    SIRIUSi-XHS-PWR TECHNICAL REFERENCE MANUAL SIRIUSi-XHS-PWR V20-1 

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    Table of contents  

    1. About this document 4 1.1. Legend 4 1.2. Online versions 5 

    1.2.1. SIRIUS-XHS-PWR technical reference manual 5 1.2.2. DewesoftX® tutorials 5 

    2. Getting started 6 2.1. Software installation 6 2.2. Connecting SIRIUSi-XHS-PWR 7 

    2.2.1. Network settings 8 2.2.1.1. Troubleshooting 8 

    2.2.2. DewesoftX® Settings SIRIUSi-XHS-PWR 9 2.2.3. Channel Setup SIRIUSi-XHS-PWR 15 

    2.3. Simple Measurement 16 2.3.1. Help - Manual 16 2.3.2. Analogue channel setup 17 2.3.3. Sample rate 17 2.3.4. Measurement Mode 18 2.3.5. Analyse Mode 18 

    2.4. Advanced configuration 19 2.5. Firmware upgrade 20 2.6. Licensing 21 2.7. Troubleshooting 22 

    3. System overview 23 3.1. Main features 24 

    3.1.1. Hybrid ADC technology 25 3.1.2. Perfect synchronization 25 3.1.3. Perfect amplifiers for power measurements 25 3.1.4. E-mobility power device 26 3.1.5. Open interface protocols 26 3.1.6. Power analyzer 26 3.1.7. Data Acquisition Software Included for Free 27 

    3.2. Dimensions 27 3.3. System specifications 28 

    4. Module overview 29 4.1. Voltage measurement 29 

    4.1.1. Voltage measurement: Specifications 29 4.2. Current measurements 30 

    4.2.1. Current measurement: Specifications 30 

    5. Safety instructions 31  

      SIRIUSi-XHS PWR V20-1 2/36   

     

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    5.1. General Safety Instructions 31 5.1.1. Environmental Considerations 31 5.1.2. Product End-of-Life Handling 31 5.1.3. General safety and hazard warnings for all Dewesoft systems 31 

    6. Notice 34 6.1. Warranty Information 34 6.2. Calibration 34 6.3. Support 34 6.4. Service/repair 34 6.5. Restricted Rights 34 6.6. Printing History 35 6.7. Copyright 35 6.8. Trademarks 35 6.9. Documentation version 36 

        

      

    SIRIUSi-XHS PWR V20-1 3/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    1. About this document  This is the Technical Reference Manual for SIRIUSi-XHS-PWR Version V20-1  SIRIUS-XHS-PWR (SIRIUS-XHS-1xHV-1xDCCT) is a high performance, high speed line of SIRIUS® real time                         data acquisition hardware used for the most demanded power measurements, specially designed for                         E-mobility applications. Each system also includes a professional license for our award-winning                       DewesoftX® data acquisition software.  The manual is divided into several chapters. You will find: 

    ● A detailed description of the SIRIUS-XHS-PWR hardware ● A description of the mechanical connections and variants of current ranges ● A comprehensive introduction to the configuration of the modules using DewesoftX® ● Detailed technical data: Specifications, etc. 

      

    1.1. Legend  The following symbols and formats will be used throughout the document. 

     Important Gives you important information about a subject. Please read carefully!  

     Hint Gives you a hint or provides additional information about a subject.  

     Example Gives you an example to a specific subject.  

     Safety symbols in the manual: 

     Warning  Calls attention to a procedure, practice, or condition that could cause the body injury or death  

     Caution Calls attention to a procedure, practice, or condition that could possibly cause damage to                           equipment or permanent loss of data. 

      

    SIRIUSi-XHS PWR V20-1 4/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    1.2. Online versions 

    1.2.1. SIRIUS-XHS-PWR technical reference manual The most recent version of this manual can be downloaded from our homepage: https://download.dewesoft.com/list/manuals-brochures/hardware-manuals In the Hardware Manuals section click the download link for the SIRIUSi-XHS-PWR technical reference                           manual. 

    1.2.2. DewesoftX® tutorials The DewesoftX® tutorials document, provides basics and additional information and examples for                       working with DewesoftX® and certain parts of the program. The latest version of the DewesoftX® tutorials can be found here: https://download.dewesoft.com/list/manuals-brochures/software-manuals In the Software Manuals section click the download link of the DewesoftX® tutorials entry. 

     Important Read safety instructions first in chapter 5. Safety instructions.   

