Upload
ngohuong
View
213
Download
0
Embed Size (px)
Citation preview
PRESENTATIONS
Training Workshop on Laboratory Methods and Procedures
Caribbean Coastal Pollution Project (CCPP)
Assessment, Monitoring and Management of Persistent Organic Pollutants (POP) and Persistent Toxic Substances (PTS) in the Coastal Ecosystems of the Wider Caribbean Region
19‐20 January 2009 Reef Yucatan Hotel, Merida, Mexico
Day 1: Monday, 19 Jan Ken Drouillard, GLIER, University of Windsor, Canada
Quality Systems Documentation Overview of Analytical Methods for Stockholm Convention Persistent Organic Pollutants (POPs)
Chris Metcalfe, Trent University and Watershed Sciences Centre, Canada Chromatographic Techniques, Limits of Detection and Calibration Limits of Detection
Nargis Ismail, GLIER, University of Windsor, Canada How to Validate Your Method
Ken Drouillard, GLIER, University of Windsor, Canada Quality Control Charts and Evaluation of Data Integrity
Gerardo Gold Bouchot, CINVESTAV, Mexico & Nargis Ismail, GLIER, University of Windsor, Canada Micro‐extraction Technique
Ken Drouillard, GLIER, University of Windsor & Chris Metcalfe, Trent University and Watershed Sciences Centre, Canada
Clean‐up of POPs from Biological Samples
Day 2: Tuesday, 20 Jan Nargis Ismail, GLIER, University of Windsor, Canada
GC‐Methods and instrument maintenance
Ken Drouillard, GLIER, University of Windsor, Canada Data Reporting & Data Management
TABLE: Determination of PCBs and Pesticides in Tuna Homogenate (Iaea‐435) ‐ Dec 2008
TABLE: List of CBs, OCs and PCB’s, providing Instrument Detection Limit (IDL), Calibration (Working) Linear Range and Coefficient of Determination (R2).
Quality Systems DocumentationDocumentation
•What is its function and why is it needed?
•Elements of a Quality System Manual and Accessory Documents
•The GLIER QA Manual Example
(Provided as supplementary materials)
Quality Systems Documentation• Combined these Documents Demonstrate:
– how to produce valid resultshow to produce valid results– show others that the lab is capable of doing so– act as the ‘corporate memory’ for all details and procedures related to the laboratory
– used as a training manual to new staff– provide baseline on which to seek and pdemonstrate improvements in policies through time
– necessary component towards seeking laboratory accreditation
• Laboratory Competence– Personnel in a laboratory have specific knowledge and skills related to the science underlying their testing procedures
– Staff can demonstrated this knowledge
– Procedures conform to the requirements of the science
• A competent laboratory has:– People with the skills and knowledgePeople with the skills and knowledge
– Environment with the facilities and equipment
– Quality control and procedures to produce valid results
• Quality – Degree to which a set of procedures fulfills objective measurement requirementsmeasurement requirements
– For analytical chemistry • ability of methods to produce precise and accurate results
• ability to demonstrate that the method was implemented under controlled conditions during a set of the analyses for unknown samplesunknown samples
The Quality System Documentation Should Adhere To the Following Principles:
• Capacity – Resources (Personnel with skills), facilities & equipment,Resources (Personnel with skills), facilities & equipment, quality control, procedures to produce quality results
• Exercise of Responsibility– Persons in organization have authority to make decisions and perform functions within the scope of the work
S i ifi h d• Scientific Method – accepted scientific approaches utilized
Principles (Cont.): • Objectivity of Results
– Test results are mainly based on measurable quantities with repeatable measurement characteristicsrepeatable measurement characteristics
• Impartiality of Conduct– Pursuit of competent results through scientific approaches are the overriding influence on the work of persons performing the tests
T bilit f M t• Traceability of Measurement– Results are based on a recognized system of measurements
– Instrumentation used in such measurements are calibrated and functioning correctly
Principles (Cont.):
• Repeatability of Test– A test that produces objective results will produce the same results during subsequent re‐testing
• Transparency of the Process– Processes used to produce objective results are open to internal and external scrutiny
– Improvement of documentation, methods, equipment can be identified and strived for
Quality Systems Documents• Quality System Manual (QSM)
– First level document that describes the policy implementation of the Lab Quality Management SystemSystem
• Management Requirements• Technical Requirements• Revision History
• Quality System Procedures (QSP)– Second level document detailing instructions & procedures in QSMprocedures in QSM
• Standard Operating Procedures (SOPs)– Laboratory testing procedures
• Related Procedures– Additional procedures common across different SOPs
1) Quality System Manual (QSM)A) Management Requirements– Organization and Management
• Legal identification of laboratory and organization• Legal identification of laboratory and organization• Scope of laboratory activities governed by QSM• Authorization of QSM by Laboratory Heads
– Quality Policy• Policy statements:
– Top management commitment to implementing a quality management system and continual improvement
– Roles and responsibilities of laboratory personnel» Technical management and quality management
• Document and Data control– Statement indicating that all QSM documents are current and have been reviewed by top management and technical management
– Identification and location of supporting procedures
1) Quality System Manual (QSM)A) Management Requirements– Control of Non‐conforming testing work
• Statement of policy on what happens when any non‐conformance is found.
– Corrective and Preventative Action System– Statement of policy to implement continual improvement in
laboratory procedures– Control of Quality Records
– All records, technical and non‐technical generated within the scope of the laboratory operations
– Records provide information sufficient for traceability:– all actions from sample receipt to report– responsibility for all involved in the process
– Audits ‐ internal and external of QSM– Management Review
‐statement of frequency of management review of quality systems
1) Quality System Manual (QSM)B) Technical Requirements
– Personnel– Controlled Environment and Facilities– Method Validation and Calibration– Equipment– Measurement Traceability– Handling of Test Items– Quality Assurance of Test Results– Results Reporting, Data Storage and Security
C) QSM Revision History
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Technical Requirements
l
‐Each sub‐section reflects a stand‐alone document subject to revision‐Describes Purpose, Scope and Authorities and Responsibilities for tasks
– Personnel
– Controlled Environment and Facilities
– Method Validation and Calibration
– Equipmentq p
– Measurement Traceability
– Handling of Test Items
– Quality Assurance of Test Results
– Results Reporting
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
Personnel• Authorities and Responsibilities:
– Laboratory Head– Laboratory Head• Responsible for Management of Lab Operations
– Quality Manager• Ensures the Implementation of QSM and QSP• Reviews documentation, data quality, internal audits, proficiency testing, implementation of continual improvement policies
– Laboratory Supervisor• laboratory productivity, competency of technical personnel/training, safety, laboratory productivity, competency of technical personnel/training, safety,method validation, equipment use and maintenance, data reporting, data security and storage
– Laboratory Technician• Implementation of Standard Operating Procedures• Instrument calibration and log book recording, sample processing, instrument analysis, implementation of related procedures
• Personnel (Cont.)
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
• Technician Competence, Qualification & Training– Describe education and experience requirements for all personnel descriptions
– In‐house training program for technicians
– Analyst Proficiency testing• Analyst must demonstrate proficiency prior to implementing assigned tasks by satisfactorily analyzing quality control sample
• Analyst testing results are documented
• Environmental ConditionsF iliti d t t t t ti
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Facilities are adequate to carry out testing– Descriptions of
• Laboratory space – Criteria for lighting, temperature, humidity, air quality
• Safety considerations – Use/availability of fumehoods, venting, solvent cabinets,
h i l t h / h t ti fi tchemical storage, emergency showers/eyewash stations, first aid kits, emergency exits, other institutional safety procedures
• Security Considerations– Documentation storage areas, sample storage areas, preparation laboratory, instrument access
• Test Method and Method Calibration
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Method Selection• Statement of persons responsible for selecting methods adopted within SOP. That such methods are based on published international, national and/or regional standards or deemed fit for testing criteria
– Analytical MethodsAnalytical Methods• Index of all SOPs and Related Procedures Implemented by Lab
• Test Method and Method CalibrationV lid ti
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Validation • SOPs & related procedures are validated for control samples
• Documentation of performance criteria for accuracy, recovery, precision, detection limits, calibration and linearity
• Frequency of method validation (re validation• Frequency of method validation (re‐validation exercises) for control samples
• Review of method performance– Periodic examination of internal quality control, method calibration, method quality control data, performance history of validation data
• EquipmentEquipment inventory
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Equipment inventory– Records and Procedures
• Logs of: instrument, date of commission, history of modifications/repairs, calibration history, performance history
– Out of Service/Return to Service– Calibration Status
• SOPs provide specific details and include:• Calibration blank• Adequate number of standards used to define calibration, curve fitting procedures and statistical measurements of curve fit
• Low standard detection limit• Recovery standards, Certified Reference Tissues, Sample Duplicates
• Measurement and Traceability– Origins and certificates of analytical standards, certified
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
reference materials– Instrument calibration
• Periodic scheduled calibration against known standards (Cross check standards)
• Calibration conducted routinely prior to each use (i.e. Working standards)
– Frequency of calibration for:• Balances, volumetric glassware, temperature of storage ( l / d d) h h(sample/standard) storage areas, gas chromatographs
– Quality Control Samples:• Reference control standard (calibration accuracy); reference material (method accuracy); sample duplicate (method precision); standard recovery (method recovery); method blank (biological contamination)
– Level of Control Effort:• Frequency of analysis of QC samples relative to unknown samples
• Handling of test items– Sample reception procedures and forms
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Sample storage and Disposal– Chain of Custody
• Quality Assurance– QA/QC Sample Preparation & Effort of Control Sample Analyses– Evaluation of Interferences– Statistical Control
• Control Charting and evaluation of control chart dataControl Charting and evaluation of control chart data• Non‐conformances and trending monitored for analytical standards, reference materials, blanks, and standard recoveries
– Proficiency testing• Record of participation in interlaboratory comparisons and testing results
• Handling of test items– Sample reception procedures and forms
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– Sample storage and Disposal– Chain of Custody
• Quality Assurance– QA/QC Sample Preparation & Effort of Control Sample Analyses– Evaluation of Interferences– Statistical Control
• Control Charting and evaluation of control chart dataControl Charting and evaluation of control chart data• Non‐conformances and trending monitored for analytical standards, reference materials, blanks, and standard recoveries
– Proficiency testing• Record of participation in interlaboratory comparisons and testing results
• Test Reports
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
Results Reporting
• Test Reports– QA signoff
– Flagged results
– ND, <DL
– Sig. digits
Selected Procedures
Selected Procedures
Selected Procedures
Selected Procedures
Selected Procedures
Selected Procedures
More Information:
• Canadian Association for Laboratory A dit ti (f l CAEAL)Accreditation (formerly CAEAL)– http://www.cala.ca/T27‐17025_Handbook.pdf
• GLIER – Quality System Manual Example– Electronic files supplied
Overview of Analytical Methods for Stockholm Convention PersistentStockholm Convention Persistent
Organic Pollutants (POPs)
Ken DrouillardGLIER, University of Windsor
Main Source for Presentation Materials:
Muir, D.C.G., E.D. Sverko. 2006. Analytical u , .C.G., . . S e o. 006. a yt camethods for PCBs and organochlorine pesticides in enviroinmental monitoring ans surveillance: a critical appraisal. Anal. Bioanal. Chem. 386: 769‐789
File available in Electronic PDF format.
