ACQUITY UPLC I-Class System Pinnacle of UPLC Performance Why introducing ACQUITY UPLC I-Class to the UPLC Family? to the UPLC Family? ©2011 Waters Corporation 1 Henrik Andersen

ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

  • Upload
    others

  • View
    55

  • Download
    1

Embed Size (px)

Citation preview

Page 1: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

ACQUITY UPLC I-Class SystemPinnacle of UPLC Performance

Why introducing ACQUITY UPLC I-Class to the UPLC Family?to the UPLC Family?

©2011 Waters Corporation 1

Henrik Andersen

Page 2: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

ACQUITY UPLC H-ClassACQUITY UPLC

©2011 Waters Corporation 2

Page 3: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Instrument Contribution to Extra-Column Effects

22det

2,det

2,

2,

2,

2,

2, Fectorectorvpostcolumnvcolumnvprecolumnvinjectorvtotalv ⋅+++++= τσσσσσσ

Injectionvolume

+injectorband-

Tubingbetweeninjector

and column

Columnvolume

+frits

Tubingbetweencolumn

and detector

Band-spreadinginside thedetector

cell

Time-basedBand-

spreadingin the

Detectorband-spreading

column detector cell+

tubing

Detector(SamplingRate; TimeConstant)

Engineering developments have specifically improved dispersion I j fi i fl h li f— Injector, fittings, flow path, sealing surfaces

— Reduced tubing volumes

— Improved flow cell dispersion

©2011 Waters Corporation 3

— Higher pressure/backpressure resistance

Page 4: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

System Band Spreading:Effect of Tubing ID and Length

Tubing with a large ID increases the overall system volume and contributes to band spreading

g g

Excessive tubing lengths contribute to band spreadingDecreasing tubing ID and length increases system pressure

Smallest ID is best

Flow Direction

0.020"0.0025" 0.040"

Shortest length is best

©2011 Waters Corporation 4

Page 5: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Extra Colum VolumeContribution Of Tubing ID

Tubing ID Volume/foot Volume/100 cm

0 040’’ 250 µl 824 µl

g

0.040 250 µl 824 µl

0.020’’ 62 µl 206 µl

0.009’’ 12 µl 42 µl

0.005’’ 3.5 µl 12.5 µl

0.004’’ 2.5 µl 8.2 µl

0.003’’ 1.4 µl 4.6 µlµ µ

How does it translate to chromatographic performances?— Peak capacity is increased (with decreasing ID)Peak capacity is increased (with decreasing ID)

— Peak shape is improved

— Overal chromatographic performances are strongly impacted

Higher pressure limit is just a tool a technology enabler

©2011 Waters Corporation 5

Higher pressure limit is just a tool, a technology enabler

Page 6: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

ACQUITY UPLC I-Class SystemWhat is it?

New cells for TUV and PDA— 500 nl/10 mm

— 250 nl/10 mm (for 1mm ID columns)

Best inlet for all MS detectors

Compatible with FLR & ELSD

Column ovens— Single or multiple column

N APH ith ll ID t bi

Sample manager (SM-FTN or SM-FL)

— Less dispersion volume

— New APH with smaller ID tubing

Binary solvent manager (BSM)

Less dispersion volume

— Improved carry over

©2011 Waters Corporation 6

— 1200 bars 18000 PSI

— New tubing with smaller ID

Page 7: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

I-Class FTN vs. I-Class FL:Comparison Summaryp y

ACQUITY I-Class FTN I-Class FL

Dwell Volume* 120 µL 100 µL 95 µL

Bandspread* 12 µL <9 µL <7 µL Bandspread* 12 µL <9 µL <7 µL

Carryover 0.005 % 0.001 % 0.002 %

Precision <0.3% Full loop <1.0% PLUNO

<1% 0.2 to 1.9 μL<0.5% 2 to 10 μL

<0.3% Full loop <1.0% PLUNO

Cycle Time < 15 sec(with load ahead) <30 sec < 15 sec

(with load ahead)