       

      

    SIRIUSi-XHS PWR V20-1 5/36    

    https://download.dewesoft.com/list/manuals-brochures/hardware-manualshttps://download.dewesoft.com/list/manuals-brochures/software-manuals

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2. Getting started This chapter will help you to install the software, connect your SIRIUSi-XHS system to the PC via                                 ethernet and will show you how to configure DewesoftX®.  To follow these steps, you need the following items:  

    ● your brand new SIRIUSi-XHS-PWR system (included in the shipment)  ● your network cable (included in the shipment)  ● your PC with Windows 10 (older versions like Windows® 7 may also work) 

    2.1. Software installation  For optimal working, we recommend that you install the latest version of Dewesoft. If you already have                                 installed the older version Dewesoft is recommended that you find the newest version on the website                               under Support/Downloads/DewesoftX section. You can also check if a newer version is available in                           software.  

     Image 1: Check for update 

     

       

      

    SIRIUSi-XHS PWR V20-1 6/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2.2. Connecting SIRIUSi-XHS-PWR In this chapter, you can see the basic instructions for connecting SIRIUSi-XHS-PWR devices. Advanced                           connections are described in the following chapters.  

    STANDALONE UNIT  

     Image 2: Connection of SIRIUSi-XHS-PWR standalone device to PC 

      

    PoE and MULTIPLE UNITS 

     Image 3: Connection of multiple SIRIUSi-XHS-PWR units to PC using PoE 

      

    SIRIUSi-XHS PWR V20-1 7/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2.2.1. Network settings SIRIUSi-XHS-PWR is a network device with static IP. The ethernet port on your SBOX, data logger or PC should be configured in a way to match the IP address of the SIRIUSi-XHS-PWR unit.  

     Example SIRIUSi-XHS-PWR has the IP address 192.168.10.1 Your network adapter should be set to: - IP: 192.168.10.x, where x is number from 2-255 - Subnet mask: 255.255.255.0 

     

    2.2.1.1. Troubleshooting If Dewesoft doesn’t detect your devices you should try pinging them via command prompt. In command prompt write: ping IP replace IP with IP that is set on your device. The result should be something like this:  

     Image 4: ping IP in Command Prompt 

     

       

      

    SIRIUSi-XHS PWR V20-1 8/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2.2.2. DewesoftX® Settings SIRIUSi-XHS-PWR SIRIUSi-XHS-PWR devices have static IP. When adding a new device open DewesoftX® and go to Options->Settings as shown in Image 5. Click on the Options button at the top right, and then on the Settings item in the pop-up to open the                                         DewesoftX® settings dialogue.   

     Image 5: Settings 

       

      

    SIRIUSi-XHS PWR V20-1 9/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

        

    Image 6: DewesoftX® settings 

    Adding a new XHS device is done by clicking +, this will show the Add device window where you can                                       find a Real-time controller. Sirius-XHS is an RT device so you should add it by clicking + on RT Device. If                                         you want to use two devices you should add two.  

      

    SIRIUSi-XHS PWR V20-1 10/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

     Image 7: Adding RT device 

      Support of RT-devices is still an Experimental feature. If you do not see this option when adding                                 a device, the option needs to be enabled under Advanced -> Experimental settings.  

      

    SIRIUSi-XHS PWR V20-1 11/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

     Image 8: Enable experimental feature 

      After that you will see a new Sirius XHS or Dewesof RT connection in the device tree.  

      

    SIRIUSi-XHS PWR V20-1 12/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

     Image 9: Two RT devices 

      If the text by the device is disconnected, you should click on the device (Dewesoft RT Connection) , set                                     the correct IP and refresh.  

      

    SIRIUSi-XHS PWR V20-1 13/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

     Image 10: Setting IP 

      If everything is set correctly both devices should be detected as Sirius-XHS devices.  

      

    SIRIUSi-XHS PWR V20-1 14/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

     Image 11: Devices overview 

     Before measurement you need to set synchronization. This is done by selecting Local system and                             selecting time source as External, External (Clock provider) as PTP and Dewesoft DAQ Devices (Clock                             slave) as PTP.  

     Image 12: Setting synchronization 

     

       

      

    SIRIUSi-XHS PWR V20-1 15/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2.2.3. Channel Setup SIRIUSi-XHS-PWR In the channel setup you can see a preview of the connected devices on the left side.   

    When you click on a connector in the image the corresponding channel in the Channel setup                               grid will automatically be selected. This also works the other way around: when you select a                               channel (or multiple channels) in the setup grid, the corresponding connectors in the image                           will be highlighted.  

     The sampling rate will be set for all connected SIRIUSi-XHS-PWR units: of course only up to the max.                                   sampling rate of the individual units.  In channel setup mode you can set the wanted sample rate, range etc. and select channels you want to                                     measure.  

     

     Image 13: Channel setup 

      

    2.3. Simple Measurement This chapter describes measurement basics, how to configure SIRIUSi-XHS-PWR and gives some details                         on the measurement setup. 