Stockholm Convention “POPs”(The Dirty Dozen)
• Organochlorine Pesticides– Aldrin– ChlordaneChlordane– DDT– Dieldrin– Endrin– Heptachlor– Mirex– Toxaphene
• Industrial Chemicals• Industrial Chemicals– Polychlorinated Biphenyls (PCBs)– Hexachlorobenzene
• Industrial By‐Products– Dioxins– Furans
Challenges and Issues• Environmental analytical chemistry methods developed over past 30‐40 years
• Large numbers of methodsLarge numbers of methods – different selectivity and sensitivities
• GC‐MSD and GC‐ECD common
– different environmental matrices• All POPs relatively hydrophobic, most likely detected in organic phases: soils sediments tissue samples relative to air or waterphases: soils, sediments, tissue samples relative to air or water
• Instrument availability & limitations– E.g. High Resolution GC‐MSD for dioxins & furans, toxaphene
• High costs of ‘best practice’ techniques– Availability of analytical standards
Challenges and Issues
– Use of 13C‐labelled recovery standards– Certified reference materials– Costs of implementing QA/QC programs (personnel, time and consumables)
M t t d l b t dit ti• Movement towards laboratory accreditation– Standardized SOPs– Participation in Interlaboratory comparison exercises
Analytes of Concern• PCBs
– Total PCBs, sum PCBs (which congeners?)• IADN (Integrated Atmospheric Deposition Network – GreatIADN (Integrated Atmospheric Deposition Network Great Lakes) includes 100 congeners
• UNEP/World Bank/GEF Projects: 15‐30 congeners• Environment Canada – 42 or 71 congeners• Key Issue – guidelines based on sum or total PCBs, while analytical methods vary in # congeners analyzed
– Dioxin‐like PCBs (NO‐PCBs and MO‐PCBs)• NO‐PCBs: IUPAC #’s: 81, 77, 126, 169• MO‐PCBs: IUPAC #’s: 105, 114, 118, 123, 156, 157, 167, 189• Key issue – dioxin‐like activity often mostly related to NO‐and MO‐PCBs. Analytical methods require specialized separation procedures for testing (Particularly NO‐PCBs)
Analytes of Concern• UNEP Global POPs Monitoring workshop categorized essential PCBs to be monitored in human tissues and food products:human tissues and food products:– IUPAC #: 28/31, 52, 101/90, 118, 138, 153, 180– Most robust interlab comparisons
• UNEP Recommended PCBs for Sum PCB estimation in environmental matrices
IUPAC # 8/5 18 28/31 44 49 52 95/66 87 99– IUPAC #: 8/5, 18, 28/31, 44, 49, 52, 95/66, 87, 99, 101, 105/132, 110, 118, 128, 146, 149, 151, 153, 138/163, 156, 183, 187, 201/157, 170, 180, 194, 195, 206, 209
– Congeners in bold present in Quebec Ministry Analytical Standard
Analytes of Concern• OC Pesticides
– Chlordane• Essential: cis‐ and trans‐chlordane, cis‐ and trans‐nonachlor, oxychlordane• Recommended: MC4, MC5, MC6
– Heptachlor• Essential: Heptachlor, heptachlorepoxide
– DDT• Essential: 4,4’‐DDE, 4,4’‐DDD, 4,4’‐DDT• Recommended: 2,4’‐DDE, 2,4’‐DDD, 2,4’‐DDT
– Mirex• Essential: Mirex• Recommended: photomirexi ld i– Dieldrin
– Endrin– Aldrin– Toxaphene – few analytical standards available: Parlar (P) 26, 50, 62 most
common congeners analyzed for– Dioxins and Furans – several congeners – requires HR‐GC/HR‐MSD
No Consensus SOPs in Place for PCB/OCs
• UNEP POPs workshop on Global monitoring:– ‘Given broad range of technical expertise for analysis of PCBs and OCPs, as evident from the extensive international participation in interlaboratory calibration projects for these compounds, no single detailed, step‐by‐step analytical method can be recommended Insteadanalytical method can be recommended. Instead laboratories would use methods best‐suited to their situation and take part in international interlaboraotry comparisons to verify their work’
Internet Accessible PCB/OC SOPs• U.S. EPA & U.S. Geological Survey
– www.nemi.gov• Japan Environment AgencyJapan Environment Agency
– www.env.go.jp/en/index.html• Oslo‐Paris Commission
– www.ospar.org• Helsinki Commission
– www.helcom.fi• International Organization for Standardization
– www.iso.org• Association of Official Analytical Chemists International
– www.aoac.org• Several Others
Sample storage and handling• Norstrom and Won (1985); Wells et al (1997)
• Losses of more volatile OC‐pesticides (tetrachlorobenzenes and HCB) and reaction of DDT during freeze‐drying
• Recommend avoiding freeze drying• Recommend avoiding freeze drying
• Kiriluk et al (1996)– No effect of PCBs in fish when stored at ‐20 or ‐80oC
• Boer and Smedes (1997)– Reductions y1pin extractable lipids of samples stored at ‐20oC after 2 years, affects expression of lipid normalized residuesnormalized residues
– Recommended sample storage at ‐70oC when samples will be stored for > 2 year time period
•Short term storage of biological samples (< 2 years) for PCBs/OCs should be frozen at ‐20oC
Sample storage and handling• Storage containers:
– Preferred: Soap‐washed, Solvent (acetone, followed by hexane)‐rinsed glass jars fitted withfollowed by hexane) rinsed glass jars fitted with Teflon lined caps or hexane‐rinsed foil liner under lid
– Wrap excised sample in hexane‐rinsed foil and place inside a ziplock bag or whirlpack bag
Clear polyethylene bags or polypropylene jars for– Clear polyethylene bags or polypropylene jars for temporary storage i.e. sample transfer from field to laboratory
• Sample dissection – use of solvent‐cleaned stainless steel tools
Sample Preparation
• Sample homogenization:– Large Samples:
• Care in cleaning between samples
– Small samples & homogenate subsamples:• Mortar and pestle with Na2SO42 4
• Na2SO4 – ACS‐grade, muffled at 450oC overnight and maintained at 110oC until use
• Other desiccants: Celite; Hydromatrix must be solvent extracted prior to use
• Samples are with a non‐interfering chemical with similar elution and chromatographic properties of analytes. Unversally adopted for PCBs/OCs
• GC‐MSD
Sample Recovery Spikes
GC MSD– Isotopically labelled PCBs or OC‐pesticides– Identical physical properties to native analytes, similar
adsorption/partitioning properties– Requires MSD to distinguish native and labelled analyte
• GC‐ECD– Environmentally rare chemicals that respond well to ECD and
have similar physical properties of PCB/Ocshave similar physical properties of PCB/Ocs– PCBs #30, 204– Pentachloronitrobenzene– endrin ketone– 1,3,5‐tribromobenzene
• When to spike?
Analyte Extraction• Considerations
– Degree of adsorption/partitioning to sample matrix components (e.g. soils, blood proteins, tissue lipids)Extraction solvents (Pesticide analysis grade)– Extraction solvents (Pesticide analysis grade)
• Polar:apolar binary mixture more efficient for recovery of OCs from tissue samples of low lipid content and from sediments
– Acetone/hexane– Hexane/dichloromethane– Acetone/dichloromethane
– Solvent/matrix contact time and processing timeAmount of solvents utilized per sample– Amount of solvents utilized per sample
• Tissues: cold column, sonication, soxhlet, microwave‐assisted, supercritical fluid, pressurized liquid extraction
• Sediments: soxhlet, microwave‐assisted, pressurized liquid
Extraction
Extraction – blood and milk• Blood
– plasma preferred, then serum, then whole blood– Liquid‐liquid (shake flask), vortex unit, sonication + centrifugation &
dryingdrying• ethanol/hexane• methanol/hexane• Alcohol shown to be important in disrupting lipoproteins increasing extraction
efficiency– Solid phase extraction (C18‐cartridges) and elution with hexane,
hexane/DCM
• MilkLi id Li id i ( h k fl k)– Liquid‐Liquid extraction (shake flask) common
• Acetone/hexane• Dichloromethane/hexane
– Waring blender with acetonitrile and potassium oxalate + extraction by SPE (C18 cartridge)
Tissue Lipid Content• >95% of PCB/OCs in biological samples contained within neutral lipids (triglycerides) of samples with > 1% total lipid content
i f li id li d O id f l• Expression of lipid normalized POP residues useful to understand environmental fate and food web biomagnifications, not relevant for fish advice information
• Aliquot of sample extracts (usually 10%) removed following extraction to determine lipids by weighing
• Total lipids: chloroform:methanol:water (Bligh and Dyer 1959)– chloroform:methanol:water (Bligh and Dyer 1959)
– Propanol:cyclohexane:water (Smedes 1999)• Neutral lipids:
• Hexane:DCM or Extraction solvent combination
• Blood plasma lipids (neutural lipids):– Colorimetric approaches (phospho‐vanillin reaction)
Lipid Content
Li id t t i diff t ti f k ll
•Total lipids sometimes determined for other purposes, i.e. to establish health and energy density of animal•For tissues with low lipid content (<1%); total lipids ≠ neutral lipids!
Lipid content in different tissues of muskellunge determined by chloroform/methanol extraction and by hexane/DCM extraction
Drouillard et al. 2004. Chemosphere 55:395
Clean‐up• Clean‐up involves separating analytes from coextractedmatrix components that can interfere with analysis– Lipids, pigments, hydrophobic proteins, sulfur, other contaminants
– Separation of co‐eluting analytes into separate fractions for examination by GC/ECD
• Common clean‐up methods– Gel permeation chromatography (GPC) – size exclusion– Silica gel (removal lipids and separation)– Alumina (separation)– Florisil (separation)– Activated copper (sulfur removal)
Method 1668
Method 1668
EPA1668
Env. Can.
