Upload
lamnguyet
View
238
Download
0
Embed Size (px)
Citation preview
QUAL
-QA-
TEM
P-02
20_r
ev15
Part Number: RAN322D
Date: 24/3/14ECR: 5180
ABX Pentra 60Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2
Tn
TN
Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN322D
Objective Version update for the ABX Pentra 60 with or without Takasago valves
Intervention
informationIntervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory
Tracking number TN-RAN322D Intervention estimated duration: 0h30
User process impact YES NO Potential patient result impact YES NO
User training needed YES NO User documentation update YES NO
Applied in factory on instrument serial #
since instrument serial # 12829
Note history Note index Reason for change
A to B
Software
version
Note index Previous New Mini. requirements
A to B 2.3.0 2.3.2
B to C 2.3.2 2.3.3
C to D 2.3.3 3.0.0
New parts Parts designation Previous New Reason for change
KIT V3.0.0-1 + V3.0.0-2 P60OT
XEC130AS XEA769DS Software update
New
adjustments
Adjustments Previous New Reason for change
New
documents
Documentation Previous New Reason for change
User manual multilanguages CD-ROM
RAX058F RAX058G Update
Additionnal
requirements
Designation Part number Link
2/4
Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN322D
1. Modifications and evolutionsRefer to addendum RAM264AEN provided in the kit.
2. Software installation
1- Switch the instrument off and remove the left-hand side panel.2- Using an antistatic mat, replace the instrument EPROM (refer to the diagram on the following page).
Replace the EPROM by the correct one, depending on whether the isntrument is equipped with Takasago valves ornot. The EPROM V3.0.0-1 must be used for instruments not equipped with Takasago valves, and the EPROM V3.0.0-2 VTK must be used for instruments equipped with Takasago valves.
3- Re-install the left-hand side panel.
Part detailskit or Assy Designation
Component p/n
Component designation QtyAvaila-ble for order
XEA769DSKIT V3.0.0-1 + V3.0.0-2 P60OT
Yes
EPROM V3.0.0-1 1 No
EPROM V3.0.0-2 VTK 1 No
RAN322D V3.0.0 P60OT technical note 1 No
RAX058G Documentation CD-ROM P60 OT 1 Yes
RAM264AEN Addendum P60 OT V3.0.X EN 1 No
For instruments equipped with a software version below V2.0.1, it is requested to record the instrument settings (internal values: motor positions, calibration, etc.) before installation.
Without Takasago valves With Takasago valves
3/4