    2.3.1. Help - Manual Note that this document is just a quick start guide. For detailed information about DewesoftX® consult                               the Manual. To open the manual press the F1 button or click on the Options button and then select                                     Manual from the pop-up menu.   

      

    SIRIUSi-XHS PWR V20-1 16/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    Image 14: Help - Manual 

     When DewesoftX® has started up, you will be in Measure mode and see the Setup files list. Click on Ch.                                       setup (on the right of Setup files) to switch to the Channel setup mode.   

    Image 15: Setup files 

    2.3.2. Analogue channel setup In the analogue channel setup screen you can see all channels of your connected SIRIUSi-XHS-1xHV-HA                             units. Per default only the first channel will be set to Used. Unused channels will not show up in                                     measure mode and can thus not be used for display, calculations or storing: thus, we will also set the                                     other channels to the used. Then you can click into the selected region to toggle Used/Unused for all                                   channels at once. The selected channels will also be highlighted in the small preview image of the                                 device.  When you press the Setup button of a channel (the column at the right edge of the channel table – not                                         shown in this screen-shot), you can change all the settings of the channel amplifier.  You can also change the sample rate of the SIRIUSi-XHS-1xHV-HA.

    2.3.3. Sample rate One of the most important settings is the sample rate. The sample rate defines how many data points,                                   SIRIUSi-XHS-PWR will transfer to DewesoftX®. Higher sample rate also means that more data needs to                             be transferred via ethernet to your computer. 

      

    SIRIUSi-XHS PWR V20-1 17/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      The sampling speed mainly depends on your application. To display your signal in a time domain with a                                   good time resolution, you should sample 10 to 20 times faster than the frequency of the signal that you                                     want to measure, e.g. 1 kS/s for a 50 Hz sine-wave. If you have a lot of high frequency components, it                                         may be necessary to sample 100 times faster, e.g. 5 kS/s for the 50 Hz sine-wave, or even more. If you                                         display only the frequency domain (FFT analysis), a 2.5 times faster sampling would be sufficient (125 S/s                                 for the 50 Hz sine-wave). The higher the sampling rate, the better the time resolution. But also the file                                     size will increase.   

    Image 16: Sample rate 

    2.3.4. Measurement Mode A click on Measure (at the right side of Ch. setup in Image 11) will take you to the Recorder screen                                         measure mode where you can already see live data. 

      Hint When switching to Measure mode the data will not be stored automatically.  

    In measure mode you can have several measurement screens. DewesoftX® will create 2 default screens:                             Recorder and Custom but you can also create new screens or change the widgets of the existing                                 screens as you like.   The most important sections of the Measure mode:  Recorder widget displays all your measurement channels. You can use the channel-selector list to                           assign measurement channels to the instruments. Each instrument has different settings. To start storing the data to a file, press the Store button. When you are done, press the Stop button to                                         stop recording.  Now DewesoftX® has created a datafile with all the data that you have seen during the recording                                 session. You can now click the Analysis button (on the left-top of the screen to the right of the Measure                                       button) to go to Analysis mode. 

    2.3.5. Analyse Mode When you have just stopped a measurement, DewesoftX® will automatically open the last recorded                           data file in Review mode, so that you can start the analysis right away.   

      

    SIRIUSi-XHS PWR V20-1 18/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    Image 17: Analyse mode 

    The Review mode is much like the measurement mode. You will see the same measurement screens,                               the channel-selector list and the properties of the currently selected instrument.  Differences are:  (1) you have additional tool-buttons  (2) there is a Signal overview window which will show you the whole data of one selected channel of the                                       data file   Now you can use the cursors to analyse your data, zoom in and out of the data, click Offline math to add                                           computations based on your data, etc. You can also change the design of your measurement screens,                               print reports based on your data and export the data to other file formats for further analysis.   

    2.4. Advanced configuration Note, that the DewesoftX® launcher has already done the hardware setup for you – you can check this                                   in the Settings dialogue. Click the Settings button (1) – and then click the Settings Menu item (2).   

      

    SIRIUSi-XHS PWR V20-1 19/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    Image 18: Open settings dialogue 

    2.5. Firmware upgrade ● Download the Dewesoft upgrade package (.dxu file) from the Dewesoft downloads page under 

    the section Drivers.  ● Copy the file into the Firmwares folder of your Dewesoft installation (e.g. 

    D:\DewesoftX\System\X3\Firmwares).  ● Connect the Dewesoft instrument to the PC and run Dewesoft X3.  ● Go to settings under the Update tab:

     Image 19: Firmware update 

     ● If the firmware package isn’t selected, select it by pressing the button and find the folder with 

    the firmware file in it.  

