Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
1
CCE of heavily irradiated Si diodes operated with increased free carrier concentration
and under forward bias
Ljubljana group:
I. Mandić, M. Batič, V. Cindro, I. Dolenc, G. Kramberger, M. Mikuž, M. Zavrtanik
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
2
Motivation:
• LHC upgrade relevant fluences: 1016 n/cm2
• standard Si detectors operated in standard waywill be very difficult if not impossible to use
check the alternative way of operation with heavily irradiatedstandard detectors:
forward biasA. Chilingarov et al., NIM A399 (1997) 35.
increased carrier concentration with DC light illuminationV. Eremin et al., NIM A360 (1995) 458G. Lutz, NIM A377 (1996) 234.G. Kramberger et al., NIMA A497 (2003) 440.
cryogenic temperaturesV. Palmieri et al., NIM A413 (1998) 475.K. Borer et al., NIM A440 (2000) 5.
measure CCE with 90Sr source
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
3
Samples and Setup:
• 4 diodes: S = 0.5x0.5 cm2, D = 300 µm, ρ = 15 kΩcm
• irradiated with neutrons in TRIGA reactor in Ljubljana.Equivalent fluences: 5x1014, 1x1015, 4x1015 and 8x1015 n/cm2
• diodes were kept for 14 days at room temperature after irradiation.After that they were stored at T = -17°C
• CCE was measured on our TCT setup:• liquid nitrogen cryostat which allows illumination of diodes with laser• 90Sr source with set of collimators• trigger provided by scintilator + PMT (rate ~ 5 events/s)• MITEQ AM-1309 wide band current amplifier• digital oscilloscope • pulses were stored in computer and integrated in 18 ns time window offline• spectra of integrated pulses were fitted with convolution of Landau and Gaussian distribution
• collected charge measured with unirradiated diode used as normalizationto calculate CCE
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
4
Setup:
red lasercontinuousred lasercontinuous
optical cryostatJanis VPF−100
attenuator
beam positioner
red light (DC)
Sr 90 source
lead
block
PM tube
block
LeadScintilator
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
5
Reverse bias, DC illumination of n side with red light (hole injection):
• injection of holes into irradiated detector influences theoccupation probability of deep levels:
it reduces the concentration of negative space charge (trapping of holes)lower full depletion voltage effective space charge sign can be reinverted to positive
More in: G. Kramberger et al., Nucl. Instr. and Meth. A497 (2003) 440.
• Measurements of CCE with DC light illumination were done with high light intensity (in “saturated” mode)
once the light intensity is high enough high field region moves from n-side top-side (the space charge sign is re-inverted).
increasing the light intensity beyond this point doesn’t change the spacecharge much, because the e-h pairs created in the low field region recombine
Current vs. light intensity curve saturates
CCE doesn’t depend much on light intensity once it is high enough
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
6
CCE vs. bias voltage for standard operation, DC light (hole injection) and forward biasT=-30˚C Highest CCE at given voltage measured with forward bias!
Voltage (V)0 50 100 150 200 250 300
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1(a)
2A/cmµNo DC light, j = 4.4
2A/cmµDC light, j = 100
2A/cmµForward bias, j = 680
14Fluence 5x10
Voltage (V)0 50 100 150 200 250 300
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1(b)
2A/cmµNo DC light, j = 6.0
2A/cmµDC light, j = 72
2A/cmµForward bias, j = 272
14Fluence 10x10
Voltage (V)0 50 100 150 200 250 300
CC
E
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
(c)
2A/cmµNo DC light, j = 10
2A/cmµDC light, j = 48
2A/cmµForward bias, j = 80
14Fluence 40x10
Voltage (V)0 50 100 150 200 250 300
CC
E
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
(d)
2A/cmµNo DC light, j = 18
2A/cmµDC light, j = 40
2A/cmµForward bias, j = 72
14Fluence 80x10
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
7
Current measurements at different temperatures and modes of operation
Temperature (K)100 120 140 160 180 200 220 240
Cu
rren
t (A
)
-910
-810
-710
-610
-510
-410(a) Forward Bias
DC holesNormal operation
2 n/cm14Fluence: 5x10
Uforward= 80 V Ureverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
Cu
rren
t (A
)
-910
-810
-710
-610