Instrumental Analysis
• GC‐ECD, GC‐MSD most common for PCB/OCs• GC Injection Port
– Splitless most common, on‐column injection (less common)– pressure programming can improve chromatography
• GC Columns– Open tubular capillary columns– DB‐5 columns, 30 m or 60 m (0.25 mm i.d. x 0.1 µm film thickness) commonly used) y
• 30 m column shows co‐elution of several PCBs; 60 m can resolve greater numbers of PCBs
– DB‐XLB (J&W Scientific); HT‐8 (SGE Inc.) considered advanced in separation of PCB congeners but less commonly adopted
Co‐eluting PCB congeners on 30m DB‐column As determined by GC‐ECD
60 m column:28, 31 resolved33,20 resolved149, 118 resolved153 resolved from 132/105
71/ 41/ 64 – co‐elute101/90 – co‐elute151/82 – co‐elute77/110 – co‐elute (fractionated on florisil)101/132 – co‐elute135/144 – co‐elute156‐171 – co‐elute195/208 – co‐elute
‐Co‐eluting congeners, particularly for mixedCo eluting congeners, particularly for mixed homologue groups, can lead to error in quantitation when the proportion of congeners in the standard differs from the proportion of congeners in the sample!
‐Occasional confirmation by GC‐MSD is useful
Quebec Ministry of Environment StandardGC‐ECD with 60 m DB‐5 column
Co‐elution of OCs with PCB congeners(60 m DB‐5 column, GC‐ECD)
Without Fractionation• Heptachlor epoxide / oxychlordane/ PCB 70• p,p’‐DDE/dieldrin/PCB 87
– p,p’‐DDE/dieldrin readily resolved on new 60 m column• o’p‐DDE/PCB110• Endosulfan II/PCB118• p,p’‐DDE with cis‐nonachlor (Resolved on our system)• Trans‐nonachlor with PCB 99 (Resolved on our system)• PCB 180 and PBDE 47
• **Some issues with supplied standards: mixed heptachlor epoxide+ oxychlordane and p,p’‐DDE/dieldrin could result in quantitationerrors for these compounds unless very good chromatography is realized
GC‐ECD of supplied OC‐StandardMixture (60 m DB‐5 column)
Fractionation Alumina or Florisil
• Can further separate PCBs/OC pesticides for cleaner chromatography removing some (not all) co‐elution issues highlighted previously
• Additional processing time, i.e. solvent reduction and instrument analysis of multiple fractions per sample
• Florisil Cleanup (3 fractions)– Fraction 1:
• PCBs – all congeners (except NO‐PCBs), HCB (F1,F2), Heptachlor (F1, F2), Aldrin (F1, F2), Endosulfan II (f1, f2), DDE (F1, F2 – both isomers), DDD (F1, F2) ; mirex (F1)
– Fraction 2:• HCHs (100 F2), HCB (F1,F2), Heptachlor (F1,F2), heptachlor epoxide (Isomer a) Aldrin (F1,
F2), oxychlordane (F2, F3), trans‐chlordane (F2); Endosulfan I (F1+F2), cis‐chlordane (F2); DDE (F1, F2), DDD (F1, F2); DDT (F2)
– Fraction 3:• Endrin, heptachlor epoxide (F2, F3) heptachlor epoxide (isomer B), oxychlordante (F2, F3),
Dieldrin (F3), methoxychlor (F3)
Chromatographic Resolution• Good instrument maintenance
– Diligence in good sample clean‐up • removal of lipids, H2O, sulfur and co‐contaminants
– Maintain clean injection port, gases (change of gas filters), frequent change of inlets, septa
– Pressure programming, good gas pressure and no leaks– Frequent checks of column resolution
• E.g. peak separation 28/31E.g. peak separation 28/31• good peak shape and consistent response factor for p,p’DDE
– Column duty cycle – may need to change column 2 – 3 x per year
Instruments ‐ Detectors
• GC‐ECD GC‐MSD (LR, EI ‐ SIM)‐high sensitivity (detection 0.05 pg/g) ‐moderate sensitivity (DL 0.3 pg/g)
‐moderate selectivity ‐high selectivity few co‐eluting peaks‐useful for peak identification
‐simple maintenance and robust ‐can be finicky, frequent autotuningvery consistent baseline & response required, response can be variable
ease of training and software use moderate training required‐ease of training and software use ‐moderate training required
‐low cost ($30 – 45 k) ‐moderate cost ($90 – 130 k)
‐limited use for other analytes ‐useful for OPs and other pesticides
Quality Assurance
• Maturity of PCB/OC methods requires adopting sound QA measures with results reportingU f tifi d l ti l t d d• Use of certified analytical standards
• Surrogate spike recoveries• Laboratory blanks
– clean vegetable oil or suitable matrix• Certified reference materials (or in‐house reference homogenates)homogenates)
• Use of quality control charts• Participation in inter‐laboratory comparisons
Accuracy of CRM Analyses
• Mean of replicates (n=7) concentration measured in CRM not
significantly different then certified value (t‐test)g y ( )
• QUASIMEME
– Inter‐laboratory variation 15‐20% for PCB concentrations are common
• European Commission
t bl PRC/OC ( l ) i CRM f ll– acceptable PRC/OCs (mean values) in CRM as follows– Conc. Range < 1ng/g ‐50% to + 20% certified value– 1 – 10 ng/g ‐30% to + 10% certified value– > 10 ng/g ‐20% to + 10% certified value
POPs Training WorkshopJanuary, 2009
Chromatographic Techniques,Limits of Detection
and Calibration
Presentations by C. Metcalfe
Introduction toChromatographic Techniques
Chromatography
• Separation of individual substance from mixtures • Extension of “partitioning theory”
• Applications in environmental analytical chemistry:
• Preparatory - Preparation of samples for analysis
• Analytical - Identification and quantitation of individual “analytes”
Partitioning Theory
• Partitioning between two phases reaches equilibrium• Partition coefficient (K) = conc in phase 1/conc in phase 2
solidgas
liquidliquid
liquidgas
solidliquid
Phase 2Phase 1
No analytical applications
Partitioning• Multiple partitioning of multiple solutes (1, 2, 3, 4, etc.)
between two immiscible phases• Separation of solutes occurs because of differences in
partition coefficients:• a) Solute 1: K1 = 2• b) Solute 2: K2 = 1• c) Solute 3: K3 = 0.5• etc.
Elution Chromatography
ColumnPacking = Stationary Phase(solid)
Solvent=Mobile Phase(liquid)
Eluate
SoluteA&B
Solute B
Solute A
Chromatogram• Detector signal vs. time
0
50
100
150
200
250
300
0 5 10 15 20 25 30time
concentration
AB
Characteristics of a Chromatogram• retention time: identifies substance• peak area: quantifies substance• resolution: separation of peaks• peak shape (width, sharpness): ideal peak is narrow and sharp
Retention time [min]5 64
4.3
5.2
Solute A
Solute B
Classification of Chromatographic Methods• Nature of the mobile phase
• Gas chromatography• Liquid chromatography
• Physical configurations• Planar - Paper chromatography, thin-layer
chromatography• Column – Packed column, open tubular column
• Sample development – Elution chromatography (only type used in analytical applications)
• Application• Preparative – prepare samples for analysis• Analytical – identification and quantification
• Retention mechanism• Adsorption• Partition• Ion exchange• Size exclusion
Stationary phase = solid
Adsorption ChromatographyMainly used in preparative
Applications(e.g. Silica gel, Florisil)
Stationary phase = liquid
Partition ChromatographyMainly used in analytical Applications (GC, HPLC)
Stationary phase – Cation or anion exchange resin
Ion Exchange ChromatographyPreparative applications (e.g. solid phase extraction),
Analytical applications (ion chromatography)
Mobile anions held near cations covalently bound to stationary phase
Anion Exchange Resin
Stationary phase = Porous Gel
Size Exclusion ChromatographyMainly preparative applications
(e.g. gel permeation chromatography)
Smallmolecules are retained
Analytical ChromatographyGas Chromatography
• Used for analytes that are volatile at high temperatures (200-300oC)
• Stationary phase: liquid adsorbed to a “solid support”• Mobile phase: inert gas (nitrogen, helium)• Main separation factor: temperature• Typical column: open tubular (capillary)• Common analyte selective detectors
– Flame photometric (FPD) – e.g. Organophosphate pesticides
– Electron capture (ECD) – e.g. PCBs and OC insecticides
– Mass selective (MSD) – e.g. range of analytes
Gas Chromatography
Open Tubular (Capillary) Column
Length – 10-60 m
Support coated open tubular column(SCOT)
Wall coatedOpen tubular column(WCOT)
Gas flow
Silica column
Determination of PCB congenersby GC-ECD
Cl Cl
209 congeners
Cl
Analytical ChromatographyHigh Performance (or Ultra High Pressure)
Liquid Chromatography• Used for analytes that are not volatile, or not suitable for GC
analysis (e.g. heat labile, polar, ionic)• Stationary phase: liquid adsorbed to a “solid support”• Mobile phase: liquid solvent• Main separation factor: pressure• Typical column: packed column • Common detectors
– Ultraviolet– Fluorescence - e.g. aromatic hydrocarbons – Mass selective (MSD) – e.g. range of analytes
• LC-MS• LC-MS/MS: High sensitivity and selectivity
Liquid Chromatography
Three different separation modesMobile Phases/Stationary Phases
• Normal phase high performance-liquid chromatography (HPLC)– Mobile phase: non-polar solvent– stationary phase: polar (SiO2, Al2O3)– used for: polar analytes
• Reversed-phase HPLC– mobile phase: polar solvent– stationary phase: non-polar (end-capped SiO2/Al2O3, organic
polymers)– used for: non-polar analytes
• Ion chromatography– mobile phase: water (buffer)– stationary phase: ion exchange resin (cation exchange, anion
exchange)– used for: cations (Na+, K+, Ca++), anions (Cl-, F-)
HPLC-Fluorescence: PAH analysisanthracene
pyrene
PAHs = polycyclic aromatic hydrocarbons(environmental carcinogens)
Reverse phase mode
Determination of Anions by Ion Chromatography with Conductivity
Detector (IC-CD)
Limits of Detection
Detection TheoryIf we analyze many replicates of a sample,
we will generate concentration data that are distributed (statistically) normally, and we can calculate a mean concentration and a standard deviation about the mean value. Variability about the mean is caused by:
• Instrumental variability:– Detector signal “noise”– Injection volume variations– Temperature program variations, etc.