      SIRIUSi-XHS PWR V20-1 20/36   

     

    https://download.dewesoft.com/list/drivers/firmware

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    ● Select the device you want to upgrade and start the firmware upgrade by pressing the “Upgrade” button.  

    2.6. Licensing SIRIUSi-XHS-1xHV-HA or any other Dewesoft device already comes with an embedded Dewesoft license. You can check the license details with all the available options in the Licensing tab (1) by pressing the three dotted button (6). However if the user decides to upgrade the license with an additional extension, Dewesoft will require a new license registration. The registration can be made online (2) or offline by importing an offline license (5) in case the system doesn’t have an internet connection. Offline license can be pre-registered on a different PC with the internet connection. If needed, the license can also be written on the actual device (5).   Active and embedded licenses are seen under Active licenses tab (5). If the license is recognized as none active, it usually means that the wrong license was entered.   

     Image 20: Licensing 

      All licenses regarding SIRIUSi-XHS-PWR will only work when the SIRIUSi-XHS-PWR system is                       connected to your PC and the device has been activated in the hardware setup.  

       

      

    SIRIUSi-XHS PWR V20-1 21/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    2.7. Troubleshooting If your SIRIUSi-XHS-PWR device is not found by DewesoftX®:  

    ● If you did not restart Windows after the software installation, restart now ● Make sure that you have started DewesoftX® version X3 SP12 or higher ● Make sure that the external power supply is connected and okay  ● Disconnect the network cable and reconnect it. If this does not work, try to connect the network                                 

    cable to another Ethernet port of your PC  ● Try to restart DewesoftX®  ● Try to restart the PC 

       

      

    SIRIUSi-XHS PWR V20-1 22/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3. System overview      

    SIRIUS-XHS  

    Sirius XHS is the first device ever with Hybrid ADC technology capable of doing both high bandwidth transient recording and very high dynamic alias-free acquisition - software-selectable per channel. 

    Modern interfaces and protocols allow open and flexible connectivity.     

     Image 21: SIRIUSi XHS 1xHV1xDCCT device 

      

       

      

    SIRIUSi-XHS PWR V20-1 23/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3.1. Main features  

    ● HYBRID ADC TECHNOLOGY: offers everything you ever wanted out of a high-end data                         acquisition device; software selectable per channel: 

    ○ HIGH SAMPLING RATE HIGH BANDWIDTH: 15 MS/sec sampling rate captures even the                       shortest transients; 

    ○ HIGH DYNAMIC ALIAS-FREE: Alias free filtering allows perfect acquisition of signals with                       up to 160 dB dynamic, similar performance than our best Dual-core Sirius, better than 24                             bit. 

    ● PERFECT SYNCHRONIZATION: even though users can select some channels to be high                       bandwidth and some to be alias-free, filtering is made in the way that all signals are perfectly                                 time aligned with zero phase shift. 

    ● MODERN DATA INTERFACES: GLAN interface allows the distribution of devices with PTPv2                       synchronization; the USB3 interface allows the fastest data transfers with a quick connection to                           the computer. 

    ● OPEN PROTOCOLS: data is transferred to any host using open OPC UA industry-standard                         protocol; in parallel the data is available over XCP protocol, allowing connection to ECU                           calibration software packages. 

    ● HIGH ISOLATION: High channel-to-channel and channel-to-ground isolation prevents damage                 to the systems from excessive voltage and avoids ground loops. 

    ● EXTREME CONDITIONS: SIRIUSi-XHS-1xHV-1xHA comes in ruggedized chassis and can                 withstand the harshest environmental conditions operating in temperature range from -40 °C to                         +80 °C. All the units are dust- and waterproof with IP67 rating. 

    ● ADDITIONAL SAFETY: SIRIUSi-XHS-1xHV-1xHA includes digital interlock safety solution on the                   chassis covers and on the cabling. Additionally, multiple temperature sensors are distributed                       inside the chassis to monitor the temperature and to initiate shutdown in case of overheating. 

    ● SOFTWARE INCLUDED WITH FREE LIFETIME UPGRADES: The easy-to-use but rich in                     functionality, award-winning DEWESoft X3 software is included. All upgrades to the software are                         free forever with no hidden licensing costs. 

