4
QUAL-QA-TEMP-0220_rev15 Part Number: RAN322D Date: 24/3/14 ECR: 5180 ABX Pentra 60 Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 Tn TN

Tn TNits-information.horiba-abx.com/_basedoc/hematology/abx... · 2016-09-05 · Tn TN. Kit Version 3.0.0 ... Objective Version update for the ABX Pentra 60 with or without Takasago

Embed Size (px)

Citation preview

QUAL

-QA-

TEM

P-02

20_r

ev15

Part Number: RAN322D

Date: 24/3/14ECR: 5180

ABX Pentra 60Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2

Tn

TN

Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN322D

Objective Version update for the ABX Pentra 60 with or without Takasago valves

Intervention

informationIntervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory

Tracking number TN-RAN322D Intervention estimated duration: 0h30

User process impact YES NO Potential patient result impact YES NO

User training needed YES NO User documentation update YES NO

Applied in factory on instrument serial #

since instrument serial # 12829

Note history Note index Reason for change

A to B

Software

version

Note index Previous New Mini. requirements

A to B 2.3.0 2.3.2

B to C 2.3.2 2.3.3

C to D 2.3.3 3.0.0

New parts Parts designation Previous New Reason for change

KIT V3.0.0-1 + V3.0.0-2 P60OT

XEC130AS XEA769DS Software update

New

adjustments

Adjustments Previous New Reason for change

New

documents

Documentation Previous New Reason for change

User manual multilanguages CD-ROM

RAX058F RAX058G Update

Additionnal

requirements

Designation Part number Link

2/4

Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN322D

1. Modifications and evolutionsRefer to addendum RAM264AEN provided in the kit.

2. Software installation

1- Switch the instrument off and remove the left-hand side panel.2- Using an antistatic mat, replace the instrument EPROM (refer to the diagram on the following page).

Replace the EPROM by the correct one, depending on whether the isntrument is equipped with Takasago valves ornot. The EPROM V3.0.0-1 must be used for instruments not equipped with Takasago valves, and the EPROM V3.0.0-2 VTK must be used for instruments equipped with Takasago valves.

3- Re-install the left-hand side panel.

Part detailskit or Assy Designation

Component p/n

Component designation QtyAvaila-ble for order

XEA769DSKIT V3.0.0-1 + V3.0.0-2 P60OT

Yes

EPROM V3.0.0-1 1 No

EPROM V3.0.0-2 VTK 1 No

RAN322D V3.0.0 P60OT technical note 1 No

RAX058G Documentation CD-ROM P60 OT 1 Yes

RAM264AEN Addendum P60 OT V3.0.X EN 1 No

For instruments equipped with a software version below V2.0.1, it is requested to record the instrument settings (internal values: motor positions, calibration, etc.) before installation.

Without Takasago valves With Takasago valves

3/4

Kit Version 3.0.0 - Eproms V3.0.0-1 V3.0.0-2 TN: RAN322D

Without Takasago valves With Takasago valvesV3.0.0-1 V3.0.0-2 - VTK

U43 EVEN U42 ODD

4/4