• Method variability:– Recoveries during sample prep– Dilution errors during sample prep– Calibration errors during analysis– Matrix effects during analysis, etc.
Concentration
Mean
-4 4-2 20
+/- 3 SD
99% confidence limits about the meancan be calculated statistically (i.e. students t values), or approximated as 3x the standarddeviation about the mean.
Chromatographic Detection Limits
“Noise”
Analyte “signal”
Detection limit: The smallest signal (i.e. lowest concentration) of a specific analyte that is statistically different from background ”noise”.
All electronic instruments are susceptible to noise as a result of variability inAC current, electromagnetic fields, etc. Noise will vary with the instrument and with chromatographic retention time, and may vary over time and with the operator.
Chromatographic Detection Limits• Instrumental detection limit (IDL): The smallest signal of a particular
analyte that can be distinguished from background noise for a particular instrument.
Typically determined by replicate injections of an analytical standard dissolved in an appropriate carrier solvent into the chromatographic instrument.– IDL should always be below the MDL
• Method detection limit (MDL): The smallest signal of a particular analytethat can be distinguished from background noise for a particular instrument using a particular analytical method.
Typically determined by spiking an analytical standard into an environmental matrix (e.g. water, wastewater, tissues, sediments, soil) and preparing an extract that is injected in an appropriate carrier solvent into the chromatographic instrument:– Limit of detection (LOD): Defined statistically– Limit of quantitation (LOQ): Defined statistically
Note: LOD and MDL are often used synonymously
Chromatographic Detection LimitsSignal to Noise Ratio (S/N): The relative strength of an analytical signal (S) relative to the average strength of the instrument noise (N).
Max
Min
Mean
Convenient method for determining S/N: Divide the arithmetic mean of the signal for a series of replicate analyses by the standard deviation of the replicate results- tells you how much the noise is affecting your results
Chromatographic Detection Limits
• Limit of Detection (LOD): The lowest concentration that can be determined to be statistically different from a blank with a specific level of confidence (i.e. 99%). LODs are typically matrix, method and analyte specific.
Above the LOD, the analyte can be reported as detected< LOD = report analyte as “not detected” (nd); note that we can never
say that the concentration = zero• Limit of Quantitation (LOQ): The concentration above which
quantitative results may be reported with a specific level of confidence (i.e. >99%). LOQs are typically matrix, method and analyte specific.
Above the LOQ, the analyte can be quantified> LOD and < LOQ = report analyte as “present” or “detected”> LOQ = report analyte concentration in the sample matrix (e.g. mg/L)
Determining the LOD Method 1: Replicate Analyses
• The calculation of an LOD is based upon the variability (i.e. precision) observed between replicate analyses (e.g. n = 7) of an identical concentration of the analyte spiked into an appropriate sample matrix.
• Since precision will vary for different concentrations of the analyte, the initial spike level selected for LOD determination is important.
• The spike level should be selected as equivalent to an instrumental S/N in the range of 2.5 to 5.
• At least one method blank (i.e. unspiked) should be analyzed with each set of samples used to determine the LOD.
• In order for the LOD to be valid, the recoveries of the analyte using the analytical method should be “reasonable” (i.e. >75%) and reproducible.
Method 1: Replicate AnalysesPROCEDURE
• Estimate an appropriate spiking level by determining the analyte concentration that corresponds to an instrument S/N of 2.5 to 5.
• Spike the analyte into an appropriate sample matrix1 (e.g. water, tissue, sediment), and prepare at least 7 aliquots for analysis using an appropriate method.
• Analyze the replicate samples and determine the mean and the standard deviation (SD; for n -1 d.f.) for the concentrations determined in the samples.
• Computation method 1a:LOD = SD x 3LOQ = SD x 10
• Computation method 1b (US EPA):LOD = t (n-1, 0.99) x SD; where t = Students’ t value for 99% level confidenceLOQ = LOD x 3.3
1) If possible, choose a sample matrix for spiking that has non-detectable levels of the target analyte.
Method 1: Replicate AnalysesPROCEDURE
Analysis of lindane spiked into fish tissue at 0.21 ng/g wet weight (n=9):
Sample # Conc (ng/g) % Recovery1 0.23 1102 0.21 1003 0.24 1144 0.19 905 0.18 866 0.23 1107 0.22 1058 0.17 819 0.16 76
Mean = 0.20 ng/gSD = 0.029S/N = 6.89
Mean recovery = 97%
Computational method 1a:LOD = 0.029 x 3 = 0.087 LOQ = 0.029 x 10 = 0.29
Computational method 1b:LOD = 2.896 x 0.029 = 0.084LOQ = 0.084 x 3.3 = 0.28
Students’ t values at 99% confidence level:n n-1 t-value7 6 3.1438 7 2.9989 8 2.89610 9 2.82111 10 2.764
Method 1: Replicate AnalysesPROCEDURE
The five point check:• Does the spike level exceed 10x the
LOD? If so, the spike level was too high.• Is the LOD higher than the spike level? If
so, the spike level was too low.• Is the S/N in the appropriate range (i.e.
<10)?• Are the recoveries of the analytes
reasonable? • Does the LOD meet regulatory
requirements (e.g. fish consumption advisories)?
Lindane spiked into fish at 0.21 ng/g:Mean = 0.20 ng/gSD = 0.029S/N = 6.89
Mean recovery = 97%Range = 76-114%
Computational method 1a:LOD = 0.087 LOQ = 0.29
Computational method 1b:LOD = 0.084LOQ = 0.28
Determining the LOD Method 2: Serial Dilutions
• The calculation of an LOD is based upon analysis of serial dilutions of a standard of the target analyte spiked into an appropriate sample matrix.
• Choose an upper limit of the dilution series that is at least 10x (i.e. one order of magnitude) greater than the IDL.
• Continue the dilution series until the target analyte can no longer be detected (i.e. S/N = 1).
• Analyze all points in the dilution series at least 3 times.• At least one replicate of the method blank (i.e. unspiked) should be
analyzed.• In order for the LOD to be valid, the recoveries of analyte using the
analytical method should be “reasonable” (i.e. >75%) and reproducible.
Method 2: Serial DilutionsPROCEDURE
Spiking Concentration
Sign
al MethodBlank
LOD using Method 1:Lowest concentration that Is statistically different from the blank
LOD using Method 2:Concentration at which there is an asymptote (change in slope) in the serial dilution regression
0 10 20
LOQ = 3.3 x LOD
Uses of Serial Dilution Method
• Confirm that the LODs and LOQs determined using the Replicate Analysis method are reasonable.
• Estimate the LODs when it is not possible to find a matrix blank that is not contaminated with significant quantities of the target analyte, such as:– marine mammal blubber– human blood serum– domestic wastewater
The Sample Matrix• “Matrix effects” are defined as signal suppression or enhancement as
a result of co-extractives that are present in the sample .• The composition of some sample matrices are highly variable (e.g.
domestic wastewater, blood serum), and theoretically, new MDLsshould be determined for every new set of samples.
• However, for practical reasons, it is usually assumed that the MDLsdetermined for a representative sample matrix can be applied to all samples of a similar type.
• Gas chromatography tends to be less susceptible to matrix effects because of the high degree of sample clean-up and because only volatile compounds are analyzed.
• Liquid chromatography with some detection systems (e.g. fluorescence, MS) is more susceptible to matrix effects.
• Internal standards using stable isotope surrogates (e.g. deuterated or 13C-labelled analytes) can be used to compensate for these effects.
Analytical Calibration
Calibration Procedures• External standards: Known concentrations of analytes are analyzed
separately from the sample for quantitation of the analytes in the samples. – Reference standards can be purchased for most contaminants under regulatory
control (e.g. POPs)
– Certified standards are not available for “emerging contaminants” and must be made up gravimetrically by researchers
• Internal standards: Known concentrations of surrogate analytes that do not interfere with the analytes are analyzed with the sample.]
– Mass spectrometric detectors: Can use stable isotope labelled surrogates (i.e. deuterated or 13C-labelled compounds).
– Other detectors: Can use structurally similar surrogates that are not detected in environmental samples.
• Standard additions: Known concentrations of the analytes are added to the sample – a technique rarely used for chromatographic analysis.
External Standard CalibrationChoice depends on:• Linear range of response of the detector over a range of
analyte concentrations:– GC-ECD: Narrow linear range– GC-MS: Wide linear range
• Time required for analysis (e.g. some chromatographic runs may take >60 min).
A) Single Point Calibration:Assumes that the response is linear at concentrations below
the calibration point, and that the calibration line passes through the origin.
Concsample = Concstd x Response sample
Response std
ConcR
espo
nse
Calibration stdat upper limitof linear range
Unknown
External Standard CalibrationB) Two Point Calibration:• High concentration – Near upper limit of linear
range• Low concentration – Near LOQ
C) Multi-point Calibration:Three to six calibration points over the linear range,
with the lowest near the LOQ
Concentrations of analytes are determinedby interpolating from a linear regressionline.
Conc
Res
pons
e
Unknown
Internal Standard CalibrationStandard added to the sample that does not interfere with detection of
the analyte, but has similar or identical chromatographic properties as the analyte (i.e. analytical surrogate):
• Typically added at a constant concentration to all samples; used in conjunction with external calibration method
• Can be added to the sample:– Before sample preparation – “Method” IS that compensates for
recoveries and sample preparation errors– After sample preparation – “Instrument” IS that compensates for
instrumental variability (e.g. injection volumes)– Should report whether data are adjusted for detection of the
internal standards.
How to Validate Your Method•Approaches to Method Validation•Approaches to Method Validation•Frequency•Troubleshooting
Nargis Ismail
GLIER Laboratory – University of Windsor 1
Nargis IsmailGlier
Why Do We Need To Validate A Method?
• Method validation provides evidence that anl ti l th d h i d th h thanalytical method when carried through the
whole process in exactly the same mannerproduces results that fit for the purpose.
• Establish specificity, range and linearity of thetechnique.E t bli h Q tit ti li it IDL d MDL
GLIER Laboratory – University of Windsor 2
• Establish Quantitation limit; IDL and MDL
Validation Protocol
By carrying at least 6 replicates of asuitable clean Matrix* spiked with thesuitable clean Matrix spiked with thetarget Analytes atleast three levels ofconcentration (or a Reference Sample)carried through all the sampleprocessing steps.