     

       

      

    SIRIUSi-XHS PWR V20-1 24/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3.1.1. Hybrid ADC technology User can select for each channel two modes of operation: 

    a) High bandwidth mode: with more than 3 MHz bandwidth and 15 Ms/sec sampling rate, XHS can                               perfectly acquire impulse, step and square signals without any ringing or overshoot. Such a                           mode is perfect for transient recording and power analysis. Such acquisition mode is typically in                             SAR ADCs. 

    b) Alias free mode: up to 1 Ms/sec data can be acquired with extremely high dynamic, similar to                                 our dual 24 bit SIRIUS dual-core. The data is totally alias-free, so all higher frequencies are fully                                 rejected. Such a mode is perfect for sound, vibration, and general data recording applications.                           Such acquisition mode is typically in Sigma-Delta ADCs. 

    Usually, you would need two totally separate devices for the above-mentioned applications. But on                           SIRIUS XHS you can select channel per channel, depending on the application, the appropriate mode of                               ADC operation.   

    3.1.2. Perfect synchronization If you mix channels from Sigma-Delta and SAR devices (high bandwidth and alias-free), then channels are delayed due to SIgma-delta filtering group delay. But with SIRIUS XHS, due to the nature of the chosen filter, both modes are perfectly aligned and synchronized.  As on any Dewesoft devices, data can be synchronized to other data sources, such as vehicle bus interface, GPS, IMU, video, and others.  Devices can be synchronized between each other using PTPv2 mechanism, IRIG or PPS signal. This also provides absolute time synchronization using in-house time servers or GPS.   

    3.1.3. Perfect amplifiers for power measurements  SIRIUS XHS HV: Highly isolated CATII 1000 V, from 200 V to 2000 V range, >3 MHz bandwidth with down to 0.03% accuracy, this device is perfect for direct connection of high voltage signal.  SIRIUS XHS HA: DCCT technology, up to 1000 A range, >500 kHz bandwidth, with down to 0.01% gain linearity, this device is perfect for direct connection to the in-vehicle high current lines.  

       

      

    SIRIUSi-XHS PWR V20-1 25/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3.1.4. E-mobility power device The SIRIUS XHS 1xHV,1xDCCT device is well suited for in-vehicle measurements. In-built patented DC-CT current transducer allows very precise current measurements even in most demanding applications such as very high current peaks as well as the leakage current measurement.  The device is perfect to be installed in a vehicle directly in the power lines. Its IP67 protection allows the usage in harsh environments.  

    3.1.5. Open interface protocols 

    In today’s world of open toolchain and intercommunication, each device should be compliant with standard protocols. All protocols can be used at the same time.  OPC UA: is the industry standard. Actually, it is more than a standard, it is a perfect framework where the device can be described and set up in any system, including SCADA, MES, ERP, mobile devices and others.  XCP: Starting with version 1.4., XCP became a very powerful interface protocol in the automotive industry for data exchange. In the modern age of e-mobility, the required sampling rates are much higher than ever and 1 GBIT XCP interface allows data transfers with as high as 5 Ms/sec.  CAN/CAN FD: Vehicle bus systems can be read through the CAN FD interface while the measured data can be also transmitted through this “golden” standard protocol.  

    3.1.6. Power analyzer 

    Dewesoft software will automatically calculate and store not only raw data from the voltage and current transducers but also all the power parameters: P, Q, S, D, Cos φ, power factor, P, Q, cos φ for each harmonic and other relevant power parameters.  Each parameter can visually be displayed in a time recorder, digital or analog display.  Using a computer graphics card as a multi-threaded calculation platform, the Dewesoft power module can calculate all those parameters up to a full speed of 15 Ms/sec.  

       

      

    SIRIUSi-XHS PWR V20-1 26/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3.1.7. Data Acquisition Software Included for Free 

    All Dewesoft data acquisition systems are bundled with award winning Dewesoft X Professional DAQ software. Dewesoft X is the world's most advanced and easy-to-use data acquisition and analysis software. Dewesoft X's flexibility and power will help you unleash the DAQ system to its full potential and gives you many advantages over other DAQ systems. Functionalities like plug-and-play, hardware auto-detection, smart TEDS sensors, advanced storing and data analysis features will take your measurement and analysis needs to a whole new level.   

    3.2. Dimensions  TECHNICAL DRAWING  

     Image 22: Dimensions 

     

       

      

    SIRIUSi-XHS PWR V20-1 27/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    3.3. System specifications   

    Power 

    Power supply 

     9 - 36 V DC PoE (RJ45) PWR IN (Lemo 1B 2-pin) 

    Power consumption  Typ. 30 W 

    Environmental 

    Operating Temperature  -40 to 70 °C 

    Storage Temperature  -40 to 85 °C 

    IP rating  IP67 

    Shock & Vibration  TBD 

    Interfaces 

    Ethernet GbE (XCP, OPC UA) incl. IEEE1588v2 synchronization (PTP) (RJ45) 

    CAN  CAN-FD, CAN 2.0 (DSUB9) 