-510
-410
(b)
2 n/cm14Fluence: 10x10
Uforward= 120 V Ureverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
Cu
rren
t (A
)
-910
-810
-710
-610
-510
-410
(c)
2 n/cm14Fluence: 40x10
Uforward= 250 V Ureverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
Cu
rren
t (A
)
-910
-810
-710
-610
-510
-410
(d)
2 n/cm14Fluence: 80x10
Uforward= 250 V Ureverse = 250 V
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
8
• CCE vs. temperature
Temperature (K)100 120 140 160 180 200 220 240
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
(a)
Forward BiasDC holesStandard operationDC electrons
2n/cm14Fluence: 5x10
Uforward = 80 VUreverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
(b)
2n/cm14Fluence: 10x10
Uforward = 120 VUreverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
(c)
2n/cm14Fluence: 40x10
Uforward = 250 VUreverse = 250 V
Temperature (K)100 120 140 160 180 200 220 240
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
(d)
2n/cm14Fluence: 80x10
Uforward = 250 VUreverse = 250 V
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
9
• CCE under forward bias doesn’t change much with temperature
• Hole injection:transition from beneficial to harmful with falling T at low fluence
• Standard mode or electron injection:improvement with falling T
both transitions occur between 180 and 220 K
These results agree with Space Charge Sign Inversion occurringin this temperature range.Temperature of transition agrees with temperatures ofSCSI and CCE transitions reported in the RD39 paper: E. Verbitskaya et al., NIM A514 (2003) 47-61.
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
10
Attempt of CCE calculation
• two dominant deep traps:
300x10-151x10-150.070.52donor
2x10-1515x10-1510.57acceptor
σh (cm2)σe (cm2)gt (cm-1)Et - Ev (eV)
• parameters of the traps chosen from parameterintervals of dominant electron and hole trap from the paper:G. Kramberger et al., “Determination of the effective electronand hole trap in neutron-irradiated silicon detector” NIM A516 (2004) 109
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
11
• effective space charge and trapping times were calculated using equations for defects in general stationary state:
)exp(,1/
1
TkEE
ncncncpc
Pb
itt
tipn
tinpt
−=
+
+
+=
−
χχχ
• contribution of the trap to the space charge is given by:
)(0 teqtt PkgeQ −Φ= k = 0,1 (donor, acceptor)
• effective trapping times for electrons and holes are obtained from:
eqhthhthhh
eqeteethee
gPv
gPv
Φ=
Φ−=
στ
στ
/1)1(/1
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
12
• Free carrier concentration was estimated from current measurements.In the case of standard and forward bias operation it was assumed (simplification!) that electrons and holes contribute equally to the current. Therefore concentration of electrons and holes was estimatedas:
evSIn⋅⋅
=2 hvS
Ip⋅⋅
=2
• In the case of DC light illumination the additional carrier densitywas estimated from measurement of current increase, taking intoaccount the surface of the illuminated area and the side of illumination.
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
13
Forward bias:
• electric field in forward bias approximated with E = Vbias/D in the whole detectorvolume
• at high fluences forward bias I-V dependence is almost ohmic and one can calculate resistance of the material as R = Vbias/I
)-1 Aµ1/I (0.05 0.1 0.15 0.2 0.25 0.3
U/I
(MO
hm
)
2
4
6
8
10
12
14
2 n/cm158x10
2 n/cm154x10
2 n/cm151x10
2 n/cm145x10
Forward Bias T = -30 C
Resistance vs. 1/I in forward bias
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
14
Resistance at high fluence very high: 100 MΩcm at 8x1014 n/cm2
also in reverse bias there should be significant voltage dropand therefore electric field in the undepleted region because of current flow: E = ρ ·j, (ρ = resistivity, j = current density)
“undepleted” bulk should contribute to CCE
• in reverse bias: Ubias = Ujunction + I·R, (R – resistance, I – current)
• in the junction w=√(2εε0Ujunction/e0Neff) el. field is Ej ~ Neff·x
• total field E = Ej + j ·ρ inside the junction and E = j ·ρ elsewhere
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
15
CCE vs bias voltage at T = -30CFull markers: measurements, empty markers: calculation
Voltage0 50 100 150 200 250 300
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
, T = -30 C2 n/cm15 = 1x10Φ