User Simplicity 3 injection modesloop changes

Single injection mode

3 injection modesloop changesloop changes mode loop changes

Sample organizer available with both configurations— 18 shelves

©2011 Waters Corporation 7

*Measured with a complete system – Measurement on module only doesn’t make sense

Page 8: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

NEW Column Heater

Supports one UPLC column ith g a d col mnwith guard-column

From ambiant +5°C to 90°CNew 0.003” ID tubing (APH)g ( )

Robust 18K psi fitting

Excellent method transfer between UPLC systemsbetween UPLC systems

©2011 Waters Corporation 8

Page 9: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Column Managerg

Two (2) Columns plumbed right or l ft lleft only

From 4°C to 90°CNew 0.003” ID tubing (APH)New 0.003 ID tubing (APH)

Robust 18 K PSI fitting

Excellent Method transfer between UPLC systems

Also available for 2D Technology

©2011 Waters Corporation 9

Page 10: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

ACQUITY UPLC I-Class SystemOptical Detection (TUV & PDA)p ( )

500nL flow cell has been optimized for reduced dispersion— Flow cell body is the sameFlow cell body is the same

— Inlet and outlet tubing i.d. has been reduced

o .0025” ID inlet PEEK tubing

0035” ID outlet PEEK tubingo .0035 ID outlet PEEK tubing

250nL / 183 µm version for 1mm chemistry

ELSD & FLR are compatible with I-Class

©2011 Waters Corporation 10

Page 11: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Lower Dispersion I-Class versus Competitor Designp g

Competitive UHPLC System: 21μL extra column band spread

90.00

AU

36.00

54.00

72.00

Average Peak Capacity at 5σ : 42

0.00

18.00

Minutes0.07 0.14 0.21 0.28 0.35 0.42

AU

0.078

0.104

0.130

Average Peak Capacity at 5σ: 70

0.000

0.026

0.052

0.00 0.07 0.14 0.21 0.28 0.35 0.42

©2011 Waters Corporation 11

Minutes

ACQUITY UPLC I-Class: 5μL extra column band spread

Page 12: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Fastest Possible Scenario (1 mm Column)ACQUITY vs ACQUITY IClassQ Q

Instrument Method Name: 30sec_10_95p600mL_90C ACQUITY

0.45

7

0.48

3

0.088

0.110ACQUITY UPLC

0.15

8

0.20

5

0.35

1

0.41

9

0.50

3

AU

0.022

0.044

0.066 Average peak capacity at 5σ = 57

0.000

Instrument Method Name: 20sec_NO hold 10_95_p904mL_90C

0.17

0

0.24

1 0.26

1

0.088

0.110ACQUITY UPLC – I CLASS SM

0.09

8

0.11

7

0.21

3

0.27

6

AU

0.022

0.044

0.066 Average peak capacity at 5σ = 70

0.000

Minutes0.00 0.10 0.20 0.30 0.40 0.50 0.60 0.70

Ballistic gradient on 1mm x 50mm BEH C18

©2011 Waters Corporation 12

Due to the 15k psi limit, with the ACQUITY, the flow rate is 0.6 ml/mn

With the ACQUITY I-Class the flow rate is 0.9 ml/mn

Page 13: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Comparison Of Isocratic Separations2.1 mm Columns – 0.8 ml/mn

Competitive design;21μL measured band spread

Instrument Method Name: SampleName: TCA 50 inj 6

0.44

8

0.55

1 0.68

4

0.93

4

0

48.00

60.00

21μL measured band spreadAverage USP Plates = 6016

0

1.31

1.62

8

2.46

0mAU

0.00

12.00

24.00

36.00

I t t M th d N TCA i ti 50 50 0 8 S l N t 50

ACQUITY; 12μL measured band spreadAverage USP Plates = 7163

Instrument Method Name: TCA isocratic 50_50 0p8 SampleName: tca 50

0.44

4

0.55

0

0.68

5

0.93

3

1.32

0

.646

85

AU

0 024

0.036

0.048

0.060

ACQUITY I-Class FL; 1

2.48

0.000

0.012

0.024

Instrument Method Name: TCA isocratic 50_50 0p8 SampleName: tca 50

5 501

360 08

0.10Q ;

