kit includes all parts necessary for connection to speci�c GC. Please provide make and model of GC.

Software is purchased separately.

7090 SERIES

Please contact your PAC representative for more information:

Continuing research and development may result in speci�cation or appearance changes at any time.