GLIER Laboratory – University of Windsor 3
Frequency of Method Validation• When a new method developed in a lab or if the
method is imported from elsewhere. A j h i i ti th d• Any major changes in existing methods – Change of Solvents/ Absorbent– Change in instrumentation or major parts replacement– Trending failure of QA/QC and intervention
• e.g. suspected interferences or contamination
• Periodic method validation and documentation
GLIER Laboratory – University of Windsor 4
• Periodic method validation and documentation– Re-affirm performance characteristics (range and
linearity)
Analytical Method Development• Analytical standard
– commercial availability (Certified standard solution, neat sample, purity)
– commission synthesis; isolation & extraction from env. sample
• Chromatography and peak detection for analytical standards
– evaluate relative purity of standard – Examine for co-elution between analytes within the standard
» Similar number of peaks as indicated on certification papers
GLIER Laboratory – University of Windsor 5
p p» Peak quality (peak shape and tailing)» Confirm by GC-MSD (SCAN mode and SIM mode of
quantitation ion/ molecular ion or with library match)
Analytical Method Development Cont.)• Chromatography and peak detection for analytical
standards– Qualitative peak identification for each analyte in standard
» Documented chromatograph from standard supplier or with neat std.
» GC-MSD confirmation based on Molecular Ion or fragmentation pattern
– Confirm standard concentrations » Use of independent std. » Response factors of reference compounds from different
std.– Establish Linear response range of instrument for analytes
» Dilution series of standards that vary ~1000 fold» Determine IDL and/or Instrument Quantitation Limit
GLIER Laboratory – University of Windsor 6
» Determine IDL and/or Instrument Quantitation Limit– Examine for co-elution issues between standard components
and other analytes co-examined in the lab
– Establish appropriate spiking recovery standard » Native or labelled Internal recovery standard
Analytical Method Development Cont.)
GLIER Laboratory – University of Windsor 7
Figure 1: GC trace of Pesticide Mix 1 std. recorded on GC/MSD
GLIER Laboratory – University of Windsor 8
Figure 2: GC trace of Pesticide Mix 1 std. recorded on GC/ECD
GLIER Laboratory – University of Windsor 9
Figure 3: GC trace of Quebec Certified PCB std. recorded on GC/ECD
Machine Calibration
• Figure 2.ppt
GLIER Laboratory – University of Windsor 10
Performance Parameters1. Accuracy (%E)
expressed is a measure of the Bias from the target values.
2. Precision (CV)• (repeatability) is determined by analyzing at least 6 replicates of a
suitable clean Matrix* spiked with the target Analytes at three levels of concentration (or a Reference Sample) carried through all sample processing steps calculate the Standard Deviation (σ).
• It is defined by the following formula:% CV = (Standard Deviation/Mean Measured Value) x 100
GLIER Laboratory – University of Windsor 11
% CV of individual analytes is acceptable when it is <30% (EPA method 1668)
Performance Parameters (Cont.)3. Recovery• As % Recovery (% R) of recovery internal standard is
determined by analyzing at least 6 spiked samplesdetermined by analyzing at least 6 spiked samples (with Surrogate or fortified with Analyte Standard) carried through all sample processing steps and calculate mean measured value. It is defined by the following formula:% R = (Mean Measured Value/Reference Value) x 100% R of internal recovery standard is acceptable when it is ±40%
GLIER Laboratory – University of Windsor 12
% R of internal recovery standard is acceptable when it is ±40%
Performance Parameters (Cont.)4. Detection Limit (MDL), is determined by analyzing at least 6 replicates of
a suitable clean Matrix (Trioleine, corn oil or a low level sample) spiked with the target Analytes at a low Concentration level (which must be 10 times higher than the lowest level of the Calibration Standard) carriedtimes higher than the lowest level of the Calibration Standard) carried through all sample processing steps, calculate the Standard Deviation and evaluate the MDL using the following formula:
MDL = t(n-1) x σσ = the Standard Deviation of replicate analysis (ng/mL) at the lowest levelt(n-1) = the student’s distribution value for 99% (or 95%) confidence
GLIER Laboratory – University of Windsor 13
Level with (n-1) degrees of freedomn = the number of replicate analysis performed
• Determination of MDL includes adjustments of sample size and final extract volume to optimize recovery of some Analytes.
Validation Data – AIEA-435Tuna Fish Homogenate
• Figure.ppt
GLIER Laboratory – University of Windsor 14
Troubleshooting• Identify Isomers (o,p’ e,g.,
DDE, DDD, DDT) based on elution time from the column.
Coeluting PCBs Number of Chlorine atom
Table 1 :
• Read GC trace very carefully and check for any background peak that might be present in reagent blank or in method blank; overlapped with peak of interest.
28, 31 3 66, 95 4, 5 56, 60 4 71, 41, 64 4 84, 89, 101, 90 5 117, 87, 115 5 77, 110 4, 5
GLIER Laboratory – University of Windsor 15
• Co-elution of PCB congeners on 5% phenyl phase; DB-5 column are presented in Table 1.
123, 139, 149, 118 5, 5, 6, 5153, 132, 105 6, 6, 5 164, 163, 138 6 158, 129 6
How to Validate Your Method•Approaches to Method Validation•Approaches to Method Validation•Frequency•Troubleshooting
Nargis Ismail
GLIER Laboratory – University of Windsor 1
Nargis IsmailGlier
Why Do We Need To Validate A Method?
• Method validation provides evidence that anl ti l th d h i d th h thanalytical method when carried through the
whole process in exactly the same mannerproduces results that fit for the purpose.
• Establish specificity, range and linearity of thetechnique.E t bli h Q tit ti li it IDL d MDL
GLIER Laboratory – University of Windsor 2
• Establish Quantitation limit; IDL and MDL
Validation Protocol
By carrying at least 6 replicates of asuitable clean Matrix* spiked with thesuitable clean Matrix spiked with thetarget Analytes atleast three levels ofconcentration (or a Reference Sample)carried through all the sampleprocessing steps.
GLIER Laboratory – University of Windsor 3
Frequency of Method Validation• When a new method developed in a lab or if the
method is imported from elsewhere. A j h i i ti th d• Any major changes in existing methods – Change of Solvents/ Absorbent– Change in instrumentation or major parts replacement– Trending failure of QA/QC and intervention
• e.g. suspected interferences or contamination
• Periodic method validation and documentation
GLIER Laboratory – University of Windsor 4
• Periodic method validation and documentation– Re-affirm performance characteristics (range and
linearity)
Analytical Method Development• Analytical standard
– commercial availability (Certified standard solution, neat sample, purity)
– commission synthesis; isolation & extraction from env. sample
• Chromatography and peak detection for analytical standards
– evaluate relative purity of standard – Examine for co-elution between analytes within the standard
» Similar number of peaks as indicated on certification papers
GLIER Laboratory – University of Windsor 5
p p» Peak quality (peak shape and tailing)» Confirm by GC-MSD (SCAN mode and SIM mode of
quantitation ion/ molecular ion or with library match)
Analytical Method Development Cont.)• Chromatography and peak detection for analytical
standards– Qualitative peak identification for each analyte in standard
» Documented chromatograph from standard supplier or with neat std.
» GC-MSD confirmation based on Molecular Ion or fragmentation pattern
– Confirm standard concentrations » Use of independent std. » Response factors of reference compounds from different
std.– Establish Linear response range of instrument for analytes
» Dilution series of standards that vary ~1000 fold» Determine IDL and/or Instrument Quantitation Limit
GLIER Laboratory – University of Windsor 6
» Determine IDL and/or Instrument Quantitation Limit– Examine for co-elution issues between standard components
and other analytes co-examined in the lab
– Establish appropriate spiking recovery standard » Native or labelled Internal recovery standard
Analytical Method Development Cont.)
GLIER Laboratory – University of Windsor 7
Figure 1: GC trace of Pesticide Mix 1 std. recorded on GC/MSD
GLIER Laboratory – University of Windsor 8
Figure 2: GC trace of Pesticide Mix 1 std. recorded on GC/ECD
GLIER Laboratory – University of Windsor 9
Figure 3: GC trace of Quebec Certified PCB std. recorded on GC/ECD
Machine Calibration
• Figure 2.ppt
GLIER Laboratory – University of Windsor 10
Performance Parameters1. Accuracy (%E)
expressed is a measure of the Bias from the target values.
2. Precision (CV)• (repeatability) is determined by analyzing at least 6 replicates of a
suitable clean Matrix* spiked with the target Analytes at three levels of concentration (or a Reference Sample) carried through all sample processing steps calculate the Standard Deviation (σ).
• It is defined by the following formula:% CV = (Standard Deviation/Mean Measured Value) x 100
GLIER Laboratory – University of Windsor 11
% CV of individual analytes is acceptable when it is <30% (EPA method 1668)
Performance Parameters (Cont.)3. Recovery• As % Recovery (% R) of recovery internal standard is
determined by analyzing at least 6 spiked samplesdetermined by analyzing at least 6 spiked samples (with Surrogate or fortified with Analyte Standard) carried through all sample processing steps and calculate mean measured value. It is defined by the following formula:% R = (Mean Measured Value/Reference Value) x 100% R of internal recovery standard is acceptable when it is ±40%
GLIER Laboratory – University of Windsor 12
% R of internal recovery standard is acceptable when it is ±40%
Performance Parameters (Cont.)4. Detection Limit (MDL), is determined by analyzing at least 6 replicates of
a suitable clean Matrix (Trioleine, corn oil or a low level sample) spiked with the target Analytes at a low Concentration level (which must be 10 times higher than the lowest level of the Calibration Standard) carriedtimes higher than the lowest level of the Calibration Standard) carried through all sample processing steps, calculate the Standard Deviation and evaluate the MDL using the following formula:
MDL = t(n-1) x σσ = the Standard Deviation of replicate analysis (ng/mL) at the lowest levelt(n-1) = the student’s distribution value for 99% (or 95%) confidence
GLIER Laboratory – University of Windsor 13
Level with (n-1) degrees of freedomn = the number of replicate analysis performed
• Determination of MDL includes adjustments of sample size and final extract volume to optimize recovery of some Analytes.
Validation Data – AIEA-435Tuna Fish Homogenate
• Figure.ppt
GLIER Laboratory – University of Windsor 14
Troubleshooting• Identify Isomers (o,p’ e,g.,
DDE, DDD, DDT) based on elution time from the column.
Coeluting PCBs Number of Chlorine atom
Table 1 :
• Read GC trace very carefully and check for any background peak that might be present in reagent blank or in method blank; overlapped with peak of interest.