5μL measured band spreadAverage USP Plates = 11,356

0.41

5 0.5

0.63

0.89

1

1.26

2

1.56

4

2.37

3

AU

0.02

0.04

0.06

0.08

©2011 Waters Corporation 13

0.00

Minutes0.00 0.40 0.80 1.20 1.60 2.00 2.40 2.80

Page 14: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

UPLC H-Class to UPLC I-ClassMethod Transfer Examplep

ACQUITY UPLCH-ClassNot Retention Time Adjusted

ACQUITY UPLC I-ClassIncreasing regulatory demands

ACQUITY UPLC H-ClassRetention Time Adjusted

ACQUITY UPLC I-Class

©2011 Waters Corporation 14

Page 15: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Original Barbiturates Separation Original Barbiturates Separation ACQUITY UPLC HACQUITY UPLC H--Class SystemClass SystemACQUITY UPLC HACQUITY UPLC H Class SystemClass System

1 – Allobarbital2 – Phenobarbital3 – Butalbarbital4 – Mephobarbital5 – Amobarbital6 - Secobarbital

©2011 Waters Corporation 15

Page 16: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

UPLC Method on ACQUITY UPLC HUPLC Method on ACQUITY UPLC H--Class SystemClass SystemClass SystemClass System

Sample: Barbiturates Mix @ 20μM

Column: 2.1 x 50mm ACQUITY UPLC BEH C18, 1.7μmColumn: 2.1 x 50mm ACQUITY UPLC BEH C18, 1.7μm

Injection Volume: 1.5μL

Temperature: 45°C

Fl R t 0 64 L/ iFlow Rate: 0.64mL/min

Mobile Phase A: 10mM Ammonium Bicarbonate Buffer pH 10.0

Mobile Phase B: Acetonitrile

Gradient: Please refer to the table below

Detection Wavelength: 245nm

Time (min) %A %BTime (min) %A %B

Initial 90 10

1.70 68 32

©2011 Waters Corporation 16

1.71 90 10

Page 17: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Gradient Hold Insert for Gradient Hold Insert for ACQUITY UPLC IACQUITY UPLC I--ClassClassACQUITY UPLC IACQUITY UPLC I--ClassClass

©2011 Waters Corporation 17

Page 18: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

System System Volume Difference Volume Difference Not AdjustedNot Adjusted

1 2

0.28

System System Volume Difference Volume Difference Not AdjustedNot Adjusted

ACQUITY UPLC H-Class

3

4

5

6

AU

007

0.14

0.21

0.00

0.07

ACQUITY UPLC I Cl

1

2

3

4

5

6

U

0.21

0.28ACQUITY UPLC I-Class

AU

000

0.07

0.14

©2011 Waters Corporation 18

0.00

Minutes0.00 0.30 0.60 0.90 1.20 1.50

Page 19: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Adjusted Gradient Table forAdjusted Gradient Table forACQUITY UPLC IACQUITY UPLC I--Class SystemClass System

ACQUITY UPLC H-Class Gradient Table

ACQUITY UPLC IACQUITY UPLC I--Class SystemClass System

BEFORE NOW

Time (min) %A %B

Initial 90 10

1.70 68 32

Q

1.71 90 10

Gradient Hold insert of 262uL at 640uL/min Flow Rate translates to

ACQUITY UPLC I-Class Gradient Table

Flow Rate translates to 0.40 minutes

Time (min) %A %B

Initial 90 10

0.40 90 10

2 10 68 32

©2011 Waters Corporation 19

2.10 68 32

2.11 90 10

Page 20: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

System System Volume Difference Volume Difference AdjustedAdjusted

1 2

0.28

yy jj

ACQUITY UPLC H-Class

3

4

5

6

AU

007

0.14

0.21

0.00

0.07

ACQUITY UPLC I-Class

1 2

3 4

5

6

U

0.21

0.28ACQUITY UPLC I-Class

AU

000

0.07

0.14

©2011 Waters Corporation 20

0.00

Minutes0.00 0.30 0.60 0.90 1.20 1.50

Page 21: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Explosives Mix Separation run on Explosives Mix Separation run on ACQUITY UPLC HACQUITY UPLC H--Class SystemClass SystemOverlay of 6 Replicate InjectionsOverlay of 6 Replicate InjectionsOverlay of 6 Replicate InjectionsOverlay of 6 Replicate Injections