    Sync Input / Output 

    Level (Input/Output)  TTL compatible 

    Max. Output Current  ±24 mA (±50 mA for 1 sec) 

    Max. Sync-cable length  100 m (Master/Slave), 200 m (IRIG) 

    Additional Specifications 

    Interlock  YES (chassis covers + cables) 

       

       

      

    SIRIUSi-XHS PWR V20-1 28/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    4. Module overview  

    4.1. Voltage measurement  

    KEY FEATURES  

    ● Sampling rate: 15 MS/sec ● Input voltage ranges: ±2000 V ... ±20 V ● AAF: 1 MHz, 6th order  ● Bandwidth: 3 MHz ● Input impedance: 10 MΩ ● Isolation: CATII 1000V 

     

    4.1.1. Voltage measurement: Specifications  Inputs  Voltage 

    ADC Type  HybridADC - 24 bit alias free up to 1 MS/sec, 16 bit up to 15 MS/sec 

    Sampling Rate  Simultaneous 15 MS/sec 

    Filtering  AAF 1MHz (6th order Butterworth) 

    Analogue bandwidth  > 3MHz 

    Voltage ranges  ±2000 V, ±1000 V, ±400 V, ±200 V, ±100 V, ±40 V, ±20 V 

    Input Accuracy  Signal frequency  Accuracy 

      DC  ±0.03% of reading ±0.02% of range ±0.04V 

      Up to 1kHz  ±0.03% of reading ±0.02% of range 

      Up to 10 kHz  ±0.1% of reading ±0.05% of range 

      Up to 100 kHz  ±2% of reading ±0.1% of range 

      Up to 1000 kHz  ±5% of reading ±0.5% of range 

    Voltage range ±2000 V  Filter OFF  Filter ON 

    Typ. Noise floor @ 15 MS/sec / 1 MS/sec / 10kHz  81 dB / 92dB / 103 dB  87 dB / 97 dB / 108 dB 

    Typ. CMR @ 50 Hz/1 kHz/10 kHz  TBD 

    Channel Crosstalk @ 50 Hz/1 kHz  TBD 

    Gain Drift  Typical 10 ppm/K, max. 40 ppm/K 

    Offset Drift  Typical 1 mV/K + 1 ppm of range/K, max 2 mV/K + 5 ppm of range/K 

    Gain Linearity 

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    4.2. Current measurements  

    KEY FEATURES  

    ● Zero-flux transducer ● Measurement of residual current in core ● High accuracy ● High bandwidth ● Low noise ● Temperature stability ● Low power consumption 

     

    4.2.1. Current measurement: Specifications  

    Inputs  Current 

    ADC Type  HybridADC - 24 bit alias free up to 1 MS/sec, 16 bit up to 15 MS/sec 

    Sampling Rate  Simultaneous 15 MS/sec 

    Filtering  AAF 1MHz (6th order Butterworth) 

    Current Ranges  1000 A  500 A  100 A 

    Sensitivity  4 mV/A  8 mV/A  40 mV/A 

    Recommended cable size  95 mm2  50 mm2  25 mm2 

    Recommended PG cable gland  M32  M25  M16 

    Recommended grounding cable size  50 mm2  25 mm2  16 mm2 

    Offset Accuracy  ± 50 mA 

    Gain linearity  < 0.01% 

                   

      

    SIRIUSi-XHS PWR V20-1 30/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    5. Safety instructions  Your safety is our primary concern! Please be safe!  

    5.1. General Safety Instructions  

     Warning  

    The following general safety precautions must be observed during all phases of operation, service, and                             repair of this product. Failure to comply with these precautions or with specific warnings elsewhere in                               this manual violates safety standards of design, manufacture, and intended use of the product.                           Dewesoft d.o.o. assumes no liability for the customer’s failure to comply with these requirements.   All accessories shown in this document are available as an option and will not be shipped as standard                                   parts.  

    5.1.1. Environmental Considerations Information about the environmental impact of the product.  

    5.1.2. Product End-of-Life Handling Observe the following guidelines when recycling a Dewesoft system:  System and Components Recycling Production of these components required the extraction and use of natural resources. The substances                           contained in the system could be harmful to your health and to the environment if the system is                                   improperly handled at its end of life! Please recycle this product in an appropriate way to avoid an                                   unnecessary pollution of the environment and to keep natural resources.  

    This symbol indicates that this system complies with the European Union’s requirements                       according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE).                     Please find further information about recycling on the Dewesoft web site                     www.dewesoft.com   

    Restriction of Hazardous Substances This product has been classified as Monitoring and Control equipment and is outside the scope of the                                 2002/95/EC RoHS Directive. However, we take care of our environment and the product is lead-free.  