28, 31 3 66, 95 4, 5 56, 60 4 71, 41, 64 4 84, 89, 101, 90 5 117, 87, 115 5 77, 110 4, 5
GLIER Laboratory – University of Windsor 15
• Co-elution of PCB congeners on 5% phenyl phase; DB-5 column are presented in Table 1.
123, 139, 149, 118 5, 5, 6, 5153, 132, 105 6, 6, 5 164, 163, 138 6 158, 129 6
Quality Control Charts and Evaluation of Data Integrity
• Record keeping• Westgard Rules• Issuing Non‐compliance reports• Corrective Actions
Data Quality Evaluation• Laboratory Supervisor/Analyst Must Inspect Chromatgrams Carefully for Analysis Anomolies
• Qualitative/Subjective Evaluations – Important!– Experience counts!– Chromatography Peak Shape
• Peak Co‐Elutions and interferences• Excessive peak tailing
– septa leak, dirty inlet, degraded column, matrix artifacts
• Double peaks (column overload)p• Retention time drift (dirty sample – e.g. Excess lipids)• Interference peaks (matrix artifact)• Negative Baseline (matrix artifact – e.g. Hydrocarbons)
– Peak Integration Area • Computer Autocalculation Vs Manual Baseline Set
• Quantitative Assessment of data is the role of the Quality Manager
Data Quality Evaluation
• Censoring Rules– Rules of thumb to guide decisions
• Blank Adjustment Procedures:– Sample peak area 0 ‐ 30% batch blank, perform blank correction– Sample peak ares >30% batch blank, indicate INT (interference)
• Acceptable Spiking Recovery Range: e.g. (70% ‐ 130%)• Below Detection Limit: Peak < MDL or MQL, indicated <MDLBelow Detection Limit: Peak MDL or MQL, indicated MDL
• Statistical Approaches– Interpretation of Control Charts– Westgard Rules
Control Charting• Control charts provide temporal history of QA measurements
Aid i k d d i i– Aids in source track down and intervention strategies
• Database used to establish statistical criteria on which to judge batch to batch QA performance– Detect out‐of‐control samples while considering normal variability in method output
Control Charts – What to Collect?• Any Quality Control parameter can be used to establish database. Examples include:– Equivalent signal to analyte peak in sample blanks– Spiking standard recoveries in blanks, samples, CRMs– %Difference in analytical results between sample duplicates run within batches
– Deviation of lab result from certified value of analytesin SRMs
• Other parameters of potential interest:Other parameters of potential interest:– Detector response (i.e. peak areas) for standards– Instrument calibration results (e.g. Slopes and intercepts)
• Databases – ‘Build as you go’– Excel, Access, other formsDatabase generally valid when > 15 points are
Control Charts – What to Collect?
– Database generally valid when > 15 points are included
– Initial data set should be evaluated to ensure qualitative acceptability of performance measures
• Censor Rules
– Continual developmentL t d h t t d t d• Long term and short term records generated.
• Database continuously updated with ‘in‐control’ data as it is created
Example Control Chart: TBB Spike Recoveries (GLIER GC‐ECD SOP)
Example Control Chart: Reference Tissue (GLIER GC‐ECD SOP)
Evaluations using Control Charts
• The control chart database should be set up to provide an updated sample mean and standard deviation for tested protocol
• Levey‐Jennings Chart Rules:– For a given batch of samples, the QA parameter is examined to see if it
falls in the range of mean ± 3*SD of control chart value– Depending on goals, values of ±2*SD or ±1*SD may be used
• less common in Environmental Analytical Laboratories
• Example: – Control Chart Spike TBB Recovery: 95 ± 8 %p y– Sample Recovery = 120%– Control Chart Acceptable Range: 71 – 119 %– Sample recovery > acceptable range: therefore reject sample QA
Evaluations using Control Charts• The evaluation is based on t‐test statistic. Since the standard deviation is derived from data generated by the lab this means that approximately 1 in 20the lab, this means that approximately 1 in 20 samples will be rejected for QC!!!
• QC Parameters which are measured more frequently (e.g. spike recoveries performed in blanks, every sample and CRM analyzed) will fail QC more often than QC parameters measured less frequent (i.e. CRM recoveries)recoveries)
• Failures in QC are subject to corrective actions. The steps associated with corrective actions must be documented but will vary depending on degree of seriousness of the failure
Statistical Formulae’s
• Mean:
• Microsoft Excel formulae: =Average(cell range)• Microsoft Excel formulae: =Average(cell ‐ range)
• Standard Deviation
Mi ft E l Std ( ll )• Microsoft Excel: = Stdev(cell range)
• Example: = Stdev(A4..A120) or = average(a4..a120)
Westgard Rules• Multi‐rule QC procedure
– Provides high rigor in error detection and limits excessive false error hits– Provides trending interpretation of QC measures from control chart data
• Implemented across consecutive QC measurements and/or across multiple p Q / pbatches of samples– 1 x 2SD Rule
• QC in any given sample is within 2xSD of control chart mean – Sample is ‘In Control’– 1 x 3SD Rule and nested rules
• QC in any given sample is outside 3xSD of control chart mean – Reject, sample is out of control
– 22S Rule• 2 Consecutive values are outside of 2xSD – Run is out of control
– R4S Rule• The Difference (Range) between 2 controls within a batch of samples exceeds 4xSD – Run
is out of control– 41S Rules
• 4 consecutive control values on one side of the mean > 1 SD from the mean– 10x Rule
• 10 consecutive control values are on one side of the mean for the control chart
Westgard RulesQC in any given sample is> Mean± 3SD
QC in 2 consecutive sample> Mean± 2SD
The range of QC values among 4 consecutive measures exceeds 4 SD
QC values for 4 consecutive measuresare all on one side of the mean + are > mean± 2 SD
QC values for 10 consecutive measuresare all on one side of the mean
Non‐Conformance/Corrective Action Report
• Failure of any laboratory QC rule (e.g. i t l h ti W t d)censoring, control charting or Westgard)
should be followed up with a Non‐Compliance Report
• Non‐compliance report is filled out and filed in conjunction with the analytical reportj y p
Corrective Actions
• Vary depending on the severity of the QC measure and up to the discretion of themeasure and up to the discretion of the laboratory supervisor
• 13S – rule and acceptable within batch results:– QC issue communicated to client and accepted by client as acceptable measurement
– Data re‐examined for transcription errorsData re examined for transcription errors– Chromatogram re‐examined for peak anomolies– Sample re‐extracted and failed data discarded
Corrective Actions
• Failure of trending data and multiple QC t ithi d b thmeasurements within and across bathes as
per Westgard or other censoring rules– Systematic analyst error? Re‐train/re‐test analyst
– Procedural/method bias?• Examine instrument calibration & performance
• Cross validate analytical standards – ‘working std.’ issue
• Full method validation
Micro‐extraction Technique• Developed in‐house at GLIER for PCB/OC‐pesticides
– Reduced sample size requirements (0.15 – 1.0 g)• Less pooling required for small samples (e.g. Invertebrates)p g q p ( g )• Omit GPC step by extracting < 0.15 g lipid from sample
– Reduced solvent volume utilization• Cheaper consumable costs (especially when omitting GPC!)• Less time watching roto‐evaporators!
– Higher throughput• Batch of 6 sampels (+ blank, CRM) can be full extracted, clean‐Batch of 6 sampels (+ blank, CRM) can be full extracted, cleanup in 1 day
– Disadvantages• Less mass extracted, so mostly the ‘major’ OC‐pesticides/PCBs picked up
Materials‐SPE Manifold with PTFE valves (12 port unit)
‐Customized glass reservoirs (40 mL)Customized glass reservoirs (40 mL)‐Customized reservoir rack
‐25 mL glass luer‐lock syringes1 μm Glass fiber syringe filters‐1 μm Glass fiber syringe filters‐small mortar & pestles (4 oz)‐micro‐scale (5 digit) for lipids
Setup costs: ~$3000 USD
Procedure• 0.15‐1 g of sample is ground up with ~ 10 g Na2SO4 in small mortar and pestle
• Homogenate poured into 25 mL syringe containg 15 mLh DCM (1 1 / ) ( l l d)hexane:DCM (1:1 v/v) (valves closed)
• Spiking recovery std spiked into syringe and allowed to stand 1 hr
• ‐Valves opened and eluant drained • by gravity at ~ 3 drips/second into reservoirsy g y p /• ‐after solvent reaches bedding, add• 5mL additional extraction solvent & elute• ‐Apply vacuum until columns are dry
Clean‐up of POPs from Biological Samples
GPC
Florisil
Silica Gel
Clean‐up• Several clean‐up protocols available for PCB/OC‐pesticides
• Common ones include:
– Gel permeation chromatographyGel permeation chromatography
– Florisil
– Silica Gel
– Alumina
• Often GPC + one more clean‐up step included for high fat biological samples
• This presentation – detailed methods for GPC, florisil and Silica Gel.
– Not intended to be proscriptive of individual labs
Gel Permeation Chromatography• Size exclusion chromatography• Biobeads wet are packed in solvent within a chromatography
column• Beads contain pores on surface with MW cutoff of 1000• Beads contain pores on surface with MW cutoff of 1000
– Molecules > 1000 g/mol are excluded from pores. They elute from the column via the mobile phase by passing in between beads. Stationary phase is inert relative to these compounds.
• Mostly biogenic materials – e.g. Lipids, hydrophobic proteins that can be discarded
– Molecules < 1000 g/mol can ‘fit’ within pores. They exhibit a tortuous diffusive path, moving in out of pores of beads and between beads. They are eluted from the column slowly. Stationary exhibits i t ti ithinteractions with
– Yields analyte fraction
• Manual, semi‐automated, automated systems available• Since stationary phase is inert for both analyte and biogeneic
materials, the GPC column can be re‐used refused provided blank and calibration checks are routinely performed
GPC – Decision Rules (GLIER SOP)
• GPC clean‐up required when– % lipid content of a sample > 3.75 % in a 4 g sample
– Total extracted lipid content in sample > 0.15 g
• Limitations– % lipid content of sample > 16.25% in a 4 g sample
R d t ti i ht f l• Reduce extraction weight of sample
• Split sample and run across multiple GPCs
– GPC removes up to 0.5 g lipid in extracted sample• Total lipid extracted in sample should not exceed 0.65 g
MaterialsGPC Columns-50 cm x 2.2 cm i.d. Pyrex column-250 mL pressure equalizing Separatory funnel (as reservoir)Separatory funnel (as reservoir)-50 g BioBeads (S-X3, BioRad), wet packed in 50% Hexane:DCM (V/V)
We find it practical to set up a bank of 8 GPC columns. Enables full sample batch to be GPC processed simultaneouslyGPC processed simultaneously
Approx. Costs to set up bank of GPC: ~$4000.00 (glassware + Biobeads)
GPC – Maintenance and Calibration(See GLIER‐ Related Procedures SOP 02.005.1)
• Maintenance– GPC columns remain active as long as the beads are kept covered in
solvent.