Voltage0 50 100 150 200 250 300
CC
E
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
, T = -30 C2 n/cm15 = 4x10Φ
Voltage0 50 100 150 200 250 300
CC
E
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
0.45
0.5
, T = -30 C2 n/cm15 = 8x10ΦVoltage
0 50 100 150 200 250 300
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
DC holesForward bias
Normal operation
, T = -30 C2 n/cm14 = 5x10Φ
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
16
Conclusions:
• DC hole injection and forward bias operation greatly improve CCEat moderate voltages (up to 250 V) already at moderate cooling (-30°C)
• at given voltage, highest CCE achieved by operating under forward bias
• if operated at low temperature (below ~ 200 K) the problem ofhigh current in forward bias is much smaller while CCE stays constant
• increase of CCE seems to be caused by increase of sensitive depth of the detector and to lesser extent by change of trapping times
• trapping is the main limitation for using standard Si detectors at high fluences Φeq > 1015 n/cm2
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
17
Examples of signals:
• averages of 5000 events:
Time (ns)0 2 4 6 8 10 12 14 16 18
Ind
uce
d C
urr
ent
(arb
)
-0.03
-0.02
-0.01
0
0.01
0.02
0.03Reverse Bias 250 V, DC hole injection
Forward Bias 120 V
Co, T = -302 n/cm15 10⋅ = 1 Φ
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
18
• spectra of integrated pulses were fitted with convolutionof Landau and Gaussian distribution. The most probable value of the Landau distribution returned by the fit was taken as the measure of collectedcharge.
• collected charge measured with unirradiated diode was used as normalization to calculate CCE
Examples of fitted spectra:
Charge (arb)-0.1 -0.05 0 0.05 0.1 0.15 0.2 0.25 0.3
Nb
. of
entr
ies
0
50
100
150
200
250
300
350 = 0Φ
2 n/cm15 10⋅ = 1 Φ
, DC holes2 n/cm15 10⋅ = 1 Φ
CoBias = 250 V, T = -30
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
19
CCE doesn’t depend much on light intensity once it is high enough
Light Power (arb.)0 50 100 150 200 250 300 350 400 450 500
A)
µD
etec
tor
curr
ent
incr
ease
(
0
2
4
6
8
10
12
14Not Irradiated
2 n/cm14 = 5x10Φ2 n/cm15 = 1x10Φ2 n/cm15 = 4x10Φ2 n/cm15 = 8x10Φ
CoVoltage = 250 V, T = -30
Measurements of CCE with DC light illumination were done with high light intensity (in “saturated” regime)
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
20
• calculate CCE vs. fluence at -30°Cbest agreement for DC light injection and forward bias if in bothconstant field E = V/D assumed in the whole detector at all fluencesso that CCE is a function of trapping only
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
Measurements (markers):
Calculations (lines):
DC hole injection
Full depletion assumed
(a)
Bias = 250 V T = -30 C
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
CC
E0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
Measured
Calculated
(b)
Forward bias = 80 V, T = -30 C
Reverse bias +DC light Forward bias
Measured and calculated CCE
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
21
Ohmic voltage drop (I·R) and voltage drop in the junction (Vj = Vbias - I·R)
DC hole injection: Standard operation:
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
Vo
ltae
Dro
p (
V)
0
50
100
150
200
250DC hole injection
Voltage drop in the pn junction
Ohmic voltage drop
(a)
DC hole injection, Bias = 250 V T = -30 C
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
Vo
ltae
Dro
p (
V)
0
50
100
150
200
250
Standard operation
Voltage drop in the pn junction
Ohmic voltage drop
(b)
Standard operation, Bias = 250 V T = -30 C
I. Mandic, RD50 Workshop, CERN, May 5-7, 2004
22
Different calculations of CCE compared to measured CCE:
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
CC
E
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
DC hole injection
Measurements
Calculations (lines + empty markers): Full depletion assumed
Ohmic Voltage dropStandard
(b)
DC hole injection, Bias = 250 V T = -30 C
)2 n/cm14Fluence (100 10 20 30 40 50 60 70 80 90 100
CC
E
0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
0.4
Standard operation
Measurements
Calculations (lines + empty markers):
Ohmic Voltage drop
Standard
(c)
Standard operation, Bias = 250 V T = -30 C
Disagreement smaller if ohmic voltage drop is taken into account.