    5.1.3. General safety and hazard warnings for all Dewesoft systems Safety of the operator and the unit depend on following these rules.  

      SIRIUSi-XHS PWR V20-1 31/36   

     

    http://www.dewesoft.com/

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    ● Use this system under the terms of the specifications only to avoid any possible danger. ● Read your manual before operating the system. ● Observe local laws when using the instrument. ● DO NOT touch internal wiring! ● DO NOT use higher supply voltage than specified! ● Use only original plugs and cables for harnessing. ● You may not connect higher voltages than rated to any connectors. ● The power cable and connector serve as Power-Breaker. The cable must not exceed 3 meters,                             

    the disconnect function must be possible without tools. ● Maintenance must be executed by qualified staff only. ● During the use of the system, it might be possible to access other parts of a more                                 

    comprehensive system. Please read and follow the safety instructions provided in the manuals of                           all other components regarding warning and security advice for using the system. 

    ● With this product, only use the power cable delivered or defined for the host country. ● DO NOT connect or disconnect sensors, probes or test leads, as these parts are connected to a                                 

    voltage supply unit. ● Ground the equipment: For Safety Class 1 equipment (equipment having a protective earth                         

    terminal), a non-interruptible safety earth ground must be provided from the mains power                         source to the product input wiring terminals. 

    ● Please note the characteristics and indicators on the system to avoid fire or electric shocks.                             Before connecting the system, please read the corresponding specifications in the product                       manual carefully. 

    ● The inputs must not unless otherwise noted (CATx identification), be connected to the mains                           circuit of category II, III and IV. 

    ● The power cord separates the system from the power supply. Do not block the power cord, since                                 it has to be accessible for the users. 

    ● DO NOT use the system if equipment covers or shields are removed. ● If you assume the system is damaged, get it examined by authorized personnel only. ● Adverse environmental conditions are Moisture or high humidity Dust, flammable gases, fumes                       

    or dissolver Thunderstorm or thunderstorm conditions (except assembly PNA) Electrostatic                   fields, etc. 

    ● The measurement category can be adjusted depending on module configuration. ● Any other use than described above may damage your system and is attended with dangers like                               

    short-circuiting, fire or electric shocks. ● The whole system must not be changed, rebuilt or opened. ● DO NOT operate damaged equipment: Whenever it is possible that the safety protection                         

    features built into this product have been impaired, either through physical damage, excessive                         moisture, or any other reason, REMOVE POWER and do not use the product until the safe                               operation can be verified by service-trained personnel. If necessary, return the product to                         Dewesoft sales and service office for service and repair to ensure that safety features are                             maintained. 

    ● If you assume a more riskless use is not provided anymore, the system has to be rendered                                 inoperative and should be protected against inadvertent operation. It is assumed that a more                           riskless operation is not possible anymore if the system is damaged obviously or causes strange                             noises. the system does not work anymore. the system has been exposed to long storage in                               adverse environmental. the system has been exposed to heavy shipment strain. 

    ● Warranty void if damages caused by disregarding this manual. For consequential damages, NO                         liability will be assumed! 

      

    SIRIUSi-XHS PWR V20-1 32/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    ● Warranty void if damage to property or persons caused by improper use or disregarding the                             safety instructions. 

    ● Unauthorized changing or rebuilding the system is prohibited due to safety and permission                         reasons (CE). 

    ● Be careful with voltages >25 VAC or >35 VDC! These voltages are already high enough in order to                                   get a perilous electric shock by touching the wiring. 

    ● The product heats during operation. Make sure there is adequate ventilation. Ventilation slots                         must not be covered! 

    ● Only fuses of the specified type and nominal current may be used. The use of patched fuses is                                   prohibited. 

    ● Prevent using metal bare wires! Risk of short circuit and fire hazard! ● DO NOT use the system before, during or shortly after a thunderstorm (risk of lightning and high                                 

    energy over-voltage). An advanced range of application under certain conditions is allowed with                         therefore designed products only. For details, please refer to the specifications. 

    ● Make sure that your hands, shoes, clothes, the floor, the system or measuring leads, integrated                             circuits and so on, are dry. 

    ● DO NOT use the system in rooms with flammable gases, fumes or dust or in adverse                               environmental conditions. 

    ● Avoid operation in the immediate vicinity of high magnetic or electromagnetic fields,                       transmitting antennas or high-frequency generators, for exact values please refer to the                       enclosed specifications. 

    ● Use measurement leads or measurement accessories aligned with the specification of the                       system only. Fire hazard in case of overload! 

    ● Do not switch on the system after transporting it from a cold into a warm room and vice versa.                                     The thereby created condensation may damage your system. Acclimatise the system                     unpowered to room temperature. 