– The packing solvent should be similar to the extraction solvent derived from the sample extraction process
• Calibration (Preliminary)– Dilute 0.2 g vegetable oil into 4mL hexane/DCM (1:1 v/v)
– Extracts are transferred to GPC column
– Before beads dry, the initial container containing extracts are rinsed with 3 x 4 mL (12 mL total pre measured solvent) rinses ofwith 3 x 4 mL (12 mL total pre‐measured solvent) rinses of hexane/DCM (1:1 v/v) and pipetted onto the top of the column
– After final rinsing, separatory funnel is replaced on top of GPC column and reservoir filled with 288 mL hexane:DCM
– Column eluant is collected in 20 mL fractions and each fraction evaporated until dryness
– Cummulative volume containing 100% lipids is determined
• Validation (at least once per year)– Use of NIST‐SRM 1588a (cod liver oil) or triolein spiked with PCB/OC
d d
GPC – Maintenance and Calibration(See GLIER‐ Related Procedures SOP 02.005.1)
standard
– Proceed as per calibration instructions
– First fraction = cumulative eluant containing lipids. Typically this reflects the first 120 mL of eluant. The fraction is discarded
– Second fraction = remaining column eluant (120‐300 mL). This fraction is concentrated and subject to florisil or silica gel clean‐up
• Alternatives to GPC‐clean‐up (lipid removal)Alternatives to GPC clean up (lipid removal)– Acid silica gel can be used for PCBs
– Several OC‐pesticides are degraded by the acid‐clean‐up
Florisil Chromatography • Florisil – proprietary solid phase (activated MgSiO3) with high surface area
• Polar components are strongly bound to florisil, non‐Polar components are strongly bound to florisil, nonpolar compounds removed using a gradient of non‐polar to moderately polar solvents strengths
• Florisil becomes contaminated and is generally not‐reusable
• Fractionation for PCB‐OcsFractionation for PCB Ocs
– Fully activated florisil• Fraction 1 – Hexanes: PCBs and some OC‐pesticides
• Fraction 2 – 85%Hexane:15% DCM: OC‐pestides
• Fraction 3 – 50%Hexane:50% DCM – OC‐pesticides
• Fraction 4 – Toluene – NO‐PCBs
• Activation Procedure– Analytical grade Florisil purchased (Fisher Scientific)
Florisil Chromatography
Scientific)
– Florisil 60‐100 mesh is kept in drying oven at 130oC over‐night (minimum over‐night)
– One hour prior to use, sufficient quantity of Florisil is removed and placed in a dessicator to cool to room temperaturecool to room temperature
– Note other SOPs call for 1.2% deactivation, by shaking fully activiated florisil with HPLC grade water and storing in a sealed glass jar
MaterialsFlorisil Columns-25 cm x 1 cm i.d. glass column equiped with 250 mL reservoirs and PTFE stopcock-add 2 cm glass wool to bottom of column-wet pack 6 g fully activated florisil in excess of hexane-add ~2cm cap of activated Na2SO4
We find it practical to set up a bank of 8 florisil columns. Enables full sample batch to be processed simultaneously
Approx. Costs to set up bank of GPC: ~$1200.00 (glassware)
Florisil Procedure– Prepare florisil column as described
above
– Hexane is eluted until it is level with the Na2SO4 cap
– Without allowing column to dry, sample extracts (~2mL) are pipetted onto of column & eluted (~1 drip/s)
– Original container with extract rinsed with 3 x 2 mL portions of hexane (pre‐measured 6 mL total)
– Column eluted 44 mL hexane; cummulative hexane eluant = Fraction1
– Column eluted with 50 mL of 15% DCM/85% hexane (v/v) = Fraction 2
– Column eluted with 150 mL 60% DCM/40% hexane (V/v) = Fraction 3
Fractionation Florisil
• Florisil Cleanup (3 fractions)– Fraction 1:Fraction 1:
• PCBs – all congeners (except NO‐PCBs), HCB (F1,F2), Heptachlor (F1, F2), Aldrin (F1, F2), Endosulfan II (f1, f2), DDE (F1, F2 – both isomers), DDD (F1, F2) ; mirex (F1)
– Fraction 2:• HCHs (100 F2), HCB (F1,F2), Heptachlor (F1,F2), heptachlor epoxide (Isomer a) Aldrin (F1,
F2), oxychlordane (F2, F3), trans‐chlordane (F2); Endosulfan I (F1+F2), cis‐chlordane (F2); DDE (F1, F2), DDD (F1, F2); DDT (F2)
– Fraction 3:• Endrin heptachlor epoxide (F2 F3) heptachlor epoxide (isomer B) oxychlordante (F2 F3)• Endrin, heptachlor epoxide (F2, F3) heptachlor epoxide (isomer B), oxychlordante (F2, F3),
Dieldrin (F3), methoxychlor (F3)
• Silica gel used in place of florisil
• Baker Scientific, 60‐200 mesh activated silica gel
Silica Chromatography (Trent Univesity’s Procedure)
, g
• 30 cm x 1 cm i.d. Glass column plugged with glass wool
• 5 g activated silica gel wet packed into column in hexane
• Hexane eluted to column until it reaches silica gel bedding
• Sample extracts added to column with sample rinses
• Fraction 1 = elution with 40 mL hexane– PCBs, some OC‐pesticides
• Fraction 2 = elution with 70 mL hexane (50%):DCM (50%)– Remaining OC‐pesticides
GC‐ Analysis
-Florisil or Silica gel fractions are each
concentrated to 2 mL (or less)
-Sulphur containing materials clean-up
using activated copper prior to addition
to GC-vials
-Each fraction analyzed by GC-ECD
GC Methods and Instrument Maintenance
• Documented Methods for POPs (OCs/PCBs)• Basic Maintenance and Troubleshooting
GLIER - University of Windsor 1
Documented Methods for POPs (OCs/PCBs)
• Injector Temp: 250oC; 63Ni-ECD Detector Temp: 300oC.
Column: 60 m x 0 25 mm I D x 0 10 µm film thickness• Column: 60 m x 0.25 mm. I.D. x 0.10 µm film thickness DB-5 (J&W).
• 1 µl injection volume using splitless injection mode.
• Carrier Gas: (UHP) - He at approximately 22 cm/sec –
GLIER - University of Windsor 2
( ) pp ydetermined at 90oC (1 mL/min); Column Head Pressure is 22.88 psi.
• Make-up Gas: Ar/CH4 (95%/5%) at 50 mL/min.
Documented Methods for POPs (OCs/PCBs) (Cont.)
280oC5 min
90oC
200oC
20oC/ min
3oC/ min
1 min
2 min
GLIER - University of Windsor 3
42 min
Figure 1: GC Oven Temperature Program for Separation of OCs and PCBs.
Documented Methods for POPs (OCs/PCBs) (Cont.)
GLIER - University of Windsor 4
Figure 2: GC trace of Quebec Certified PCB std. recorded on GC/ECD
Documented Methods for POPs (OCs/PCBs) (Cont.)
GLIER - University of Windsor 5
Figure 3: GC trace of Pesticide Mix 1 Std. recorded on GC/ECD
Basic Maintenance of GC/ECD
UHP Compressed Gases ith in line filters• UHP Compressed Gases with in-line filters (moisture and oxygen traps)
• GC Injection Port• GC Column• GC Detector
GLIER - University of Windsor 6
• GC Detector
Basic Maintenance of GC/ECD (Cont.)
GC Injection Port• Glass liner should be replace with a new one, depend p , p
on quality and quantity of samples. • Co-Extractives present in samples, often stick to the
cold spot of the injector. The injector’s temp. must keep higher than the oven initial temp.
• Some Pesticides may partially degraded in the injector port liner for e g DDT to DDE Endrin to Endrin
GLIER - University of Windsor 7
port liner, for e.g., DDT to DDE, Endrin to Endrin Ketone). Use deactivated liner. If the degradation is >15%, replace the liner with a new one.
Basic Maintenance of GC/ECD (Cont.)
GC Column• Fused silica capillary column coated with non polar• Fused silica capillary column coated with non-polar
chemically bonded stationary phase; polysiloxanes with outer protective layer of polyimide.
• Do not exceed the upper temp. limit more than the temp. specified from the column manufacturer, this might damage the column coating and destroy the stationary phase.
GLIER - University of Windsor 8
p• Always keep GC column inside box and close both ends
with septa.• It is a good practice to keep an extra new column in
hand.
Basic Maintenance of GC/ECD (Cont.)
GC Detector (ECD)• ECD detector is highly sensitive, the entire GC system
should be leak free otherwise oxidation of 63Ni foil will occur and cause higher noise level, baseline drifting and shorter lifetime of the detector will result.
• Always pre-condition the column out of the detector (capped the detector port), while attached first end of the column to the injector.
GLIER - University of Windsor 9
• Before installing or removing a column to the ECD, lower the temp. of the detector <100oC to prevent oxidation of 63Ni foil.
Troubleshooting of GC/ECD Machine• Interferences – Sometimes Negative and
positive peaks or spikes appear due to the presence of sulfur, phthalate ester (positive p , p (ppeak) and hydrocarbon (negative peak/spikes).
• If sensitivity of peaks goes down, increase in noise level:
Action - change glass liner, cut column ~10-20 cm from injector side and condition the column overnight
GLIER - University of Windsor 10
column overnight. This problem is also related to the detector due to incorrect installation distance.
Troubleshooting of GC/ECD Machine (Cont.)
• In Standard mix, if peaks deteriorated and lost resolution – don’t waste time and change column with a new one.co u t a e o e
• Having Noisy Baseline or Column bleeding –Bake the system for an hour; Injector at 300oC, Column at 285 oC and detector at 320 oC.
• Polysiloxane - Stationary Phase in DB-5 Column – destroys in presence of moisture as well as Ni
GLIER - University of Windsor 11
destroys in presence of moisture as well as Ni foil of EDC detector.
• Samples should be free of moisture.
Data Reporting &Data Reporting &Data Management
Official Data Reports• Standardize cover page should be developed t i t t ti t tto ensure consistent reporting structure
• Pertinent information about method, MDLs, QC results
• Results with appropriate significant digits
• Data flags (i e ND’s INT other)• Data flags (i.e ND’s, INT, other)
• Authorization sign‐offs – Laboratory Supervisor and Quality Manager
Data Reporting
From Muir and Sverko 2006. Paper provided in electronic materials.