Peak 1 HMX Peak 2 RDX Peak 3 TNB Peak 4 1,3-DNB Peak 5 NB Peak 6 Tetryl Peak 7 TNTPeak 8 2-Amino-4 6-DNTPeak 8 2 Amino 4,6 DNTPeak 9 4-Amino-2,6-DNTPeak 10 2,4-DNT Peak 11 2,6-DNTPeak 12 2-NTP k 13 4 NTPeak 13 4-NTPeak 14 3-NT

©2011 Waters Corporation 21

Page 22: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

System System Volume Difference Volume Difference Not AdjustedNot Adjusted

10.80

yy jj

ACQUITY UPLC H-Class

2

3

4

6 7 2 14

AU 0.40

0.60

5 6 7 8 9

10 11 12 13

0.00

0.20

1

2

3

U

0.60

0.80 ACQUITY UPLC I-ClassLesser System Volume, hence, shorter peak retention times

4

5 6 7 8 9

10 11 12 13

14

AU

000

0.20

0.40

©2011 Waters Corporation 22

0.00

Minutes0.00 1.50 3.00 4.50 6.00 7.50

Page 23: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

System System Volume Difference Volume Difference AdjustedAdjusted

10.80

System System Volume Difference Volume Difference AdjustedAdjusted

ACQUITY UPLC H-Class

2

3

4

5 6 7 8 9 2 14

AU 0.40

0.60

5 6 7 8 9

10 11 12 13

0.00

0.20

080 1

2

3

U 040

0.60

0.80 ACQUITY UPLC I-Class4

5 6 7 8 9

10 11 12 13

14

A

0.00

0.20

0.40

©2011 Waters Corporation 23

Minutes0.00 1.50 3.00 4.50 6.00 7.50

Page 24: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Original ACQUITY UPLC Separationg Q p

Column: 2 1 x 50mm BEH C18 1 7µmColumn: 2.1 x 50mm BEH C18 1.7µmSample: Beta-BlockersInj Vol: 1µLTemp: 60°CFlow Rate: 750µL/minEluent A: 10mM Ammonium Bicarbonate pH10Eluent B: Acetonitrile

AU

0.70

Gradient: 5% to 60% B in 2 minutesDetection: 230nm 20Hz Normal

0.00

Minutes0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00

©2011 Waters Corporation 24

Minutes

Page 25: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Transferring Methods ACQUITY UPLC to ACQUITY UPLC I-Class

1.00

Q

Peaks Resolution

2-3 3.54ACQUITY UPLC

AU

0.00

0.50 3-4 4.56

7-8 2.02

AU

0.50

1.00 Peaks Resolution

2-3 3.27

3-4 5.20

7 8 2 37

ACQUITY UPLC I-Class – Same Conditions

0.00

1.00

7-8 2.37

Peaks Resolution

2 3 3 61ACQUITY UPLC I-Class – Optimized Transfer*

AU

0.00

0.50

2-3 3.61

3-4 4.80

7-8 2.13

©2011 Waters Corporation 25

Minutes0.00 0.50 1.00 1.50 2.00

*Compensated for System Volume (20µL) and Temperature Differences (8°C)

Page 26: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

Transferring Methods ACQUITY UPLC to ACQUITY UPLC I-Class

Important things to remember— Passive versus active heating

Q

Passive versus active heating

— Small system volume differences

— System dispersion differences

©2011 Waters Corporation 26

Page 27: ACQUITY UPLC I-Class System - Waters Corporation...ACQUITY UPLC I-Class System ACQUITY UPLC H-Class Gradient Table BEFORE NOW Time (min) %A %B Initial 90 10 1.70 68 32 1.71 90 10 Gradient

©2011 Waters Corporation 27