    ● Do not disassemble the system! There is a high risk of getting a perilous electric shock.                               Capacitors still might be charged, even if the system has been removed from the power supply. 

    ● The electrical installations and equipment in industrial facilities must be observed by the security                           regulations and insurance institutions. 

    ● The use of the measuring system in schools and other training facilities must be observed by                               skilled personnel. 

    ● The measuring systems are not designed for use in humans and animals. ● Please contact a professional if you have doubts about the method of operation, safety or the                               

    connection of the system. ● Please be careful with the product. Shocks, hits and dropping it from already- lower level may                               

    damage your system. ● Please also consider the detailed technical reference manual as well as the security advice of the                               

    connected systems. ● This product has left the factory in safety-related flawless and in proper condition. In order to                               

    maintain this condition and guarantee safety use, the user has to consider the security advice                             and warnings in this manual. 

      

       

      

    SIRIUSi-XHS PWR V20-1 33/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    6. Notice The information contained in this document is subject to change without notice.  Note: Dewesoft d.o.o. shall not be liable for any errors contained in this document. Dewesoft MAKES NO                               WARRANTIES OF ANY KIND WITH REGARD TO THIS DOCUMENT, WHETHER EXPRESS OR IMPLIED.                         DEWESOFT SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND                 FITNESS FOR A PARTICULAR PURPOSE. Dewesoft shall not be liable for any direct, indirect, special,                             incidental, or consequential damages, whether based on contract, tort, or any other legal theory, in                             connection with the furnishing of this document or the use of the information in this document.  

    6.1. Warranty Information The copy of the specific warranty terms applicable to your Dewesoft product and replacement parts can                               be obtained from your local sales and service office. To find a local dealer for your country, please visit                                     https://dewesoft.com/support/distributers.  6.2. Calibration Every instrument needs to be calibrated at regular intervals. We recommend annual calibration. Before                           your Dewesoft data acquisition system is delivered, it is calibrated. Detailed calibration reports for your                             Dewesoft system can be requested. We retain them for at least one year, after system delivery.  6.3. Support Dewesoft has a team of people ready to assist you if you have any questions or any technical difficulties                                     regarding the system. For any support please contact your local distributor first or Dewesoft directly.  E-mail: [email protected]  Address: Dewesoft d.o.o. Gabrsko 11a 1420 Trbovlje Slovenia  Europe Tel.: +386 356 25 300 Web: http://www.dewesoft.com The telephone hotline is available Monday to Friday from 07:00 to 16:00 CET (GMT +1:00)  

    6.4. Service/repair The team of Dewesoft also performs any kinds of repairs to your system to assure a safe and proper                                     operation in the future. For information regarding service and repairs please contact your local                           distributor first or Dewesoft directly on https://dewesoft.com/support/rma-service.  

    6.5. Restricted Rights Use Slovenian law for duplication or disclosure. Dewesoft d.o.o. Gabrsko 11a, 1420 Trbovlje, Slovenia /                             Europe. 

      

    SIRIUSi-XHS PWR V20-1 34/36    

    https://dewesoft.com/support/distributersmailto:[email protected]://www.dewesoft.com/https://dewesoft.com/support/rma-service

  • DEVICE TECHNICAL REFERENCE MANUAL 

       

    6.6. Printing History Version 1.0.0 Released 2020 Last changed: 17/09/2020 at 07:00.  6.7. Copyright Copyright © 2015-2020 Dewesoft d.o.o.  This document contains information which is protected by copyright. All rights are reserved.                         Reproduction, adaptation, or translation without prior written permission is prohibited, except as                       allowed under the copyright laws.  All trademarks and registered trademarks are acknowledged to be the property of their owners.  6.8. Trademarks We take pride in our products and we take care that all key products and technologies are registered as                                     trademarks all over the world. The Dewesoft name is a registered trademark. Product families (KRYPTON, SIRIUS, DSI, DS-NET, IOLITE) and technologies (DualCoreADC,                   SuperCounter, GrandView) are registered trademarks as well. When used as the logo or as part of any graphic material, the registered trademark sign is used as a part                                         of the logo. When used in text representing the company, product or technology name, the ® sign is not used. The                                   Dewesoft triangle logo is a registered trademark but the ® sign is not used in the visual representation                                   of the triangle logo.     

      

    SIRIUSi-XHS PWR V20-1 35/36    

  • DEVICE TECHNICAL REFERENCE MANUAL 

      

    6.9. Documentation version 

    Doc-Version  Date [dd.mm.yyyy] 

    Notes 

    1.1.0  17.09.2020  Initial version 

     

      

    SIRIUSi-XHS PWR V20-1 36/36