• Test Reports
2) Quality System Procedures (QSP)(Focus on Technical Requirements)
– QA signoff
– Flagged results
– ND, <DL
– Sig. digits
Data management• Apart from logs, control charts and other documents, the
laboratory has a responsibility to maintain data records and review data results at clients requestdata results at clients request
• Large amounts of residual material and data generated in analysis process– Remaining, unextracted sample– Sample extracts– Clean‐up extracts (post‐GC analysis)– Raw data (chromatograms and/or chromatographic files)
Peak Areas and Calc lation Spreadsheets– Peak Areas and Calculation Spreadsheets– Calculation results– Final Report
• Laboratory should develop policy and procedures for retention, storage and eventual destruction of above materials
Data storage and back‐up
• Physical samples and/or extracts kept in secure location. Often refrigerated or frozen.– E.g. Use of vial‐files
• Print‐outs of chromtographs filed with client final report
• Electronic data sets periodically backed upE t l h d d i– External hard‐drive
– Network back‐ups (ideal)
– CD’s Notoriously Insecure
DETERMINATION OF PCBs AND PESTICIDES IN TUNAHOMOGENATE (IAEA-435) - Dec 2008
Tuna Homo
AVG. S.D. RSD IDL MDL
ref. & info Mean Standard Relative Inst. Methodvalues
Compound values (ng/g) stdev n observed Deviation Standard Detection Detection
conc. Deviation Limit Limit (ng/g) (ng/g) (ng/g) (%) (ng/mL) (ng/g) PCB18/17 PCB 18 4.3 4.6 8 0.449 0.0912 20.33 0.192 0.170 PCB 31/28 2.9 1.2 8 1.593 0.1963 12.32 0.101 0.365 PCB 52 4.4 2.5 8 2.919 0.2167 7.42 0.151 0.403 PCB 49 3.1 1.7 8 1.470 0.1155 7.86 0.113 0.215 PCB 44 3.0 2.8 8 0.662 0.1225 18.50 0.093 0.228 PCB 70 4.6 2.1 8 2.010 0.2924 14.55 0.089 0.544 PCB 95 8.3 6.3 8 7.211 0.5224 7.24 0.100 0.972 PCB 101 23.0 10.0 8 15.975 1.1365 7.11 0.097 2.114 PCB 99 16.0 4.1 8 9.614 0.7400 7.70 0.080 1.376 PCB 87 6.9 3.9 8 2.645 0.5317 20.10 0.055 0.989 PCB 110 15.0 8.3 8 4.546 0.8388 18.45 0.063 1.560 PCB 151/82 12.0 2.7 8 1.381 0.0981 7.10 0.069 0.182 PCB 149 29.0 7.1 8 18.816 1.0886 5.79 0.092 2.025 PCB 118 24.0 9.3 8 16.873 0.6717 3.98 0.074 1.249 PCB153 81.0 37.0 8 55.451 4.3924 7.92 0.077 8.170 PCB 105/132 6.7 3.6 8 4.338 0.7997 18.43 0.061 1.487 PCB 138 70.0 32.0 8 44.131 3.2421 7.35 0.057 6.030 PCB 187 31.0 10.0 8 20.330 1.0187 5.01 0.072 1.895 PCB 183 11.0 2.9 8 6.954 0.6662 9.58 0.057 1.239
GLIER Laboratory – University of Windsor
PCB 128 9.5 4.7 8 2.389 0.3980 16.66 0.044 0.740
DETERMINATION OF PCBs AND PESTICIDES IN TUNAHOMOGENATE (IAEA-435) - Dec 2008 (CONT.)
PCB 195/208 2.2 1.3 8 1.821 0.1910 10.49 0.085 0.355 PCB 194 3.9 1.1 8 2.758 0.1647 5.97 0.041 0.306 PCB 206 2.8 1.4 8 2.143 0.0725 3.38 0.047 0.135 PCB 209 2.0 0.8 8 1.398 0.1108 7.93 0.053 0.206 mirex 8 1.951 0.0773 3.96 0.045 0.144 HCB 2 6 1 0 8 1 813 0 1535 8 46 0 031 0 285HCB 2.6 1.0 8 1.813 0.1535 8.46 0.031 0.285heptachlor 1.1 2.1 8 0.266 0.0222 8.36 0.041 Aldrin 1.1 0.6 8 0.281 0.0454 16.15 0.084 endosulfan I 7.7 6.1 8 0.809 0.1008 12.45 0.191 trans nonachlor 27.0 9.0 8 0.030
' DDE 91 0 58 0 8 116 824 11 5911 9 92 0 030 19 544pp'-DDE 91.0 58.0 8 116.824 11.5911 9.92 0.030 19.544op'-DDE 2.9 2.3 8 8.340 0.3941 4.73 0.733 Endosulfan II 17.0 17.0 8 pp'-DDD 12.0 7.4 8 9.772 0.6563 6.72 0.034 1.221 o,p-DDD 2.5 0.5 8 1.140 a-BHC 0.8 0.7 8 0.029 0.0046 15.75 0.020 0.009 b-BHC 1.3 1.4 8 0.106 0.0222 20.94 0.062 0.041g-BHC 1.1 0.9 8 0.215 0.0296 13.74 0.024 0.055 Trans-chlordane 0.5 0.3 8 0.825 0.0638 7.73 0.027 0.119 cis-Chlordane 8.0 1.4 8 4.751 0.2957 6.22 0.029 0.550cis-nonachlor 13.0 3.5 8 0.028 pp'-DDT 18.0 10.0 8 30.121 2.7117 9.00 0.083 5.044 op'-DDT 7.8 4.8 8 0.742 Heptachloro Epoxide isomer A 2.8 2.5 8 0.979 0.1243 12.70 0.033 0.231Oxychlordane 8 2.130 0.2164 10.16 0.032 0.402 heptachlor epoxide isomer B 8 Dieldrin 3.80 2.1 8 3.161 0.1281 4.05 0.030 0.238 Endrin 12.0 12.0 8 0.390 0.0646 16.58 0.120 methoxychlor
12 50
GLIER Laboratory – University of Windsor
TBB 12.50 (82.01)
PCB 30 21.04
Table. List of CBs, OCs and PCB’s, providing Instrument Detection Limit (IDL), Calibration (Working) Linear Range and Coefficient of Determination (R2).
Parameter Rt (min) IDL Calibration/Working Range ng/mL R2
ng/mL (Levels):L0; L1; L2; L3; L4; L5; L6
(May/2008)
Hexachlorobenzene (HCB)
11.922 0.022 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9994
Octachlorosty(OCS)
11.698 0.017 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9995rene (OCS)
pp’-DDE 19.304 0.019 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9983
Mirex 19.478 0.028 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9978
a-BHC 11.698 0.014 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9985
b BHC 12 273 0 044 0 ;0 2; 1 0; 2 0; 10 0; 20 0; 60 0 0 9996b-BHC 12.273 0.044 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9996
g-BHC 12.464 0.016 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9987
oxy-Chlordane
17.054 0.020 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9971
trans-Chl d
17.903 0.018 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9981Chlordane
cis-Chlordane 18.515 0.019 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9989
pp’-DDD 18.715 0.026 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9984
cis-Nonachlor 21.395 0.018 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9990
pp’ DDT 22 868 0 026 0 ;0 2; 1 0; 2 0; 10 0; 20 0; 60 0 0 9977pp -DDT 22.868 0.026 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9977
HC Epoxide 16.996 0.018 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9971
Dieldrin 17.054 0.017 0 ;0.2; 1.0; 2.0; 10.0; 20.0; 60.0 0.9978
PCB #18/17 12.66/12.72 0.131 0.08-249.20 0.9978
PCB #28/31 13.89/13.94 0.065 0.11-351.10 0.9988PCB #28/31 13.89/13.94 0.065 0.11 351.10 0.9988
PCB #33 14.231 0.077 0.0634-198.2 0.9986
PCB #52 14.977 0.098 0.0634-198.4 0.9986
PCB #49 15.131 0.073 0.0633-198.0 0.9987
PCB #44 15.685 0.060 0.0633-198.0 0.9988
GLIER Laboratory – University of Windsor
PCB #74 16.964 0.050 0.064-200.0 0.9991
PCB #70 17.084 0.057 0.0634-198.2 0.9994
PCB #95 17 262 0 068 0 0320-100 2 0 9985
(Cont.)
PCB #95 17.262 0.068 0.0320-100.2 0.9985
PCB #101 18.141 0.061 0.0643-201.0 0.9989
PCB #99 18.36 0.050 0.0634-198.0 0.9989
PCB #87 19.238 0.035 0.0638-199.4 0.9988
PCB #110 19.667 0.039 0.0637-199.0 0.9990
PCB #151/82 20.198 0.043 0.0801-250.5 0.9990
PCB #149 20.713 0.056 0.0633-197.8 0.9981
PCB #118 20.791 0.046 0.0641-200.2 0.9988
PCB #153 21.793 0.046 0.0635-198.4 0.9987
PCB # 105/132 21.949 0.038 0.0478-149.5 0.9979
PCB #138 23.098 0.035 0.0645-201.6 0.9990
PCB #158 23.221 0.029 0.0161-50.20 0.9977
PCB #187 23.906 0.043 0.0635-198.4 0.9988
PCB #183 24.152 0.035 0.0633-198.0 0.9961
PCB #128 24.406 0.027 0.0643-200.8 0.9988
PCB #177 25.298 0.037 0.0640-200.0 0.9988
PCB #171/156 25.532 0.027 0.1271-397.4 0.9987
PCB #180 26.398 0.029 0.0638-199.4 0.9986
PCB #191 26.742 0.025 0.0640-200.0 0.9987
PCB #170 27.871 0.032 0.0634-198.0 0.9989
PCB #201 28.328 0.025 0.0480-150.0 0.9988
PCB #195/208 30 05/30 11 0 042 0 1281 400 4 0 9985PCB #195/208 30.05/30.11 0.042 0.1281-400.4 0.9985
PCB #194 31.135 0.023 0.0641-200.2 0.9974
PCB #205 31.425 0.023 0.0638-199.4 0.9985
PCB #206 33.067 0.025 0.064-200.0 0.9983
PCB #209 34 643 0 029 0 064-200 0 0 9984
GLIER Laboratory – University of Windsor
PCB #209 34.643 0.029 0.064-200.0 0.9984